WAVEJET 354 TOUCH [TELEDYNE]

DGTL O-SCOPE 500MHZ 4CH 2GS/S;
WAVEJET 354 TOUCH
型号: WAVEJET 354 TOUCH
厂家: TELEDYNE TECHNOLOGIES INCORPORATED    TELEDYNE TECHNOLOGIES INCORPORATED
描述:

DGTL O-SCOPE 500MHZ 4CH 2GS/S

文件: 总8页 (文件大小:4275K)
中文:  中文翻译
下载:  下载PDF数据表文档文件
WaveJet Touch  
Oscilloscopes  
350 MHz / 500 MHz  
Key Features  
350 MHz and 500 MHz bandwidths  
Up to 2 GS/s sample rate  
of touch operation in a portable design. With up to 5 Mpts of memory  
and 2 GS/s, every detail of the waveform can be captured and easily  
measured. The all-inclusive WaveJet Touch delivers maximum value for  
minimum investment.  
Long Memory – up to 5 Mpts  
7.5” touch screen display  
26 measurement parameters  
Replay mode  
Standard Pass/Fail Mask and  
Measurement testing  
Touch Screen Control  
Portable Performance  
Standard I2C, SPI, and UART  
serial triggers  
Touch screen operation simplifies  
how all aspects of the oscilloscope  
are controlled, increasing productivity  
and decreasing setup time. Intuitively  
touch controls instead of navigating  
confusing soft keys.  
The small form factor and light-weight  
design make it easy to carry and use  
anywhere. Coupled with an ultra-quick  
boot time, the WaveJet Touch is ready  
to use when needed. The WaveJet  
Touch is on and acquiring signals in  
less than 5 seconds, enabling  
Standard USB Host, USB Device,  
GPIB, and LAN connectivity  
Multi-language user-interface  
and help  
Shorten Debug Time  
Ultra-quick boot time  
measurements to begin immediately.  
Intricate signals can be quickly acquired  
by combining a long capture time with  
a variety of complex triggers, including  
triggers for I2C, SPI, and UART.  
Flexible Connectivity  
Easily document results by saving  
screenshots directly to a memory stick  
or printer. Standard GPIB, LAN, and  
USB connections enable easy remote  
control. Using Teledyne LeCroy’s  
WaveStudio the WaveJet Touch can  
quickly connect to a PC.  
Pass/fail mask and measurement  
testing enables deeper analysis and  
performs characterization and valida-  
tion testing with ease. Scroll back in  
time to view previous waveforms and  
isolate anomalies using Replay mode.  
ADVANCED TOOLS FOR WAVEFORM ANALYSIS  
Long Capture Time  
Long memory is critical for maintain-  
ing high sample rate during long data  
captures. The standard 5 Mpts of  
memory eliminates the tradeoff  
between long capture and high  
sample rate. The long memory  
makes the WaveJet Touch a great  
tool for viewing signals that have low  
frequency and high frequency com-  
ponents. The built-in zoom function  
provides an easy way to get a close  
look at all of the signal details.  
Advanced Triggering  
Digital Filtering  
Multiple Acquisition Modes  
Equipped with a variety of complex  
triggers, the WaveJet Touch can  
easily capture intricate signals.  
Standard I2C, SPI, and UART triggers  
allow for comprehensive protocol  
debug. A pattern logic trigger can trig-  
ger the oscilloscope depending upon  
the state of the four channel inputs.  
Digital filtering is available on each  
channel. The Low-Pass, High-Pass,  
and Simple Moving Average filters  
allow the oscilloscope to isolate only  
the desired frequencies. By imple-  
menting the filter directly on each  
channel the math trace is still avail-  
able for additional analysis.  
In addition to normal sampling,  
the high resolution, peak detect,  
and average acquisition modes make  
it easy to capture a wide range of  
signal types. An Equivalent time  
sampling rate of 100 GS/s allows for  
reconstruction of repetitive signal  
with extreme precision.  
2
Waveform Math and Analysis  
Pass/Fail Testing  
Automatic Measurements  
Math operators include basic arith-  
metic functions and advanced FFT,  
derivative, and integral functions.  
