5992-0193EN [KEYSIGHT]

M9290A CXA-m PXIe X-Series Signal Analyzer;
5992-0193EN
型号: 5992-0193EN
厂家: Keysight Technologies    Keysight Technologies
描述:

M9290A CXA-m PXIe X-Series Signal Analyzer

文件: 总18页 (文件大小:1450K)
中文:  中文翻译
下载:  下载PDF数据表文档文件
Keysight Technologies  
M9290A CXA-m  
PXIe X-Series Signal Analyzer  
10 Hz to 3.0, 7.5, 13.6 or 26.5 GHz  
Data Sheet  
Table of Contents  
Technical Specifications...................................................................3  
Definitions and conditions........................................................3  
Recommended best practices in use ......................................3  
Frequency and Time Specifications.........................................4  
Amplitude Accuracy and Range Specifications ......................6  
Dynamic Range Specifications ................................................9  
PowerSuite Measurement Specifications..............................12  
Tracking Generator Specifications ........................................13  
General Specifications ...........................................................14  
Inputs and Outputs.................................................................14  
I/Q Analyzer............................................................................15  
System Requirements ............................................................15  
Software .........................................................................................16  
Related Literature...........................................................................17  
Deploy a smaller  
microwave footprint  
In test system development, one  
of your crucial requirements is  
doing more in less space—but this  
often means tradeoffs between  
footprint and precision in signal  
analysis. That is, until now: we’ve  
expanded the Keysight X-Series  
with the CXA-m, a PXIe signal  
analyzer that offers fully specified  
performance up to 26.5 GHz. It  
lets you handle RF and microwave  
signals in four slots, and you  
can leverage your existing code.  
Choose the CXA-m and deploy a  
smaller microwave footprint.  
03 | Keysight | M9290A CXA-m PXIe X-Series Signal Analyzer - Data Sheet  
Technical Specifications  
Get more information  
Definitions and conditions  
Temperatures referred to in this document are defined as follows:  
– Full temperature range = Individual module temperature of 5 to 68 °C, as reported by  
the module, and environment temperature of 0 to 55 °C.  
– Controlled temperature range = Individual module temperature of 25 to 40 °C, as  
reported by the module, and environment temperature of 20 to 30 °C.  
This CXA-m PXIe X-Series signal  
analyzer data sheet is a summary  
of the complete specifications  
and conditions available in the  
CXA-m PXIe Signal Analyzer  
Specification Guide. The CXA-m  
PXIe Signal Analyzer Specifica-  
tion Guide can be obtained on  
the web at:  
Specifications describe the warranted performance of calibrated instruments. Data  
represented in this document are specifications under the following conditions unless  
otherwise noted.  
– It is within its calibration cycle  
www.keysight.com/find/  
cxa-m_manuals  
– Under auto couple control, except when Auto Sweep Time Rules = Accy  
– The analyzer has been stored at an ambient temperature within the allowed  
operating range for at least two hours before being turned on; if it had previously  
been stored at a temperature range inside the allowed storage range, but outside  
the allowed operating range  
For ordering information, refer to  
the CXA-m PXIe Signal Analyzer  
Configuration Guide  
– The analyzer has been turned on at least 30 minutes with Auto Align set to normal,  
or, if Auto Align is set to off or partial, alignments must have been run recently  
enough to prevent an Alert message; if the Alert condition is changed from Time and  
Temperature to one of the disabled duration choices, the analyzer may fail to meet  
specifications without informing the user  
(5992-0193EN).  
95th percentile values indicate the breadth of the population (approx. 2σ) of perfor-  
mance tolerances expected to be met in 95 percent of the cases with a 95 percent  
confidence, for any ambient temperature in the range of 20 to 30 °C. In addition to the  
statistical observations of a sample of instruments, these values include the effects of  
the uncertainties of external calibration references. These values are not warranted.  
These values are updated occasionally if a significant change in the statistically observed  
behavior of production instruments is observed.  
Typical describes additional product performance information that is not covered by  
the product warranty. It is performance beyond specifications that 80 percent of the  
units exhibit with a 95 percent confidence level over the temperature range 20 to 30 °C.  
Typical performance does not include measurement uncertainty.  
Nominal values indicate expected performance, or describe product performance that is  
useful in the application of the product, but are not covered by the product warranty.  
Recommended best practices in use  
– Use slot blockers and EMC filler panels in empty module slots to ensure proper  
operating temperatures. Keysight chassis and slot blockers optimize module  
temperature performance and reliability of test.  
– Set chassis fan to high at environmental temperatures above 45°C.  
04 | Keysight | M9290A CXA-m PXIe X-Series Signal Analyzer - Data Sheet  
Frequency and Time Specifications  
Frequency range  
DC coupled  
AC coupled  
Option F03  
Option F07  
Option F13  
Option F26  
10 Hz to 3.0 GHz  
10 Hz to 7.5 GHz  
10 Hz to 13.6 GHz  
10 Hz to 26.5 GHz  
10 MHz to 3.0 GHz  
10 MHz to 7.5 GHz  
10 MHz to 13.6 GHz  
10 MHz to 26.5 GHz  
Band  
LO multiple (N)  
0
1
2
3
4
4
5
6
7
8
9
10  
1
2
2
2
2
4
4
4
4
4
4
4
10 Hz to 3.08 GHz  
2.95 to 7.575 GHz  
7.45 to 9.55 GHz  
9.45 to 12.60 GHz  
12.50 to 13.05 GHz  
12.95 to 13.80 GHz  
13.40 to 15.55 GHz  
15.45 to 19.35 GHz  
19.25 to 21.05 GHz  
20.95 to 22.85 GHz  
22.75 to 24.25 GHz  
24.15 to 26.55 GHz  
Frequency reference  
Accuracy  
± [(time since last adjustment x aging rate) + temperature stability + calibration accuracy]  
Aging rate  
Option PFR  
Standard  
± 1 x 10-7 / year  
± 1.5 x 10-7 / 2 years  
Option PFR  
± 1 x 10-6 / year  
Temperature stability  
20 to 30 °C  
Full temperature range  
Achievable initial calibration accuracy  
Standard  
± 2 x 10-6  
± 2 x 10-6  
Standard  
± 1.4 x 10-6  
± 1.5 x 10-8  
± 5 x 10-8  
Option PFR  
± 4 x 10-8  
Example frequency reference accuracy  
(with Option PFR)  
1 year after last adjustment  
Residual FM  
= ± (1 x 1 x 10-7 + 5 x 10-8 + 4 x 10-8)  
= ± 1.9 x 10-7  
Option PFR  
Standard  
≤ 0.25 Hz p-p in 20 ms nominal  
≤ 10 Hz p-p in 20 ms nominal  
Frequency readout accuracy (start, stop, center, marker)  
± (marker frequency x frequency reference accuracy + 0.25 % x span + 5 % x RBW + 2 Hz + 0.5 x horizontal resolution 1)  
Marker frequency counter  
Accuracy  
Delta counter accuracy  
Counter resolution  
± (marker frequency x frequency reference accuracy + 0.100 Hz)  
± (delta frequency x frequency reference accuracy + 0.141 Hz)  
0.001 Hz  
Frequency span (FFT and swept mode)  
Range  
0 Hz (zero span), 10 Hz to maximum frequency of instrument  
2 Hz  
Resolution  
Accuracy  
Swept  
± (0.25 % x span + horizontal resolution)  
± (0.10 % x span + horizontal resolution)  
FFT  
1. Horizontal resolution is span/(sweep points – 1)  
05 | Keysight | M9290A CXA-m PXIe X-Series Signal Analyzer - Data Sheet  
Frequency and Time Specifications (Continued)  
Sweep time and triggering  
Range  
Span = 0 Hz  
Span ≥ 10 Hz  
Span ≥ 10 Hz, swept  
Span ≥ 10 Hz, FFT  
Span = 0 Hz  
1 μs to 6000 s  
1 ms to 4000 s  
± 0.01 % nominal  
± 40 % nominal  
± 1 % nominal  
Accuracy  
Trigger  
Trigger delay  
Free run, video, external, RF burst, periodic timer  
Span = 0 Hz or FFT  
Span ≥ 10 Hz, swept  
Resolution  
–150 to +500 ms  
1 μs to 500 ms  
0.1 μs  
Time gating  
Gate methods  
Gate length range (except method = FFT)  
