PXI-5422 [NI]
200 MS/s, 16-Bit Arbitrary Waveform Generator; 200 MS / s的16位任意波形发生器型号: | PXI-5422 |
厂家: | National Instruments |
描述: | 200 MS/s, 16-Bit Arbitrary Waveform Generator |
文件: | 总3页 (文件大小:370K) |
中文: | 中文翻译 | 下载: | 下载PDF数据表文档文件 |
200 MS/s, 16-Bit Arbitrary Waveform Generator
NEW
NI PXI-5422
• 16-bit resolution, 200 MS/s
sampling rate
Operating Systems
• Windows 2000/NT/XP
• 8, 32, 256, or 512 MB of
onboard memory
Recommended Software
• NI Modulation Toolkit
• LabVIEW
• LabWindows/CVI
• Measurement Studio
• 80 MHz analog bandwidth
• Multimodule synchronization
with <20 psrms skew
• Time and frequency-domain
waveform creation software
• Function generator emulation mode
for easy standard waveform generation
• External clock and reference inputs
• 16-bit LVDS digital pattern output
(32, 256, and 512 MB models)
Driver Software (included)
• NI-FGEN
• LabVIEW Express VIs
• FGEN Soft Front Panel
• NI Analog Waveform Editor
(with 32, 265 and 512 MB models)
Calibration
• Gain, offset and timing
self-calibration
• 2 year external calibration cycle
Applications
Communications
Digitally modulated I/Q signals
Direct to IF modulated signals
Semiconductor
Imaging sensors (CCD, CMOS)
Display devices (LCD)
Biomedical devices
Mixed signal devices
Automotive
Telematics systems
Antilock brake systems
Aerospace/Defense
Avionics
Description
Timing and Synchronization
The National Instruments PXI-5422 is a 200 MS/s arbitrary waveform • Multi-instrument synchronization
generator featuring 16-bit resolution and up to 512 MB of onboard
memory in a compact, 1 slot 3U PXI module. Because the NI PXI-5422 • 3 sample clock sources – Divide-by-N,
uses the PCI-based PXI platform, waveforms can be downloaded at up to High-Resolution, and External.
84 MB/s, far faster than traditional GPIB-based instruments. Using the • 1.06 µHz sample rate resolution
Synchronization and Memory Core (SMC) architecture of the PXI-5422, with high-resolution clock
you can create mixed signal test systems by synchronizing the generator • Phase lock to external reference or
with <20 psrms of skew
with digitizers and digital waveform generator/analyzers or synchronize
multiple arbitrary waveform generators to form a phase-coherent
multichannel generator.
the PXI 10 MHz reference clock
Radar
Scientific Research
Analytical instruments
Software
• NI Analog Waveform Editor for
Exceptional Time and Frequency-Domain Performance
• 1.8 ns rise time, 6% pulse aberration
creating frequency and time-domain signals
• NI-FGEN instrument driver with LabVIEW Express VIs
for function and arbitrary waveform generation
• FGEN Soft Front Panel for interactive control
• -60 dBc (0.1%) total harmonic distortion (THD) at 10 MHz
• -139 dBm/Hz average noise density
• -81 dBc intermodulation distortion (IMD) with intermediate
frequency (IF) optimized direct path
• < 1.3 psrms jitter -138 dBc/Hz phase noise (10 MHz carrier,
10 kHz offset)
Ordering Information
NI PXI-5422 ..............................................................779087-0M1
Includes SMB112 cable, NI-FGEN, and FGEN Soft Front Panel.
32, 256, and 512 MB models include NI Analog Waveform Editor.
1M (onboard memory): 1 (8 MB), 2 (32 MB), 3 (256 MB), 4 (512 MB)
Triggering and Sequencing
• Four triggering modes – single, continuous, stepped, and burst
Recommended PXI Switch
NI PXI-2593 ..................................................................778793-01
• Up to 2 million waveform segments
• Up to 3 million sequence instructions (links and loops)
• Segment looping up to 16,777,216 times or infinitely
BUY NOW!
For complete product specifications, pricing, and accessory
information, call (866) 265-9891 (U.S. only) or go to ni.com
and search on a four-digit model number listed above.
200 MS/s, 16-Bit Arbitrary Waveform Generator
Specifications
General
Onboard Clock (Internal VCXO)
Frequency accuracy.......................................... 25 ppm
Number of channels.........................................
