TPHPU63A-108 [TOTAL-POWER]

Medical Power Supplies;
TPHPU63A-108
型号: TPHPU63A-108
厂家: TOTAL POWER INTERNATIONAL    TOTAL POWER INTERNATIONAL
描述:

Medical Power Supplies

文件: 总2页 (文件大小:340K)
中文:  中文翻译
下载:  下载PDF数据表文档文件
Total Power International, Inc.  
63W Desk Top Type  
Medical Power Supplies  
TPHPU63A SERIES  
Features:  
Wide Operating Voltage 90 to 264 VAC,47 to 63 Hz  
IEC-320-C14 Input Inlet  
Single Output  
Over Voltage and Over Load protection  
Class I  
Energy Star 2.0, Efficiency level V  
ANSI/AAMI ES 60601-1: 2005(UL/cUL 3rd Edition)  
EN 60601-1:2006 (TUV/T-mark 3rd Edition)  
Input to Output : 2MOPP  
2 year warranty  
Electrical Characteristics:  
Safety Approvals Input Voltage Range  
100~240VAC  
Vin  
Application:  
Operate Voltage Range  
90~264VAC  
47~63Hz  
- Medical Touch Panel PC  
- Patient Monitoring System  
- Ultrasound System  
Input Frequency  
fin  
Output Power Range  
Output Voltage Range  
Output Current Range  
Input Current (Low Line)  
Input Current (High Line)  
See rating chart  
See rating chart  
See rating chart  
Po  
Vo  
Io  
Io=Full load, Vin=100VAC  
Io=Full load, Vin=240VAC  
Io=Full load, 25°C, Cool start, Vin=115VAC  
Io=Full load, 25°C, Cool start, Vin=230VAC  
Io=Full Load, Vin=230VAC  
Io=Full Load  
1.62A  
Iil  
0.72A  
Iih  
37A (max.)  
74A (max.)  
Low Line Inrush Current  
High Line Inrush Current  
Efficiency  
Ir  
Safety Approvals:  
87~92%  
Eff  
Line Regulation  
1% (max.)  
5% (max.)  
REG-i  
REG-o  
OVP  
EN 60601-1  
IEC 60601-1  
P
S
Load Regulation  
E
Vin=230VAC  
Over Voltage Protection  
112~132%  
110~150%  
4mS (max.)  
12mS (min.)  
RoHS  
2002/95/EC  
COMPLIANT  
OCP Over Current Protection  
Time of Transient Response  
Hold-Up Time  
Io=Full Load to Half Load, Vin=100VAC  
Io=Full Load, Vin=110VAC  
Io=Full Load, Vin=100VAC  
Full Load, Vin=90VAC  
Ttr  
Th  
Ts  
Start Up Time  
0.3~2S  
Vp-p Ripple & Noise(Peak to Peak)  
1% (max.)  
0.25mA (max.)  
Safety Ground Leakage Current  
Ilk  
Vin=240VAC/60Hz  
All output  
Temperature Coefficient  
TC  
0.04%/°C  
Operating Altitude  
Up to 3000m  
See rating chart  
6115VDC (min.)  
OA  
Pno  
Vps  
No-Load Power Consumption  
Dielectric Withstanding Voltage  
No load, Vin=230VAC  
Primary to secondary  
Dielectric Withstanding Voltage  
for Primary to PE  
Vpg  
Ris  
Primary to PE  
2350VDC (min.)  
Test Voltage=500VDC  
(min.)  
50M  
Isolation Resistance  
Note: The specifics for testing the energy efficiency of TPHPU63 Series are outlined in a separate document titled  
"Test Method for Calculating the Energy Efficiency of Single-Voltage External Ac-Dc and Ac-Ac Power Supplies  
(August 11, 2004)," which is available on the ENERGY STAR Website.  
Environmental  
Operating Temperature  
See derating curve  
-40~85°C  
To  
Ts  
Storage Temperature  
Operating Humidity  
0~95%  
Ho  
Storage Humidity  
0~95%  
Hr  
Operating Temperature at 25°C, Calculated per MIL-HDBK-217F  
Derate linearly from 100% load at 40°C to 50% load at 70°C  
0.1M Hrs (min.)  
MTBF  
Pd  
Total Power International, Inc. Toll Free: 877-646-0900(U.S.) PH# (978) 453-7272  
www.total-power.com sales@total-power.com  
Total Power International, Inc.  
63W Desk Top Type  
Medical Power Supplies  
TPHPU63A SERIES  
Output Voltage And Current Rating Chart ( Single Output ) :  
Total  
Regulation  
Max.  
Output Power  
Output Voltage  
Output Current  
Pno (max.)  
Model Number  
12 VDC  
15 VDC  
18 VDC  
24 VDC  
30 VDC  
36 VDC  
48 VDC  
5.25 A  
4.20 A  
3.50 A  
2.62 A  
2.10 A  
1.75 A  
1.31 A  
5%  
5%  
5%  
3%  
3%  
3%  
3%  
63W  
63W  
63W  
63W  
63W  
63W  
63W  
0.5W  
0.5W  
0.5W  
0.5W  
0.5W  
0.5W  
0.5W  
TPHPU63A-105  
TPHPU63A-106  
TPHPU63A-107  
TPHPU63A-108  
TPHPU63A-109  
TPHPU63A-110  
TPHPU63A-111  
TPHPU63A-105~107 are required to use AWG#16/2C/4FT output cable.  
TPHPU63A-108~110 are required to use AWG#18/2C/6FT output cable.  
The regulation and efficiency will be changed by modified output cable.  
Mechanical Specifications:  
Note:  
1. Dimensions are shown in mm.  
2. Weight: 330~380gs approx.  
3. Optional output connector:  
See page Appendix.  
Derating Curve :  
120%  
1. Operating Temperature: 0 to 70°C  
2. Derate linearly from 100% load at 40°C to 50%  
load at 70°C  
100%  
80%  
60%  
40%  
20%  
0%  
0
10  
20  
30  
40  
50  
60  
70  
80  
Temperature (deg c)  
Total Power International, Inc. Toll Free: 877-646-0900(U.S.) PH# (978) 453-7272  
www.total-power.com sales@total-power.com  

