PXI-4065 [NI]

Low-Cost 6 1⁄2-Digit Multimeters; 低成本6 1/2位数字万用表
PXI-4065
型号: PXI-4065
厂家: National Instruments    National Instruments
描述:

Low-Cost 6 1⁄2-Digit Multimeters
低成本6 1/2位数字万用表

文件: 总6页 (文件大小:426K)
中文:  中文翻译
下载:  下载PDF数据表文档文件
1
Low-Cost 6 ⁄ -Digit Multimeters  
2
NI PXI-4065, NI PCI-4065, NI PCIe-4065,  
NI USB-4065 NEW!  
1
• Basic 6 ⁄  
2-digit digital multimeters  
Operating Systems  
• Windows Vista/XP/2000  
(DMMs) for PXI, PCI, PCI Express,  
and USB  
• 7 built-in measurement modes –  
AC/DC voltage, AC/DC current, 2- or  
4-wire resistance, and diode test  
Recommended Software  
• LabVIEW  
• LabVIEW Real-Time Module1  
• LabWindows/CVI  
• Measurement Studio  
• LabVIEW SignalExpress  
300 VDC/Vrms isolation  
1
• 10 readings/s maximum at 6 ⁄  
2
digits  
• Ideally suited for OEMs, educational  
labs, and production test  
Other Compatible Software  
• Microsoft Visual Basic  
• C/C++  
For increased accuracy, resolution,  
measurement speed, and functionality,  
consider NI 407x FlexDMM devices.  
Driver Software (included)  
• NI-DMM  
1PXI-4065 only  
• NI-DMM Soft Front Panel  
• DMM/Switch Express VI  
Overview  
National Instruments 4065 devices are low-cost 6 ⁄  
Efficient Mixture of Accuracy, Resolution,  
and Speed  
1
2
-digit PXI, PCI,  
1
PCI Express, and USB DMMs for the measurement of voltage, current,  
and resistance, or for diode test. They are ideally suited for OEMs,  
educational laboratories, or other cost-conscious test and measurement  
facilities. With 300 VDC/Vrms of isolation, current measurements up to  
NI 4065 DMMs offer 6 ⁄  
2
digits of resolution at up to 10 readings/s and  
up to 3000 readings/s at lower resolutions. By trading off speed for  
resolution, you can maximize measurement quality and throughput.  
NI 4065 DMMs also offer comparable accuracy over a 24-hour cycle  
or one year of an external calibration to other GPIB, LAN, or USB-based  
3 A, and 2- or 4-wire resistance measurements, NI 4065 DMMs provide  
1
a complete multimeter solution for basic 6 ⁄  
2
-digit measurement needs.  
For higher-performance requirements, consider using NI 407x PCI  
and PXI FlexDMM devices. They combine industry-leading accuracy and  
Digits  
6 ⁄2  
Resolution (bits)  
Reading Rate (S/s)  
1
22  
18  
15  
10  
1500  
3000  
1
5 ⁄  
2
1
1
resolution at 6 ⁄  
2
and 7 ⁄  
2
digits with 1.8 MS/s digitizer capability for faster  
1
4 ⁄  
2
sampling rates in production test environments.  
Table 1. NI 4065 Reading Rates versus Resolution  
Flexible Software Improves Usability  
1
6 ⁄  
2-digit DMMs on the market. With an NI 4065, there is no need to  
You can program NI 4065 DMMs through a standard application  
programming interface (API) in NI LabVIEW and LabWindows/CVI or  
Microsoft programming languages. For simple benchtop measurements,  
sacrifice basic measurement quality for a low selling price.  
For higher resolution, accuracy, or speed, consider using one of the  
1
NI 407x 6 ⁄  
2
-digit FlexDMM devices. The PCI-4070 can accurately measure  
1
the NI-DMM Soft Front Panel provides the common display and interface 100 readings/s at 6 ⁄  
2
digits with 31 ppm of DC voltage accuracy over a  
found on traditional DMMs to make simple measurements quickly and  
easily. When combined with switches, the DMM/Switch Express VI can  
rapidly develop a high-channel-count data-logging system with a single  
function call in LabVIEW. In LabVIEW SignalExpress, the DMM/Switch  
Express VI sets up a live data logger without any programming.  
two-year period. This represents a 3X improvement over the NI 4065  
DMMs for accuracy and noise. The PXI-4071 is also available with  
1
measurements at 7 ⁄  
2
digits up to 1000 V. For more information on the  
complete line of National Instruments DMMs, visit ni.com/multimeters.  
1
Low-Cost 6 ⁄  
2-Digit Multimeters  
Integration with Switching  
NI PXI-4065  
You can easily combine the NI PXI-4065 DMM with NI PXI switches in the  
low-cost NI PXI-1033 integrated chassis and controller to create high-  
1
channel-count, 6 ⁄  
