M9380A [KEYSIGHT]
M9380A PXIe CW Source;型号: | M9380A |
厂家: | Keysight Technologies |
描述: | M9380A PXIe CW Source |
文件: | 总17页 (文件大小:2651K) |
中文: | 中文翻译 | 下载: | 下载PDF数据表文档文件 |
Keysight Technologies
M9380A PXIe CW Source
1 MHz to 3 GHz or 6 GHz
Data Sheet
Overview
– Aerospace and defense
– Interference injection
– LO substitution
Introduction
With high-power levels and accurate measurements, the M9380A CW
source provides the Keysight Technologies, Inc. quality and performance
in the PXI form factor—a trusted Keysight product with global services and
support, fast repair and a wide scope of calibration utilities.
– Wireless component test
Product description
The M9380A PXIe CW source is a compact modular solution that
provides frequency coverage from 1 MHz to 3 or 6 GHz. A typical M9380A
configuration includes three individual PXIe modules—M9310A source
output, M9301A synthesizer and the M9300A frequency reference—
designed for fast data interfaces and high-speed automated test systems.
Instrument control is provided through a soft front panel and programmatic
interfaces tuned to your application development environment of choice.
Product features
– Frequency coverage from 1 MHz to 3 or 6 GHz
– Output power of +18 dBm across the frequency range
– Output power level of +19 dBm from 1 MHz to 5 GHz
– Better than ±0.4 dB absolute amplitude accuracy
– License key upgraded frequency range and output power
– One day startup assistance
– Chassis slot compatibility: PXIe slot
Uncompromising values
– Keeps costs manageable—purchase what you need today and
easily upgrade later using license-key upgrades without returning your
modules to Keysight.
– Reduces development time and simplifies integration into existing test
environments with multiple drivers and programmatic interfaces.
– Reduces startup time with Keysight IO libraries easy configuration,
one-step software install, and integrated instrument level CW source
soft front panel.
Figure 1. M9380A PXIe CW source with three
modules consisting of the M9310A PXIe source
output, M9301A PXIe synthesizer, and M9300A
PXIe frequency reference.
– Fast repair turnaround time with calibrated core exchange strategy.
Applications
03 | Keysight | M9380A PXIe CW Source 1 MHz to 3 GHz or 6 GHz - Data Sheett
Technical Specifications and Characteristics
Block diagram
(+13 dBm sine
M9301A PXIe Synthesizer
M9300A PXIe Frequency Reference
all 1001M0H0z OMutsH) z
Out 1
3 dB
RF/LO Out 1A
External Trigger Trig 2 (front)
to PXI Backplane
100 MHz PLL
100 MHz
Out 2
100 MHz
Out 3
3 dB
3 dB
3 dB
3 GHz to 6 GHz
RF/LO Out 1B
RF/LO Out 2A
RF/LO Out 2B
Banded
Switched
Filter
Divide
by 1-2-4-8
100 MHz
(1 to 110 MHz)
Ref In
Bank
PLL
100 MHz
Out 4
10 MHz OCXO
TIMEBASE
850 MHz
Divide
by 1 or 2
100 MHz
Out 5
(RF Outputs: 187.5 MHz to 6 GHz)
(-10 dBm to +10 dBm)
3 dB
100 MHz In (+13 dBm Typical)
10 MHz
10 MHz Out
Divide
by 10
3 dB
100 MHz
10 MHz
OXCO Out
Out
(10 MHz, +10 dBm
to 16 dBm, sine)
(+13 dBm
Typical)
M9310A PXIe Source Output
RF
LO
IF
RF In
ALC and
Power
Control
Pulse
Modulator
(1 MHz to 6 GHz)
(-5 dBm to +5 dBm)
Pulse Mod Trigger
Trig 1 (Front Panel)
ALC Hold Trigger
PXI Backplane
(-130 dBm to +19 dBm, 1 MHz to 5 GHz)
(-130 dBm to +18 dBm, >5 GHz to 6 GHz)
ALC
Coupler
Log Amp
RF Out
Limiter
6 GHz
100 MHz In
(+10 dBm typical)
100 MHz Out
(+10 dBm typical)
Figure 2. M9380A PXIe CW source with three modules consisting of the M9300A PXIe frequency reference, M9301A PXIe synthesizer and M9310A PXIe source output.
04 | Keysight | M9380A PXIe CW Source 1 MHz to 3 GHz or 6 GHz - Data Sheett
Technical Specifications and Characteristics
Definitions for specifications
Temperatures referred to in this document are defined as follows:
– Full temperature range = Individual module temperature of ≤ 75 °C, as reported by
the module, and environment temperature of 0 to 55 °C.