Advanced math can be performed on  
the results of basic math functions.  
The scaling and offset of the math  
trace can easily be adjusted with its  
own dedicated knobs.  
Waveforms can be tested against a  
mask or measurement parameters  
can be assigned pass/fail criteria for  
validation testing. On a pass/fail  
condition the oscilloscope can be  
configured to: stop the acquisition,  
beep, save the waveform, take a  
screen capture or output a pulse to  
another instrument.  
Measure waveforms with up to  
26 different automatic parameters  
and min/max statistics. Display up to  
four measurements simultaneously.  
All instance measurements allow  
the oscilloscope to measure all  
occurrences of a parameter in a  
single acquisition.  
Replay Mode Waveform Playback  
Scroll back in time using Replay mode to view previous waveforms and isolate anomalies. Use cursors and measurement  
parameters to quickly find the source of problems. Replay mode is always active so there is no need to worry about turning it on.  
Go Back in Time to Identify the Source of a Problem  
3
WAVEJET TOUCH  
The WaveJet Touch delivers the  
performance, features, and touch  
screen user interface to simplify  
operation and shorten debug time.  
4
1
1. 7.5” Touch Screen – Directly touch the  
display for a simplified user experience.  
2. Ultra-quick Boot Time – The WaveJet  
Touch is on and ready to acquire  
waveforms in less than 5 seconds.  
3. Connectivity – Documenting results is  
easy using the front-mounted USB port.  
Simply press the Print button on the front  
panel to quickly save screen images.  
4. Portability – The small footprint and light  
weight of the WaveJet Touch means it  
is easy to carry and use anywhere, even  
when bench space is limited.  
5. Local Language User Interface and  
Help – Select from 7 different language  
preferences.  
3
2
6. Auto Setup – Quickly configure vertical,  
horizontal, and trigger settings with a push  
of a button.  
4
7. Replay Control – Rotate to see a history of  
captured waveforms.  
6
8. Independent Vertical Controls – Quickly  
5
change the vertical scale of any channel.  
9
7
9. Push Knobs – Push the Offset knobs to  
automatically zero the channel offset or push  
the Delay knob to center the trigger point on  
the screen.  
10. GPIB and LAN Ports – Standard connectivity  
for easy remote control and data transfer.  
11. USB Host Port – Can be configured to print a  
hardcopy or an interface for remote control.  
8
12. Auxiliary Output – Send a trigger out or  
pass/fail pulse to another instrument.  
10  
11 10  
12  
5
ACCESSORIES  
LogicStudio  
The WaveJet Touch can be paired with Teledyne LeCroy’s LogicStudio 16 to turn your PC into a mixed signal oscilloscope with  
tools for capturing, viewing, and measuring analog, digital, and serial signals in one place. LogicStudio offers 16 channels,  
Soft Carrying Case  
Rack Mount Kit  
The small form factor of the WaveJet Touch lends itself to  
being conveniently transported. The WJT-SOFTCASE provides  
an easy way to carry the oscilloscope and all of its accesso-  
ries. It is equipped with a custom foam insert to ensure the  
oscilloscope is secure and protected during transport.  
The WaveJet Touch can easily be integrated into a test rack  
using the WJT-RACK accessory. The sturdy rack mount is  
simple to assemble and converts the oscilloscope to a  
6U rack-mounted package. With standard pass/fail mask  
and measure testing, the WaveJet Touch is a natural fit for  
production line testing.  
