Multi-Purpose Application Board for DMS-30 & DMS-40 Series Meters  
FEATURES  
Accommodates 3½ and 4½ digit meters  
Gain (span) and offset (zero) adjustments for  
DMS-30 meters  
PRODUCT OVERVIEW  
The DMS-EB Application Board covers many common panel meter applications while providing  
simple signal and power I/O terminations (via a standard MOLEX connector) for DATEL's DMS-30 and  
DMS-40 Series, LED/LCD Display, Digital Panel Voltmeters. The unique plug-and-play design of the  
DMS-EB makes changing from application to application and meter to meter quick, easy and inexpensive.  
The highly versatile DMS-EB has provisions for such common applications as attenuation of high  
input voltages, gain (span) and offset (zero) adjustments, 9/12V battery operation, and direct decimal  
point placement. The DMS-EB is compatible with all four versions of the DMS-30 and DMS-40 Series  
( 200mV, 2V, 20V and 200V) making it interchangeable from meter to meter and application to  
application. Solder gaps conveniently placed on the board are easily bridged or cut configuring the board  
to meet each application's specific requirements.  
Input divider network for high voltages  
Operate meter and board from +5V supply or  
9/12V battery  
Solder gaps for decimal placement  
Same size as DMS-30/40 Series meters  
2.02" (51.31mm)L x 0.83" (21.08mm)W  
This low-cost addition to DATEL's expanding line of miniature meters is evidence of our continuing  
commitment to offer the most versatile, easy-to-use, panel meter products.  
SIMPLIFIED SCHEMATIC  
NOTE:  
Components are user-supplied and not part  
of the standard product.  
Figure 1. DMS-EB Schematic Diagram for DMS-30 Meters  
For full details go to  
www.murata-ps.com/rohs  
www.murata-ps.com  
Technical enquiries email: sales@murata-ps.com, tel: +1 508 339 3000  
MPM_DMS-EB.B01 Page 1 of 4  
DMS-EB  
Multi-Purpose Application Board for DMS-30 & DMS-40 Series Meters  
Close SG6 when measuring floating inputs with a 9V-powered meter  
(floating inputs are described in Technical Note 7). SG6 provides a bias  
reference for the input signal by tying LO to the meter's Analog Com-  
mon terminal (pin 10).  
Ordering Information  
DMS-EB-C Application board with mating connector  
and terminals  
DMS-BZL1-C DMS-30/40 bezel assembly  
DMS-BZL2-C DMS-30/40 bezel assembly with  
sealing gasket  
3. Input Resistor Dividers: Use 1%, or better, metal film resistors for  
R1 and R2 in DMS-30 applications, and 0.5%, or better, in DMS-40  
applications. Using tight-tolerance resistors in input divider networks  
allows most final calibration adjustments to be made with the meter's  
rear calibration pots through the access holes provided. Make sure  
the resistor power and voltage ratings are adequate for the given  
application.  
Functional Specifications  
(TA = +25°C)  
Input Supply Range  
See applicable meter's data sheet  
4. Using U1 (LM7805CT): The power supply voltage range listed in the  
Battery Operation section is rated for a 150mA LED meter or a 35mA  
backlit LCD meter. A 10μF/35V tantalum capacitor (C2) should be  
installed with the polarized end next to the + symbol on the DMS-EB.  
This is especially important if the power supply is located far from U1.  
Input Supply Range  
DMS-30/40PC-X-XS  
DMS-30/40PC-X-XL  
DMS-30/40LCD-X-5  
DMS-30/40LCD-X-5B  
U1 (LM7805CT) Installed:  
+7.5 to +12.6V  
+7.5 to +18.0V  
+7.5 to +32.0V  
+7.5 to +12.6V  
Operating and Storage Temperature  
Humidity  
See applicable meter's data sheet  
0 to 95%, non-condensing  
5. DMS-40 Meters: When the DMS-EB is used with DMS-40PC or DMS-  
40LCD meters, be sure to open SG13. The RN-DMS resistor acces-  
sory kit comprised of resistor R3 and potentiometers R4 (Gain/Span)  
and R5 (Offset/Zero), cannot be used with DMS-40 series meters. To  
change the input range on DMS-40LCD meters from the low range to  
the high range, close SG16.  
Dimensions  
2.02" (51.31mm)L x 0.83" (21.08mm)W  
J1 Connector & Wire Information  
Terminal Type  
Crimp Tool  
Wire Size  
Insulation Diameter  
Stripping Length  
DATEL P/N 4400-01032-0  
DATEL P/N 39-2099000  
22 to 26 AWG  
0.062" (1.57mm) max.  
0.100" to 0.125" (2.54 to 3.17mm)  
6. Calibration Potentiometer Adjustment Holes: Two overlapping  
holes on the DMS-EB provide access to the ¾-turn calibration poten-  
tiometer on DMS-30 meters and the 3-turn potentiometer on DMS-40  
meters. The adjustment range of both pots is very narrow. See the  
applicable product data sheet for more information on each meter’s  
calibration potentiometers.  
Introduction  
As shipped, the DMS-EB is configured for 5V-powered DMS-30 3½  
digit meters operating in the single-ended mode (SG5 closed) using the  
