TMC40 [TRUMPOWER]
30-48 WATT MEDICAL & ITE SWITCHING POWER SUPPLIES; 30-48 WATT医疗及ITE开关电源![TMC40](http://pdffile.icpdf.com/pdf1/p00140/img/icpdf/TMC40_773392_icpdf.jpg)
型号: | TMC40 |
厂家: | ![]() |
描述: | 30-48 WATT MEDICAL & ITE SWITCHING POWER SUPPLIES |
文件: | 总2页 (文件大小:133K) |
中文: | 中文翻译 | 下载: | 下载PDF数据表文档文件 |
![](http://public.icpdf.com/style/img/ads.jpg)
TRUMPower
30-48 WATT MEDICAL & ITE SWITCHING POWER SUPPLIES
TMC40 SERIES
DESCRIPTION
The TMC40 series of compact, open PCB constructed, AC-DC
switching power supplies are capable of delivering 30-48
watts of continuous output power at convection cooling. They
operate at 90-264 VAC input voltage without the need of
voltage selection, and are suited for medical, information
technology and industrial applications. Approval to both
EN60601-1 and EN60950-1 safety standards improves
design-in time and reduces end equipment compliance costs.
FEATURES
z
z
z
z
z
z
z
Medical and ITE approvals
Compact size 2” x4” x1.18”
Single, dual and triple outputs
Full-range input 90-264 VAC
Low earth leakage current
Level B emissions
SAFETY STANDARD APPROVALS
UL 2601-1
CSA C22.2 No. 601.1
RoHS compliant
TÜV EN60601-1
INPUT SPECIFICATIONS
Input voltage:
Input frequency:
Input current:
90-264 VAC
47-63 Hz
GENERAL SPECIFICATIONS
0.9 A (rms) for 100 VAC
0.5 A (rms) for 240 VAC
Switching frequency:
Efficiency:
62±5 KHz
Earth Leakage current:
150 µA max. @ 264 VAC, 63 Hz
80-88% typical except TMC40-T31-3 and
TMC40-T31-5 at 75% typical
12 msec minimum at 110 VAC
Hold-up time:
Line regulation:
Inrush current:
±0.5% maximum at full load
OUTPUT SPECIFICATIONS
Output voltage/current:
25 A @ 115 VAC, or 50 A @ 230 VAC,
at 25 oC cold start
See rating chart
Maximum output power: See rating chart
Withstand voltage:
MTBF:
4000 VAC from input to output
2% peak to peak on 5.0 V model,
1500 VAC from input to ground
Ripple and noise:
1% peak to peak on other models.
Provided on output #1 only; set at
112–132% of its nominal output voltage
All outputs protected to short circuit
conditions
500 VAC from output to ground
400,000 hours at full load at 25 oC ambient,
Overvoltage protection:
calculated per MIL-HDBK-217F
Overcurrent protection:
EMC Performance
EN55011 / EN55022:
FCC:
Class B conducted, class B radiated
Class B conducted, class B radiated
Class B conducted, class B radiated
Harmonic distortion, class A and D
Line flicker
Temperature coefficient:
Transient response:
All outputs ±0.04% /°C maximum
Maximum excursion of 4% or better on all
models, recovering to 1% of final value
within 500 µs after a 25% step load
change
VCCI:
EN61000-3-2:
EN61000-3-3:
EN61000-4-2:
EN61000-4-3:
EN61000-4-4:
EN61000-4-5:
EN61000-4-6:
EN61000-4-8:
EN61000-4-11:
ESD, ±8 KV air and ±6 KV contact
Radiated immunity, 3 V/m
Fast transient/burst, ±2 KV
Surge, ±1 KV diff., ±2 KV com.