– Controlled temperature range = Individual module temperature of ≤ 55 °C, as
reported by the module, and environment temperature of 20 to 30 °C.
Specifications describe the warranted performance of calibrated instruments. Data
represented in this document are specifications unless otherwise noted under the
following conditions.
– Calibrated instruments have been stored for a minimum of 2 hours within the full
temperature range
– 45 minute warm-up time
– Calibration cycle maintained
– When used with Keysight M9300A frequency reference
– When used with Keysight interconnection cables
Characteristics describe product performance that is useful in the application of the
product. Characteristics are often referred to as Typical or Nominal values and are itali-
cized.
– Typical describes characteristic performance, which 80% of instruments will meet
when operated within the controlled temperature range.
– Nominal describes representative performance that is useful in the application of the
product when operated within the controlled temperature range.
Recommended best practices in use
– Use slot blockers and EMC filler panels in empty module slots to ensure proper
operating temperatures.
– Keysight chassis and slot blockers optimize module temperature performance and
reliability of test.
– At environment temperatures above 45 °C, chassis fan should be set to high.
Additional information
– All graphs contain measured data from one unit and is representative of product
performance at the controlled temperature range unless otherwise noted.
– The specifications contained in this document are subject to change.
05 | Keysight | M9380A PXIe CW Source 1 MHz to 3 GHz or 6 GHz - Data Sheett
Technical Specifications and Characteristics (continued)
Frequency
Frequency range
Option F03
1 MHz to 3 GHz
1 MHz to 6 GHz
0.01 Hz
Option F06
Resolution
Frequency switching speed 1
≤ 5 ms, nominal
Frequency reference (M9300A PXIe frequency reference module)
Reference outputs
100 MHz Out (Out 1 through Out 5)
Amplitude
Connectors
Impedance
10 MHz Out
Amplitude
Connectors
Impedance
OCXO Out
Amplitude
Connectors
Impedance
≥ 10 dBm
13 dBm, typical
5 SMB snap-on
50 Ω, nominal
9.5 dBm, nominal
1 SMB snap-on
50 Ω, nominal
11.5 dBm, nominal
1 SMB snap-on
50 Ω, nominal
1. Mean time from IVI command to carrier frequency settled within 1 ppm or 1 kHz whichever is greater and amplitude settled within 0.2 dB (at the
controlled temperature range) or within 0.5 dB (at the full temperature range). If the ALC is off, the settle limit is 0.5 dB above +10 dBm, (at the
controlled temperature range). Simulataneous carrier frequency and amplitude switching.
06 | Keysight | M9380A PXIe CW Source 1 MHz to 3 GHz or 6 GHz - Data Sheett
Technical Specifications and Characteristics (continued)
Frequency (continued)
Frequency accuracy
Same as accuracy of internal time base or external reference input
Internal timebase
Accuracy
± (time since last adjustment x aging rate)
± temperature effects
± calibration accuracy
Frequency stability
Aging rate
Daily
< ±0.5 ppb/day, after 72 hour warm-up
< ±0.1 ppm/year, after 72 hours warm-up
< ±0.6 ppm/10yrs, after 72 hours warm-up
±5 x 10 -8
Yearly
Total 10 years
Achievable initial calibration accuracy (at time of shipment)
Temperature effects
20 to 30 °C
< ±10 ppb
< ±50 ppb
Full temperature range
Warm up
5 minutes over +20 to +30 °C, with respect to 1 hour
< ±0.1 ppm
15 minutes over +20 to +30 °C, with respect to 1 hour
< ±0.01 ppm
External reference input
Frequency
1 MHz to 110 MHz, sine wave
±1 ppm, nominal
Lock range
Amplitude
0 to 10 dBm, nominal
1 SMB snap-on
Connector
Impedance
50 Ω, nominal
07 | Keysight | M9380A PXIe CW Source 1 MHz to 3 GHz or 6 GHz - Data Sheett
Technical Specifications and Characteristics (continued)
Amplitude
Output parameters
Settable range
Standard
Option 1EA
+10.7 to -130 dBm
+20 to -130 dBm
Resolution
ALC on 1
0.02 dB, nominal
0.3 dB, nominal
ALC off
Maximum output power
Frequency
Standard
+10 dBm
+10 dBm
Option 1EA
+19 dBm
+18 dBm
1 MHz to 5 GHz
> 5 to 6 GHz
Absolute level accuracy in CW mode [ALC on] 2
< Max power to -20 dBm
< -20 to -110 dBm
< -110 to -120 dBm
<-120 to -130 dBm
Frequency
1 MHz to 3 GHz
±0.4 dB
±0.5 dB
±0.7 dB
±0.8 dB, nominal
±0.15 dB, typical
±0.15 dB, typical
±0.25 dB, typical
> 3 to 6 GHz
±0.5 dB
±0.6 dB
±1.0 dB
±0.8 dB, nominal
±0.15 dB, typical
±0.25 dB, typical
±0.5 dB, typical
1. Settable to 0.01 dB.
2. Specifications apply at the controlled temperature range. For temperatures outside this range, absolute level accuracy degrades by ± 0.02 dB/°C.
-120 dBm Output Power Normalized Response
1
0.75
0.5
0.25
0
-0.25
-0.5
-0.75
-1
0
1
2
3
4
5
6
Frequency (GHz)
Figure 3. Output power normalized response at –120 dBm.
08 | Keysight | M9380A PXIe CW Source 1 MHz to 3 GHz or 6 GHz - Data Sheett
Technical Specifications and Characteristics (continued)
Amplitude (continued)
Amplitude Repeatability +10 dBm ALC On
0.05
850 MHz
0.04
0.03
0.02
0.01
0
1900 MHz
2200 MHz
3500 MHz
5800 MHz
-0.01
-0.02
-0.03
-0.04
-0.05
0
20
40
60
80
100
120
Elapsed Time (minutes)
Figure 4. Amplitude repeatability at various carrier frequencies. Repeatability measures the ability of the
instrument to return to a given power setting after a random excursion to any other frequency and power setting.
It should not be confused with absolute level accuracy.
Relative Level Accuracy
ALC On
0.6
0.85 GHz
1.8 GHz
2 GHz
0.4
2.2 GHz
3.5 GHz
5.8 GHz
0.2
0
-0.2
-0.4
-0.6
-120
-100
-80
-60
-40
-20
0
20
Power Level (dBm)
Figure 5. Relative level accuracy at various carrier frequencies.
09 | Keysight | M9380A PXIe CW Source 1 MHz to 3 GHz or 6 GHz - Data Sheett
Technical Specifications and Characteristics (continued)
Amplitude (continued)
VSWR
1 MHz to 6 GHz
< 1.5:1, nominal
Maximum reverse power
1 MHz to 6 GHz
1 W, nominal
Max DC voltage
25 VDC, nominal
VSWR
1.50
1.45
1.40
1.35
1.30
1.25
1.20
1.15
1.10
1.05
1.00
0 dB Aꢀn
30 dB Aꢀn
Bypass Path
0
1
2
3
4
5
6
Frequency (GHz)
Figure 6. Measured VSWR from 1 MHz to 6 GHz.
10 | Keysight | M9380A PXIe CW Source 1 MHz to 3 GHz or 6 GHz - Data Sheett
Technical Specifications and Characteristics (continued)
Spectral purity
Phase noise at 20 kHz offset
1 GHz
2 GHz
3 GHz
6 GHz
-122 dBc/Hz, typical
-117 dBc/Hz, typical
-112 dBc/Hz, typical
-108 dBc/Hz, typical
Single Sideband Phase Noise in CW Mode
Normal Loop Bandwidth
-20
-30
-40
1GHz
-50
2GHz
2.4GHz
3GHz
-60
-70
5.8GHz
6GHz
-80
-90
-100
-110
-120
-130
-140
-150
-160
10 Hz
100 Hz
1 kHz
10 kHz
100 kHz
1 MHz
10 MHz
Frequency Offset From Carrier
Figure 7. Single sideband phase noise in normal loop bandwidth from 10 Hz to 10
MHz, offset at 1, 2, 2.4, 3, 5.8, and 6 GHz.
Single Sideband Phase Noise in CW Mode
Narrow Loop Bandwidth
-20
-30
-40
1GHz
-50
2GHz
-60
-70
2.4GHz
3GHz
5.8GHz
6GHz
-80
-90
-100
-110
-120
-130
-140
-150
-160
10 Hz
100 Hz
1 kHz
10 kHz
100 kHz
1 MHz
10 MHz
Frequency Offset From Carrier
Figure 8. Single sideband phase noise in narrow loop bandwidth from 10 Hz to 10 MHz,
offset at 1, 2, 2.4, 3, 5.8, and 6 GHz.
