User's Manual  
CFM/CMM Thermo Anemometer  
Model AN100  
CFM/CMM  
Thermo-Anemometer  
AN100  
Introduction  
Congratulations on your purchase of the Extech AN100 CFM/CMM Thermo  
Anemometer. This instrument measures Air Velocity, Air Flow (volume) and  
Temperature. The large, easy-to-read backlit LCD includes primary and secondary  
displays plus numerous status indicators. The meter is shipped fully tested and  
calibrated and with proper use will provide years of reliable service.  
Meter Description  
1. LCD Display  
2. Vane  
3. Rubber Boot  
4. Meter  
5. Power On/Off  
6. Area  
7. AVG  
8. Backlight  
9. Temperature Buttons  
10. Airflow/Velocity Buttons  
Keypad  
AUTO POWER OFF  
Press to turn the meter ON or OFF  
MAX/MIN Used to record and store the highest, lowest and average airflow or  
velocity readings.  
(LEFT) also serves as change decimal point button in AREA mode  
UNITS Press to select the mode of operation. In FLOW mode, the meter displays  
air volume. In VELOCITY mode, the meter displays air speed.  
(UP) also serves as increase number button in AREA mode.  
AVG Used to Average multiple readings in FLOW or VELOCITY mode. Up to 20  
readings can be averaged.  
HOLD Press to freeze the displayed reading. Press again to unlock display. This  
button also functions as the RIGHT scroll button in AREA mode and RECALL  
mode.  
AREA Press and hold to manually enter the area of a duct in CFM or CMM mode  
NEXT In AREA mode, used to select memory locations 1-8.  
Press to turn the backlight on/off  
MAX/MIN (Temperature) Used to record and store the highest, lowest readings for  
air temperature.  
°C °F HOLD (Temperature) Press to freeze the displayed temperature reading.  
Press again to unlock the display. Press and hold for 3 seconds to switch between  
°C and °F. Meter will beep twice to indicate change.  
Battery compartment located on the back of the instrument. The rubber protective jacket  
must be removed from the meter to access the compartment.  
2
Model AN100 Version 2.0 02/06  
Display Layout  
MAX (top of LCD): Max Hold function engaged for the Air Temperature function  
HOLD (top of LCD): Data Hold function engaged for the Air Temperature function  
VEL: indicates that meter is in air velocity mode  
FLOW: indicates that meter is in air flow mode  
MAX (bottom of LCD): Max Hold for the IR Temperature and RH function  
HOLD (bottom of LCD): Data Hold for the IR Temperature function and RH function  
oC / oF: Temperature units of measure  
CFM/CMM: airflow units of measure  
Ft2, m2: units for area dimensions  
m/s, ft/min, km/h, MPH, knots: air velocity units of measure  
X10, X100: multipliers for air flow readings  
AVG: air averaging mode  
RECORD: indicates that min/max function is running (top for temp, bottom for air)  
Large LCD digits at center of display for Relative Humidity and IR Temperature  
Smaller LCD digits at top, right of display for Probe Temperature  
: Low battery indicator  
3
Model AN100 Version 2.0 02/06  
Operation  
Connecting the Vane  
1. The vane plug is inserted in the meter’s sensor jack at the top of the meter. The plug  
and jack are keyed so that the plug can only fit in the jack one way.  
2. Turn the plug carefully until it lines up with the jack and then firmly push the plug in  
place. Do not apply undue force or try to twist the plug side-to-side.  
3. If the vane is not connected to the meter or if the sensor is defective, the LCD display  
will indicate OL in place of a Temperature reading.  
Air Velocity Measurements (Single Point)  
Side view of Vane  
1. Turn on the meter using the ON/OFF  
button.  
2. Press UNITS button to select the desired unit of  
measure. NOTE: At power up the meter will  
display the last unit of measure previously  
entered.  
Arrow  
airflow  
3. Place the sensor in the air stream. Ensure that  
the air enters the vane as indicated by the arrow  
sticker placed inside the vane.  