meter's internal reference (SG13 closed). This configuration is the most  
common and works well for most simple voltage measurements.  
Many of the applications described below apply to both DMS-30  
and DMS-40 meters. However, Application 5, installing the span (R4) and  
offset (R5) adjust potentiometers, can only be used with DMS-30 meters.  
Also note that some of the other applications require the installation of  
R4 and R5. See Technical Note 5 when using the DMS-EB with DMS-40  
meters.  
7. Input Configurations: The DMS-EB is supplied with SG5 closed. SG5  
connects J1, pin 2 (LO) to power return J1, pin 6 (GND) and configures  
the meter for single-ended operation, that is, the low side of the input  
is at system ground or zero Volts. SG5 must also be closed when  
measuring floating inputs. Floating inputs are signals which have no  
electrical connection, or reference to, the power supply which powers  
the meter.  
In applications in which input LO is connected to ground elsewhere in  
the system, and SG5 is closed, unwanted ground-loop induced display  
errors may develop. Ground loops, a condition in which display drive  
currents also flow through the input signal wiring, can cause unstable  
display readings. If this occurs, try opening SG5 to see if the display  
stabilizes.  
Please read all of the following technical notes and perform any  
required modifications before soldering the DMS-EB onto the meter  
(aligning pin 1 to pin 1). After soldering, carefully connect +5Vdc to +V  
(J1, pin 5) and power return to –V (J1, pin 6).  
TECHNICAL NOTES  
SG5 must be open when measuring differential inputs that are refer-  
enced to the same supply which powers the meter. DMS Application  
Note 2 (Input Configurations, Power Supplies and Ground Loops)  
provides a detailed discussion of input configurations.  
1. LCD Backlighting: For DMS-30LCD backlit models, close SG15. For  
DMS-40LCD backlit models close SG18. See the DMS-30LCD or the  
DMS-40LCD data sheets for limitations on backlighting 9V-powered  
meters.  
8. Soldering: DATEL recommends the use of "no-clean" solders when  
installing or making modifications to the DMS-EB.  
2. 9V-Powered, LCD Display Models: DMS-30LCD-X-9 and DMS-  
40LCD-X/X-9 meters cannot be used in single-ended input configura-  
tions, i.e., LO (J1, pin 2) cannot be connected to GND (J1, pin 6). When  
operating from a 7.5 to 14Vdc power source with a single-ended  
input, use a +5V-powered meter and install U1 (see Battery Operation  
section and Technical Note 4).  
www.murata-ps.com  
Technical enquiries email: sales@murata-ps.com, tel: +1 508 339 3000  
MPM_DMS-EB.B01 Page 2 of 4  
DMS-EB  
Multi-Purpose Application Board for DMS-30 & DMS-40 Series Meters  
APPLICATIONS  
1. Decimal Point Placement: DATEL ships the DMS-EB with all decimal  
Using Full Range Span Adjustment (DMS-30 Only)  
point solder gaps (SG1, SG2, SG3 and SG17) open. To enable a specific  
decimal point, close its respective solder gap with solder. When reas-  
signing decimal places for subsequent applications, remember to  
open previously closed solder gaps.  
5.1. For DMS-30-1 and -2 models: Configuring the DMS-EB for span  
adjustment is simply a matter of installing resistor R3 and potenti-  
ometer R4, opening SG13 and then closing SG8, SG11 and SG12.  
5.2. For DMS-30-0 models: Span adjustment requires that you open  
SG8, SG11 (if closed) and SG13 and then close SG7, SG9 and  
SG12. Then install R3 and R4.  
Close SG1 for 1.999 (DP1)  
Close SG2 for 19.99 (DP2)  
Close SG3 for 199.9 (DP3)  
Adjust span as desired. DMS-30-1 and -2 span adjustment range  
is typically +10% and –5%. The DMS-30-0 has a much wider span  
adjust, however, it should be limited to 10% for optimum accuracy.  
Close SG17 for 1999.9 (DP4 on DMS-40)  
2. Display Test: Tie J1, pin 4 (TEST OUT) to pin 3 (TEST IN) to test the  
display of the DMS-30. Do not leave the meter in the test mode for  
more than 10 seconds. Display test is not available on backlit models.  
6. Measuring Input Voltages Greater than 1.999Vdc (Use DMS-30-1  
Only): If possible, the resistors used for R1 and R2 should be 1%  
metal-film types with TCR's less than 100ppm/°C. More information  
on selecting 1% resistors can be found in Application Note 14.  
3. Battery Operation: Open SG4 and install U1 (LM7805CT) with its  
metal tab facing down as shown Figure 2. Allowable input power  
ranges (J1, pins 5 and 6) are as follows:  
CAUTION: Do not exceed 250Vdc input signals.  
DMS-30/40PC-X-XS  
DMS-30/40PC-X-XL  
DMS-30/40LCD-X-5  
DMS-30/40LCD-X-5B  
+7.5 to +12.6V  
+7.5 to +18.0V  
+7.5 to +32.0V  
+7.5 to +12.6V  
6.1. Open SG10. For applications in which input voltages exceed  
100Vdc, also cut the 2 adjoining traces.  