2-digit data-logging systems with up to 588 channels.  
You can configure these systems, controllable from any PCI Express or  
ExpressCard slot on a PC or laptop computer, with more than 100  
available PXI switch topologies. For further system expansion, you can  
add more than 3,000 channels with each additional 18-slot PXI chassis.  
Figure 3. Use the USB-4065 with USB-controlled switching for a complete  
DMM/switch solution.  
Ordering Information  
Figure 1. Expand channel count with PXI switching.  
NI PXI-4065............................................................................780011-01  
NI PCI-4065............................................................................779770-01  
NI PCIe-4065..........................................................................779771-01  
NI USB-4065, PCI-4065 and PCIe-4065  
You can integrate NI 4065 DMMs with stand-alone switching using the  
4- or 12-slot USB Switch Mainframes from National Instruments. You can  
configure these switch systems, controllable from any available USB port  
on a PC or laptop computer, with more than 50 available SCXI switch  
topologies. For existing SCXI switch systems, the NI 4065 DMMs can also  
provide control via the AUX connector on the front of the DMM. Switching  
offers a simple method of channel expansion for data-logging systems  
based on an NI 4065.  
NI USB-4065..........................................................................780152-01  
Includes the standard P-1 probe set and NI-DMM software.  
Accessories for PXI  
NI PXI-1033 (chassis and integrated MXI Express controller,  
1 PXI Express port, 3 m cable) ..............................................779756-01  
NI PXI-2527 (300 V, 64-channel multiplexer) ........................778572-27  
Accessories for USB, PCI, and PCI Express  
NI USB Switch Mainframe (4-slot) ......................................778570-01  
NI SCXI-1127 (300 V, 64-channel multiplexer)......................776572-27  
General Accessories  
P-1 Probe Set (standard)........................................................761000-01  
P-2 Probe Set (additional connectors) ..................................184698-01  
P-3 Probe Set (banana plug to bare wire) ............................185692-01  
CSM-10 A (10 A current shunt) ............................................777488-02  
BUY NOW!  
For complete product specifications, pricing, and accessory  
information, call 800 813 3693 (U.S.) or go to ni.com/multimeters.  
Figure 2. Expand channel count with USB-controlled switching.  
BUY ONLINE at ni.com or CALL 800 813 3693 (U.S.)  
2
1
Low-Cost 6 ⁄  
2-Digit Multimeters  
Specifications  
DC Specifications  
Resolution (Digits)  
6 ⁄2  
Reading Rate1 (S/s)  
0.6 (0.5)  
6 (5)  
Aperture Time (NPLC)  
RMS Noise2 (ppm of range)  
1
100  
10  
0.06  
0.2  
10 (8.33)  
30 (25)  
6
2
0.25  
0.4  
1
5 ⁄  
2
60 (50)  
900  
1,500  
3,000  
1
0.55  
1.7  
2.5  
0.06  
0.04  
0.02  
1
4 ⁄  
2
11.5  
1 Specified for 60 Hz and (50 Hz) operation  
2 Measured on the 10 V range  
DC Voltage (ppm of reading + ppm of range)1  
Range  
100 mV3  
1 V  
10 V  
100 V  
Resolution  
100 nV  
1 µV  
10 µV  
100 µV  
Input Resistance4  
>10 G, 10 MΩ  
>10 G, 10 MΩ  
>10 G, 10 MΩ  
10 MΩ  
24-Hour2, Tcal 1 °C  
90-Day, Tcal 5 °C  
65 + 35  
1-Year, Tcal 5 °C  
Tempco (0 to 55 °C)  
5 + 2  
30 + 30  
20 + 6  
15 + 5  
20 + 6  
20 + 20  
90 + 35  
90 + 7  
90 + 6  
110 + 7  
110 + 20  
65 + 7  
65 + 6  
75 + 7  
75 + 20  
5 + 0.2  
5 + 0.2  
6 + 0.2  
6 + 0.5  
300 V  
1 mV  
10 MΩ  
1ppm (part per million) = 0.0001%  
2Relative to external calibration source  
3With offset nulling  
4Default input resistance is 10 M  
Tcal = temperature at which last external calibration was performed NI factory calibration is 23 ˚C 1 ˚C  
Tempco = temperature coefficient  
DC Current (ppm of reading + ppm of range)  
Range  
10 mA  
100 mA  
1 A  
Resolution  
10 nA  
100 nA  
1 µA  
Burden Voltage (typical)  
24-Hour2, Tcal 1 °C  
50 + 100  
90-Day, Tcal 5 °C  
300 + 200  
300 + 50  
800 + 100  
12001 + 200  
