PXI-4072 [NI]

61⁄2-Digit Digital Multimeter, 1.8 MS/s Isolated Digitizer, and LCR Meter; 61/2位数字万用表, 1.8 MS / s的隔离数字化仪和LCR表
PXI-4072
型号: PXI-4072
厂家: National Instruments    National Instruments
描述:

61⁄2-Digit Digital Multimeter, 1.8 MS/s Isolated Digitizer, and LCR Meter
61/2位数字万用表, 1.8 MS / s的隔离数字化仪和LCR表

文件: 总5页 (文件大小:718K)
中文:  中文翻译
下载:  下载PDF数据表文档文件
1
6 ⁄ -Digit Digital Multimeter, 1.8 MS/s  
2
Isolated Digitizer, and LCR Meter  
NI PXI-4072  
• Multifunction device  
Operating Systems  
• Windows Vista/XP/2000  
• Linux®  
1
• 6 ⁄  
2-digit digital multimeter  
• 1.8 MS/s isolated digitizer  
• LCR meter (inductance,  
capacitance, and resistance)  
• 20 built-in measurements  
• 10- to 23-bit flexible resolution  
• Built-in self-calibration  
Recommended Software  
• LabVIEW  
• LabVIEW Real-Time Module  
• LabWindows/CVI  
• LabVIEW SignalExpress  
300 VDC/Vrms isolation  
• 425 Vp AC measurements  
Driver Software (included)  
• NI-DMM driver  
• DMM Express VI for LabVIEW  
• DMM Soft Front Panel  
Calibration  
• Gain, temperature, and offset  
self-calibration  
• 2-year external calibration cycle  
Mode  
Capacitance  
Ranges  
300 pF, 1 nF, 10 nF, 100 nF, 1 µF, 10 µF  
100 µF, 1000 µF, 10,000 µF  
10 µH, 100 µH  
Reading Rate (S/s)  
Overview  
20  
3
40  
20  
3
The National Instruments PXI-4072 FlexDMM, based on the NI 4070  
architecture, combines the functionality found in three common test  
instruments – a digital multimeter, an LCR meter, and a digitizer. This  
functionality provides engineers with the 20 most common test functions  
in a compact 3U PXI module, including voltage, current, capacitance,  
inductance, temperature, and resistance. Integrating these measurements  
in a PXI module reduces test system size and cost, increases throughput,  
and shortens test development time, making the NI PXI 4072 an  
excellent fit for use in automated tests on both the production floor and  
Inductance  
1 mH, 10 mH  
100 mH, 1 H, 5 H  
Table 1. PXI-4072 Key Specifications  
in an R&D environment.  
1
The PXI-4072 offers a DC reading rate from 10 kS/s at 4 ⁄  
2
digits  
to 5 S/s at 7 digits. As a digitizer, the FlexDMM can acquire both AC-  
and DC-coupled voltage and current waveforms up to 300 V and 1 A  
input at a maximum sampling rate of 1.8 MS/s. As an LCR meter, the  
FlexDMM delivers 0.25 percent basic accuracy for both inductance and  
capacitance at measurement rates up to 40 S/s. Using the intuitive  
NI-DMM instrument driver, you can quickly change from function to  
function in a single application programming interface (API).  
Ordering Information  
NI PXI-4072............................................................................778270-01  
Includes the P-1 Probe Set, NI-DMM, DMM Express VI, and DMM Soft Front Panel.  
Recommended Switching and Accessories  
NI PXI-2527  
300 V multiplexer switch ................................................ 778572-27  
NI PXI-2530  
128x1 multiplexer switch..................................................778660-01  
P-1 Probe Set (standard probe) ............................................761000-01  
