User's Manual  
Heavy Duty Dissolved Oxygen Meter  
Model 407510  
Introduction  
Congratulations on your purchase of Extech's Heavy Duty Dissolved Oxygen / Temperature  
Meter which simultaneously displays Dissolved Oxygen or Oxygen in air, and Temperature.  
Units of measure are mg/l for Dissolved Oxygen, % for Oxygen in air, and oC/oF for  
temperature. Oxygen is measured with a remote sensor that also contains a thermistor for  
measuring temperature. Advanced features include RS 232 PC serial interface, data hold, auto  
shut off, and Min/Max/Avg recording.  
Meter Description  
1. LCD Display  
2. POWER ON/OFF key  
3. Data HOLD key  
4. °C/°F selection key  
5. LCD Contrast Adjust  
6. RECORD key  
7. RECALL key  
8. Factor adjust key  
9. Probe head  
10. Sensor body  
11. Sensor plug  
12. Sensor jack  
13. RS-232 output jack  
14. Battery cover (rear)  
15. Altitude compensation key  
16. O2 Cal key  
17. Zero key  
18. O2 calibration key  
19. % Salt key  
2
407510 V7.0 10/06  
Getting Started  
Note: The dissolved oxygen probe assembly may be shipped to you  
from the factory dry. If the probe does not contain electrolyte  
solution, refer to the Electrolyte Replacement section of this manual  
for the instructions for adding the electrolyte solution prior to initial  
use.  
Filling the Electrolyte Container Prior to Initial Use  
1. Unscrew and remove the electrolyte container from the tip of  
the probe.  
2. Fill the electrolyte container with new electrolyte.  
3. Screw the electrolyte container back onto the electrode  
holder and probe handle. Excess Electrolyte may be forced  
out through the threads. Wipe clean.  
4. Lightly tap the probe to remove any air bubbles.  
Calibration  
For first time use or after long periods (several days) of non-use, calibrate the instrument using  
the following steps. The procedure should be performed in an open, well-ventilated area.  
1. Ensure that the sensor is DISCONNECTED from the meter.  
2. Turn ON the meter.  
3. Slide the O2/DO selector to the O2 position.  
4. Press the zero key to null the meter (the display should indicate "OO").  
5. Connect the DO sensor to the top of the meter (remove the plastic probe head  
protective cap).  
6. Allow the meter to stand for at least five minutes until the display stabilizes.  
7. Press the O2 Cal key and the display will indicate approx. 20.9 (typical O2 in air).  
3
407510 V7.0 10/06  
Dissolved Oxygen (DO) Measurements  
Measurement Preparation  
Before taking measurements, perform the calibration described above if necessary. In addition,  
perform the % SALT and ALTITUDE compensation adjustments as described below:  
% SALT Compensation  
Slide the O2/DO switch to the DO position. If the measured liquid is pure water or is known  
not to have a salt content, set the % Salt factor to 0%.  
1. Press the % Salt key. The display will indicate 0% on the middle portion of the LCD  
display and S on the bottom portion of the LCD.  
2. Pressing the Factor Adj. key once will add 1% to the salt factor.  
3. Press the Factor Adj. key until the correct factor is displayed.  
4. Press the % Salt key to store the factor.  
NOTE: 1% = 10 ppt (parts per thousand)  
ALTITUDE Compensation  
The meter has a default calibration for sea level operation (0 meters). Any significant  
deviation from sea level requires a compensation adjustment. Perform the following steps to  
do so:  
1. Press the ALTITUDE key and the display will indicate a “0” in the middle of the LCD  
and an “H” on the lower portion of the LCD.  
2. Press the Factor Adj. key once to add 100 meters, twice to add 200 meters, and so on.  
Maximum compensation is 3900 meters.  
3. Once the desired compensation is displayed, press the ALTITUDE key again to store  
the value.  
Taking Measurements  
1. Immerse the probe in the solution under test. For best automatic temperature  
compensation, immerse the probe to a depth of at least 4" (10 cm).  
2. Wait until the display stabilizes.  
3. The velocity of the liquid coming into contact with the probe must be at least 0.6 to 1  
ft/min (0.2 to 0.3 m/s). If the solution is standing, stir the solution with the probe or use  
an agitator.  
4. Rinse the probe with clean water after each use.  
Oxygen in Air (O2) Measurements  
Follow the same steps above for DO measurements except for the variations described below:  
1. Slide the O2/DO switch to the O2 position.  
2. The display will indicate the oxygen air value in %.  
Temperature Measurements  
The lower display will indicate the temperature of the measured solution. Press the oC/oF key to  
change the temperature units.  
4
407510 V7.0 10/06  
Advanced Features  
Data Hold  
Press the Data Hold key to freeze the displayed value. The LCD will display DH for Data Hold  
mode along with the held reading. Press the Hold key again to release the data hold function.  
MIN/MAX/AVG Data Recording  
When selected, the Data Recording function records the Min, Max, and Average readings. To  
start a data recording session:  
1. Press the RECORD key once. The REC indicator will appear on the display and the meter  
will begin to record the lowest (min), highest (max), and average (avg) readings.  
2. To recall the data, press the RECALL key once. The MAX indicator will appear along with  
the highest reading recorded since the RECORD key was pressed.  
3. Press the RECALL key again to view the MIN value.  
4. Press the RECALL key again to view the AVG reading.  
5. To exit the Record mode, press the RECORD key again. The display indicators REC,  
MIN, MAX, and AVG will disappear.  
Auto Power OFF  
The meter includes an Auto shut off feature that preserves battery life. The meter will  
automatically turn off if no function button is pressed in any 10 minute period. To disable this  
feature press the RECORD key to engage the record function.  
LCD Contrast Adjust  
It may be necessary to adjust the display contrast due to a change in viewing angle or voltage  
drift. Use the LCD Contrast adjustment located on the right side of the meter to set the  
preferred contrast.  
CPU Reset  
Certain improper keystroke sequences can cause the meter to “hang” and become inoperable.  
The following procedure will reset the meter.  
1. Slide the O2/DO switch to the O2 position.  
2. Turn the meter OFF and disconnect the DO probe.  
3. Press and HOLD the O2 CAL key and then press the POWER key.  
4. Release the O2 CAL key and then press the ZERO key.  
5. Connect the DO probe.  
6. Wait several minutes for the reading to stabilize and then press the O2 CAL key.  
RS-232 PC Interface  
The meter includes a RS-232 serial data port. This interface was designed to operate with the  
Extech Data Acquisition Software/Hardware kit Part No. 407001. This kit enables the user to  
capture, store and display readings using a PC.  
5
407510 V7.0 10/06  
Probe Head Replacement  
The probe head on the tip of the probe holds a thin Teflon® membrane and an electrolyte  
container. Calibration or reading errors could indicate loss of electrolyte or a dirty/damaged  
membrane. The membrane is very delicate and can easily be damaged if it comes in contact  
with solid objects. Replacement probe heads are available from your local distributor.  
Replacement procedure  
1. Unscrew the probe head and electrolyte container from the probe handle and discard.  
2. Fill the new probe head/electrolyte container with electrolyte.  
3. Screw the probe head onto the probe handle. Excess electrolyte may be forced out  
through the threads. Wipe clean.  
4. Lightly tap the probe to remove any air bubbles.  
1. Probe handle  
2. Temperature sensor  
3. Probe Head  
6
407510 V7.0 10/06  
Battery Replacement  
The low battery indication LBT appears on the LCD when the battery runs low.  
To replace the battery:  
1. Remove the meter's rubber protective cover to access the rear battery compartment.  
2. Remove the battery compartment cover using a small coin or screwdriver and remove the  
9V battery.  
3. Replace the 9V battery and reinstall the cover and holster.  
Specifications  
General Specifications  
Circuit  
Custom one-chip LSI microprocessor circuit  
Display  
Dual function 0.5" (13 mm) 1999 count LCD w/ contrast adjust  
Measurements  
Dissolved O2: mg/l (milligrams per liter); Oxygen in air: %;  
Temperature: °C/°F  
Data hold  
Data hold key freezes displayed reading  
Remote Polarographic Dissolved Oxygen Sensor  
Temperature sensor: Precision thermistor  