1-Year, Tcal 5 °C  
500 + 200  
500 + 50  
1000 + 100  
12001 + 200  
Tempco (0 to 55 °C)  
30 + 20  
<60 mV  
<0.6 V  
<0.35 V  
<1 V  
100 + 40  
500 + 60  
10001 + 200  
30 + 5  
65 + 10  
65 + 20  
3 A  
3 µA  
1Add 650 ppm/A of reading for currents above 1.5 A  
2Relative to external calibration source  
Resistance (4- and 2-Wire) (ppm of reading + ppm of range)  
Range  
100 Ω  
1 kΩ  
Resolution  
100 µΩ  
1 mΩ  
Test Current  
1 mA  
1 mA  
24-Hour1, Tcal 1 °C  
30 + 30  
90-Day, Tcal 5 °C  
95 + 40  
1-Year, Tcal 5 °C  
110 + 40  
Tempco (0 to 55 °C)  
8 + 3  
8 + 1  
20 + 6  
95 + 10  
110 + 10  
10 kΩ  
100 kΩ  
1 MΩ  
10 MΩ  
100 MΩ  
10 mΩ  
100 mΩ  
1 Ω  
10 Ω  
100 Ω  
100 µA  
10 µA  
5 µA  
500 nA  
20 + 6  
20 + 6  
20 + 10  
150 + 10  
2000 + 20  
95 + 10  
95 + 10  
110 + 12  
400 + 12  
6000 + 40  
110 + 10  
110 + 10  
125 + 12  
500 + 12  
8 + 1  
8 + 1  
10 + 1  
30 + 2  
400 + 4  
500 nA || 10 MΩ  
8000 + 40  
1Relative to external calibration source  
Specifications are for 4-wire measurements. For 2-wire measurements, perform offset nulling or add 200 mto reading.  
Diode Test1  
Range  
Resolution  
Test Current  
Accuracy  
10 V  
10 µV  
100 µA, 1 mA  
Add 50 ppm of reading and 50 ppm of range to 10 VDC voltage specifications.  
1Can be used to test p-n junctions, LEDs, or zener diodes up to 10 V.=  
BUY ONLINE at ni.com or CALL 800 813 3693 (U.S.)  
3
1
Low-Cost 6 ⁄  
2-Digit Multimeters  
DC Functions General Specifications  
Effective common-mode rejection ratio (CMRR)  
(1 kresistance in LO lead).................................................. >150 dB (DC, 50 and 60 Hz 1%) with second-order DC noise rejection,  
10 PLC aperture  
Maximum 4-wire lead resistance.......................................... Use the lesser of 10% of range or 1 kΩ  
Overrange............................................................................... 105% of range except 300 V and 3 A range  
DC voltage input bias current................................................ <40 pA at 23 °C (typical)  
AC Specifications  
Digits  
6 ⁄2  
Desired Bandwidth  
10 Hz to 100 kHz  
100 Hz to 100 kHz  
500 Hz to 100 kHz  
Recommended Reading Rate  
1
1 S/s  
10 S/s  
100 S/s  
1
5 ⁄  
2
1
4 ⁄  
2
AC Voltage (% of reading + % of range)  
Range (Peak Voltage)  
200 mV ( 320 mV)  
2 V ( 3.2 V)  
20 V ( 32 V)  
300 V ( 425 V)  
Frequency  
10 to 40 Hz  
>40 Hz to 20 kHz  
>20 to 50 kHz  
>50 to 100 kHz  
24-Hour, Tcal 1 °C  
1.5 + 0.04  
90-Day, Tcal 5 °C  
2 + 0.05  
1-Year, Tcal 5 °C  
2 + 0.05  
Tempco (0 to 55 ˚C)  
0.01 + 0.003  
0.005 + 0.003  
0.01 + 0.003  
0.2 + 0.04  
0.3 + 0.04  
1.5 + 0.08  
0.2 + 0.05  
0.3 + 0.05  
1.5 + 0.08  
0.2 + 0.05  
0.3 + 0.05  
1.5 + 0.08  
0.02 + 0.005  
Tempco = temperature coefficient  
AC Current (% of reading + % of range)  
Range (Peak Current)  
10 mA ( 15 mA)  
Frequency  
10 to 40 Hz  
24-Hour, Tcal 1 °C  
90-Day, Tcal 5 °C  
1-Year, Tcal 5 °C  
Tempco (0 to 55 ˚C)  
1.6 + 0.05  
2.1 + 0.05  
2.1 + 0.05  
0.015 + 0.03  
100 mA ( 150 mA)  
500 mA ( 750 mA)  
3 A ( 4.2 A)  
>40 Hz to 5 kHz  
0.3 + 0.05  
0.3 + 0.06  
0.3 + 0.06  
0.015 + 0.03  
Tempco = temperature coefficient  
High Crest Factor Additional Error1  
Crest Factor  
1 to 3  
3 to 4  
Additional Error (% of reading)  
0.05%  
0.1%  
1%2  
4 to 5  
1Applicable for non-sine wave signals up to the rated peak voltage/current or bandwidth  
2For frequencies above 2 kHz  
AC Functions General Specifications  
Input impedance .................................................................... 10 Min parallel with 200 pF  
Input coupling ........................................................................ AC coupling  
Maximum volt-hertz product ................................................. >3 x 107 V-Hz  
Maximum DC voltage component......................................... 250 V  
CMRR (1 kresistance in LO lead) ...................................... >70 dB (DC to 60 Hz)  
Overrange............................................................................... 105% of range except 300 V, 3 A range  