Gate delay range  
Gated LO; gated video; gated FFT  
100.0 ns to 5.0 s  
0 to 100.0 s  
Gate delay jitter  
33.3 ns p-p nominal  
Sweep (trace) point range  
All spans  
1 to 40001  
Resolution bandwidth (RBW)  
Range (–3.01 dB bandwidth)  
Bandwidth accuracy (power)  
1 Hz to 3 MHz (10 % steps), 4, 5, 6, 8 MHz  
1 Hz to 750 kHz  
820 kHz to 1.2 MHz  
1.3 to 2.0 MHz  
± 1.0 % (± 0.044 dB) nominal  
± 2.0 % (± 0.088 dB) nominal  
± 0.13 dB nominal  
2.2 to 3 MHz  
± 0.22 dB nominal  
4 to 8 MHz  
± 0.32 dB nominal  
Bandwidth accuracy (–3.01 dB)  
RBW range  
1 Hz to 1.3 MHz  
± 2 % nominal  
Selectivity (–60 dB/–3 dB)  
EMI bandwidth (CISPR compliant)  
EMI bandwidth (MIL STD 461E compliant)  
Analysis bandwidth 1  
4.1:1 nominal  
200 Hz, 9 kHz, 120 kHz, 1 MHz  
10 Hz, 100 Hz, 1 kHz, 10 kHz, 100 kHz, 1 MHz  
(Option EMC)  
(Option EMC)  
Maximum bandwidth  
Option B25  
Standard  
25 MHz  
10 MHz  
Video bandwidth (VBW)  
Range  
Accuracy  
Measurement speed 2  
1 Hz to 3 MHz (10 % steps), 4, 5, 6, 8 MHz, and wide open (labeled 50 MHz)  
± 6 % nominal  
Local measurement and display update rate  
Remote measurement and LAN transfer rate  
Marker peak search  
11 ms (90/s) nominal  
6 ms (167/s) nominal  
5 ms nominal  
Center frequency tune and transfer  
Measurement/mode switching  
22 ms nominal  
75 ms nominal  
1. Analysis bandwidth is the instantaneous bandwidth available around a center frequency over which the input signal can be digitized for further analysis or  
processing in the time, frequency, or modulation domain.  
2. Sweep points = 101.  
06 | Keysight | M9290A CXA-m PXIe X-Series Signal Analyzer - Data Sheet  
Amplitude Accuracy and Range Specifications  
Amplitude range  
Measurement range  
Preamp off  
Displayed average noise level (DANL) to +30 dBm  
Displayed average noise level (DANL) to +23 dBm  
Preamp on  
Input attenuator range  
Standard  
0 to 70 dB in 10 dB steps  
Option FSA  
0 to 70 dB in 2 dB steps, 7.5 GHz  
0 to 70 dB in 10 dB steps, 7.5 to 26.5 GHz  
Maximum safe input level  
Average total power  
+30 dBm (1 W)  
Input attenuation ≥ 10 dB, preamp off  
Input attenuation ≥ 20 dB, preamp on  
+30 dBm (1 W)  
Peak pulse power  
+50 dBm (100 W)  
DC volts  
< 10 μs pulse width, < 1 % duty cycle, and input attenuation ≥ 30 dB  
AC coupled  
DC coupled  
Display range  
Log scale  
± 50 Vdc  
± 0.2 Vdc  
0.1 to 1 dB/division in 0.1 dB steps  
1 to 20 dB/division in 1 dB steps (10 display divisions)  
10 divisions  
Linear scale  
Scale units  
dBm, dBmV, dBμV, dBmA, dBμA, V, W, A  
Frequency response  
Specification  
95th percentile (≈ 2σ)  
(10 dB input attenuation, 20 to 30 °C, σ = nominal standard deviation)  
9 kHz to 10 MHz  
10 MHz to 3 GHz  
3 to 13.6 GHz  
13.6 to 19.3 GHz  
19.3 to 24.2 GHz  
± 0.50 dB  
± 0.65 dB  
± 1.30 dB  
± 1.50 dB  
± 2.20 dB  
± 2.50 dB  
± 0.4 dB  
± 0.5 dB  
± 0.8 dB  
± 1.0 dB  
± 1.3 dB  
± 1.3 dB  
24.2 to 26.5 GHz  
Preamp on (Option P03, P07, P13, P26)  
100 kHz to 10 MHz  
± 0.5 dB  
10 MHz to 3 GHz  
± 1.0 dB  
3 to 7.5 GHz  
± 1.2 dB  
7.5 to 13.6 GHz  
± 1.0 dB  
13.6 to 21 GHz  
± 1.2 dB  
21 to 24.2 GHz  
± 1.8 dB  
24.2 to 26.5 GHz  
± 2.4 dB  
Input attenuation switching uncertainty  
Specifications  
± 0.3 dB  
Additional information  
± 0.15 dB typical  
± 0.30 dB nominal  
± 0.50 dB nominal  
± 0.70 dB nominal  
Attenuation > 2 dB, preamp off  
Relative to 10 dB  
(reference setting)  
50 MHz (reference frequency)  
100 kHz to 3.0 GHz  
3.0 to 7.5 GHz  
7.5 to 26.5 GHz  
07 | Keysight | M9290A CXA-m PXIe X-Series Signal Analyzer - Data Sheet  
Amplitude Accuracy and Range Specifications (Continued)  
Total absolute amplitude accuracy  
(10 dB attenuation, 20 to 30 °C, 1 Hz ≤ RBW ≤ 1 MHz, input signal –10 to –50 dBm, all settings auto-coupled except Auto Swp Time = Accy, any  
reference level, any scale, σ = nominal standard deviation)  
At 50 MHz  
± 0.40 dB  
At all frequencies  
100 kHz to 3 GHz  
Preamp on (Option P03/P07/P13/P26)  
Precision amplitude accuracy (Option PAA)  
Preamp off (10 dB attenuation)  
