15EHT1 [TE]
High Performance RFI Power Line Filters for Medical Equipment; 高性能RFI电源线滤波器的医疗设备型号: | 15EHT1 |
厂家: | TE CONNECTIVITY |
描述: | High Performance RFI Power Line Filters for Medical Equipment |
文件: | 总2页 (文件大小:71K) |
中文: | 中文翻译 | 下载: | 下载PDF数据表文档文件 |
CORCOM Product Guide
High Performance RFI Power Line Filters for Medical Equipment
HT Series
HT6
UL Recognized
CSA Certified
VDE Approved
HT7
HT3
HT1
HT Series
Specifications
Maximum leakage current,each line-to-ground
The HT series power line filters provide superior common-
mode and premium differential-mode attenuation to RFI
noise on the frequency range from 10kHz to 30MHz, for
IEC 60601 and UL2601 patient-connected equipment, as
well as all other electronic equipment that cannot tolerate
leakage current.
@ 120 VAC 60 Hz:
@ 250 VAC 50 Hz:
Hipot rating (one minute):
line-to-ground
2µA
5µA
1500 VAC
1450 VDC
50/60 Hz
line-to-line
They are a size- and cost-effective solution to the problem
of meeting both medical leakage current limits and
conducted emissions limits in equipment with high noise
sources (such as switching power supplies). HT series
filters can meet the following Conducted Emissions
standards:
Operating frequency:
Rated voltage (max.):
250 VAC
Minimum insertion loss in dB:
Line-to-ground in 50 ohm circuit
Current
Frequency–MHz
• European Normalized Standards based on EN55022,
Level A (in many cases, Level B also).
Rating .01 .02 .05 .08 .15 .5
1
2
5 10 20 30
• FCC Part 15, Computing Devices, Class B.
3A
6A
22 32 36
-
49 46 40 30 22 12 12 12
• FCC Part 18, ISM Equipment, Industrial and
Consumer.
16 23 32 41 46 41 33 26 15
9
6
8
5
2
8
2
10A
15A
9
4
15 24 30 36 42 34 22 11 12
18 22 27 41 34 22 12 12
Note: When using the models with an IEC connector,
remember that the leakage current of the companion line
cord, GA400, is more significant than that of the filter -
which may disqualify this line cord from use in patient care
applications.
9
Line-to-line in 50 ohm circuit
Current
Frequency–MHz
For more information on UL2601, see Appendix C.
Rating .01 .02 .05 .08 .15 .5
1
2
5 10 20 30
Electrical Schematics
3 & 6 Amp
3A
6A
3
4
7
1
1
8
30
-
61 70 65 65 48 40 32 32
14 45 51 70 70 65 55 47 37 37
17 32 52 70 70 70 65 55 40 35
LINE
LOAD
10A
15A
12 16 15 10 51 70 70 70 70 70 65 55
10 Amp
LINE
LOAD
LOAD
15 Amp
LINE
Resistor location for reference only.
70
Catalog 1654001
Revised 05-07
Dimensions are in inches and
millimeters unless otherwise
specified. Values in italics
are metric equivalents.
Dimensions are shown for
USA Cust. Svc.: 1-800-468-2023
CORCOM Prods: 1-847-680-7400
www.corcom.com
www.tycoelectronics.com
reference purposes only.
Specifications subject
to change.
Germany:
49-89-6089-0
CORCOM Product Guide
High Performance RFI Power Line Filters for Medical Equipment (Continued)
HT Series
HT7 & HT7M (with metric insert)
Case Styles
HT1 (3, 6 and 10 Amp)
E
D
B
B
E
C
A
Tapped insert
No. 6-32 x 1/4 (2) - T7
No. M-3 x 5 (2) - T7M
C
D
A
Typical dimensions:
Recommended Panel Cutout
Terminals: .250 [6.35] (5) Holes: .07 [1.8 ] Dia.(4) Slot: .07 x .16 [1.8 x 4.1 ]
Mounting holes: .188 [4.78 ] Dia. (2)
1.575
40.01
0.235
5.96
HT1 (15 Amp)
R
0.910
23.1
0.157
3.98
Dia.
1.295
32.89
E
D B
Panel Cutout (back mount)
Tolerance ±.005 [0.13]
Case Dimensions
2.00±.01
50.80±.25
C
A
A
B
C
D
E
(max)
± .015
± .38
Typical dimensions:
Part No.
(max) (max) (max)
Terminals: .250 [6.35] (5) Holes: .07 [1.8 ] Dia.(4) Slot: .07 x .16 [1.8 x 4.1 ]
Mounting holes: .203 x .156 [5.16 x 3.96 ] Dia. (4)
3.56
90.4
2.55
64.8
2.15
54.6
2.15
54.6
2.25
57.2
2.27
57.7
2.27
57.7
3.12
79.2
3.12
79.2
1.81
46.0
1.81
46.0
1.78
45.2
1.8
2.938 3.38
74.63 85.9
2.938 3.38
74.63 85.9
1.575 0.63
40.01 16.0
4.063 4.47
103.2 113.5
4.063 4.47
103.2 113.5
3EHT1, 6EHT1
HT3
3EHT3, 6EHT3
Blue
Blue
3EHT7, 3EHT7M, 3.52
6EHT7, 6EHT7M 89.4
Green/Yellow
Brown
†
B
Brown
4.69
119.1
3.69
93.7
5.45
138.4
5.95
151.1
10EHT1
45.7
1.8
A
D
E
C
10EHT3
45.7
2.18
55.4
2.18
55.4
3.5
88.9 100.6
3.5 3.96
3.96
15EHT1
Typical dimensions:
Wire leads: 4.0 [101.6 ] Min.
Mounting holes: .188 [4.78 ] Dia. (2)
15EHT6
HT6
88.9 100.6
† ±.02 [±.5]
Part Numbers
E
D B
6EHT1
6EHT3
6EHT7
6EHT7M
10EHT1
10EHT3
15EHT1
15EHT6
2.00±.01
50.80±.25
Line Cord No.
GA400
A
C
Typical dimensions:
Terminals: No. 8-32 (5)
Torque: 18±2 in.lb
1/2
7
foot, 3-conductor line cord to mate with T7 models
Mounting holes: .203 x .156 [5.16 x 3.96 ] Dia. (4)
71
Catalog 1654001
Revised 05-07
Dimensions are in inches and
millimeters unless otherwise
specified. Values in italics
are metric equivalents.
Dimensions are shown for
reference purposes only.
Specifications subject
to change.
USA Cust. Svc.: 1-800-468-2023
CORCOM Prods: 1-847-680-7400
www.corcom.com
www.tycoelectronics.com
Germany:
49-89-6089-0