11 | Keysight | M9380A PXIe CW Source 1 MHz to 3 GHz or 6 GHz - Data Sheett
Technical Specifications and Characteristics (continued)
Spectral purity (continued)
Broadband noise floor
Range
1 MHz to 6 GHz
< -140 dBc/Hz, nominal, at +10 dBm output power level
Harmonics
Range
≤ 0 dBm
< -43 dBc
< -29 dBc
< -35 dBc
≤ +10 dBm
1 MHz to < 400 MHz
400 MHz to 1.5 GHz
> 1.5 GHz to 3 GHz
-46 dBc, typical
-31 dBc, typical
-39 dBc, typical
< -35 dBc
< -27 dBc
< -30 dBc
-37 dBc, typical
-29 dBc, typical
-33 dBc, typical
Nonharmonics 1
Nonharmonic miscellaneous spurious 2
Nonharmonic HET band mixing spurs (0 dBm)
Nonharmonic Frac-N
< -70 dBc, nominal
< -67 dBc, nominal
< -66 dBc, nominal
Subharmonics
1 MHz to 6 GHz
none
Analog modulation
Pulse parameters
Pulse on/off ratio 1 MHz to 400 MHz
Pulse on/off ratio > 400 MHz to 6 GHz
Pulse rise/fall time
> 85 dB, typical
> 95 dB, typical
< 10 ns, nominal
System requirements
Topic
Windows 7 requirements
Operating systems
Processor speed
Windows 7 (32-bit and 64-bit)
1 GHz 32-bit (x86), 1 GHz 64-bit (x64)
(no support for Itanium 64)
Available memory
4 GB minimum
8 GB or greater recommended
Available
disk space 3
1.5 GB available hard disk space, includes:
– 1 GB available for Microsoft .NET framework 3.5 SP1 4
– 100 MB for Keysight IO libraries suite
Video
Support for DirectX 9 graphics with 128 MB graphics memory recommended
(Super VGA graphics is supported)
Browser
Microsoft Internet Explorer 7.0 or greater
Version 16.3.16603.3 or later
Keysight IO libraries
1. Non-harmonics include mixing spurs for frequencies below 400 MHz, synthesizer spurs, and other miscellaneous chassis and power supply products, for
offsets >10 kHz.
2. With a Keysight M9036A embedded controller.
3. Because of the installation procedure, less memory may be required for operation than is required for installation.
4. NET framework runtime components are installed by default with Windows Windows 7. Therefore, you may not need this amount of available disk space.
12 | Keysight | M9380A PXIe CW Source 1 MHz to 3 GHz or 6 GHz - Data Sheett
Technical Specifications and Characteristics (continued)
Environmental and physical specifications
Temperature
Humidity 1
Operating
Non-operating (storage)
0 to 55 °C
-40 to +70 °C
Type tested at 95%, +40 °C
(non-condensing)
Altitude
Connectors
EMC
Up to 15,000 feet (4,572 meters)
SMA female
RF OUT
Complies with European EMC Directive
2004/108/EC
– IEC/EN 61326-2-1
– CISPR Pub 11 Group 1, class A
– AS/NZS CISPR 11
– ICES/NMB-001
This ISM device complies with Canadian ICES-001.
Cet appareil ISM est conforme a la norme NMB-001 du Canada.
Warm-up time
45 minutes
Size
M9300A
M9301A
M9310A
1 PXIe slot
1 PXIe slot
1 PXIe slot
Dimensions
Module
M9300A
M9301A
M9310A
Length
Width
22 mm
22 mm
22 mm
Height
210 mm
210 mm
210 mm
130 mm
130 mm
130 mm
Weight
M9300A
M9301A
M9310A
0.551 kg (1.215 lbs)
0.535 kg (1.179 lbs)
0.551 kg (1.215 lbs)
Power drawn from chassis
M9300A
M9301A
M9310A
≤ 18 W
≤ 25 W
≤ 28 W
1. Samples of this product have been type tested in accordance with the Keysight environmental test manual and verified to be robust against the
environmental stresses of storage, transportation and end-use--those stresses include but are not limited to temperature, humidity, shock, vibration,
altitude and power-line conditions. Test methods are aligned with IEC 60068-2 and levels are similar to MIL-PRF-28800F Class 3.
13 | Keysight | M9380A PXIe CW Source 1 MHz to 3 GHz or 6 GHz - Data Sheett
Software
Instrument connection software
Keysight IO library
The IO suite offers a single entry point for connection to the most
Free software download at
common instruments including AXIe, PXI, GPIB, USB, Ethernet/LAN,
RS-232, and VXI test instruments from Keysight and other vendors.
It automatically discovers interfaces, chassis, and instruments. The
graphical user interface allows you to search for, verify and update
IVI instrument and soft front panel drivers for modular and traditional
instruments. The IO suite safely installs in side-by-side mode with NI
I/O software.
www.keysight.com/find/iosuite
Module setup and usage
Keysight soft front
panel
The PXI module includes a soft front panel (SFP), a software-based
graphical user interface (GUI) which enables the instrument’s
capabilities from your PC.
Included on CD-ROM shipped with
module or online
Programming
Driver
Development environments
lVI-COM
IVI-C
Visual Studio (VB.NET, C#, C/C++), VEE
LabVIEW, LabWindows/CVI, MATLAB
Included on CD-ROM shipped with
module or online
LabVIEW
MATLAB
Programming assitance
Command expert
Assists in finding the right instrument commands and setting correct
parameters. A simple interface includes documentation, examples,
syntax checking, command execution and debug tools to build
sequences for integration in Excel, MATLAB, Visual Studio, LabVIEW,
VEE, SystemVue.
Free software download at
www.keysight.com/find/
commandexpert
Programming
examples
Each module includes programming examples for Visual Studio.net,
LabVIEW, MATLAB, LabWindows, and Keysight VEE Pro.
Included on CD-ROM shipped with
module or online at
www.keysight.com/find/m9380a
14 | Keysight | M9380A PXIe CW Source 1 MHz to 3 GHz or 6 GHz - Data Sheett
Setup and Calibration Services
Assistance
One day startup
assistance
Gain access to a technical expert who will help you get started quickly with the
Included in base configuration
Included in base configuration
M9380A CW source and its powerful software tools. The flexible instruction
format is designed to get you to your first measurements and familiarize you with
ways to adapt the equipment to a specific application.
Calibration and traceability
Factory calibration
The M9380A CW source ships factory calibrated with an ISO-9002,
NIST-traceable calibration certificate.
Calibration cycle
Calibration sites
A one year calibration cycle is recommended.
– At Keysight worldwide service centers
– On-site by Keysight
More information visit
www.keysight.com/find/infoline
– By self-maintainers
N7800A
The M9380A CW source is supported by Keysight’s Calibration and adjustment
Licensed software. For more information,
Calibration and
adjustment software
software. This is the same software used at Keysight service centers to automate visit
calibration. The software offers compliance tests for ISO 17025:2005,
ANSI/NCSL Z540.3-2006, and measurement uncertainty per ISO Guide to
Expression of Measurement Uncertainty.
www.keysight.com/find/calibrationsoftware
Keysight calibration
status utility
The Keysight calibration status utility helps ensure your M9380A is calibrated by Included in base configuration
managing the calibration interval and providing messages regarding instrument
and module calibration status.
15 | Keysight | M9380A PXIe CW Source 1 MHz to 3 GHz or 6 GHz - Data Sheett
Support
Support
Core exchange
program
Keysight’s replacement core exchange program allows fast and easy module repairs. A replacement
core assembly is a fully functioning pre-calibrated module replacement that is updated with the
defective module serial number, allowing the replacement module to retain the original serial number.
For qualified
self-maintainers in
US only
Self-test utility
A self-test utility runs a set of internal tests which verifies the health of the modules and reports their
status.