Conducted immunity, 3 Vrms
Magnetic field immunity, 3 A/m
Voltage dips,
ENVIRONMENTAL SPECIFICATIONS
Operating temperature:
Storage temperature:
-10oC to +70oC
-40oC to +85oC
Relative humidity:
5% to 95% non-condensing
Derating:
30% reduction for 500 ms
60% reduction for 100 ms
>95% reduction for 10 ms
Derate from 100% to +50oC linearly to
50% at +70oC
Tumbler Technologies + TRUMPower
3350 Scott Blvd., Bldg. 13, Santa Clara, California 95054, USA
Phone: 408-988-6616 ● Fax: 408-988-6622 ● email: sales@trumpower.com ● Website: www.TRUMPower.com
UNIVERSAL INPUT
TMC40 MEDICAL & ITE SERIES
OUTPUT VOLTAGE/CURRENT RATING CHART
Maximum
Output #1
Output #2
Imin. Imax. Tol.
(N/A)
Output #3
Imin. Imax. Tol.
(N/A)
Output
P/N
Vnom.
Imin.
Imax. Tol.
Vnom.
Vnom.
Power
TMC40-S05
TMC40-S12
TMC40-S15
TMC40-S24
TMC40-S48
TMC40-D23
TMC40-D25
TMC40-T31
TMC40-T31-3
TMC40-T31-5
TMC40-T32
TMC40-T39
5 V
12 V
15 V
24 V
48 V
+5 V
+5 V
+5 V
+3.3 V
+5 V
+5 V
+5 V
0 A
0 A
8.0 A 2%
3.5 A 2%
3.0 A 2%
2.0 A 2%
1.0 A 2%
6 A 3%
6 A 3%
6 A 3%
6 A 3%
6 A 3%
6 A 3%
6 A 3%
40 W
42 W
45 W
48 W
48 W
40 W
40 W
40 W
30 W
30 W
40 W
40 W
(N/A)
(N/A)
(N/A)
(N/A)
(N/A)
(N/A)
(N/A)
(N/A)
(N/A)
(N/A)
0 A
0 A
0 A
0.5 A
0.5 A
0.5 A
0.8 A
0.5 A
0.5 A
0.5 A
+12 V
+24 V
+12 V
+5 V
0.1 A
2 A
1 A
5%
5%
5%
5%
5%
5%
5%
0.1 A
0.1 A
0.1 A
0 A
2 A
-
12 V
0 A
0.3 A
0.3 A
0.3 A
0.3 A
0.3 A
4%
4%
4%
4%
4%
2 A
+12 V
+12 V
0 A
0 A
0 A
0 A
+3.3 V
+15 V
+24 V
1.5 A
0.1 A 1.5 A
0.1 A 1 A
-
15 V
12 V
-
NOTE: 1. Safety approvals are for PCB form only. To order unit with cover fitted, add suffix “C”.
MECHANICAL SPECIFICATIONS
NOTES:
1. Dimensions shown in inches [mm]
2. Tolerance 0.02 [0.5] maximum
3. Input connector mates with Molex housing
09-50-3031 and Molex 2878 series crimp
terminal.
4. Output connector mates with Moles housing
09-50-3061 and Moles 2878 series crimp
terminal.
5. Ground tab is 0.25 [6.35] x 0.032 [0.8]
6. To ensure compliance with level B
emissions, connect the two “
*” marked
mounting holes with metallic standoffs to
chassis.
7.
Weight: 205 grams (0.45 lbs.) approx.
PIN CHART
PIN
1
2
3
4
5
6
MODEL
TMC40-S05
TMC40-S24
TMC40-S12
TMC40-S48
TMC40-S15
+V1
+V1
RTN
RTN
N.C
N.C
TMC40-D23 TMC40-D25
TMC40-T31 TMC40-T32
TMC40-T31-3 ,TMC40-T31-5
+V1
+V1
+V1
+V1
+V1
+V1
RTN
RTN
RTN
RTN
RTN
RTN
N.C
-V3
+V3
+V2
+V2
+V2
TMC40-T39
Tumbler Technologies + TRUMPower
3350 Scott Blvd., Bldg. 13, Santa Clara, California 95054, USA
Phone: 408-988-6616 ● Fax: 408-988-6622 ● email: sales@trumpower.com ● Website: www.TRUMPower.com
相关型号: