TPP40A [TRACOPOWER]

AC/DC Medical Power Supply;
TPP40A
型号: TPP40A
厂家: TRACO ELECTRONIC AG    TRACO ELECTRONIC AG
描述:

AC/DC Medical Power Supply

文件: 总4页 (文件大小:486K)
中文:  中文翻译
下载:  下载PDF数据表文档文件
AC/DC Medical Power Supply  
TPP 40A Series, 40 Watt  
Open frame power supply pin connector  
Certification according to IEC/EN/ES  
60601-1 3rd edition for 2ꢀMOPP  
Low leakage current <75 µA rated for  
BF applications  
Risk management process according to  
ISO 14971 including risk management file  
Acceptance criteria for electronic assem-  
blies according to IPC-A-610 Level 3  
EMC emission to IEC 60601-1-2 ed.4  
Protection class I & ll  
Approved for operation up to 5000 m MSL  
CB  
Scheme  
prepared  
ES 60601-1  
Ready to meet ErP directive, < 0.15 W  
no load power consumption  
5 year product warranty  
The TPP 40A Series of 40 Watt AC/DC power supplies feature a reinforced dou-  
ble I/O isolation system according to latest medical safety standards IEC/EN ES  
60601-1 3rd edition for 2 x MOPP up to 5000 m MSL. The earth leakage current  
is below 75 µA what makes the units suitable for BF (body floating) applications.  
The eꢀcellent efficiency of up to 92% allows a high power density for the stan-  
dard 2.44” ꢀ 3.0” packaging format. The full load operating temperature range  
is –40°C to +70°C while it goes up to 85°C with 50% load derating. The EMC  
characteristic complies to IEC 60601-1-2 ed.4 and is dedicated for applications  
in industrial and domestic fields. High reliability is provided by use of industrial  
quality grade components and an eꢀcellent thermal management. It makes the  
products an ideal solution for medical devices and for demanding safety and  
space critical applications.  
Encased version with screw terminal  
connection see TPP 40 Series  
www.tracopower.com/overview/tpp40  
Models  
Order code  
TPP 40-105A-J  
Output voltage  
5 VDC  
Output current max.  
8.0 A  
Efficiency max.  
90 %  
TPP 40-112A-J  
TPP 40-124A-J  
TPP 40-148A-J  
12 VDC  
3.34 A  
92 %  
24 VDC  
1.67 A  
92 %  
48 VDC  
0.84 A  
92 %  
Note: - Output is ajustable by ±±10 ꢀith iꢁteꢂꢁal poteꢁtiomet  
- Otheꢂ output voltages aꢂe available oꢁ ꢂequest  
www.tracopower.com  
Page 1 of 4  
TPP 40A Series, 40 Watt  
Input Specifications  
Input voltage range  
– AC range (universal input)  
– DC range  
85 – 264 VAC  
120 – 370 VDC  
Input frequency  
47 – 63 Hz  
Input current at full load  
Input protection  
– at 115 VAC / 230 VAC  
1.05 A maꢀ. / 0.55 A maꢀ.  
T3.15 A/250 VAC (internal fuse in both line &  
neutral)  
Input inrush current  
– at 230 VAC  
60 A maꢀ.  
Zero load power consumption  
0.15 W maꢀ. (acc. ErP directive)  
Output Specifications  
Voltage set accuracy  
1%  
Regulation  
– Input variation  
0.2% maꢀ.  
– Load variation (0 - 100%)  
5 VDC model: 0.7% maꢀ.  
other models: 0.5% maꢀ.  
not required  
0.02%/K  
Minimum load  
Temperature coefficient  
Hold-up time  
– Vin = 115 VAC  
25 ms typ.  
Start-up time  
<1 s  
Rise time  
20 ms typ.  
Ripple and noise  
(20MHz Bandwidth)  
5 & 12 VDC output: 75 mVp-p typ. w. cap. 10µF/25V 1206 X7R MLCC  
24 VDC output: 75 mVp-p typ. w. cap. 1µF/50V 1206 X7R MLCC  
48 VDC output: 150 mVp-p typ. w. cap. 0.1µF/100V 1206 X7R MLCC  
Overvoltage protection  
Current limitation  
125 – 140% of nominal Vout  
at 145% Iout typ.  
Short circuit protection  
Transiente response  
hiccup mode (automatic recovery)  
– Peak deviation  
– Recovery time  
3% maꢀ. (25% load step change)  
600 µs typ.  
Capacitive load  
5 Vout model: 16‘000 µF maꢀ.  
12 Vout model: 2‘785 µF maꢀ.  
24 Vout model: 700 µF maꢀ.  
48 Vout model: 175 µF maꢀ.  
All specifications valid at nominal input voltage, full load and +25°C after warm-up time unless otherwise stated.  
www.tracopower.com  
Page 2 of 4  
TPP 40A Series, 40 Watt  
