PXI-5124 [NI]

150 MHz, 200 MS/s, 12-Bit Digitizer; 150兆赫, 200 MS / s的12位数字化仪
PXI-5124
型号: PXI-5124
厂家: National Instruments    National Instruments
描述:

150 MHz, 200 MS/s, 12-Bit Digitizer
150兆赫, 200 MS / s的12位数字化仪

文件: 总3页 (文件大小:250K)
中文:  中文翻译
下载:  下载PDF数据表文档文件
150 MHz, 200 MS/s, 12-Bit Digitizer  
NI PXI-5124  
• 2 channels simultaneously sampled  
at 12-bit resolution  
Operating Systems  
• Windows 2000/NT/XP  
• 200 MS/s real-time and 4.0 GS/s  
random interleaved sampling  
• 150 MHz bandwidth  
• 200 mV to 20 V input range  
• >75 dBc SFDR  
Recommended Software  
• LabVIEW  
• LabWindows/CVI  
• Measurement Studio  
• 8, 32, 256, or 512 MB of memory  
per channel  
Driver Software (included)  
• NI-SCOPE driver  
• Edge, window, hysteresis, video,  
and digital triggering with  
50 ps timestamping  
• LabVIEW Express VIs  
• SCOPE Soft Front Panel  
• NI Spectral Measurements Toolkit  
(with 32, 256 and 512 MB models)  
Calibration  
• Gain, offset, frequency response  
and timing self-calibration  
• 2 year external calibration cycle  
Overview  
Triggering, Clocking, and Synchronization  
The National Instruments PXI-5124 high-resolution digitizer • Edge, window, hysteresis, video, digital, triggering with  
features two 200 MS/s simultaneously sampled inputs channels with 50 ps timestamping  
12-bit resolution, 150 MHz bandwidth, and up to 512 MB of • Pre and posttrigger acquisition in single and multiple-record mode  
memory per channel in a compact 3U PXI module. The NI PXI-5124 • Internal 200 MHz clock or external clock from 50 to 210 MHz  
uses the high-speed PCI bus and the scatter-gather bus mastering of • Phase lock to PXI 10 MHz reference or external reference from  
the NI MITE ASIC to move data to the computer at speeds up to 100  
times faster than traditional instrument interfaces, thereby  
1 to 20 MHz  
Applications  
Communications  
xDSL  
Wireless communications  
Baseband I & Q  
dramatically decreasing overall test time. With the Synchronization Software  
and Memory Core (SMC) architecture of the PXI-5124, you can • IVI-compliant NI-SCOPE driver for  
create mixed-signal systems using signal generators and digital  
waveform generator/analyzers or build a high-channel-count  
digitizer with subnanosecond synchronization between channels.  
LabVIEW, LabWindows/CVI, and  
Microsoft C++ and Visual Basic with  
more than 50 built-in measurements  
• SCOPE Soft Front Panel  
Consumer Electronics  
DVD, DVD-R, and PVR  
Set top box  
Gaming console  
Biomedical and Scientific Research  
Ultrasonic medical imaging  
Mass spectrometry  
Particle physics  
Dual 200 MS/s, 12-Bit Input Channels  
for Time and Frequency Analysis  
for interactive control  
• Spectral Measurements Toolkit  
for sophisticated frequency-domain  
measurements in communications,  
signal intelligence, and  
Aerospace/Defense  
• 150 MHz input bandwidth with anti-alias and noise filters  
Emulation of IC communications  
Consumer Electronics  
RADAR, SONAR, and LIDAR  
Satellite  
• >75 dBc spurious free dynamic range (SFDR)  
• 4.0 GS/s equivalent time sampling for repetitive signals  
• Independent channel selectable 200 mVpp to 20 Vpp input ranges  
• Independent channel selectable 50 or 1 Minput impedance  
• 2-year calibration cycle and 0 to 55 ˚C operating temperature  
Signal intelligence  
avionics applications  
Ordering Information  
NI PXI-5124 ..............................................................778757-0M1  
Includes NI-SCOPE driver and SCOPE Soft Front Panel. 32, 256, and 512 MB  
models include NI Spectral Measurements Toolkit.  
Deep Onboard Memory  
• 8, 32, 256, or 512 MB of memory per channel  
• Capture more than 1 million triggered waveforms with multiple  
record hardware rearm  
1M (memory per channel): 1 (8 MB), 2 (32 MB), (256 MB), 4 (512 MB)  
• Stream data continuously from onboard memory to host memory  
or disk  
Recommended PXI Switch  
NI PXI-2593 ..................................................................778793-01  
150 MHz, 200 MS/s, 12-Bit Digitizer  
Specifications  
These specifications are valid for 0 to 55 ˚C, unless otherwise stated.  