9 kHz to 10 MHz  
10 MHz to 3 GHz  
3 to 7.5 GHz  
7.5 to 9.55 GHz  
9.55 to 12.55 GHz  
12.55 to 13.6 GHz  
13.6 to 15.55 GHz  
15.55 to 22.8 GHz  
22.8 to 24.2 GHz  
± (0.40 dB + frequency response)  
± 0.60 dB (95th Percentile ≈ 2σ)  
± (0.36 dB + frequency response) (95th percentile)  
Specification  
± 0.70 dB  
± 0.85 dB  
± 0.95 dB  
± 0.90 dB  
± 0.95 dB  
± 0.90 dB  
± 0.95 dB  
± 1.10 dB  
± 1.20 dB  
± 1.30 dB  
Additional information  
± 0.31 dB typical  
± 0.40 dB typical  
± 0.44 dB typical  
± 0.38 dB typical  
± 0.44 dB typical  
± 0.32 dB typical  
± 0.35 dB typical  
± 0.50 dB typical  
± 0.50 dB typical  
± 0.57 dB typical  
24.2 to 26.5 GHz  
Preamp on (0 dB attenuation)  
100 kHz to 10 MHz  
10 MHz to 3 GHz  
3 to 7.5 GHz  
7.5 to 9.55 GHz  
9.55 to 12.55 GHz  
12.55 to 19.3 GHz  
19.3 to 21 GHz  
21 to 22.8 GHz  
22.8 to 24.2 GHz  
± 0.30 dB typical  
± 0.65 dB typical  
± 0.73 dB typical  
± 0.49 dB typical  
± 0.71 dB typical  
± 0.65 dB typical  
± 0.80 dB typical  
± 1.20 dB typical  
± 1.39 dB typical  
± 1.66 dB typical  
24.2 to 26.5 GHz  
Input voltage standing wave ratio (VSWR) (≥ 10 dB attenuation)  
10 MHz to 3 GHz  
3 to 7.5 GHz  
7.5 to 13.6 GHz  
13.6 to 24.2 GHz  
24.2 to 26.5 GHz  
< 1.2 nominal  
< 1.4 nominal  
< 1.6 nominal  
< 1.8 nominal  
< 2.2 nominal  
Resolution bandwidth switching uncertainty (referenced to 30 kHz RBW)  
1 Hz to 3 MHz RBW  
4, 5, 6, 8 MHz RBW  
Reference level  
± 0.15 dB  
± 1.0 dB  
Range  
Log scale  
Linear scale  
Accuracy  
–170 to +23 dBm in 0.01 dB steps  
Same as log (707 pV to 3.16 V)  
0 dB  
Display scale switching uncertainty  
Switching between linear and log  
Log scale/div switching  
0 dB  
0 dB  
08 | Keysight | M9290A CXA-m PXIe X-Series Signal Analyzer - Data Sheet  
Amplitude Accuracy and Range Specifications (Continued)  
Display scale fidelity  
–80 dBm ≤ input mixer level < –10 dBm  
± 0.15 dB total  
Trace detectors  
Normal, peak, sample, negative peak, log power average, RMS average, and voltage average  
Preamplifier (Option P03/P07/P13/P26)  
Frequency range  
Option P03  
Option P07  
Option P13  
Option P26  
100 kHz to 26.5 GHz  
10 MHz to 3 GHz  
3 to 26.5 GHz  
100 kHz to 3.0 GHz  
100 kHz to 7.5 GHz  
100 kHz to 13.6 GHz  
100 kHz to 26.5 GHz  
+20 dB nominal  
Gain  
Noise figure  
10 dB nominal  
DANL + 176.24 dB nominal  
09 | Keysight | M9290A CXA-m PXIe X-Series Signal Analyzer - Data Sheet  
Dynamic Range Specifications  
1 dB gain compression (two-tone)  
Total power at input mixer  
+6 dBm nominal  
+4 dBm nominal  
–15 dBm nominal  
–19 dBm nominal  
Preamp off  
10 MHz to 7.5 GHz  
7.5 to 26.5 GHz  
10 MHz to 7.5 GHz  
7.5 to 26.5 GHz  
Preamp on  
Displayed average noise level (DANL)  
(Input terminated, sample or average detector, averaging type = Log, 0 dB input attenuation, IF Gain = High, 20 to 30 °C)  
Parentheses indicate typical performance  
Preamplifier OFF  
–95 dBm nominal  
–110 dBm nominal  
–115 dBm nominal  
(–125) dBm  
–144, (–148) dBm  
–148, (–150) dBm  
–146, (–149) dBm  
–141, (–145) dBm  
–144, (–147) dBm  
–136, (–140) dBm  
–142, (–145) dBm  
–132, (–138) dBm  
–134, (–139) dBm  
–132, (–137) dBm  
–128, (–133) dBm  
Preamplifier ON  
10 Hz  
100 Hz  
1 kHz  
9 kHz to 1 MHz  
1 to 10 MHz  
10 MHz to 1.5 GHz  
1.5 to 4.5 GHz  
4.5 to 7 GHz  
7 to 9.5 GHz  
9.5 to 13 GHz  
13 to 14.5 GHz  
14.5 to 19.3 GHz  
19.3 to 23 GHz  
23 to 24 GHz  
24 to 26.5 GHz  
Spurious responses  
Residual response  
(Input terminated and 0 dB attenuation)  
–154, (–158) dBm  
–160, (–163) dBm  
–160, (–163) dBm  
–157, (–161) dBm  
–158, (–160) dBm  
–156, (–160) dBm  
–158, (–161) dBm  
–153, (–157) dBm  
–152, (–157) dBm  
–150, (–155) dBm  
–144, (–149) dBm  
200 kHz to 26.5 GHz (swept)  
Zero span or FFT or other  
frequencies  
–90 dBm  
–100 dBm nominal  
Tuned frequency (f)  
Excitation Freq  
Mixer level  
Response  
Image responses (Second mixer)  
10 MHz to 7.5 GHz  
7.5 GHz to 19.3 GHz  
19.3 GHz to 21 GHz  
21 GHz to 22.8 GHz  
22.8 GHz to 24.2 GHz  
24.2 GHz to 26.5 GHz  
