PXI-4204 [NI]
DAQ for High-Voltage Measurements 16-Bit, 200 kS/s, 8 Analog Inputs; DAQ用于高电压测量16位, 200 kS / s的, 8个模拟输入型号: | PXI-4204 |
厂家: | National Instruments |
描述: | DAQ for High-Voltage Measurements 16-Bit, 200 kS/s, 8 Analog Inputs |
文件: | 总4页 (文件大小:109K) |
中文: | 中文翻译 | 下载: | 下载PDF数据表文档文件 |
DAQ for High-Voltage Measurements
16-Bit, 200 kS/s, 8 Analog Inputs
New
NI PXI-4204
• 8 analog inputs at 200 kS/s,
1ꢀ-bit resolution
Operating Systems
• Windows 2000/NT/XP
•
100 V input range
Recommended Software
• Programmable gain (0.5, 1, 10, 100)
per channel
• Programmable 2-pole Butterworth
filter (ꢀ Hz or 10 kHz) per channel
• Differential simultaneous
sampling inputs
• Combicon screw terminal
direct connectivity
• NI-DAQ driver simplifies
configuration and measurements
• LabVIEW
• LabWindows/CVI
• Measurement Studio for
Visual Studio.NET
Other Compatible Software
• Visual Basic, C/C++, and C#
Driver Software (Included)
• NI-DAQ 7
Device
NI PXI-4204
Bus
Analog Inputs
8 DI
Resolution
Sampling Rate
200 kS/s
Input Range
Triggers
Filter Settings
2-pꢁle Butterwꢁrth
(6 Hz ꢁr 10 kHz)
PXI
16 bits
0ꢀ. tꢁ 100 ꢂ
Digital (2)
Table 1. NI PXI-4204 Channel, Speed, and Resolution Specifications
Driver Software
Overview and Applications
NI-DAQ 7 is the robust driver software included with all
National Instruments data acquisition and signal conditioning
products. This easy-to-use software tightly integrates the full
functionality of your DAQ hardware to LabVIEW, LabWindows/CVI,
and Measurement Studio for Visual Studio.NET. High-performance
features include multidevice synchronization, networked
measurements, and DMA data management. Bundled with
NI-DAQ 7, the Measurement & Automation Explorer (MAX) utility
simplifies the configuration of your measurement hardware with
device test panels, interactive measurements, and scaled I/O
channels. NI-DAQ also provides numerous example programs for
LabVIEW and other application development environments to get
you started with your application quickly.
The National Instruments PXI-4204 module delivers accurate data
acquisition for input ranges up to 100 V. With this module you can
handle a broad variety of applications including:
• 42 V automotive applications
• High-voltage, multichannel data logging applications
• Fuel cell and battery test applications
Features
The NI PXI-4204 is a full-featured data acquisition module with a
100 V input range, 1ꢀ-bit accuracy, and software-selectable filter
and gain settings per channel. Programmable filter and gain settings
ensure that the PXI-4204 achieves maximum accuracy over the
entire 100 V input range. In addition, the PXI-4204 is designed to
work with LabVIEW 7 Express and NI-DAQmx. The DAQ Assistant
in LabVIEW 7 can configure the PXI-4204 and acquire data through
a menu-based window, eliminating the need to manually program
the device.
For high-channel-count, high-voltage applications, explore NI signal
conditioning hardware. Visit ni.com/sigcon
Ordering Information
NI PXI-4204 ......................................................778745-4204
Includes NI-DAQ driver software.
For more information on extended warranty and value-
added services, visit ni.com/services
DAQ for High-Voltage Measurements
16-Bit, 200 kS/s, 8 Analog Inputs
Specifications – PXI-4204
Complete Accuracy Tables
Input Characteristics
SSH Disabled
Absolute Accuracy
Noise + Quantization (mV)
Absolute Accuracy
at Full Scale
Percent Error
at Full Scale
(ꢀ)
Nominal Range (V)
ꢀ of Reading
24 Hrs
Offset
(mV)
16ꢀ8
10ꢀ4
4ꢀ7
Temp
Drift (ꢀ/°C)
0ꢀ0027
0ꢀ002.
0ꢀ0027
Positive FS
Negative FS
1 Year
0ꢀ070
0ꢀ038
0ꢀ068
0ꢀ068
Single Pt.
9ꢀ34
Averaged
0ꢀ92
(mV)
88
30
8ꢀ.
4ꢀ8
100
.0
.
-100
-.0
-.
0ꢀ063
0ꢀ031
0ꢀ061
0ꢀ061
0ꢀ088
0ꢀ060
0ꢀ17
0ꢀ97
4ꢀ68
0ꢀ.7
0ꢀ69
0ꢀ40
0ꢀ.
-0ꢀ.
4ꢀ1
0ꢀ49
0ꢀ40
0ꢀ0027
SSH Enabled
Nominal Range (V)
Positive FS
100
.0
.