RMS Noise (20 MHz filter enabled)  
Full Scale Input Range  
200 mV  
400 mV  
1 V  
2 V  
4 V  
50  
1 MΩ  
Acquisition System  
ꢀ4 µVrms  
188 µVrms  
470 µVrms  
ꢀ40 µVrms  
1.88 mVrms  
4.7 mVrms  
104 µVrms  
1ꢀ2 µVrms  
480 µVrms  
ꢀ60 µVrms  
1.ꢀ2 mVrms  
4.8 mVrms  
ꢀ.4 mVrms  
Number of channels......................................... 2 simultaneously sampled  
Resolution......................................................... 12 bits  
Bandwidth (-3 dB)  
Full Scale Input Range  
400 mV, 1 V, 2 V, 5 V, 10 V, 20 V  
200 mV  
50 Ω  
150 MHz  
85 MHz  
1 MΩ  
145 MHz1  
75 MHz  
10 V  
20 V (1 Monly)  
Bandwidth limit filters  
(software selectable) ....................................... 20 MHz noise (2-pole Bessel)  
60 MHz antialias (4-pole elliptical)  
Intermodulation distortion3 (IMD).................... -75 dBc  
Phase noise density (10 MHz input) ................ <-100 dBc/Hz at 100 Hz  
<-120 dBc/Hz at 1 kHz  
Maximum sampling rate .................................. 200 MS/s real-time, 4 GS/s random interleave sampling  
Onboard sample memory ................................. 8, 32, 256, or 512 MB per channel  
(4, 16, 128, 256 million samples)  
<-130 dBc/Hz at 10 kHz  
Acquisition Modes  
Real-time sampling rate................................... 200 MS/s to 3.052 kS/s sample rate  
Pre and posttrigger data points2...................... 0 to 100% of full record length  
Random interleave sampling (RIS)................... 4 GS/s to 400 MS/s sample rate (repetitive signals only)  
Memory per  
channel (MB)  
Maximum number  
of records  
Timebase System  
Multiple records acquisition  
(0 to 100% pre and posttrigger data)  
8
32  
32,768  
131,072  
Total sample clock jitter4 ................................. 1 psrms  
Internal  
256  
512  
1,048,576  
2,0ꢀ7,152  
Internal sample clock frequency...................... 200 MS/s sampling rate with  
decimation by n where 1n 65,535  
Timebase accuracy........................................... 25 ppm ( 0.0025%)  
Input impedance............................................... 50 and 1 M || 25 pF, software selectable  
Full-scale input range....................................... 50 : 200 mV, 400 mV, 1 V, 2 V, 4 V, 10 V  
1 M: 200 mV, 400 mV, 1 V, 2 V, 4 V, 10 V, 20 V  
External  
External clock sources...................................... CLK IN (SMB connector), PXI star  
External clock range......................................... 50 MHz to 210 MHz, variable with  
decimation by n where 1n 65,535  
Vertical offset ranges....................................... 50% of full scale input range  
Maximum input overload ................................. 50 : 7 Vrms with peaks 10 V  
1 M: peaks 42 V  
External reference sources .............................. CLK IN (SMB connector), PXI backplane 10 MHz  
External reference range.................................. 1 to 20 MHz in 1 MHz increments  
External clock/reference amplitude................. Sine wave: 0.65 to 2.8 Vpp (0 to 13 dBm)  
Square wave: 0.2 to 2.8 Vpp  
Input coupling................................................... AC, DC, GND (AC coupling on 1 Monly)  
AC coupling cutoff frequency (-3 dB)............... 12 Hz  
Accuracy  
DC accuracy (0 V offset setting)  
External clock/reference impedance ............... 50 , AC coupled  
Full Scale Input Range  
200 mV, 400 mV  
1 V, 2 V  
50 and 1 MΩ  
Trigger System  
0.65% of Input 1.3 mV  
0.65% of Input 1.5 mV  
0.65% of Input 10.0 mV  
Modes............................................................... Edge, Hysteresis, Window, Video, Digital, Immediate, Software  
Source............................................................... CH 0, CH 1, TRIG, PXI_Trig <0:6>, PFI <0:1>, PXI Star, Software  
Slope................................................................. Rising or falling  
4 V, 10 V, 20 V  
Hysteresis......................................................... Fully programmable  
Video trigger..................................................... Negative sync of NTSC, PAL, and SECAM standards  
Video trigger types........................................... Any line, specific line, specific field  