10 MHz to 26.5 GHz  
Mixer level  
f + 1645 MHz  
f - 1645 MHz  
f + 1645 MHz  
f - 1645 MHz  
f + 1645 MHz  
f - 1645 MHz  
−10 dBm  
−10 dBm  
−10 dBm  
−10 dBm  
−10 dBm  
−10 dBm  
−10 dBm  
−70 dBc (−80 dBc typical)  
−70 dBc (−80 dBc typical)  
−70 dBc (−80 dBc typical)  
−70 dBc (−80 dBc typical)  
−70 dBc (−80 dBc typical)  
−70 dBc (−80 dBc typical)  
−64 dBc typical  
LO-related spurious  
Other spurious responses  
IF feedthrough  
Response  
−10 dBm  
−75 dBc (−80 dBc typical)  
−70 dBc (−80 dBc nominal)  
−70 dBc (−80 dBc nominal)  
First RF order (f ≥ 10 MHz from carrier) −10 dBm  
High RF order (f ≥ 10 MHz from carrier) −30 dBm  
Second harmonic distortion (SHI)  
Source frequency  
10 MHz to 3.75 GHz  
3.75 to 13.25 GHz  
SHI (nominal)  
+50 dBm  
+62 dBm  
10 | Keysight | M9290A CXA-m PXIe X-Series Signal Analyzer - Data Sheet  
Dynamic Range Specifications (Continued)  
Third-order intermodulation distortion (TOI)  
Parentheses indicate typical performance  
Preamp off  
10 MHz to 2 GHz  
2 to 3 GHz  
3 to 7.5 GHz  
7.5 to 13.6 GHz  
13.6 to 26.5 GHz  
10 MHz to 26.5 GHz  
+12 dBm, (+16) dBm  
+12 dBm, (+17) dBm  
+12 dBm, (+16) dBm  
+11 dBm, (+15) dBm  
+10 dBm, (+14) dBm  
–8 dBm nominal  
(Two –20 dBm tones at input mixer  
spaced by 100 kHz, 0 dB attenuation,  
20 to 30 °C  
Option P03/P07/P13/P26  
Preamp on  
(Two –45 dBm tones at the preamp input,  
spaced by 100 kHz, 0 dB attenuation,  
20 to 30 °C)  
Figure 1. Nominal dynamic range for Band 0,  
for second and third order distortion,  
10 MHz to 3 GHz  
Figure 2. Nominal dynamic range, for  
second and third order distortion, 3 GHz to  
26.5 GHz  
11 | Keysight | M9290A CXA-m PXIe X-Series Signal Analyzer - Data Sheet  
Dynamic Range Specifications (Continued)  
Phase noise  
Offset  
Specification  
Typical  
Noise sidebands (20 to 30 °C, CF = 1 GHz)  
100 Hz  
1 kHz  
–90 dBc/Hz nominal  
–105 dBc/Hz  
–102 dBc/Hz  
–106 dBc/Hz  
–108 dBc/Hz  
–130 dBc/Hz  
10 kHz  
100 kHz  
1 MHz  
10 MHz  
–110 dBc/Hz  
–110 dBc/Hz  
–132 dBc/Hz  
–145 dBc/Hz nominal  
Figure 3. Nominal phase noise at different  
center frequencies  
12 | Keysight | M9290A CXA-m PXIe X-Series Signal Analyzer - Data Sheet  
PowerSuite Measurement Specifications  
Channel power  
Amplitude accuracy, W-CDMA or IS95  
(20 to 30 °C, attenuation = 10 dB)  
Occupied bandwidth  
± 1.23 dB (± 0.62 dB 95th percentile)  
Frequency accuracy  
± [span/1000] nominal  
Adjacent channel power  
Accuracy, W-CDMA (ACLR)  
(at specific mixer levels and ACLR ranges)  
MS  
Adjacent  
Alternate  
± 0.21 dB  
± 0.89 dB  
± 0.25 dB  
± 0.67 dB  
BTS  
Dynamic range (typical)  
Without noise correction  
With noise correction  
1 to 6  
–68 dB  
–73 dB  
–72 dB  
–78 dB  
Offset channel pairs measured  
Multiple number of carriers measured  
Power statistics CCDF  
Histogram resolution  
Harmonic distortion  
Up to 12  
0.01 dB  
10th  
Maximum harmonic number  
Results  
Fundamental power (dBm), relative harmonics power (dBc), total harmonic distortion in %  
Intermod (TOI)  
Measure the third-order products and intercepts from two tones  
Burst power  
Methods  
Results  
Power above threshold, power within burst width  
Single burst output power, average output power, maximum power, minimum power within burst, burst  
width  
Spurious emission  
W-CDMA (1 to 3.0 GHz) table-driven spurious signals; search across regions  
Dynamic range  
75.0 dB  
(79.1 dB typical)  
Absolute sensitivity  
−82.5 dBm  
(−86.5 dBm typical)  
Spectrum emission mask (SEM)  
cdma2000® (750 kHz offset)  
Relative dynamic range (30 kHz RBW)  
Absolute sensitivity  
73.0 dB  
−99.7 dBm  
± 0.11 dB  
(78.1 dB typical)  
(−102.7 dBm typical)  
Relative accuracy  
3GPP W-CDMA (2.515 MHz offset)  
Relative dynamic range (30 kHz RBW)  
Absolute sensitivity  
77.5 dB  
−97.7 dBm  
± 0.15 dB  
(82.4 dB typical)  
(−101.7 dBm typical)  
Relative accuracy  
13 | Keysight | M9290A CXA-m PXIe X-Series Signal Analyzer - Data Sheet  
Tracking Generator Specifications  
Output frequency  
Frequency range  
Option T03  