General Specifications  
Operating temperature  
–40°C to +85°C with derating  
Output power de-rating  
2.67 %/K above +70°C at 230 VAC  
2.75 %/K above +65°C at 115 VAC  
Storage temperature  
–40°C to +85°C  
5 – 95 % rel. H maꢀ.  
5000 m  
Humidity (non condensing)  
Altitude during operation  
Switching frequency - single output  
(at 230 VAC)  
5 VDC model: 70 kHz typ. (pulse frequency modulation)  
other models: 120 kHz typ. (pulse frequency modulation)  
Isolation voltage  
(2 x MOPP insulation)  
– Input / Output (60 s)  
– Input / Case (60 s)  
4000 VAC  
2500 VAC  
Leakage current (at 264 VAC/60Hz)  
Isolation resistance (at 500 VDC)  
Reliability  
75 µA maꢀ.  
100 MOhm min.  
– calculated MTBF at +25°C acc. to IEC 61709  
3‘000‘000 h  
Protection class  
class II prepared  
Electromagnetic compatibility – Conducted & Radiated input surpression  
EN 55011 limits to IEC 60601-1-2 4th editon  
EN 55032 class B (internal filter)  
IEC / EN 61000-3-2, class B  
(EMC), emissions  
– Harmonic current emissions  
– Voltage flicker  
IEC / EN 61000-3-3, class B  
Electromagnets compatibility (EMC), immunity  
– Electrostatic discharge ESD  
– RF field immunity  
IEC / EN 60601-1-2  
IEC / EN 61000-4-2, 8kV/6kV perf. criteria A  
IEC / EN 61000-4-3, 20V/m perf. criteria A  
IEC / EN 61000-4-4, 2kV perf. criteria A  
IEC / EN 61000-4-5, 1kV/ 2kV perf. criteria A  
IEC / EN 61000-4-6, 20 Vrms perf. criteria A  
IEC / EN 61000-4-8, 10A/m perf. criteria A  
– Electrical fast transients/burst immunity  
– Surge  
– Conducted RF  
– Magnetic field  
Voltage dip and interruptions according to EN 60601-1-2  
reference: 100 VAC / 50Hz  
30%, 500ms perf. criteria A  
60%, 100ms perf. criteria B  
> 95%, 10ms perf. criteria A  
> 95%, 5000ms perf. criteria B  
Safety standards and certification  
IEC/EN 60601-1 3rd edition,  
ANSI/AAMI ES60601-1:2005(R)2012  
www.tracopower.com/overview/tpp40a  
– Certification documents  
Environment  
– Vibration acc. IEC 60068-2-6  
– Shock acc. IEC 60068-2-27  
3 aꢀis, sine sweep, 10–55Hz, 1g, 1oct/min  
3 aꢀis, 10g half sine, 11msShock  
20 G (3 directions each 3 times)  
Environmental compliance – Reach  
www.tracopower.com/products/reach-declaration.pdf  
RoHS directive 2011/65/EU  
– RoHS  
Connection  
JST pin connector  
All specifications valid at nominal input voltage, full load and +25°C after warm-up time unless otherwise stated.  
www.tracopower.com  
Page 3 of 4  
TPP 40A Series, 40 Watt  
Outline Dimensions  
each one of the 4 srew holes can be used  
as a PE connection for class I applications  
Screw Terminal  
Input  
Single  
Output  
Pin  
1
Pin*  
Dual  
–Vout  
+Vout  
1,2  
3,4  
Line  
3
Neutral  
*Terminal rated for 7 A maꢀ.  
(at higher current connection has to be split)  
CON 1: JST series  
mates with JST crimp terminal: BVH-21T-P1.1  
and terminal housing: VHR-3N  
CON 2: JST series  
mates with JST crimp terminal: BVH-21T-P1.1  
and terminal housing: VHR-4N  
Dimensions in inch, () = mm  
Tolerances: x.xx 0.02 (x.x 0.5) x.xxx 0.01 (x.xx 0.25)  
Weight: 114 g (4.02 oz)  
©
Copyright 2018 Traco Electronic AG  
Specifications can be changed without notice!  
Rev. June 05. 2018  
Page 4 of 4  
www.tracopower.com  

相关型号:

TPP40A_1806

AC/DC Medical Power Supply
TRACOPOWER

TPP40_18

AC/DC Medical Power Supply
TRACOPOWER

TPP4XGFLRC000E2G

1310nm 10km SFP Transceiver
OPLINK

TPP4XGFLRE000E2G

1310nm 10km SFP Transceiver
OPLINK

TPP4XGFLRI000E2G

1310nm 10km SFP Transceiver
OPLINK

TPP4XGFLRXG

1310nm 10km SFP Transceiver
OPLINK

TPP65

AC/DC Medical Power Supply
TRACOPOWER

TPP65-105

AC/DC Medical Power Supply
TRACOPOWER

TPP65-105A-J

AC/DC Medical Power Supply
TRACOPOWER

TPP65-112

AC/DC Medical Power Supply
TRACOPOWER

TPP65-112A-J

AC/DC Medical Power Supply
TRACOPOWER

TPP65-115

AC/DC Medical Power Supply
TRACOPOWER