ALM65 [XPPOWER]
AC-DC Power Supplies;![ALM65](http://pdffile.icpdf.com/pdf2/p00326/img/icpdf/ALM65_2006170_icpdf.jpg)
型号: | ALM65 |
厂家: | ![]() |
描述: | AC-DC Power Supplies |
文件: | 总4页 (文件大小:762K) |
中文: | 中文翻译 | 下载: | 下载PDF数据表文档文件 |
AC-DC Power Supplies
ALM65 Series
65 Watts
• Medical & IT Safety Approvals
• Energy Efficiency Level VI
• 4th Edition Medical EMC
• IP32 Environmental Rating
• Class I and Class II Versions
• <0.21 W Standby Power
• 0 °C to 60 °C Operation
• Low Earth Leakage Current
• 3 Year Warranty
Dimensions:
ALM65:
The ALM65 series of medical external power supplies is fully approved to international medical
safety standards. It has been designed with very high efficiency and low standby power, enabling it
to meet the latest environmental legislation. The unit has a fully sealed enclosure complying with
IP32 and a smooth surface finish making it easier to wipe down in a clinical setting.
4.94 x 2.19 x 1.32” (125.5 x 55.5 x 33.5 mm)
Models & Ratings
Output Power
Output Voltage
Output Current
Total Regulation
Efficiency(1)
Model Number(2,3,4)
12.0 V
15.0 V
19.0 V
24.0 V
48.0 V
5.40 A
4.30 A
3.40 A
2.70 A
1.35 A
88%
88%
88%
88%
88%
ALM65US12
ALM65US15
ALM65US19
ALM65US24
ALM65US48
65 W
5%
Notes
1. Typical average of efficiencies measured at 25%, 50%, 75% and 100% load and
115VAC input.
3. For optional input connector retention clip add suffix ‘-A’ to the model number,
e.g. ALM65US24-A (not available for C2 versions)
2. For class II versions, add suffix 'C2-8' to the end of the part number e.g.
ALM65US24C2-8.
4. 5.5 x 2.1 mm output connector add suffix B1 to the part number,
e.g. ALM65US24-B1, ALM65US24C2-8B1, ALM65US24-AB1.
Mechanical Details
Class I
IEC 320-C14
4.941 (125.5)
35.433 0.8 (900 20)
1.512
(38.4)
ø 0.685
(17.4)
1.319 (33.5)
2.185 (55.5)
0.177 (4.50)
1.969 0.394 (50 10)
Class II
IEC 320-C8
0.374 0.012 (9.50 0.30)
1.319 (33.5)
0.039 (1.0)
RETURN ( - )
4.488 (114.0)
1.732 (44.0)
OUTPUT ( + )
Notes
1. All dimensions shown in inches (mm). Tolerance is 0.02 (0.5) maximum, except
output cable length.
5. For UK mains lead order part UK-MAINS-IEC for C14 versions, or UK-MAINS-8
for C8 versions.
2. Output connector: 5.5 outer diameter, 2.5 mm inner diameter with centre positive. 6. For US mains lead order part US-MAINS-IEC for C14 versions, or US-MAINS-8
3. Weight: 0.68 lbs (308 g) approx.
for C8 versions.
4. For European mains lead order part EU-MAINS-IEC for C14 versions,
or EU-MAINS-8 for C8 versions.
www.xppower.com
1
AC-DC Power Supplies
ALM65 Series
Input
Characteristic
Minimum
Typical
Maximum
264
Units
Notes & Conditions
Derate linearly from 100% load at 90 VAC to 80% load
at 80 VAC
Input Voltage
80
47
VAC
Input Frequency
Input Current
Inrush Current
63
Hz
A
A
1.2/0.6
Measured at 115/230 VAC
115/230 VAC, cold start at 25 °C
EN61000-3-2 Class A
264 VAC, 60 Hz
60/120
Power Factor
Earth Leakage Current
No Load Input Power
Input Protection
250
0.21
µA
W
T3.15A/250 VAC internal fuse in both line
Output
Characteristic
Min.
Typ.
Max.
Units
Notes & Conditions
Output Voltage
Initial Set Accuracy
Minimum Load
Start Up Delay
Start Up Rise Time
Hold Up Time
12
48
2
VDC
%
See Models and Ratings table
At 50% load
No minimum load required
2
40
s
ms
ms
%
20
Full load and 230 VAC
Line Regulation
Total Regulation
0.5
5
%
Maximum deviation, recovering to less than 1% within
500 µs for 25% step load
20 MHz bandwidth, measured with 20 MHz Bandwidth and
10 µF electrolytic in parallel with 0.1 µF ceramic capacitor.