1
DAC resolution ................................................. 16 bits
Maximum sampling rate .................................. 200 MS/s
Output paths..................................................... 1. Main Output Path setting with driver selected Low
Gain Amplifier or High Gain Amplifier
PLL reference clock sources............................. PXI_CLK10, CLK IN
Digital Data and Control, DDC (optional front panel connector)
Data output signals.......................................... 16 LVDS data lines (ANSI/TIA/EIA-644 compliant)
2. Direct Path optimized for IF applications
Start Trigger
Recommended Maximum Output Frequencies
Sources ............................................................ PFI <0:3>, PXI_TRIG<0:6>, PXI Star Trigger, Software, Immediate
Direct Path................................................. 80 MHz
Low Gain Path ........................................... 80 MHz
High Gain Path........................................... 40 MHz up to 12 Vpp (80 MHz up to 8 Vpp
Modes............................................................... Single, Continuous, Stepped, Burst
Markers
)
Destinations ..................................................... PFI <0:1>, PFI <4:5>, PXI_TRIG <0:6>
Analog Output
Quantity............................................................ 1 Marker per Segment
Amplitude range (full scale)
Waveform and Instruction Memory Utilization
Main output path....................................... 12 Vpp to 5.64 mVpp (50
Direct path................................................. 1 Vpp to 0.707 Vpp (50
Ω
load)
Ω
load)
8 MB Standard
32 MB Option
256 MB Option
512 MB Option
Offset range ..................................................... 50ꢀ of Amplitude Range (Signal plus offset not to
exceed amplitude range)
Onboard Memory Size 8,388,608 bytes 33,554,432 bytes 268,435,456 bytes 536,870,912 bytes
Output modes................................................... Function Generator Emulation, Arbitrary Waveform
and Arbitrary Sequence
Loop count........................................................ 1 to 16,777,215. Burst Trigger: Unlimited
Output impedance............................................ 50
DC Accuracy ..................................................... 0.4ꢀ of Amplitude 0.05ꢀ of offset 1 mV
AC amplitude accuracy .................................... 1.0ꢀ of Amplitude 1 mV at 50 kHz
Ωor 75Ω, software selectable
Output filter...................................................... Software enabled seven-pole elliptical analog filter
available on Main Output Path
Memory Limits
Arbitrary Waveform
Mode Maximum
Waveform Memory
Arbitrary Sequence
Mode Maximum
Waveform Memory1
Arbitrary Sequence
Mode Maximum
Waveforms1
Arbitrary Sequence
8 MB Standard
4,194,176
Samples
32 MB Option
16,777,088
Samples
256 MB Option
134,217,600
Samples
512 MB Option
268,435,328
Samples
Low Gain Passband Flatness (Typical)
5.000
4,194,120
Samples
16,777,008
Samples
134,217,520
Samples
268,435,200
Samples
0.000
-5.000
65,000
Samples
262,000
Samples
2,097,000
Samples
4,194,000
Samples
-10.000
-15.000
-20.000
-25.000
104,000
Samples
418,000
Samples
3,354,000
Samples
6,708,000
Samples
Mode Maximum
Segments in a Sequence2
Refer to detailed specifications for all trigger modes.
1Condition One or two segments in a sequence
2Condition: Waveform memory is <4,000 samples.
0.00
10.00
20.00
30.00
40.00
50.00
60.00
70.00
80.00
90.00
100.00
Frequency (MHz)
Power
Rise time .......................................................... < 1.8 ns for Main Output Low Gain Path
Fall time............................................................ < 2.1 ns for Main Output Low Gain Path
Pulse Aberration............................................... < 6 ꢀ for Main Output Low Gain Path
+3.3 VDC
2.0 A
2.0 A
+5 VDC
2.0 A
3.0 A
+12 VDC
0.46 A
0.46 A
-12 VDC
0.01 A
0.01 A
Total Power
22.2 W
27.2 W
Typical operating conditions
Maximum overload
Spectral Characteristics
Frequency
Direct Path
Low Gain Path
Comments
Total Harmonic Distortion
(THD)
Amplitude –1
dBFS
Physical
20 kHz
1 MHz
-85 dBc (0.006ꢀ) -81 dBc (0.009ꢀ)
Dimensions....................................................... Single 3U PXI slot
-87 dBc
-67 dBc
-80 dBc
-65 dBc
-60 dBc
-63 dBc
-60 dBc
-78 dBc
-63 dBc
-55 dBc
2nd through 6th
harmonics
Each tone is
-7 dBFS. 200kHz
spacing
Front panel connectors
10 MHz
10 MHz
40 MHz
70 MHz
CH0 ............................................................ SMB (Jack)
CLK IN........................................................ SMB (Jack)
PFI 0 ........................................................... SMB (Jack)
PFI 1 ........................................................... SMB (Jack)
Intermodulation Distortion
(IMD)
Digital data and control ................................... 68-pin VHDCI Female Receptacle
Typical
Environment
Average Noise Density
Operating temperature..................................... 0 to +55 °C (Meets IEC-60068-2-1 and IEC-60068-2-2)
Storage temperature........................................ -25 to +85 °C (Meets IEC-60068-2-1 and IEC-60068-2-2)