Min/Max/Avg readings stored for later recall  
Meter turns off after 10 minutes  
Sensor Structure  
Data recording  
Auto power off  
Sample rate  
0.4 Sec (approx.)  
Data Output  
RS 232 PC serial interface  
Operating conditions Temperature: 32 to 122 °F (0 to 50 °C); Humidity: < 80% RH  
Battery power 9V alkaline battery  
Power Consumption Approx. 6.6 mA DC  
Weight  
0.74 lbs. / 335 g (including batteries & probe)  
Dimensions  
Main instrument: 7.1 x 2.8 x1.3" (180 x 72 x 32mm)  
Probe: 4.9" x 0.8" length/diameter (125 x 20mm)  
Carrying case, 9V battery, and 5 membranes  
Accessories  
Optional  
Membranes, data acquisition software & RS232 interface cable  
accessories  
Electrical Specifications  
Measurement  
Range  
0.0 - 19.9 mg/L  
0 - 100% O2  
32 to 122 oF  
0 to 50 oC  
Resolution  
0.1mg/L  
0.1 % O2  
0.1oF/C  
Accuracy  
+ 0.4 mg/L  
+ 0.7% O2  
±1.5 oF  
Dissolved Oxygen  
Oxygen in air  
Temperature  
(probe only)  
+ 0.8 oC  
Compensation  
adjustments  
Temperature (automatic)  
Salt  
32 to 122 oF (0 to 50oC)  
0 to 39% (1% = ppt)  
0 to 3900 meters  
Altitude  
7
407510 V7.0 10/06  
Calibration and Repair Services  
Extech offers repair and calibration services for the products we sell. Extech also provides  
NIST certification for most products. Call the Customer Service Department for information on  
calibration services available for this product. Extech recommends that annual calibrations be  
performed to verify meter performance and accuracy.  
Warranty  
EXTECH INSTRUMENTS CORPORATION warrants this instrument to be free of defects in  
parts and workmanship for three years from date of shipment (a six month limited warranty  
applies to sensors and cables). If it should become necessary to return the instrument for  
service during or beyond the warranty period, contact the Customer Service Department at  
(781) 890-7440 ext. 210 for authorization or visit our website www.extech.com for contact  
information. A Return Authorization (RA) number must be issued before any product is  
returned to Extech. The sender is responsible for shipping charges, freight, insurance and  
proper packaging to prevent damage in transit. This warranty does not apply to defects  
resulting from action of the user such as misuse, improper wiring, operation outside of  
specification, improper maintenance or repair, or unauthorized modification. Extech specifically  
disclaims any implied warranties or merchantability or fitness for a specific purpose and will not  
be liable for any direct, indirect, incidental or consequential damages. Extech's total liability is  
limited to repair or replacement of the product. The warranty set forth above is inclusive and no  
other warranty, whether written or oral, is expressed or implied.  
Support line (781) 890-7440  
Technical support: Extension 200; E-mail: support@extech.com  
Repair & Returns: Extension 210; E-mail: repair@extech.com  
Product specifications subject to change without notice  
For the latest version of this User’s Guide, Software updates, and other  
up-to-the-minute product information, visit our website: www.extech.com  
Extech Instruments Corporation, 285 Bear Hill Rd., Waltham, MA 02451  
Copyright © 2006 Extech Instruments Corporation. All rights reserved including the right  
of reproduction in whole or in part in any form.  
8
407510 V7.0 10/06  

相关型号:

78041-0002

6.35mm (.250") Pitch, EXTreme PowerPlus™ Header, All Power, Through Hole, Vertical, 3 Circuits, with Peg, Lead-free
MOLEX

780417A

Heavy Duty Dissolved Oxygen Meter
EXTECH

780418

Heavy Duty Dissolved Oxygen Meter
EXTECH

78042FG

8-BIT SINGLE-CHIP MICROCOMPUTER
NEC

78043

Wire And Cable
ALPHAWIRE

78044F

Common to 78K/0 Series
NEC

78044H

Common to 78K/0 Series
NEC

78045

Heat Sink,
AAVID

78045-0660

Board Connector, 6 Contact(s), 2 Row(s), Female, 0.079 inch Pitch, Crimp Terminal, Locking, Black Insulator, Plug
MOLEX

78045-1060

2.00mm (.079) Pitch, Milli-Grid™ Crimp Housing, 10 Circuits, with Center Polarization Key and side Locking Ramps, Leadfree
MOLEX

78045-1260

Board Connector, 12 Contact(s), 2 Row(s), Female, 0.079 inch Pitch, Crimp Terminal, Locking, Black Insulator, Plug
MOLEX

78046

Wire And Cable
ALPHAWIRE