6.2. Configure the DMS-EB for span adjust using the procedure revi-  
ously outlined for the DMS-30-1 version.  
6.3. Calculate values for R1 and R2 as follows:  
U1  
Metal tab  
points down  
R1  
SG1ꢀ  
SG3  
SG2  
SG1  
xx.xx  
xxx.x  
x.xxx  
To measure input voltages  
greater than 1ꢀꢀ0dcꢁ  
open SG1ꢀ and also cut  
these two etch paths.  
Figure 2. Installing U1  
4. Differential Input Signals (Signal Referenced to Power Source):  
Open SG5. Apply the input signal to pin 1, (+) IN HI, and pin 2, (–) IN  
LO, of J1.  
Figure 3. Installing R1  
A typical value for R1 is 1MΩ. The sum of R1 + R2 should be  
between 50kΩ and 10MΩ.  
Note: Common mode voltage must not exceed 2Vdc. Differential  
signals are defined as inputs whose two terminals are both above and  
below J1, pin 6 (GROUND).  
R2 = (FSI x R1) / ( |VIN| - FSI)  
Where:  
5. Span (Gain) and Zero (Offset) Adjustments: Locations are included  
on the DMS-EB to add user-supplied potentiometers R4 and R5, and  
resistor R3. Recommended values for these components are as fol-  
lows: R3: 10.5KΩ, 1%, 1/8W metal film; R4: 2KΩ, 10-20-turns; R5:  
50KΩ, 10-20 turns. Vishay's T93YA series potentiometers are suitable  
for this application. Make sure to use only RoHS-compliant compo-  
nents and solders when making modifications to DMS-EB.  
FSI = The attenuated voltage, between the meter's (–) IN LO and  
(+) IN HI pins, needed to achieve the desired display readings  
VIN = Input voltage at J1 (between pins 1 and 2)  
Once installed, these components permit the span and zero adjust-  
ments described in the following sections. R3 and R4 provide span  
adjust, while R5 applies and offset voltage to (-) IN LO (J1, pin 2).  
www.murata-ps.com  
Technical enquiries email: sales@murata-ps.com, tel: +1 508 339 3000  
MPM_DMS-EB.B01 Page 3 of 4  
DMS-EB  
Multi-Purpose Application Board for DMS-30 & DMS-40 Series Meters  
APPLICATIONS  
7.1. Install R2: R2 = FSR / (IMAX. x 10000) where FSR is the desired  
Example  
meter reading (0 -1999) and IMAX is between 0 -1 Ampere.  
VIN is 0-5Vdc, and the display reading must be "000" to "1500."  
7.2. Open SG5.  
1. Assume R1 = 1.0MΩ.  
7.3. Configure the DMS-EB for span adjust using the procedure previ-  
ously outlined for the DMS-30-0 version.  
2. R2 = (FSI x R1) / ( |VIN| - FSI)  
R2 = (1.500 x 1,000,000) / (5.0 - 1.500)  
R2 = 428,571Ω or 432kΩ  
7.4. Apply input signal (current) to (+) IN HI and (–) IN LO.  
7.5. Adjust R4 for desired full scale readings.  
3. Calibrate the meter using a known voltage source. Adjust R4 to  
compensate for variations in R1 and R2.  
7.6. Enable the appropriate decimal point using either SG1, SG2, SG3  
or SG-17.  
7. Current Measurements (Use 200mV Model, DMS-30-0): The follow-  
ing instructions are for measuring current derived from the DMS-  
30/40's power source. When measuring current that is floating with  
respect to J1, pin 6 (POWER RETURN), leave SG5 closed.  
Example  
A 0.100 Ampere input must read "100.0" on the display.  
R2 = 1000 / (0.1 x 10000)  
R2 = 1 Ohm  
CAUTION: Do not exceed 1 Ampere input current  
Enable DP3 via SG3 for a reading of "100.0."  
HI LO TST TST +V V  
(+) (–) IN OUT  
GND  
Figure 4. DMS-EB Board  
USA:  
Mansfield (MA), Tel: (508) 339-3000, email: sales@murata-ps.com  
Toronto, Tel: (866) 740-1232, email: toronto@murata-ps.com  
Canada:  
UK:  
Milton Keynes, Tel: +44 (0)1908 615232, email: mk@murata-ps.com  
Montigny Le Bretonneux, Tel: +33 (0)1 34 60 01 01, email: france@murata-ps.com  
München, Tel: +49 (0)89-544334-0, email: munich@murata-ps.com  
Murata Power Solutions, Inc.  
France:  
Germany:  
Japan:  
11 Cabot Boulevard, Mansfield, MA 02048-1151 U.S.A.  
Tel: (508) 339-3000 (800) 233-2765 Fax: (508) 339-6356  
www.murata-ps.com email: sales@murata-ps.com ISO 9001 and 14001 REGISTERED  
Tokyo, Tel: 3-3779-1031, email: sales_tokyo@murata-ps.com  
Osaka, Tel: 6-6354-2025, email: sales_osaka@murata-ps.com  
12/15/08  
Murata Power Solutions, Inc. makes no representation that the use of its products in the circuits described herein, or the use of other  
technical information contained herein, will not infringe upon existing or future patent rights. The descriptions contained herein do not imply  
the granting of licenses to make, use, or sell equipment constructed in accordance therewith. Specifications are subject to change without  
China:  
Shanghai, Tel: +86 215 027 3678, email: shanghai@murata-ps.com  
Guangzhou, Tel: +86 208 221 8066, email: guangzhou@murata-ps.com  
notice.  
© 2008 Murata Power Solutions, Inc.  
Singapore: Parkway Centre, Tel: +65 6348 9096, email: singapore@murata-ps.com  
www.murata-ps.com  
Technical enquiries email: sales@murata-ps.com, tel: +1 508 339 3000  
MPM_DMS-EB.B01 Page 4 of 4  