Option T07  
Option T13  
2 MHz to 3 GHz  
2 MHz to 7.5 GHz  
2 MHz to 13.6 GHz  
2 MHz to 26.5 GHz  
10 Hz  
Option T26  
Resolution  
Output power level  
Range  
2 MHz to 10 GHz  
10 to 20 GHz  
20 to 26.5 GHz  
—35 to 0 dBm  
—35 to —5 dBm  
—40 to —12 dBm  
0.1 dB  
Resolution  
Absolute accuracy (at 50 MHz, –15 dBm)  
Output flatness (Referenced to 50 MHz, –15 dBm)  
2 MHz to 7.5 GHz  
7.5 to 13.6 GHz  
13.6 to 23 GHz  
23 to 26 GHz  
26 to 26.5 GHz  
Level accuracy  
2 MHz to 7.5 GHz  
7.5 to 13.6 GHz  
13.6 to 23GHz  
± 1.0 dB typical  
95th percentile (≈ 2σ)  
± 1.0 dB  
± 1.2 dB  
± 1.8 dB  
± 0.3 dB nominal  
± 0.3 dB  
± 0.3 dB  
± 0.6 dB  
± 1.2 dB  
± 2.3 dB  
Nominal  
± 0.8 dB  
± 0.9 dB  
± 1.5 dB  
± 1.8 dB  
± 2.9 dB  
± 2.5 dB  
23 to 26 GHz  
26 to 26.5GHz  
Output power sweep  
Range  
2 MHz to 10 GHz  
10 to 20 GHz  
20 to 26.5 GHz  
—35 to 0 dBm  
—35 to —5 dBm  
—40 to —12 dBm  
0.1 dB  
Resolution  
Maximum safe reverse level  
Average total power  
AC coupled  
+ 30 dBm (1 W)  
± 50 Vdc  
Phase noise  
Noise sidebands (Center Frequency = 1 GHz)  
Offset  
Typical  
10 kHz  
100 kHz  
1 MHz  
—98 dBc/Hz  
—105 dBc/Hz  
—122 dBc/Hz  
Dynamic range  
Maximum output power – displayed average 110 dBc nominal  
Output VSWR  
2 MHz to 7 GHz  
7 to 23 GHz  
< 1.7:1 nominal  
< 2.5:1 nominal  
< 3.5:1 nominal  
23 to 26.5 GHz  
14 | Keysight | M9290A CXA-m PXIe X-Series Signal Analyzer - Data Sheet  
General Specifications  
Temperature range  
Operating  
Storage  
EMC  
0 to 55 °C  
–40 to 70 °C  
Complies with European EMC Directive 2004/108/EC  
– IEC/EN 61326-1  
– CISPR Pub 11 Group 1, class A  
– AS/NZS CISPR 11  
– ICES/NMB-001  
This ISM device complies with Canadian ICES-001  
Cet appareil ISM est conforme à la norme NMB-001 du Canada  
Environmental stress  
Samples of this product have been type tested in accordance with the Keysight Environmental Test Manual and verified to be robust against the environ-  
mental stresses of storage, transportation, and end-use; those stresses include, but are not limited to, temperature, humidity, shock, vibration, altitude,  
and power line conditions; test methods are aligned with IEC 60068-2 and levels are similar to MILPRF-28800F Class 3.  
Power requirement  
Power drawn from chassis  
Weight  
≤ 65 W  
Net  
Shipping  
1.9 kg (4.2 lbs)  
4.2 kg (9.3 lbs)  
Dimensions  
Height  
Width  
132 mm (5.2 in)  
82 mm (3.2 in)  
202 mm (8.0 in)  
Length  
Calibration cycle  
The recommended calibration cycle is one year; calibration services are available through Keysight service centers  
Inputs and Outputs  
RF input  
Connector  
RF output  
Connector  
10 MHz in  
Connector  
10 MHz out  
Connector  
Trigger in  
Connector  
Trigger out  
Connector  
Analog out  
Connector  
Noise source  
3.5mm female, 50 Ω nominal  
3.5mm female, 50 Ω nominal  
SMB male, 50 Ω nominal  
SMB male, 50 Ω nominal  
SMB male, 10 kΩ nominal  
SMB male, 50 Ω nominal  
SMB male, 50 Ω nominal  
For use with Keysight 346 and SNS series noise sources  
15 | Keysight | M9290A CXA-m PXIe X-Series Signal Analyzer - Data Sheet  
I/Q Analyzer  
Frequency  
Frequency span  
Standard instrument  
Option B25  
10 Hz to 10 MHz  
10 Hz to 25 MHz  
Resolution bandwidth (spectrum measurement)  
Range  
Overall  
100 MHz to 3 MHz  
Span = 1 MHz  
Span = 10 kHz  
Span = 100 Hz  
Window shapes  
50 Hz to 1 MHz  
1 Hz to 10 kHz  
100 MHz to 100 Hz  
Flat top, Uniform, Hanning, Gaussian, Blackman, Blackman-Harris, Kaiser Bessel (K-B 70 dB, K-B 90 dB and K-B 110 dB)  
Analysis bandwidth  
Standard instrument  
Option B25  
10 Hz to 10 MHz  
10 Hz to 25 MHz  
IF frequency response (standard 10 MHz IF path)  
IF frequency response (demodulation and FFT response relative to the center frequency, 20 to 30 °C)  
Center frequency (GHz)  
≤ 3.0  
Span (MHz)  
≤ 10  
Max. error  
± 0.40 dB  
RMS (nominal)  
0.03 dB  
3.0 < f ≤ 26.5  
≤ 10  
0.10 dB  
IF phase linearity (deviation from mean phase linearity, nominal)  
Center frequency (GHz)  
≤ 3.0  
3.0 < f ≤ 7.5  
Span (MHz)  