P-2 Probe Set (additional probe) ..........................................184698-01  
P-3 Probe Set (banana plug to bare wire) ............................185692-01  
10 A current shunt, CSM-10A ..............................................777488-02  
BUY NOW!  
For complete product specifications, pricing, and accessory information,  
call 800 813 3693 (U.S.) or go to ni.com/dmm.  
1
6 ⁄ -Digit Digital Multimeter, 1.8 MS/s Isolated Digitizer, and LCR Meter  
2
Specifications  
Specifications subject to change without notice.  
For the most current and complete specifications, visit ni.com/modularinstruments.  
DC Functions  
DC Voltage ±(ppm of reading ꢀ ppm of range)  
Tempco/°C  
Without  
Self-Cal  
4 + 5  
2 + 1  
1 + 1  
With  
Input  
24 Hr  
Tcal ±1 °C  
10 + 10  
6 + 2  
4 + 2  
6 + 2  
90-Day  
Tcal ±± °C  
30 + 20  
20 + 6  
20 + 6  
30 + 6  
2-Year  
Tcal ±± °C  
40 + 20  
25 + 6  
25 + 6  
35 + 6  
Self-Cal  
0.3 + 0.3  
0.3 + 0.3  
0.3 + 0.3  
0.3 + 0.3  
0.3 + 0.3  
Range  
100 mV  
1 V  
Resolution  
100 nV  
1 µV  
Resistance  
>10 GΩ,10 MΩ  
>10 GΩ,10 MΩ  
>10 GΩ,10 MΩ  
10 MΩ  
10 V  
10 µV  
100 V  
300 V  
100 µV  
1 mV  
4 + 1  
4 + 3  
10 MΩ  
6 + 6  
30 + 20  
35 + 20  
DC Current ±(ppm of reading ꢀ ppm of range)  
Range  
20 mA  
200 mA  
1 A  
Resolution  
10 nA  
100 nA  
1 µA  
Burden Voltage (typical)  
<20 mV  
Noise (ppm of range rms)  
2-Year  
Tempco/°C  
20  
3
3
400 + 75  
400 + 20  
500 + 20  
8 + 1  
8 + 0.2  
8 + 0.4  
<200 mV  
<800 mV  
Resistance (4- and 2-wire) ±(ppm of reading ꢀ ppm of range)  
Tempco/°C  
Without  
Self-Cal  
8 + 1  
8 + 0.1  
8 + 0.1  
8 + 0.5  
8 + 1  
30 + 3  
200 + 10  
With  
Self-Cal  
0.8 + 1  
0.8 + 0.1  
0.8 + 0.1  
0.8 + 0.5  
0.8 + 1  
Max Test  
Voltage  
100 mV  
1 V  
1 V  
1 V  
10 V  
10 V  
10 V  
24 Hr  
Tcal ±1 °C  
15 + 10  
12 + 2  
12 + 2  
15 + 2  
20 + 2  
100 + 2  
900 + 20  
90-Day  
Tcal ±± °C  
50 + 10  
50 + 3  
50 + 3  
50 + 6  
60 + 10  
200 + 10  
1,800 + 40  
2-Year  
Tcal ±± °C  
80 + 10  
80 + 3  
80 + 3  
80 + 6  
90 + 10  
400 + 10  
2,000 + 40  
Range  
100 Ω  
1 kΩ  
Resolution  
100 µΩ  
1 mΩ  
Test Current  
1 mA  
1 mA  
10 kΩ  
100 kΩ  
1 MΩ  
10 mΩ  
100 mΩ  
1 Ω  
100 µA  
10 µA  
10 µA  
10 MΩ  
100 MΩ  
10 Ω  
100 Ω  
1 µA  
1 µA || 10 MΩ  
30 + 3  
200 + 10  
Diode Test  
Range  
Resolution  
Test Current  
Accuracy  
10 V  
10 µV  
1 µA, 10 µA, 100 µA, 1 mA  
Add 20 ppm of reading to 10 VDC voltage specifications  
AC Functions  
Digits  
6 ⁄2  
Reading Rate  
0.25 S/s  
2.5 S/s  
25 S/s  
100 S/s  
Bandwidth  
1 Hz to 300 kHz  
10 Hz to 300 kHz  
100 Hz to 300 kHz  
400 Hz to 300 kHz  
20 to 300 kHz  
1
1
6 ⁄  
2
1
6 ⁄  
2
1
6 ⁄  
2
1
5 ⁄  
2
1.0 kS/s  
AC Voltage 2-Year ± (% of reading ꢀ % of range), 23 °C ±± °C  
Range (rms)  
50 mV  
500 mV  
5 V  
Peak Voltage  
105 mV  
1.05 V  
Resolution  
100 nV  
1 µV  
1 to 40 Hz  
0.1 + 0.04  
40 Hz to 20 kHz 20 to ±0 kHz ±0 to 100 kHz  
100 to 300 kHz  
0.05 + 0.04  
0.09 + 0.04  
0.5 + 0.08  
3 + 0.1  
0.1 + 0.01  
0.05 + 0.02  
0.09 + 0.02  
0.5 0.02  
3 + 0.05  