High frequency reject filter.............................. 50 kHz software selectable  
Low frequency reject filter............................... 50 kHz software selectable  
Sensitivity......................................................... CH0 and CH1: 2.5% FS up to 50 MHz  
decreasing to 10% FS at 150 MHz  
TRIG: 2.5% up to 100 MHz decreasing to 10% at 200 MHz  
Level accuracy.................................................. CH0, CH1: 4.7% FS up to 10 MHz  
TRIG: 3.5% FS up to 10 MHz  
Passband flatness (referenced at 50 kHz)  
Full-scale input range  
50 and 1 MΩ  
Filters Off  
400 mV, 1 V, 2 V, 5 V, 10 V, 20 V  
0.5 dB, DC to 20 MHz  
1 dB, 20 MHz to 50 MHz  
1.7 dB, 50 MHz to 100 MHz  
0.6 dB, DC to 20 MHz  
1.5 dB, 20 MHz to 40 MHz  
-1 dB to +2 dB, DC to 55 MHz  
200 mV  
Antialias Filter On  
All ranges  
AC amplitude accuracy (50 kHz) ...................... 50 : 0.06 dB  
1 M: 0.0ꢀ dB  
Channel-to-channel crosstalk .......................... -85 dB at 10 MHz  
Time resolution................................................. 50 ps with time-to-digital converter enabled  
Holdoff5 ............................................................ 2 µs to 171.7ꢀ s, software selectable  
External Trigger Channel (TRIG)  
Spectral Characteristics (typical)  
Impedance........................................................ 1 M|| 22 pF  
Vertical Range .................................................. 5 V  
Coupling............................................................ AC, DC  
Dynamic Performance (50 input impedance with 10 MHz, -1 dBFS input signal)  
Full Scale Input Range  
SFDR  
75 dB  
75 dB  
72 dB  
72 dB  
65 dB  
65 dB  
THD  
SNR  
57 dB  
58 dB  
58 dB  
58 dB  
SINAD  
57 dB  
58 dB  
58 dB  
58 dB  
200 mV  
400 mV  
1 V  
2 V  
4 V  
-74 dBc  
-74 dBc  
-72 dBc  
-72 dBc  
-63 dBc  
-63 dBc  
Power Requirements (typical)  
+3.3 VDC  
1.3 A  
+5 VDC  
1.7 A  
+12 VDC  
130 mA  
-12 VDC  
270 mA  
Total Power  
17.6 W  
10 V  
Environment  
Operating temperature6 ................................... 0 to 55 ºC (Meets IEC-60068-2-1 and IEC-60068-2-2)  
Storage temperature........................................ -40 to 71 ºC (meets IEC-60068-2-1 and 60068-2-2)  
Relative humidity ............................................. 10 to ꢀ0 %, noncondensing (meets 60068-2-56)  
Dynamic Performance (1 Minput impedance with 10 MHz, -1 dBFS input signal)  
Full Scale Input Range  
SFDR  
70 dB  
70 dB  
70 dB  
70 dB  
67 dB  
60 dB  
60 dB  
THD  
SNR  
53 dB  
55 dB  
57 dB  
57 dB  
56 dB  
SINAD  
53 dB  
55 dB  
57 dB  
57 dB  
56 dB  
200 mV  
400 mV  
1 V  
2 V  
4 V  
-68 dBc  
-68 dBc  
-68 dBc  
-67 dBc  
-66 dBc  
-58 dBc  
-58 dBc  
Calibration  
Self-calibration................................................. Gain, offset, frequency response,  
triggering, and timing for all input ranges  
External calibration interval............................. 2 years  
10 V  
20 V  
Certification and Compliances  
CE Mark compliance  
SFDR = Spurious-free dynamic range  
THD = Total harmonic distortion  
For detailed specifications on power, environmental, safety, and physical dimensions see PXI-5124  
detailed specifications.  
SNR = Signal-to-noise ratio, excluding distortion (antialias filter enabled)  
SINAD = Signal-to-noise and distortion (antialias filter enabled)  
1.Bandwidth on the 1 Minput is 145 MHz for 0 to 40 ºC and 135 MHz for 40 to 55 ºC  
2. NI PXI-5124 is also capable of continous acquisition  
3. Measured on ranges up to 2 V on 50 input with two tones at 10.2 MHz and 11.2 MHz, each at -7 dBFS  
4. Includes effects of converter aperture and clock circuitry jitter from 100 Hz to 100 kHz  
5. Time-digital converter disabled  
6. 0 to 45 ºC in PXI-101x and 1000/B chassis  
2
National Instruments • Tel: (800) 433-3488 • Fax: (512) 683-9300 • info@ni.com • ni.com  
NI Services and Support  
NI has the services and support to meet your  
needs around the globe and through the  
application life cycle – from planning  
and development through deployment  
and ongoing maintenance. We offer  