相关型号:

DMS-EB-RMS

AC-to-RMS Converter Application Board for DMS-30PC/LCD Meters
ETC

DMS-EB2

Multi-Purpose Application Board for DMS-20PC/LCD Meters
MURATA

DMS-LM2

Self-Powered LED Display High-Voltage, 50/60/400Hz AC Line Monitor
MURATA

DMS-LM3

Self-Powered LED Display 480V AC Line Monitor
MURATA

DMS-PS-CM

High Efficiency, AC/DC Instrumentation Power Supplies with Individually Isolated Outputs
MURATA

DMS-PS1-CM

High Efficiency, AC/DC Instrumentation Power Supplies with Individually Isolated Outputs
MURATA

DMS-PS1-CM-C

0-5V and 0-10V Input 3½ Digit, LED Display Process Control Monitors
MURATA

DMS-PS10-CM

High Efficiency, AC/DC Instrumentation Power Supplies with Individually Isolated Outputs
MURATA

DMS-PS11-CM

Converter
ETC

DMS-PS12-CM

Converter
ETC

DMS-PS2-CM

High Efficiency, AC/DC Instrumentation Power Supplies with Individually Isolated Outputs
MURATA

DMS-PS3-CM

High Efficiency, AC/DC Instrumentation Power Supplies with Individually Isolated Outputs
MURATA