General Specifications  
External calibration interval .................................................. 1 year recommended  
Input protection  
Resistance, diode .............................................................. Up to 300 VDC  
DC V, AC V ......................................................................... Up to 300 VDC, 300 Vrms, 450 Vp  
DC I and AC I ..................................................................... F 3.15 A 250 V fast-acting fuse, user-replaceable  
BUY ONLINE at ni.com or CALL 800 813 3693 (U.S.)  
4
1
Low-Cost 6 ⁄  
2-Digit Multimeters  
Maximum common-mode voltage......................................... 300 VDC or Vrms  
Input terminals....................................................................... Gold-plated low-thermal emf solid copper  
Dimensions ............................................................................ PXI: 3U, 1 slot, PXI/CompactPCI module; 21.6 by 2.0 by 13.0 cm (8.5 by 0.8 by 5.1 in.)  
PCI/PCI Express: 1 slot, PCI or PCI Express board; 18.3 by 12 cm (7.2 by 4.72 in.)  
USB: 17.8 by 10.4 by 3.3 cm (7.0 by 4.1 by 1.3 in.)  
Operating temperature .......................................................... PXI: 0 to 55 °C; PCI/PCI Express: 0 to 40 °C; USB: 0 to 45°C  
Storage temperature.............................................................. -40 to 70 °C  
Relative humidity................................................................... Up to 95% at 40 °C  
Measurement category.......................................................... CAT II  
Pollution degree.....................................................................  
2
Approved at altitudes ............................................................ Up to 2000 m  
Safety  
An NI 4065 meets the requirements of the following standards of safety and electrical equipment for measurement, control, and laboratory use:  
• IEC 61010-1, EN 61010-1  
• UL 61010-1, CSA 61010-1  
Note: For UL and other safety certifications, refer to the product label, or visit ni.com/certification, search by model number or product line,  
and click the appropriate link in the Certification column.  
Electromagnetic Compatibility  
This product is designed to meet the requirements of the following standards of EMC for electrical equipment for measurement, control,  
and laboratory use:  
• EN 61326 EMC requirements; Minimum Immunity  
• EN 55011 Emissions; Group 1, Class A  
• CE, C-Tick, ICES, and FCC Part 15 Emissions; Class A  
Note: For EMC compliance, operate this device according to product documentation.  
CE Compliance  
This product meets the essential requirements of applicable European Directives, as amended for CE marking, as follows:  
• 2006/95/EC; Low-Voltage Directive (safety)  
• 2004/108/EC; Electromagnetic Compatibility Directive (EMC)  
Note: Refer to the Declaration of Conformity (DoC) for this product for any additional regulatory compliance information. To obtain the DoC for this  
product, visit ni.com/certification, search by model number or product line, and click the appropriate link in the Certification column.  
Waste Electrical and Electronic Equipment (WEEE)  
EU Customers: At the end of their life cycle, all products must be sent to a WEEE recycling center. For more information about WEEE recycling  
centers and National Instruments WEEE initiatives, visit ni.com/environment/weee.htm.  
BUY ONLINE at ni.com or CALL 800 813 3693 (U.S.)  
5
NI Services and Support  
NI has the services and support to meet  
Local Sales and Technical Support  
your needs around the globe and through  
the application life cycle – from planning  
and development through deployment  
and ongoing maintenance. We offer  
services and service levels to meet  
customer requirements in research,  
design, validation, and manufacturing.  
Visit ni.com/services.  
In offices worldwide, our staff is local to the country, giving you access  
to engineers who speak your language. NI delivers industry-leading  
technical support through online knowledge bases, our applications  