≤ 10  
≤ 10  
Peak-to-peak  
0.5 °  
0.5 °  
RMS  
0.2 °  
0.4 °  
Data acquisition (standard 10 MHz IF path)  
Time record length  
Sample rate  
4,000,000 IQ sample pairs  
90 MSa/s  
ADC resolution  
16 Bits  
Option B25 25 MHz analysis bandwidth  
IF frequency response (demodulation and FFT response relative to the center frequency, 20 to 30 °C)  
Center frequency (GHz)  
≤ 3.0  
3.0 < f ≤ 26.5  
Span (MHz)  
10 to ≤ 25  
10 to ≤ 25  
Max. error  
± 0.45 dB  
RMS (nominal)  
0.03 dB  
0.20 dB  
IF phase linearity (deviation from mean phase linearity, nominal)  
Center frequency (GHz)  
0.02 ≤ f < 3.0  
3.0 < f ≤ 7.5  
Span (MHz)  
10 to ≤ 25  
10 to ≤ 25  
Peak-to-peak  
1 °  
1 °  
RMS  
0.3 °  
0.5 °  
Data acquisition (B25 IF path)  
Time record length  
IQ analyzer  
4,000,000 IQ sample pairs  
90 MSa/s  
Sample rate  
ADC resolution  
16 Bits  
System Requirements  
Operating system  
Processor speed  
Windows 7 (32 & 64 bit)  
1.86 GHz minimum  
2.4 GHz recommended  
4 GB minimum  
Available memory  
8 GB recommended  
4 GB  
Available disk space  
Video  
Browser  
Support for DirectX 9 graphics with 128 MB graphics recommended (SuperVGA supported)  
Microsoft Internet Explorer 7.0 or greater  
16 | Keysight | M9290A CXA-m PXIe X-Series Signal Analyzer - Data Sheet  
Software  
Instrument connection software  
Keysight IO library The IO library suite offers a single entry point for connection to  
the most common instruments including AXIe, PXI, GPIB, USB,  
Free software download at  
www.keysight.com/find/  
Ethernet/LAN, RS-232, and VXI test instruments from Keysight and iosuite  
other vendors. It automatically discovers interfaces, chassis, and  
instruments. The graphical user interface allows you to search for,  
verify, and update IVI instrument and soft front panel drivers for  
modular and traditional instruments. The IO suite safely installs in  
side-by-side mode with NI I/O software.  
Module setup and usage  
Keysight soft  
|front panel  
The CXA-m includes a soft front panel (SFP), a software-based  
graphical user interface (GUI) which enables the instrument’s  
capabilities from your PC.  
Included on CD-ROM shipped  
with module or online  
Module management  
Keysight connection expert  
Connection expert is the graphical user interface included in the  
IO libraries suite that allows you to search for, verify and update IVI  
instrument and soft front panel drivers for modular and traditional  
instruments  
Free software download at  
www.keysight.com/find/  
iosuite  
Programming  
Driver  
Development environments  
lVI-COM  
IVI-C  
Visual Studio (VB .NET, C#, C/C++), VEE, LabVIEW, LabWindows/  
CVI, MATLAB  
Included on CD-ROM shipped  
with module.  
LabVIEW  
MATLAB  
Programming assistance  
Command expert Assists in finding the right instrument commands and setting correct Free software download at  
parameters. A simple interface includes documentation, examples,  
syntax checking, command execution, and debug tools to build  
sequences for integration in Excel, MATLAB, Visual Studio, and VEE.  
www.keysight.com/find/  
commandexpert  
Signal analysis software  
X-Series  
measurement  
applications  
Provides measurements for analog demodulation, noise figure,  
phase noise, vector signal analysis, and others.  
Licensed software. For more  
information, visit  
www.keysight.com/  
find/x-series_apps  
89600 VSA  
SystemVue  
89600 VSA software sees through the complexity of emerging and  
existing industry standards, serving as your window into complex  
signal interactions.  
Licensed software. For more  
information, visit  
www.keysight.com/find/vsa  
SystemVue is a system-level EDA platform for designing communi-  
cations and defense systems. Used with the M9290A, SystemVue  
Licensed software. For more  
information, visit  
enables you to create model-based design validation tests to ensure www.keysight.com/find/  
consistency from design to manufacturing. systemvue  
17 | Keysight | M9290A CXA-m PXIe X-Series Signal Analyzer - Data Sheet  
Related Literature  
