TM200-S48 [TRUMPOWER]

200 WATT MEDICAL POWER SUPPLIES; 200瓦医疗电源
TM200-S48
型号: TM200-S48
厂家: Tumbler Technologies + TRUMPower    Tumbler Technologies + TRUMPower
描述:

200 WATT MEDICAL POWER SUPPLIES
200瓦医疗电源

医疗
文件: 总2页 (文件大小:168K)
中文:  中文翻译
下载:  下载PDF数据表文档文件
TRUMPower  
200 WATT MEDICAL POWER SUPPLIES  
DESCRIPTION  
(MDD)  
TM200 SERIES  
Comprising single and multiple output models  
with 200 watts of continuous output power, the  
TM200 series is specially designed for medical  
applications. Auto-ranging input allows them to  
operate from any line voltage throughout the  
world without a selector strap. All auxiliary  
outputs are with magnetic amplifier or linear  
regulator to keep regulation. The units are  
constructed on a printed circuit board with a  
U-bracket for mechanical support and heat  
sinking.  
A cover-and-fan assembly can be  
added during manufacturing  
.
Safety Standard Approvals:  
FEATURES  
Low safety ground leakage current  
Meet EN 55011 and FCC class B  
Auto-ranging input  
UL 2601-1  
Short-circuit protection with auto-recovery  
Power Fail DetectPFDsignal  
100% burn-in at full rated load  
Optional cover-and-fan assembly  
CSA C22.2 No. 601.1  
TÜV EN60601-1  
INPUT SPECIFICATIONS  
ENVIRONMENTAL SPECIFICATIONS  
Input voltage:  
90 to 130 VAC, or 180 to  
Operating temperature:  
Storage temperature:  
Relative humidity:  
Derating:  
0ºC to +70ºC  
264 VACauto-ranging)  
-40ºC to +85ºC  
Input frequency: 47 to 63 Hz  
Input current: 4.5Armsfor 115VAC  
2.6Armsfor 230VAC  
5% to 95% non-condensing  
Derate from 100% at +50ºC  
linearly to 50% at +70ºC  
25 CFM forced air, provided  
on covered version, to be  
provided for U-bracket  
version by user  
Leakage current: 90 µA max. @ 115VAC, 60Hz  
Cooling:  
15 µA max. @ 230VAC, 50Hz  
OUTPUT SPECIFICATIONS  
Output voltage/currentSee rating chart  
GENERAL SPECIFICATIONS  
Switching frequency64KHz ±5KHz  
Efficiency:  
70% minimum on all models  
Total output power:  
Ripple and noise:  
200 watts maximum  
Hold-up time:  
Line regulation:  
Inrush current:  
12 msec minimum at 110VAC  
±0.5% maximum at full load  
20 amps @ 115VAC, or 40 amps @  
230VAC, at 25ºC cold start  
4000VAC from input to output  
1500VAC from input to ground  
500VAC from output to ground  
450,000 hours at full load at 25ºC  
ambient, calculated per  
2% peak to peak maximum  
Overvoltage protectionProvided on output #1 only; set at  
112 – 132% of its nominal output  
voltage  
Withstand voltage:  
MTBF:  
Overcurrent protectionAll outputs protected to short  
circuit conditions  
Temperature coefficientAll outputs ±0.04%/ºC maximum  
MIL-HDBK-217F  
Transient response:  
PFD signal:  
Maximum excursion of 4% or  
better on all models, recovering to  
1% of final value within 500µs  
after a 25% step load change  
EMC Performance (EN60601-1-2: 2001)  
EN55011:  
Class B conducted, Class B radiated  
FCC:  
VCCI:  
Class B conducted, Class B radiated  
Class B conducted, Class B radiated  
Harmonic distortion, Class A  
Line flicker  
ESD, ±8KV air and ±4KV contact  
Radiated immunity, 3V/m  
EN61000-3-2:  
EN61000-3-3:  
EN61000-4-2:  
EN61000-4-3:  
EN61000-4-4:  
EN61000-4-5:  
EN61000-4-6:  
EN61000-4-8:  
EN61000-4-11:  
TTL logic high for normal  
operation and TTL logic low upon  
loss of input power. This signal  
appears at least 1 ms prior to +5V  
output dropping 5% below its  
nominal value. This signal also  
provides a minimum delay of 100  
ms after +5V is within regulation.  
Fast transient/burst, ±1KV  
Surge, ±1KV diff, ±2KV com.  
Conducted immunity, 3Vrms  
Magnetic field immunity, 1A/m  
Voltage dips, 30% reduction for 500ms  
and >95% reduction for 10ms  
Interruptions, >95% for 5000ms  
Tumbler Technologies + TRUMPower  
3350 Scott Blvd., Bldg. 13, Santa Clara, California 95054, USA  
Phone: 408-988-6616 Fax: 408-988-6622 email: sales@trumpower.com Website: www.TRUMPower.com  
MEDICAL GRADE  
TM200 SERIES  
OUTPUT VOLTAGE/CURRENT RATING CHART  
Maximum  
Output #1  
Output #2  
Output #3  
Vnom. Imin. Imax. Tol.  
Output #4  
Vnom. Imin. Imax. Tol.  
Output  
Power  
200W  
Product No. Vnom. Imin. Imax. Tol. Vnom. Imin. Imax. Tol.  
TM200-S05  
TM200-S12 12V  
TM200-S15  
5V  
3.0A  
1.2A  
40A 2%  
(N/A)  
(N/A)  
(N/A)  
(N/A)  
(N/A)  
(N/A)  
(N/A)  
(N/A)  
(N/A)  
(N/A)  
16.7A 2%  
200W  
15V  
1.0A  
0.6A  
0.5A  
0.5A  
3.0A  
3.0A  
3.0A  
3.0A  
3.0A  
3.0A  
3.0A  
3.0A  
3.0A  
3.0A  
3.0A  
3.0A  
13.4A 2%  
8.4A 2%  
6.7A 2%  
4.2A 2%  
(N/A)  
(N/A)  
(N/A)  
(N/A)  
(N/A)  
(N/A)  
(N/A)  
(N/A)  
(N/A)  
(N/A)  
(N/A)  
(N/A)  
(N/A)  
(N/A)  
(N/A)  
(N/A)  
(N/A)  
(N/A)  
200W  
200W  
200W  
200W  
200W  
200W  
200W  
200W  
200W  
200W  
200W  
200W  
200W  
200W  
200W  
200W  
TM200-S24 24V  
TM200-S30 30V  
TM200-S48 48V  
TM200-D23 +5V  
TM200-D24 +5V  
TM200-D25 +5V  
TM200-T30 +5V  
TM200-T31 +5V  
TM200-T32 +5V  
TM200-T33 +5V  
TM200-Q40 +5V  
TM200-Q41 +5V  
TM200-Q42 +5V  
TM200-Q44 +5V  
TM200-Q45 +5V  
30A 2% +12V 0A  
30A 2% +15V 0A  
30A 2% +24V 0A  
30A 2% +12V 0A  
30A 2% +12V 0A  
30A 2% +15V 0A  
30A 2% +15V 0A  
30A 2% +12V 0A  
30A 2% +15V 0A  
30A 2% +12V 0A  
30A 2% +12V 0A  
30A 2% +12V 0A  
8A  
4%  
4%  
4%  
4%  
4%  
4%  
4%  
4%  
4%  
4%  
4%  
4%  
6A  
4A  
8A  
8A  
6A  
6A  
8A  
6A  
8A  
8A  
8A  
-5V  
-12V  
-15V  
-12V  
-12V  
-15V  
-12V  
-15V  
-12V  
0A  
0A  
0A  
0A  
0A  
0A  
0A  
0A  
0A  
6A 4%  
4A 4%  
4A 4%  
4A 4%  
4A 4%  
4A 4%  
4A 4%  
4A 4%  
4A 4%  
5V  
24V  
12V  
15V  
24V  
0A  
0A  
0A  
0A  
0A  
6A  
4A  
4A  
4A  
4A  
4%  
4%  
4%  
4%  
4%  
Notes: (1) Add suffix “C” for enclosed format, e.g. TM200-Q45C.  
(2) Peak current is 12A on +12V, 9A on +15V and 6A on +24V.  