engineers, and access to 14,000 measurement and automation  
professionals within NI Developer Exchange forums. Find immediate  
answers to your questions at ni.com/support.  
SERVICE  
NEEDS  
We also offer service programs that provide automatic upgrades to  
your application development environment and higher levels of technical  
support. Visit ni.com/ssp.  
Training and Certification  
NI training is the fastest, most certain route to productivity with our  
products. NI training can shorten your learning curve, save development  
time, and reduce maintenance costs over the application life cycle. We  
schedule instructor-led courses in cities worldwide, or we can hold a  
course at your facility. We also offer a professional certification program  
Hardware Services  
NI Factory Installation Services  
NI Factory Installation Services (FIS) is the fastest and easiest way to  
use your PXI or PXI/SCXI combination systems right out of the box.  
that identifies individuals who have high levels of skill and knowledge on Trained NI technicians install the software and hardware and configure  
using NI products. Visit ni.com/training.  
the system to your specifications. NI extends the standard warranty by  
one year on hardware components (controllers, chassis, modules)  
purchased with FIS. To use FIS, simply configure your system online  
Professional Services  
Our Professional Services Team is composed of NI applications engineers, with ni.com/pxiadvisor.  
NI Consulting Services, and a worldwide National Instruments Alliance  
Calibration Services  
NI recognizes the need to maintain properly calibrated devices for  
Partner program of more than 600 independent consultants and  
integrators. Services range from  
start-up assistance to turnkey  
system integration.  
high-accuracy measurements. We provide manual calibration  
procedures, services to recalibrate your products, and automated  
calibration software specifically designed for use by metrology  
laboratories. Visit ni.com/calibration.  
Visit ni.com/alliance.  
OEM Support  
We offer design-in consulting and product integration assistance if you  
want to use our products for OEM applications. For information about  
special pricing and services for OEM customers, visit ni.com/oem.  
Repair and Extended Warranty  
NI provides complete repair services for our products. Express repair  
and advance replacement services are also available. We offer  
extended warranties to help you meet project life-cycle requirements.  
Visit ni.com/services.  
ni.com • 800 813 3693  
*351556A-01*  
351556A-01  
2007-8778-101-D  
National Instruments • info@ni.com  
© 2008 National Instruments Corporation. All rights reserved. CVI, FlexDMM, LabVIEW, Measurement Studio, MXI, National Instruments, National Instruments Alliance Partner, NI, ni.com, SCXI,  
and SignalExpress are trademarks of National Instruments. The mark LabWindows is used under a license from Microsoft Corporation. Other product and company names listed are trademarks or  
trade names of their respective companies. A National Instruments Alliance Partner is a business entity independent from NI and has no agency, partnership, or joint-venture relationship with NI.  

相关型号:

PXI-4072

61⁄2-Digit Digital Multimeter, 1.8 MS/s Isolated Digitizer, and LCR Meter
NI

PXI-4110

3 independent DC power supplies/16-bit voltage setpoint and current limit
NI

PXI-4132

100 V Precision Source Measure Unit (SMU)
NI

PXI-4204

DAQ for High-Voltage Measurements 16-Bit, 200 kS/s, 8 Analog Inputs
NI

PXI-5124

150 MHz, 200 MS/s, 12-Bit Digitizer
NI

PXI-5142

14-Bit 100 MS/s Digitizer with Onboard Signal Processing
NI

PXI-5404

9 kHz to 105 MHz sine wave/DC to 100 MHz clock/1.07 μHz frequency resolution
NI

PXI-5422

200 MS/s, 16-Bit Arbitrary Waveform Generator
NI

PXI-5441

100 MS/s, 16-bit Arbitrary Waveform Generator with Onboard Signal Processing
NI

PXI-5670

250 kHz to 2.7 GHz/8, 32 or 256 MB memory
NI

PXI-5671

250 kHz to 2.7 GHz/16-bit resolution, 100 MS/s arbitrary waveform generation
NI

PXI-6052E

Converter IC
ETC