At turn on / turn off
Transient Response
Ripple and Noise
4
%
1.5
% pk-pk
Overshoot
10
175
170
%
%
%
Overload Protection
Overvoltage Protection
Short Circuit Protection
Thermal Protection
Temperature Coefficient
Patient Leakage Current
115
145
Recycle mains to reset
Trip and restart (hiccup), auto resetting
Measured internally, auto resetting
0.05
%/°C
µ A
100
264 VAC, 60 Hz
Environmental
Characteristic
Minimum
Typical
Maximum
Units
Notes & Conditions
Operating Temperature
Cooling
0
+60
°C
Derate from 100% load at 40 °C to 50% load at 60 °C
Natural convection
Operating Humidity
Storage Temperature
Operating Altitude
Shock
5
-40
90
+80
5000
%RH
°C
m
Non-condensing
IEC68-2-27, 30 g, 11 ms half sine, 3 times in each of 6 axes
Vibration
IEC68-2-6, 10-500 Hz, 2 g 10 mins/sweep, 60 mins for each of 3 axes
Derating Curve
100%
80%
60%
40%
20%
0
60
20
40
Ambient Temperature (ºC)
www.xppower.com
2
AC-DC Power Supplies
ALM65 Series
General
Characteristic
Minimum
Typical
Maximum
Units
%
Notes & Conditions
See Models and Ratings table and curves.
2 x MOPP
Efficiency
88
4000
1500
500
VAC
VAC
VAC
kHz
Isolation: Input to Output
Input to Ground
1 x MOPP (Class I versions only)
Class I versions only
PWM
Output to Ground
Switching Frequency
Power Density
65 KHz
4.55
3
W/in
Mean Time Between Failure
Weight
>300
kHrs
lb (g)
MIL-HDBK-217F at 25 °C GB
0.68 (308)
Efficiency Curves
ALM65US12
100%
90%
80%
70%
60%
50%
40%
30%
20%
10%
0%
115V
230V
10%
20%
30%
40%
50%
60%
70%
80%
90% 100%
Load
ALM65US24
100%
90%
80%
70%
60%
50%
40%
30%
20%
10%
0%
115V
230V
10%
20%
30%
40%
50%
60%
70%
80%
90% 100%
Load
EMC: Emissions
Phenomenon
Standard
Test Level
Notes & Conditions
Conducted & Radiated
Emissions
Harmonic Current
Voltage Flicker
EN55022
EN61000-3-2
EN61000-3-3
Level B
Class A
www.xppower.com
3
AC-DC Power Supplies
ALM65 Series
EMC: Immunity
Phenomenon
Standard
Test Level
Criteria
Notes & Conditions
ESD
EN61000-4-2
EN61000-4-3
EN61000-4-4
EN61000-4-5
EN61000-4-6
EN61000-4-8
3
A
A
A
A
A
15 kV air/ 8 kV contact
80-2700 MHz. IEC60601-1-2 Ed.4 at other frequencies
Radiated
EFT/Burst
Surge
Conducted
Magnetic Fields
10 V/m
3
Installation Class 3
10 V
4
A
Dip: 30% 500 ms
Dip: 60% 200 ms
Dip: 80% 5000 ms
Dip: 100% 5000 ms
Dip: 30% 25 AC Cycles
Dip: 60% 5 AC Cycles
Dip: 100% 0.5 AC Cycles
Int.: >95% 5000 ms
A/B
A/B
B
B
A
A
A
B
High Line/Low Line
High Line/Low Line
EN61000-4-11
Dips and Interruptions
230 VAC 100% load, 100 VAC 75%
230 VAC 100% load, 100 VAC 20%
At 8 angles
EN60601-1-2
Safety Approvals
Safety Agency
Safety Standard
UL60950-1 2nd Edition
EN60950-1
Notes & Conditions
UL
TUV
CB
Information Technology
IEC60950-1
CE
LVD
UL
ANSI/AAMI ES 60601-1
CSA C22.2 No. 60601
EN60601-1
CSA
TUV
CB
Medical
IEC60601-1
Others
CCC, PSE, KC & RCM
May require additional importer information
Mechanical Details
Class I
IEC 320-C14
4.941 (125.5)
35.433 0.8 (900 20)
1.512
(38.4)
ø 0.685
(17.4)
1.319 (33.5)
2.185 (55.5)
Optional Input Connector Retention (-A)
1.40 (35.7)
1.969 0.394 (50 10)
Class II
IEC 320-C8
0.177 (4.50)
0.374 0.012 (9.50 0.30)
1.319 (33.5)
0.039 (1.0)
RETURN ( - )
4.488 (114.0)
1.732 (44.0)
OUTPUT ( + )
Notes
1. All dimensions shown in inches (mm). Tolerance is 0.02 (0.5) maximum, except
output cable length.
5. For UK mains lead order part UK-MAINS-IEC for C14 versions, or UK-MAINS-8
for C8 versions.
2. Output connector: 5.5 outer diameter, 2.5 mm inner diameter with centre positive. 6. For US mains lead order part US-MAINS-IEC for C14 versions, or US-MAINS-8
3. Weight: 0.68 lbs (308 g) approx.
for C8 versions.
4. For European mains lead order part EU-MAINS-IEC for C14 versions,
or EU-MAINS-8 for C8 versions.
www.xppower.com
17 Sept 19
相关型号:
©2020 ICPDF网 联系我们和版权申明