4. View the readings on the LCD Display. The  
large main LCD display shows the Air Velocity  
reading. The upper right LCD sub-display shows  
the temperature reading.  
Air Velocity Averaging Mode  
1. To enter 20 Point Averaging Mode, press and hold the AVG button until it beeps twice.  
The AVG icon will be displayed.  
2. Take a measurement and press the AVG button. A single beep will sound and the  
HOLD icon will appear in the display.  
3. The average reading will be displayed and number of readings measured will appear in  
the upper right hand corner of the display. After 5 seconds, the display will return to the  
current reading.  
4. Repeat steps 2 - 3 until the desired number of points have been measured.  
5. To return to standard velocity measuring mode press and hold AVG button until meter  
beeps twice.  
Note: In the standard velocity measuring mode, press the AVG button once to recall the  
previous average. The average will be cleared when you enter the Averaging Mode  
again.  
4
Model AN100 Version 2.0 02/06  
Air Flow Measurements (CMM / CFM)  
1. Turn on the meter using the  
ON/OFF button  
Side view of Vane  
2. Press the UNITS button to select the desired air flow  
units: CMM (cubic meters per minute) or CFM (cubic  
feet per minute). NOTE: At power up the meter will  
display the last unit of measure previously entered.  
Arrow  
3. To begin entering the area in m2 or ft2, press and  
hold the AREA button until it beeps twice. The  
airflo  
leftmost digit of the bottom display will begin to flash.  
4. Use the (UP) button to change the flashing digit  
Use the (LEFT) button to move the decimal  
Use (RIGHT) button to select the other digits.  
After all of the digits are entered, press and hold the  
AREA button (until meter beeps twice) to save the area into memory and return to  
CFM or CMM measuring mode.  
5. Place the sensor in the air stream. Ensure that the air enters the vane as indicated by  
the arrow sticker placed inside the vane. Refer to the diagram. The large main LCD  
display shows the Air Velocity reading. The upper right LCD sub-display shows the  
temperature reading.  
The meter has 16 memory locations (8 for CFM and 8 for CMM) that can be used to store  
commonly used area sizes that you can recall at anytime.  
1. Press the AREA button until meter beeps twice. A memory location number will  
appear in the top right of the display indicating the memory location.  
2. Push the NEXT button to scroll thru and select the desired location. Once you have  
selected the desired memory location enter your dimension  
Use the (UP) button to change the flashing digit  
Use the (LEFT) button to move the decimal  
Use (RIGHT) button to select the other digits. After all of the digits are entered,  
press and hold the AREA button (until it beeps twice) to save the area into memory  
and return to CFM or CMM measuring mode.  
To select and use a previously stored dimension, press and hold the AREA button until it  
beeps twice.  
Press NEXT to scroll thru the 8 memory locations. Press and hold the AREA button until it  
beeps twice to return to CFM or CMM measuring mode.  
Air Flow Averaging Mode  
1. To enter 20 Point Averaging Mode, press and hold the AVG button until it beeps twice.  
The AVG icon will be displayed.  
2. Take a measurement and press the AVG button. A single beep will sound and the  
HOLD icon will appear in the display.  
3. The average reading will be displayed and number of readings measured will appear in  
the upper right hand corner of the display. After 5 seconds, the display will return to the  
current reading.  
4. Repeat steps 2 - 3 until the desired number of points have been measured.  
5. To return to standard airflow measuring mode press and hold AVG button until meter  
beeps twice.  
Note: In the standard velocity measuring mode, press the AVG button once to recall the  
previous average. The average will be cleared when you enter the Averaging Mode  
again.  
5
Model AN100 Version 2.0 02/06  
Data Hold (Air Velocity/Air Flow)  
1. While taking measurements, press the HOLD button to freeze the air velocity/air flow  
reading.  
2. The HOLD indicator will appear in the bottom of the LCD display.  
3. Press HOLD again to return to normal operation.  
MAX/MIN/AVG Record (Air Velocity/Air Flow)  
This allows the user to record and view the highest (MAX), lowest (MIN) and average  
(AVG) readings.  
1. Press the button MAX/MIN button. The MAX indicator and RECORD indicator along  
with the Max reading will appear on the LCD display and the meter will begin keeping  
track of the MAX, MIN and Average values.  
2. Press the MAX/MIN button again to view the minimum reading. The MIN indicator  
along with the minimum reading will appear on the LCD display.  
3. Press the MAX/MIN button again to view the average reading. The AVG indicator  
along with the average reading will appear on the LCD display.  
4. Press the MAX/MIN button again to display current readings. NOTE: the meter will  
keep recording MAX/MIN/AVG readings.  
5. To clear and stop MAX/MIN/AVG recording and return to normal operation, press and  
hold the MAX/MIN button until the meter beeps twice.  
Automatic Power OFF  
To conserve battery life, the meter automatically turns off after 20 minutes. To defeat this  
feature:  
1. Turn the meter OFF.  
2. Press and hold the  
(Backlight) key while turning the meter ON.  
3. “dis APO” will appear in the display. The AUTO POWER OFF feature will now be  
disabled.  
4. Note that AUTO POWER OFF is re-enabled each time the meter is turned on.  
5. Also note that AUTO POWER OFF is disabled in CFM/CMM or Average mode.  
Battery Replacement  
When  
appears on the LCD, the 9V battery must be replaced.  
1. Disconnect the sensor.  
2. Remove the meter’s rubber protective jacket  
3. Use a Phillips screwdriver to open the rear battery compartment  
4. Replace the 9V battery  
5. Close the battery compartment and replace the meter’s protective jacket  
6
Model AN100 Version 2.0 02/06  
Specifications  
Air Velocity  
Range  
0.40 - 30.00 m/s  
1.4 - 108.0 km/h  
80 – 5900 ft/min  
0.9 – 67.0 mph  
0.8 to 58.0 knots  
Range  
0-9999 m3/min  
0-9999 ft3/min  
Range  
Resolution  
0.01 m/s  
0.1 km/h  
1 ft/min  
0.1 mph  
0.1 knots  
Resolution  
1
Accuracy  
± (3% + 0.20 m/s)  
± (3% + 0.8 km/hr)  
± (3% + 40 ft/m)  
± (3% + 0.4 MPH)  
± (3% + 0.4 knots)  
Area  
m/s (meters per sec)  
km/h (kilometers/hour)  
ft/min (feet per minute)  
mph (miles per hour)  
knots (nautical MPH)  
Air Flow  
CMM (cubic meters/min)  
CFM (cubic ft/min)  
Air Temperature  
0 to 9.999m2  
0 to 9.999ft2  
1
Resolution  
0.1oF/C  
Accuracy  
4.0oF (2.0oC)  
14 - 140oF (-10 - 60oC)  
Circuit  
Custom LSI microprocessor circuit  
Dual function 0.5" (13 mm) 4-digit LCD  
1 reading per second approx.  
Display  
Sampling rate  
Sensors  
Air velocity/flow sensor: Conventional angled vane arms with  
low-friction ball bearing  
Temperature sensor: NTC-type precision thermistor  
Auto shut off after 20 minutes to preserve battery life  
32°F to 122°F (0°C to 50°C)  
14 to 140oF (-10 to 60oC)  
Automatic Power off  
Operating Temperature  
Storage Temperature  
Operating Humidity  
Storage Humidity  
Operating Altitude  
Battery  
<80% RH  
<80% RH  
2000 meters (7000ft) maximum  
One 9 volt (NEDA 1604) battery  
Battery life  
80 hours approx. (if the Backlight is used continuously, battery  
life will be reduced significantly)  
Battery Current  
Weight  
8.3 mA DC approx.  
1.6 lbs. (725g) including battery & probe  
Dimensions  
Main instrument: 7.0 x 2.9 x1.2" (178 x 74 x 33mm)  
Sensor Head: 2.75” (70mm) Diameter  
7
Model AN100 Version 2.0 02/06  
Warranty  
EXTECH INSTRUMENTS CORPORATION warrants this instrument to be free of defects  
in parts and workmanship for one year from date of shipment (a six month limited warranty  