相关型号:

TPHPU63A-109

Medical Power Supplies

Warning: Undefined variable $rtag in /www/wwwroot/website_ic37/www.icpdf.com/pdf/pdf/index.php on line 217
-
TOTAL-POWER

TPHPU63A-110

Medical Power Supplies

Warning: Undefined variable $rtag in /www/wwwroot/website_ic37/www.icpdf.com/pdf/pdf/index.php on line 217
-
TOTAL-POWER

TPHPU63A-111

Medical Power Supplies

Warning: Undefined variable $rtag in /www/wwwroot/website_ic37/www.icpdf.com/pdf/pdf/index.php on line 217
-
TOTAL-POWER

TPHR6503PL

SMALL SIGNAL, FET

Warning: Undefined variable $rtag in /www/wwwroot/website_ic37/www.icpdf.com/pdf/pdf/index.php on line 217
-
TOSHIBA

TPHR6503PL1

Power Field-Effect Transistor

Warning: Undefined variable $rtag in /www/wwwroot/website_ic37/www.icpdf.com/pdf/pdf/index.php on line 217
-
TOSHIBA

TPHR9003NL

Nch VDSS≤30V

Warning: Undefined variable $rtag in /www/wwwroot/website_ic37/www.icpdf.com/pdf/pdf/index.php on line 217
-
TOSHIBA

TPI

SPST THT Tactile Switches with LED

Warning: Undefined variable $rtag in /www/wwwroot/website_ic37/www.icpdf.com/pdf/pdf/index.php on line 217
-
GREATECS

TPI0505S-2W

2W 3KVDC Isolated & Regulated Single and Dual Output DC/DC Converters

Warning: Undefined variable $rtag in /www/wwwroot/website_ic37/www.icpdf.com/pdf/pdf/index.php on line 217
-
TOPPOWER

TPI0505SA-2W

2W 3KVDC Isolated & Regulated Single and Dual Output DC/DC Converters

Warning: Undefined variable $rtag in /www/wwwroot/website_ic37/www.icpdf.com/pdf/pdf/index.php on line 217
-
TOPPOWER

TPI0509S-2W

2W 3KVDC Isolated & Regulated Single and Dual Output DC/DC Converters

Warning: Undefined variable $rtag in /www/wwwroot/website_ic37/www.icpdf.com/pdf/pdf/index.php on line 217
-
TOPPOWER

TPI0509SA-2W

2W 3KVDC Isolated & Regulated Single and Dual Output DC/DC Converters

Warning: Undefined variable $rtag in /www/wwwroot/website_ic37/www.icpdf.com/pdf/pdf/index.php on line 217
-
TOPPOWER

TPI0512S-2W

2W 3KVDC Isolated & Regulated Single and Dual Output DC/DC Converters

Warning: Undefined variable $rtag in /www/wwwroot/website_ic37/www.icpdf.com/pdf/pdf/index.php on line 217
-
TOPPOWER