Relative humidity ............................................. 10 to 90ꢀ, noncondensing (Meets IEC 60068-2-56)
Amplitude Range
Average Noise Density
nV/√Hz dBm/Hz dBfs/Hz
Path
Vpk-pk
1
0.1
2
dBm
4
-16.0
10.0
25.6
Direct Path
Low Gain
Low Gain
High gain
25
14
45
-139
-144
-134
-118
-143.0
-128.0
-144.0
-143.6
Calibration
Self-calibration................................................. DC gain and offset
External calibration interval............................. 2 years
12
282
Sample Clock
Certifications and Compliances
Sources ............................................................ Internal Divide-by-N, Internal High-Resolution, External CLK
IN, External DDC Clk In, PXI star Trigger, PXI_TRIG <0:7>
Frequency resolution
Divide-by-N................................................ (200 MS/s) / N where 1
High Resolution ......................................... 1.06 µHz
CE Mark Compliance
Note
≤N 40
Unless otherwise noted, the following conditions were used for each specification:
A. Analog filter enabled
B. Signals terminated with 50
C. Direct path set to 1 Vpk-pk, Low Gain Amplifier Path set to 2 Vpk-pk,
and High Gain Amplifier Path set to 12 Vpk-pk
Ω
System Phase Noise and Jitter
System Phase
Noise Density
System Output
Jitter
< 1.3 ps rms
< 3.0 ps rms
< 1.1 ps rms
Comment
10 MHz carrier
D. Sample clock set to 200 MS/s
Divide-by-N
High Resolution
External Clock
-138 dBc/Hz (10 kHz offset)
-122 dBc/Hz (10 kHz offset)
-143 dBc/Hz (10 kHz offset)
Specifications subject to change without notice. Please see detailed specifications for more information.
2
National Instruments • Tel: (800) 433-3488 • Fax: (512) 683-9300 • info@ni.com • ni.com
NI Services and Support
NI has the services and support to meet your
needs around the globe and through the
application life cycle – from planning
and development through deployment
and ongoing maintenance. We offer
services and service levels to meet
customer requirements in research,
design, validation, and manufacturing.
Visit ni.com/services.
Local Sales and Technical Support
In offices worldwide, our staff is local to the country, giving you
access to engineers who speak your language. NI delivers industry-
leading technical support through online knowledge bases, our
applications engineers, and access to 14,000 measurement and
automation professionals within NI Developer Exchange forums.
Find immediate answers to your questions at ni.com/support.
We also offer service programs that provide automatic upgrades to
your application development environment and higher levels of
technical support. Visit ni.com/ssp.
Training and Certification
NI training is the fastest, most certain route to productivity with our Hardware Services
products. NI training can shorten your learning curve, save NI Factory Installation Services
development time, and reduce maintenance costs over the NI Factory Installation Services (FIS) is the fastest and easiest way to
application life cycle. We schedule instructor-led courses in cities use your PXI or PXI/SCXI™ combination systems right out of the
worldwide, or we can hold a course at your facility. We also offer a box. Trained NI technicians install the software and hardware and
professional certification program that identifies individuals who configure the system to your specifications. NI extends the standard
have high levels of skill and knowledge on using NI products. warranty by one year on hardware components (controllers, chassis,
Visit ni.com/training.
modules) purchased with FIS. To use FIS, simply configure your
system online with ni.com/pxiadvisor.
Professional Services
Our Professional Services Team is comprised of NI applications Calibration Services
engineers, NI Consulting Services, and a worldwide NI Alliance NI recognizes the need to maintain properly calibrated devices for
Partner Program of more than 600 independent consultants and high-accuracy measurements. We provide manual calibration
integrators. Services range procedures, services to recalibrate your products, and automated
from start-up assistance to calibration software specifically designed for use by metrology
turnkey system integration. laboratories. Visit ni.com/calibration.
Visit ni.com/alliance.
Repair and Extended Warranty
OEM Support
NI provides complete repair services for our products. Express repair
We offer design-in consulting and product integration assistance and advance replacement services are also available. We offer
if you want to use our products for OEM applications. For extended warranties to help you meet project life-cycle requirements.
information about special pricing and services for OEM customers, Visit ni.com/services.
visit ni.com/oem.
ni.com • (800) 433-3488
National Instruments • Tel: (512) 683-0100 • Fax: (512) 683-9300 • info@ni.com
© 2004 National Instruments Corporation. All rights reserved. Product and company names listed are trademarks or trade names of their respective companies.