10.5 V  
10 µV  
50 V  
300 V  
105 V  
450 V  
100 µV  
1 mV  
Tempco/°C  
0.001 + 0.001  
0.001 + 0.001  
0.001 + 0.001 0.001 + 0.001  
0.01 + 0.01  
BUY ONLINE at ni.com or CALL 800 813 3693 (U.S.)  
2
1
6 ⁄ -Digit Digital Multimeter, 1.8 MS/s Isolated Digitizer, and LCR Meter  
2
AC Current 2-Year ±(% of reading ꢀ % of range)  
Range (rms)  
10 mA  
100 mA  
1 A  
Peak Current  
20 mA  
200 mA  
2 A  
Resolution  
10 nA  
100 nA  
1 µA  
Burden Voltage (rms)  
<10 mV  
1 Hz to 20 kHz  
0.04 + 0.02  
0.04 + 0.02  
0.1 + 0.02  
Tempco/°C  
0.001 + 0.0001  
0.001 + 0.0001  
0.001 + 0.0001  
<100 mV  
<800 mV  
Frequency and Period  
Input Range  
50 mV to 300 V  
Frequency Range  
1 Hz to 500 kHz  
Period Range  
1 s to 2 µs  
Resolution  
2-Year Accuracy % of Reading  
1
6 ⁄  
2
digits  
0.01  
Capacitance and Inductance (PXI-4072 only)  
Capacitance ±(% of reading ꢀ % of range), 23 °C ±10 °C  
Range  
300 pF  
1 nF  
10 nF  
100 nF  
1 µF  
Resolution  
0.05 pF  
0.1 pF  
1 pF  
10 pF  
100 pF  
1 nF  
10 nF  
100 nF  
1 µF  
2-year  
Tempco/°C (0 to ±± °C) Effective Test Current  
Effective Frequency  
Default Model  
Parallel  
Parallel  
Parallel  
Parallel  
Series  
0.15 + 0.5  
0.15 + 0.1  
0.15 + 0.1  
0.15 + 0.1  
0.18 + 0.1  
0.18 + 0.1  
0.18 + 0.1  
0.18 + 0.1  
0.18 + 0.1  
0.01 + 0.025  
0.01 + 0.003  
0.01 + 0.001  
0.01 + 0.001  
0.01 + 0.001  
0.01 + 0.001  
0.01 + 0.001  
0.01 + 0.001  
0.01 + 0.001  
160 nA  
330 nA  
330 nA  
3.3 µA  
100 µA  
1 mA  
1 mA  
1 mA  
1 mA  
3 kHz  
3 kHz  
3 kHz  
3 kHz  
1 kHz  
1 kHz  
91 Hz  
91 Hz  
91 Hz  
10 µF  
Series  
Series  
Series  
Series  
100 µF  
1,000 µF  
10,000 µF  
Inductance ±(% of reading ꢀ % of range), 23 °C ±10 °C  
Range  
10 µH  
100 µH  
1 mH  
10 mH  
100 mH  
1 H  
Resolution  
1 nH  
10 nH  
100 nH  
1 µH  
10 µH  
2-year  
0.5 + 1  
Tempco/°C (0 to ±± °C) Effective Test Current  
Effective Frequency  
30 kHz  
Default Model  
Series  
0.01 + 0.01  
0.01 + 0.01  
330 µA  
330 µA  
330 µA  
3.3 µA  
33 µA  
0.2 + 0.1  
0.2 + 0.1  
0.15 + 0.1  
0.15 + 0.1  
0.18 + 0.1  
0.18 + 0.1  
30 kHz  
3 kHz  
3 kHz  
273 Hz  
273 Hz  
273 Hz  
Series  
Series  
Series  
Series  
Series  
Series  
0.01 + 0.001  
0.005 + 0.001  
0.005 + 0.001  
0.007 + 0.001  
0.007 + 0.001  
100 µH  
1 mH  
3.3 µA  
330 nA  
5 H  
Capacitance and Inductance, General Specifications  
Mode  
Capacitance  
Ranges  
300 pF, 1 nF, 10 nF, 100 nF, 1 µF, 10 µF  
100 µF, 1,000 µF, 10,000 µF  
10 µH, 100 µH  
Reading Rate  
20 S/s  
3 S/s  
40 S/s  
20 S/s  
3 S/s  
Inductance  
1 mH, 10 mH  
100 mH, 1 H, 5 H  
Isolated Digitizer Mode  
Acquisition System  
Available sample rates..........................................................  
Variable resolution................................................................. 10 to 23 bits  
1.8 MS/s  
n
5
,
where n = 1, 2, 3,…1.8 x 10  
Available functions ................................................................ Voltage and current  
Voltage ranges....................................................................... 100 mV to 300 V (DC or AC coupled)  