Literature  
Pub number  
M9290A CXA-m PXIe Signal Analyzer - Product Fact Sheet  
M9290A CXA-m PXIe Signal Analyzer - Configuration Guide  
M9018A PXIe 18 slot Chassis - Data Sheet  
M9037A PXIe High Performance Embedded Controller - Data Sheet  
M9036A PXIe Embedded Controller - Data Sheet  
5992-0044EN  
5992-0193EN  
5990-6583EN  
5991-3661EN  
5990-8465EN  
Web  
For more information or literature resources please visit the web:  
Product page: www.keysight.com/find/M9290A  
X-Series measurement applications: www.keysight.com/find/X-Series_Apps  
X-Series signal analyzers: www.keysight.com/find/X-Series  
18 | Keysight | M9290A CXA-m PXIe X-Series Signal Analyzer - Data Sheet  
For more information on Keysight  
Evolving Since 1939  
Our unique combination of hardware, software, services, and people can help you  
reach your next breakthrough. We are unlocking the future of technology.  
From Hewlett-Packard to Agilent to Keysight.  
Technologies’ products, applications or  
services, please contact your local Keysight  
office. The complete list is available at:  
www.keysight.com/find/contactus  
Americas  
Canada  
Brazil  
Mexico  
United States  
(877) 894 4414  
55 11 3351 7010  
001 800 254 2440  
(800) 829 4444  
Asia Pacific  
Australia  
China  
Hong Kong  
India  
Japan  
Korea  
Malaysia  
Singapore  
Taiwan  
1 800 629 485  
800 810 0189  
800 938 693  
1 800 11 2626  
0120 (421) 345  
080 769 0800  
1 800 888 848  
1 800 375 8100  
0800 047 866  
(65) 6375 8100  
myKeysight  
www.keysight.com/find/mykeysight  
A personalized view into the information most relevant to you.  
Other AP Countries  
http://www.keysight.com/find/emt_product_registration  
Register your products to get up-to-date product information and  
find warranty information.  
Europe & Middle East  
Austria  
Belgium  
Finland  
France  
Germany  
Ireland  
Israel  
Italy  
Luxembourg  
Netherlands  
Russia  
0800 001122  
0800 58580  
0800 523252  
0805 980333  
0800 6270999  
1800 832700  
1 809 343051  
800 599100  
+32 800 58580  
0800 0233200  
8800 5009286  
800 000154  
0200 882255  
0800 805353  
Opt. 1 (DE)  
Keysight Services  
www.keysight.com/find/service  
Keysight Services can help from acquisition to renewal across your  
instrument’s lifecycle. Our comprehensive service offerings—one-  
stop calibration, repair, asset management, technology refresh,  
consulting, training and more—helps you improve product quality  
and lower costs.  
Keysight Assurance Plans  
www.keysight.com/find/AssurancePlans  
Up to ten years of protection and no budgetary surprises to ensure  
your instruments are operating to specification, so you can rely on  
accurate measurements.  
Spain  
Sweden  
Switzerland  
Opt. 2 (FR)  
Opt. 3 (IT)  
0800 0260637  
Keysight Channel Partners  
www.keysight.com/find/channelpartners  
United Kingdom  
Get the best of both worlds: Keysight’s measurement expertise and  
product breadth, combined with channel partner convenience.  
For other unlisted countries:  
www.keysight.com/find/contactus  
(BP-9-7-17)  
www.keysight.com/find/CXA-m  
DEKRA Certified  
ISO9001 Quality Management System  
www.keysight.com/go/quality  
Keysight Technologies, Inc.  
DEKRA Certified ISO 9001:2015  
Quality Management System  
This information is subject to change without notice.  
© Keysight Technologies, 2014 - 2017  
Published in USA, October 17, 2017  
5992-0032EN  
www.keysight.com  

相关型号:

5992-1065EN

M9420A VXT PXIe Vector Transceiver
KEYSIGHT

5992-1086EN

M9420A VXT PXIe Vector Transceiver
KEYSIGHT

59920A-2SOC

Clock Driver
IDT

59920A-7SOC

Clock Driver
IDT

599406--5

Application Tooling for Screw-Machine Contacts
TE

599406--7

Application Tooling for Screw-Machine Contacts
TE

5994S

Panel Mounted LED Indicator Light
CML

5994S1

Panel Mounted LED Indicator Light
CML

5994S1;1;1;1

Single Color LED Array, Red, 3mm, 8 PIN
CML

5994S5

Panel Mounted LED Indicator Light
CML

5994S7

Panel Mounted LED Indicator Light
CML

5994S7;7;7;7

Single Color LED Array, Yellow, 3mm, 8 PIN
CML