Absolute Accuracy
Absolute Accuracy
at Full Scale
Percent Error
at Full Scale
(ꢀ)
ꢀ of Reading
Offset
(mV)
19ꢀ0
12ꢀ6
6ꢀ8
Noise + Quantization (mV)
Single Pt.
9ꢀ3.
4ꢀ70
0ꢀ82
Temp
Drift (ꢀ/°C)
0ꢀ0027
0ꢀ002.
0ꢀ0027
Negative FS
24 Hrs
0ꢀ08.
0ꢀ03.
0ꢀ063
0ꢀ063
1 Year
0ꢀ090
0ꢀ042
0ꢀ070
0ꢀ070
Averaged
1ꢀ02
0ꢀ73
0ꢀ60
0ꢀ60
(mV)
110
34
11
7ꢀ2
-100
-.0
-.
0ꢀ110
0ꢀ069
0ꢀ22
1ꢀ4
0ꢀ.
-0ꢀ.
6ꢀ3
0ꢀ66
0ꢀ0027
Note: Accuracies are valid for measurements following an internal calibration.
Averaged numbers assume dithering and averaging of 100 single-channel
readings. Measurement accuracies are listed for operational temperatures within
1ꢁC of internal calibration temperature and 10ꢁC of external or factory
calibration temperature.
CMRR
Module
PXI-4204
PXI-4204
Input Frequency Ranges
DC tꢁ 60 Hz
CMRR
>60 dB
>40 dB
100 Hz tꢁ 10 kHz
Filter Characteristics
Number of Channels
Filter Type .................................................. ꢀ Hz or 10 kHz, 2-pole, Butterworth
Stability
PXI-4204
8 differential
Module
PXI-4204
Input Range Gain Temperature Coefficient Offset Temperature Coefficient
100 ꢂ
2. ppm/Cꢁ 38. uꢂ/Cꢁ
Input signal ranges..................................... see accuracy tables
Resolution.................................................. 1ꢀ bits, 1 in ꢀ5,53ꢀ
Sampling Rate
Triggers
Maximum Sampling Rate .......................... 200 kS/s aggregate multi-channel
Digital Triggers
Number of triggers...............................
2
Max per channel (scanning all channels)
Purpose ................................................ Start & stop trigger, gate, clock
Source .................................................. PFI0/TRIG1 (front SMB connector),
PXI_TRIG_0…PXI_TRIG_ꢀ
Mꢁdule
PXI-4204
Single Channel
333 kS/s
Scan Rate w/SSH*
20 kS/s/ch
Scan Rate w/ꢁ SSH*
2. kS/s/ch
*SSH = Simultaneꢁus Sample and Hꢁld
(PXI trigger bus)
Slope.......................................................... Positive or negative; software
selectable
Connector
Screw Terminals
SMB Cꢁnnectꢁr
Powered On
100 ꢂDC
1. ꢂ
Powered Off
100 ꢂDC
1. ꢂ
Compatibility......................................... 5V/TTL
Response.............................................. Rising or falling edge
Pulse width........................................... 10 ns minimum
Impedance............................................ 10 kOhm
Coupling................................................ DC
Overvoltage protection
Inputs with overvoltage protection............ CH<0..7>
Input coupling ............................................ DC
Data transfers ............................................ DMA, interrupts, programmed I/O
PXI Trigger Bus
Trigger lines ..........................................
Star trigger............................................
ꢀ
1
Transfer Characteristics
Nonlinearity
Bus Interface
Range
100 ꢂ
Other Ranges
100 ꢂ, SSH Enabled
Other Ranges, SSH Enabled
Percent of Full Scale (Typical)
PXI……….……………………….Master, slave
0ꢀ02%
0ꢀ01%
0ꢀ06%
0ꢀ02%
Power Requirements
Module
PXI-4204
5 VDC ( 5ꢀ)
1 A
DNL
Module
PXI-4204
Typical
0ꢀ. LSB
Maximum
1 LSB
Calibration
Recommended warm-up time ............. 15 minutes
External calibration interval................... 1 year
Onboard calibration reference
Gain Error................................................... See Accuracy Table
Offset Error................................................ See Accuracy Table**
Level................................................ 5.000V +/- 1mV
Temperature coefficient.................. + 0.ꢀ ppm/oC
Long-term stability .......................... ꢀ ppm/Sqrt(1000h)
Amplifier Characteristics
Input Impedance
Module
PXI-4204
Input Configuration
Differential Input
Single-Ended
Normal Powered On
Powered Off
2MΩ
1MΩ
Physical
2 MΩ
1 MΩ
Dimensions (not including
connectors).......................................... 1ꢀ.0 by 10.0 cm by 2.0 cm
(ꢀ.3 x 3.9 x 0.79 in)
NMR (ꢀ0 Hz)
Analog input signal connector.................. 1ꢀ x 1 minicombicon, 3.81 mm pitch
Analog input signal mating connector ..... 1ꢀ x 1 minicombicon, 3.81 mm pitch,
28-1ꢀ AWG signal wire
PXI-4204
40 dB @ 6 Hz filter setting
**Gain Error = Actual Input Voltage * % of Reading
2
National Instruments • Tel: (800) 433-3488 • Fax: (512) 683-9300 • info@ni.com • ni.com
DAQ for High-Voltage Measurements
16-Bit, 200 kS/s, 8 Analog Inputs
Specifications
I/O Connectors
Module
Front
PXI-4204
16 screw terminals, 1 SMB
Environment:
Operating temperature......................... 0 to 55 oC
Storage temperature ............................ –20 to 70 oC
Relative humidity.................................. 10 to 90%, non-condensing
Maximum altitude ................................ 2,000m
Pollution degree (indoor use only)........