Current ranges ....................................................................... 20 mA to 1 A  
BUY ONLINE at ni.com or CALL 800 813 3693 (U.S.)  
3
1
6 ⁄ -Digit Digital Multimeter, 1.8 MS/s Isolated Digitizer, and LCR Meter  
2
DC Voltage ±(ppm of reading ꢀ ppm of range)  
Range  
100 mV  
1 V  
10 V  
100 V  
300 V  
Input Impedance Flatness Error (20 kHz) Bandwidth (-3 dB) THD 1 kHz Signal (-1 dBFS) THD 1 kHz Signal (-1 dBFS)  
>10 GΩ,1 MΩ  
>10 GΩ,1 MΩ  
>10 GΩ,1 MΩ  
1 MΩ  
-0.03 dB  
-0.03 dB  
-0.03 dB  
-0.03 dB  
-0.03 dB  
300 kHz  
300 kHz  
300 kHz  
300 kHz  
300 kHz  
-104 dB  
-109 dB  
-96 dB  
-96 dB  
-98 dB  
-78 dB  
-83 dB  
-70 dB  
-70 dB  
-72 dB  
1 MΩ  
DC Current ±(ppm of reading ꢀ ppm of range)  
Range  
20 mA  
200 mA  
1 A  
Resolution  
10 nA  
100 nA  
1 µA  
Burden Voltage (typical)  
<20 mV  
Flatness Error (20 kHz)  
0.01 dB  
Bandwidth (-3 dB)  
430 kHz  
<200 mV  
<800 mV  
0.01 dB  
0.01 dB  
430 kHz  
400 kHz  
Certifications and Compliance  
Safety  
IEC 61010-1; EN 61010-1; UL 61010-1; CSA 61010-1  
Installation category.............................................................. II  
Pollution degree.....................................................................  
Electromagnetic compatibility  
2
• EN 61326 EMC requirements; Minimum Immunity  
• EN 55011 Emissions; Group 1, Class A  
• CE, C-Tick, ICES, and FCC Part 15 Emissions; Class A  
BUY ONLINE at ni.com or CALL (800) 813 3693 (U.S.)  
4
NI Services and Support  
NI has the services and support to meet  
Local Sales and Technical Support  
your needs around the globe and through  
the application life cycle – from planning  
and development through deployment  
and ongoing maintenance. We offer  
services and service levels to meet  
customer requirements in research,  
design, validation, and manufacturing.  
Visit ni.com/services.  
In offices worldwide, our staff is local to the country, giving you access  
to engineers who speak your language. NI delivers industry-leading  
technical support through online knowledge bases, our applications  
engineers, and access to 14,000 measurement and automation  
professionals within NI Developer Exchange forums. Find immediate  
answers to your questions at ni.com/support.  
We also offer service programs that provide automatic upgrades to  
your application development environment and higher levels of technical  
support. Visit ni.com/ssp.  
Training and Certification  
NI training is the fastest, most certain route to productivity with our  
products. NI training can shorten your learning curve, save development  
time, and reduce maintenance costs over the application life cycle. We  
schedule instructor-led courses in cities worldwide, or we can hold a  
course at your facility. We also offer a professional certification program  
Hardware Services  
NI Factory Installation Services  
NI Factory Installation Services (FIS) is the fastest and easiest way to  
use your PXI or PXI/SCXI combination systems right out of the box.  
that identifies individuals who have high levels of skill and knowledge on Trained NI technicians install the software and hardware and configure  
using NI products. Visit ni.com/training.  
the system to your specifications. NI extends the standard warranty by  
one year on hardware components (controllers, chassis, modules)  
purchased with FIS. To use FIS, simply configure your system online  
with ni.com/pxiadvisor.  
Professional Services  
Our NI Professional Services team is composed of NI applications  
and systems engineers and a worldwide National Instruments Alliance  
Partner program of more than 600 independent consultants and  
integrators. Services range  
Calibration Services  
NI recognizes the need to maintain properly calibrated devices for  
high-accuracy measurements. We provide manual calibration  
procedures, services to recalibrate your products, and automated  
calibration software specifically designed for use by metrology  
laboratories. Visit ni.com/calibration.  
from start-up assistance to  
turnkey system integration.  
Visit ni.com/alliance.  
OEM Support  
We offer design-in consulting and product integration assistance if you  
want to use our products for OEM applications. For information about  
special pricing and services for OEM customers, visit ni.com/oem.  
Repair and Extended Warranty  
NI provides complete repair services for our products. Express repair  
and advance replacement services are also available. We offer  
extended warranties to help you meet project life-cycle requirements.  
Visit ni.com/services.  
ni.com • 800 813 3693  
*342574B-01*  
342574B-01  
2007-9223-101-D-A  
National Instruments • info@ni.com  
© 2007 National Instruments Corporation. All rights reserved. CVI, FlexDMM, LabVIEW, National Instruments, National Instruments Alliance Partner, NI, ni.com, SCXI,  
®
and SignalExpress are trademarks of National Instruments. The mark LabWindows is used under a license from Microsoft Corporation. Linux is the registered trademark of Linus Torvalds  
in the U.S. and other countries. Other product and company names listed are trademarks or trade names of their respective companies. A National Instruments Alliance Partner is a  
business entity independent from NI and has no agency, partnership, or joint-venture relationship with NI.  

相关型号:

PXI-4110

3 independent DC power supplies/16-bit voltage setpoint and current limit
NI

PXI-4132

100 V Precision Source Measure Unit (SMU)
NI

PXI-4204

DAQ for High-Voltage Measurements 16-Bit, 200 kS/s, 8 Analog Inputs
NI

PXI-5124

150 MHz, 200 MS/s, 12-Bit Digitizer
NI

PXI-5142

14-Bit 100 MS/s Digitizer with Onboard Signal Processing
NI

PXI-5404

9 kHz to 105 MHz sine wave/DC to 100 MHz clock/1.07 μHz frequency resolution
NI

PXI-5422

200 MS/s, 16-Bit Arbitrary Waveform Generator
NI

PXI-5441

100 MS/s, 16-bit Arbitrary Waveform Generator with Onboard Signal Processing
NI

PXI-5670

250 kHz to 2.7 GHz/8, 32 or 256 MB memory
NI

PXI-5671

250 kHz to 2.7 GHz/16-bit resolution, 100 MS/s arbitrary waveform generation
NI

PXI-6052E

Converter IC
ETC

PXI-6509

Low-Cost Industrial Digital I/O-5 V TTL/CMOS
NI