(3) Output #4 is floating. It can be connected externally for positive or negative output.  
(4) 200 watts for “C” version with a cover-and-fan assembly. 150 watts U-bracket version without moving air (maximum current of output #1  
& #2 derated to 70%), or 200 watts with 25 CFM forced air provided by user.  
(5) When the remote sense facility is not used, +Sense must be connected to +V, and –Sense to return, on P2 connector.  
(6) All models may be operated at no-load. At no-load, output voltage tolerance increases to 10%  
MECHANICAL SPECIFICATIONS  
NOTES:  
1.  
2.  
3.  
Dimensions shown in inch [mm]  
Tolerance 0.02 [0.5] maximum  
Input connector P1 is Beau Inc. P/N 72-5-03C, output connector  
P2 is Beau Inc. P/N 72-5-09C. Screws are #6-32 on 0.375 inch  
centers.  
4.  
5.  
6.  
7.  
Connector P3 mates with Molex housing 22-01-1043 and Molex  
40445 series crimp terminal.  
Connector P4 mates with Molex housing 22-01-1023 and Molex  
40445 series crimp terminal.  
P4 is for DC fan rated 24V/0.12A (models TM200-S24, -S30,  
-D25), or 12V/0.2A (other models)  
Weight:  
1.40 Kgs (3.08Lbs.) approx. for U-bracket version  
1.60 Kgs (3.52 Lbs.) approx. for “C” version  
PIN CHART  
CON  
PIN  
P1  
P2  
P3  
Product No.  
1
2
3
1, 2  
3, 4, 5  
6
7
8
9
1
2
3
4
TM200-S05 TM200-S12  
TM200-S15 TM200-S24  
TM200-S30 TM200-S48  
AC  
AC  
AC  
OUTPUT OUTPUT OUTPUT  
COM. COM.  
RET. RET.  
-SENSE RET.  
+SENSE FAN  
PFD  
LIVE NEUTRAL GROUND  
#1  
#1  
#1  
TM200-D23 TM200-D24 AC  
TM200-D25  
AC AC  
LIVE NEUTRAL GROUND  
TM200-T30 TM200-T31 AC AC AC  
OUTPUT COM. OUTPUT  
COM. COM.  
RET. RET.  
COM. COM.  
RET. RET.  
N.C.  
N.C.  
N.C.  
N.C.  
N.C.  
FAN  
FAN  
PFD  
PFD  
#1 RET. #2  
OUTPUT COM. OUTPUT OUTPUT  
#1 RET. #2 #3  
TM200-T32 TM200-T33 LIVE NEUTRAL GROUND  
TM200-Q40 TM200-Q41  
TM200-Q42 TM200-Q44  
TM200-Q45  
OUTPUT  
#4  
RETURN  
AC  
AC  
AC  
OUTPUT COM. OUTPUT OUTPUT  
#1 RET. #2 #3  
OUTPUT  
#4  
COM. COM.  
RET. RET.  
FAN  
PFD  
LIVE NEUTRAL GROUND  
Tumbler Technologies + TRUMPower  
3350 Scott Blvd., Bldg. 13, Santa Clara, California 95054, USA  
Phone: 408-988-6616 Fax: 408-988-6622 email: sales@trumpower.com Website: www.TRUMPower.com  

相关型号:

TM200-T30

200 WATT MEDICAL POWER SUPPLIES
TRUMPOWER

TM200-T31

200 WATT MEDICAL POWER SUPPLIES
TRUMPOWER

TM200-T32

200 WATT MEDICAL POWER SUPPLIES
TRUMPOWER

TM200-T33

200 WATT MEDICAL POWER SUPPLIES
TRUMPOWER

TM200CZ-24

HIGH POWER GENERAL USE INSULATED TYPE
MITSUBISHI

TM200CZ-2H

HIGH POWER GENERAL USE INSULATED TYPE
MITSUBISHI

TM200CZ-H

HIGH POWER GENERAL USE INSULATED TYPE
MITSUBISHI

TM200CZ-M

HIGH POWER GENERAL USE INSULATED TYPE
MITSUBISHI

TM200DZ-24

HIGH POWER GENERAL USE INSULATED TYPE
MITSUBISHI

TM200DZ-2H

HIGH POWER GENERAL USE INSULATED TYPE
MITSUBISHI

TM200DZ-H

HIGH POWER GENERAL USE INSULATED TYPE
MITSUBISHI

TM200DZ-M

HIGH POWER GENERAL USE INSULATED TYPE
MITSUBISHI