applies to sensors and cables). If it should become necessary to return the instrument for  
service during or beyond the warranty period, contact the Customer Service Department at  
(781) 890-7440 ext. 210 for authorization or visit our website www.extech.com for contact  
information. A Return Authorization (RA) number must be issued before any product is  
returned to Extech. The sender is responsible for shipping charges, freight, insurance and  
proper packaging to prevent damage in transit. This warranty does not apply to defects  
resulting from action of the user such as misuse, improper wiring, operation outside of  
specification, improper maintenance or repair, or unauthorized modification. Extech  
specifically disclaims any implied warranties or merchantability or fitness for a specific  
purpose and will not be liable for any direct, indirect, incidental or consequential damages.  
Extech's total liability is limited to repair or replacement of the product. The warranty set  
forth above is inclusive and no other warranty, whether written or oral, is expressed or  
implied.  
Calibration and Repair Services  
Extech offers repair and calibration services for the products we sell. Extech also  
provides NIST certification for most products. Call the Customer Service Department for  
information on calibration services available for this product. Extech recommends that  
annual calibrations be performed to verify meter performance and accuracy.  
Support line (781) 890-7440  
Technical support: Extension 200; E-mail: support@extech.com  
Repair & Returns: Extension 210; E-mail: repair@extech.com  
Product specifications subject to change without notice  
For the latest version of this User’s Guide, Software updates, and other  
up-to-the-minute product information, visit our website: www.extech.com  
Extech Instruments Corporation, 285 Bear Hill Rd., Waltham, MA 02451  
Copyright © 2006 Extech Instruments Corporation  
All rights reserved including the right of reproduction in whole or in part in any form.  
8
Model AN100 Version 2.0 02/06  
Useful Equations and Conversions  
Area equation for rectangular or square ducts  
Height (H)  
Area (A) = Width (W) x Height (H)  
Width (W)  
Area equation for circular ducts  
Area (A) = x r2  
Radius  
Where = 3.14 and r2 = radius x radius  
Cubic equations  
CFM (ft3/min) = Air Velocity (ft/min) x Area (ft2)  
CMM (m3/min) = Air Velocity (m/sec) x Area (m2) x 60  
NOTE: Measurements made in inches  
must be converted to feet or meters before using the above formulae.  
Unit of Measure Conversion Table  
m/s  
1
ft/min  
196.87  
1
knots  
1.944  
0.00987  
1
km/h  
3.6  
MPH  
2.24  
1 m/s  
1 ft/min  
1 knot  
1 km/h  
1 MPH  
0.00508  
0.5144  
0.2778  
0.4464  
0.01829  
1.8519  
1
0.01138  
1.1523  
0.6222  
1
101.27  
54.69  
87.89  
0.54  
0.8679  
1.6071  
9
Model AN100 Version 2.0 02/06  

相关型号:

AN1000

A broad portfolio of high performance, best-in-class Serial Memory Products to meet all your design requirements.
MICROCHIP

AN1000-521P

AN0625-521P
ETC

AN1001

The Next Step in Surface-Mount Power MOSFETs
VISHAY

AN1003

陶瓷天线 Multilayer Chip Antenna for 433MHz Wireless Communication
ETC

AN1004

A broad portfolio of high performance, best-in-class Serial Memory Products to meet all your design requirements.
MICROCHIP

AN1006

A broad portfolio of high performance, best-in-class Serial Memory Products to meet all your design requirements.
MICROCHIP

AN101

MOSFETs
VISHAY

AN1013

MONO CLASS_D AMPLIFIERS
STMICROELECTR

AN10145

Bi-directional low voltage translators
NXP

AN1016/D

Infrared Sensing and Data Transmission Fundamentals
ETC

AN1018

A broad portfolio of high performance, best-in-class Serial Memory Products to meet all your design requirements.
MICROCHIP

AN1020

A broad portfolio of high performance, best-in-class Serial Memory Products to meet all your design requirements.
MICROCHIP