services and service levels to meet  
customer requirements in research,  
design, validation, and manufacturing.  
Visit ni.com/services.  
Local Sales and Technical Support  
In offices worldwide, our staff is local to the country, giving you  
access to engineers who speak your language. NI delivers industry-  
leading technical support through online knowledge bases, our  
applications engineers, and access to 14,000 measurement and  
automation professionals within NI Developer Exchange forums.  
Find immediate answers to your questions at ni.com/support.  
We also offer service programs that provide automatic upgrades to  
your application development environment and higher levels of  
technical support. Visit ni.com/ssp.  
Training and Certification  
NI training is the fastest, most certain route to productivity with our Hardware Services  
products. NI training can shorten your learning curve, save NI Factory Installation Services  
development time, and reduce maintenance costs over the NI Factory Installation Services (FIS) is the fastest and easiest way to  
application life cycle. We schedule instructor-led courses in cities use your PXI or PXI/SCXI™ combination systems right out of the  
worldwide, or we can hold a course at your facility. We also offer a box. Trained NI technicians install the software and hardware and  
professional certification program that identifies individuals who configure the system to your specifications. NI extends the standard  
have high levels of skill and knowledge on using NI products. warranty by one year on hardware components (controllers, chassis,  
Visit ni.com/training.  
modules) purchased with FIS. To use FIS, simply configure your  
system online with ni.com/pxiadvisor.  
Professional Services  
Our Professional Services Team is comprised of NI applications Calibration Services  
engineers, NI Consulting Services, and a worldwide NI Alliance NI recognizes the need to maintain properly calibrated devices for  
Partner Program of more than 600 independent consultants and high-accuracy measurements. We provide manual calibration  
integrators. Services range procedures, services to recalibrate your products, and automated  
from start-up assistance to calibration software specifically designed for use by metrology  
turnkey system integration. laboratories. Visit ni.com/calibration.  
Visit ni.com/alliance.  
Repair and Extended Warranty  
OEM Support  
NI provides complete repair services for our products. Express repair  
We offer design-in consulting and product integration assistance and advance replacement services are also available. We offer  
if you want to use our products for OEM applications. For extended warranties to help you meet project life-cycle requirements.  
information about special pricing and services for OEM customers, Visit ni.com/services.  
visit ni.com/oem.  
ni.com • (800) 433-3488  
National Instruments • Tel: (512) 683-0100 • Fax: (512) 683-9300 • info@ni.com  
© 2004 National Instruments Corporation. All rights reserved. LabVIEW, LabWindows/CVI, Measurement Studio and ni.com are trademarks  
of National Instruments. Other product and company names listed are trademarks or trade names of their respective companies.  

相关型号:

PXI-5142

14-Bit 100 MS/s Digitizer with Onboard Signal Processing
NI

PXI-5404

9 kHz to 105 MHz sine wave/DC to 100 MHz clock/1.07 μHz frequency resolution
NI

PXI-5422

200 MS/s, 16-Bit Arbitrary Waveform Generator
NI

PXI-5441

100 MS/s, 16-bit Arbitrary Waveform Generator with Onboard Signal Processing
NI

PXI-5670

250 kHz to 2.7 GHz/8, 32 or 256 MB memory
NI

PXI-5671

250 kHz to 2.7 GHz/16-bit resolution, 100 MS/s arbitrary waveform generation
NI

PXI-6052E

Converter IC
ETC

PXI-6509

Low-Cost Industrial Digital I/O-5 V TTL/CMOS
NI

PXI-6529

24 or 48 channel-to-channel optically isolated inputs and solid-state relay outputs
NI

PXI-6541

100 and 50 MHz Digital Waveform Generator/Analyzers
NI

PXI-665X

PXI and PXI Express Timing and Synchronization Control
NI

PXI-6682

GPS, IEEE 1588, and IRIG-B Timing and Synchronization Modules for PXI and PXI Express
NI