Included in base
configuration
16 | Keysight | M9380A PXIe CW Source 1 MHz to 3 GHz or 6 GHz - Data Sheett
Configuration and Ordering Information
Ordering information
Software information
Model
M9380A
Description
Supported operating
systems
Microsoft Windows 7
(32/64-bit)
PXIe CW source: 1 MHz to 3 or 6 GHz
Includes:
Standard compliant
IVI-COM, IVI-C, LabVIEW, MATLAB
drivers
M9301A PXIe Synthesizer
M9310A PXIe Source Output
One day startup assistance
Module interconnect cables
Software, example programs and product
information on CD
Supported application
development
environments (ADE)
VisualStudio (VB.NET, C#, C/C++), VEE,
LabVIEW, LabWindows/CVI, MATLAB
Keysight IO libraries
(version 16.3 or newer)
Includes: VISA Libraries, Keysight
Connection Expert, IO Monitor
Base configuration
Keysight Command expert Instrument control for SCPI or IVI-COM
M9380A-F03
Frequency range: 1 MHz to 3 GHz
drivers
M9380A-300
Required for
warranted
PXIe frequency reference: 10 and 100 MHz
Adds M9300A PXIe frequency reference: 10
and 100 MHz (M9300A module can support
multiple M9380A modular instruments)
Accessories
specifications
Model
Description
Configurable options
Y1212A
Y1213A
Y1214A
Y1215A
Slot blocker kit: 5 modules
PXI EMC filler panel kit: 5 slots
Air inlet kit: M9018A 18-slot chassis
Rack mount kit for M9018A 18-slot chassis
Frequency range
M9380A-F03
1 MHz to 3 GHz
1 MHz to 6 GHz
M9380A-F06
Power
Related products
M9380A-1EA
Calibration
M9380A-UK6
High output power
Model
Description
Commercial calibration certificate with test
data for M9380A (M9301A, M9310A)
M9021A
M9045B
Y1200B
M9048A
Y1202A
M9381A
M9300A
PCIe cable interface
PCIe ExpressCard adaptor for laptop connectivity
PCIe cable for laptop connectivity
PCIe desktop adaptor for desktop connectivity
PCIe Cable for desktop connectivity
PXIe Vector Signal Generator
PXIe Frequency Reference
M9300A-UK6
Commercial calibration certificate with test
data for M9300A (module only)
Related products in recommended configuration
M9037A
PXIe embedded controller
18-slot PXIe chassis
M9018A
Recommended configuration
Advantage services:
Calibration
Keysight Advantage Services is committed to your success throughout
your equipment’s lifetime
N7800A
Calibration & adjustment software
17 | Keysight | M9380A PXIe CW Source 1 MHz to 3 GHz or 6 GHz - Data Sheet
For more information on Keysight
Evolving Since 1939
Our unique combination of hardware, software, services, and people can help you
reach your next breakthrough. We are unlocking the future of technology.
From Hewlett-Packard to Agilent to Keysight.
Technologies’ products, applications or
services, please contact your local Keysight
office. The complete list is available at:
www.keysight.com/find/contactus
Americas
Canada
Brazil
Mexico
United States
(877) 894 4414
55 11 3351 7010
001 800 254 2440
(800) 829 4444
Asia Pacific
Australia
China
Hong Kong
India
Japan
Korea
Malaysia
Singapore
Taiwan
1 800 629 485
800 810 0189
800 938 693
1 800 11 2626
0120 (421) 345
080 769 0800
1 800 888 848
1 800 375 8100
0800 047 866
(65) 6375 8100
myKeysight
www.keysight.com/find/mykeysight
A personalized view into the information most relevant to you.
Other AP Countries
www.keysight.com/find/emt_product_registration
Register your products to get up-to-date product information and
find warranty information.
Europe & Middle East
Austria
Belgium
Finland
France
Germany
Ireland
Israel
Italy
Luxembourg
Netherlands
Russia
0800 001122
0800 58580
0800 523252
0805 980333
0800 6270999
1800 832700
1 809 343051
800 599100
+32 800 58580
0800 0233200
8800 5009286
800 000154
0200 882255
0800 805353
Opt. 1 (DE)
Keysight Services
www.keysight.com/find/service
Keysight Services can help from acquisition to renewal across your
instrument’s lifecycle. Our comprehensive service offerings—one-
stop calibration, repair, asset management, technology refresh,
consulting, training and more—helps you improve product quality
and lower costs.
Keysight Assurance Plans
www.keysight.com/find/AssurancePlans
Up to ten years of protection and no budgetary surprises to ensure
your instruments are operating to specification, so you can rely on
accurate measurements.
Spain
Sweden
Switzerland
Opt. 2 (FR)
Opt. 3 (IT)
0800 0260637
Keysight Channel Partners
www.keysight.com/find/channelpartners
United Kingdom
Get the best of both worlds: Keysight’s measurement expertise and
product breadth, combined with channel partner convenience.
For other unlisted countries:
www.keysight.com/find/contactus
(BP-9-7-17)
www.keysight.com/find/m9380a
www.keysight.com/find/modular
DEꢀRA Certiꢁied
ISO9001 ꢂuꢃlitꢄ Mꢃnꢃgement Sꢄstem
www.keysight.com/go/quality
Keysight Technologies, Inc.
DEKRA Certified ISO 9001:2015
Quality Management System
This information is subject to change without notice.
© Keysight Technologies, 2017
Published in USA, December 1, 2017
5991-0283EN
www.keysight.com
相关型号:
©2020 ICPDF网 联系我们和版权申明