2
Certification and Compliance
European Compliance
Emissions................................................... EN 55011 Class A 10 meters
Immunity.................................................... EN ꢀ132ꢀ:1997 + A2:2001, Table 1
EMC........................................................... CE, C-Tick and FCC Part 15 (Class A)
Compliant
Safety......................................................... IEC ꢀ1010-1, EN ꢀ1010-1
North American Compliance
Emissions................................................... FCC Part 15A above 1 GHz
Immunity.................................................... EN ꢀ132ꢀ:1997 + A2:2001, Table 1
EMC........................................................... CE, C-Tick and FCC Part 15 (Class A)
Compliant
Safety......................................................... IEC ꢀ1010-1, UL 3111-1, UL ꢀ1010B-1,
CAN/SA C22.2 No. 1010.1
National Instruments • Tel: (800) 433-3488 • Fax: (512) 683-9300 • info@ni.com • ni.com 3
Global Services and Support
NI has the services and support to meet your needs around the Professional Services
globe and through the application life cycle – from planning and
development through deployment and ongoing maintenance –
and they are tailored for customer requirements in research,
design, validation, and manufacturing. We have direct operations
in more than 37 countries and distributors in another 12 locations.
Our local sales and support representatives are degreed
engineers, ready to partner with you to find solutions that best fit
your needs.
Our Professional Services team consists of National Instruments
Applications Engineers, NI Consulting Services, and a worldwide
Alliance Program (a network of ꢀ00 independent consultants and
integrators). Our Professional Services team can offer services
ranging from basic start-up assistance and collaborative development
with your engineers, to turnkey
system integration and
maintenance of your system.
A l l i a n c e M e m b e r
In addition to our NI Alliance
Local Sales and Technical Support
Partners, we have developed global relationships with many industry
partners that range from computer software and hardware
companies, such as Microsoft, Dell, Siemens, and Tektronix. By
collaborating with these companies, you receive a complete
spectrum of solutions – from components to turnkey systems. Find
the Alliance Program member directory at ni.com/alliance
In offices around the globe, our staff is local to the country so that
you have access to field engineers who speak your language and
are available to consult on your unique needs. We also have a
worldwide support organization staffed with Applications
Engineers trained to quickly provide superior technical assistance.
Use our online Request Support interface (ni.com/support) to
define your question, then speak to or e-mail an Applications Product Services
Engineer, or access more than 14,000 worldwide measurement
and automation professionals within NI Developer Exchange
Discussion Forums. ni.com/support also provides immediate
answers to your questions through self-help troubleshooting,
product reference, and application development resources.
For advanced technical support and software
NI GPIB products are warranted against defects in workmanship
and material for one year from the date of shipment. To help you
meet project lifecycle requirements, NI offers extended
warranties for an additional charge. NI provides complete repair
services for our products. Express repair and advanced
replacement services are also available. Or, order your software
and hardware installed in PXI and PXI/SCXI™ systems with
NI Factory Installation Services.
maintenance services, sign up for
I
Premier Support, a program that
provides expanded hours of support
availability and expedited phone
and e-mail response time (typically
four business hours).
Ordering Made Easy
Visit ni.com/products to browse product specifications, make
comparisons, or access technical representatives via online chat
or telephone. Worldwide customers can use a purchase order or
credit card to buy in local currency and receive direct shipments
from local NI offices. Our North American Customer Service
Representatives are available Monday through Friday between
7 a.m. and 7 p.m. Central Time. Outside North America, please
contact the NI office in your country.
D
Y O
L
E
P
Training and Certification
NI recognizes that both initial instruction and
ongoing education contribute to your success.
NI provides a variety of training alternatives, from self-paced tutorials
and interactive CDs, to worldwide hands-on courses taught by
experienced instructors – all designed so that you can choose how to
learn about our products. Further, NI offers certifications Order Status and Service Requests
acknowledging individual expertise in working with NI products and National Instruments brings you real-time status on current orders at
technologies. Visit ni.com/training for more information.
ni.com/status Similarly, find out the status of open technical support
incidents or hardware repair requests at ni.com/support/servicereq
(800) 433-3488
ni.com
National Instruments • Tel: (512) 683-0100 • Fax: (512) 683-9300 • info@ni.com
C This document represents a commitment from National Instruments to the environment.
© 2003 National Instruments Corporation. All rights reserved. Product and company names listed are trademarks or trade names of their respective companies.