6
SPECIFICATIONS  
WaveJet Touch 334  
WaveJet Touch 354  
Vertical  
Bandwidth (@50 Ω)  
Rise Time  
350 MHz  
1 ns (typical)  
500 MHz  
750 ps (typical)  
4
Input Channels  
Vertical Resolution  
Sensitivity  
DC Gain Accuracy  
Bandwidth Limiting Filter  
Maximum Input Voltage  
Input Coupling  
8-bits  
50 Ω: 2 mV/div – 2 V/div; 1M Ω: 2 mV/div – 10 V/div  
±(1.5% of + 0.5% of Full Scale)  
200 kHz, 2 MHz, 20 MHz, 100 MHz  
50 Ω: 5 Vrms; 1M Ω: 400 Vpk CAT I  
50 Ω: DC, GND; 1M Ω: AC, DC, GND  
Input Impedance  
Offset Range  
Offset Accuracy  
50 Ω: ±1.0 %; 1M Ω: ±1.0% || 16 pF (typical)  
2 mV/div - 50 mV/div: ±1 V, 50.2 mV/div - 500 mV/div: ±10 V, 502 mV/div - 10 V/div: ±100 V  
2 mV/div - 50 mV/div: ±(0.5% of offset value + 0.5% FS + 1 mV)  
50.2 mV/div - 500 mV/div: ±(0.5% of offset value + 0.5% FS + 10 mV)  
502 mV/div - 10 V/div: ±(0.5% of offset value + 0.5% FS + 100 mV)  
Acquisition  
Sampling Rate (Single Shot)  
Sampling Rate (Equivalent Time)  
Record Length  
2 GS/s (interleaved), 1 GS/s (all channels)  
100 GS/s  
5 Mpts/Ch (interleaved), 2.5 Mpts/Ch (all channels)  
Real Time, Peak Detect, Average, High Resolution  
Acquisition Modes  
Real Time Timebase Range  
Roll Mode Timebase Range  
Peak Detect Period  
1 ns/div - 50 s/div  
50 ms/div - 50 s/div  
1 ns  
500 ps/div - 50 s/div  
Timebase Accuracy  
10 ppm (typical)  
Probes  
Standard Probes  
Probing System  
10:1 Passive Probe (one per channel)  
BNC with Probe Sense Ring  
Trigger System  
Modes  
Sources  
Auto, Normal, Single, Stop  
Any input channel, External, Ext/10, or line  
Coupling  
Trigger Types  
I2C  
DC, AC, HFREJ, LFREJ, Noise Reject  
Edge, Edge ALT, Edge OR, Pulse Width, Period, Pulse Count, Dropout, TV, Logic, I2C, SPI, UART  
Trigger on START, RESTART, STOP, ADDR, DATA, ADDR+DATA, Data Length, or Missing ACK. 7 or 10-bit ADDR  
are supported with full Read, Write, or R/W = “Don’t Care”. DATA conditions support <=, <, =, >, >=, <>, in range,  
out of range, and don’t care (EEPROM mode only). DATA can be setup in Hexadecimal (1-5 bytes supported).  
Full range of bit rates supported for Standard, Fast, Fast-Mode Plus, and High speed mode. Bit rate is auto-  
detected. Trigger on any analog channel, EXT, or EXT/10.  
Trigger on DATA. DATA can be setup in Binary (any combination of 0, 1, or X for 4-96 bits). All bit rates are  
supported and are auto-detected. Trigger on any analog channel, EXT, or EXT/10.  
Trigger on START, STOP, DATA, or Parity ERROR. DATA can be setup in Hexadecimal (1 byte supported) or  
Binary (any combination of 0, 1, or X for 5-8 bits). All bit rates are supported between 1 kb/s and 1 Mb/s.  
Trigger on any analog channel, EXT, or EXT/10.  
SPI  
UART / RS-232  
Measure, Zoom, Math and Replay Mode  
Measure  
Base, Cycle Mean, Cycle RMS, Duty Cycle, Fall Time (90-10%), Fall Time (80-20%), Frequency, Integral, Maxi-  
mum, Mean, Minimum, Number of +Pulses, Number of -Pulse, +Overshoot, -Overshoot, Peak-Peak, Period,  
+Pulse Width, -Pulse Width, Rise Time (20-80%), Rise Time (10-90%), RMS, Skew, Skew@level, Top, Top-Base  
Zoom  
Math  
Use front panel QuickZoom button to zoom all waveforms in a separate zoom grid  
Sum, Difference, Product, Integral, Derivative, FFT (up to 8 kpts with Rectangular, Von Hann, or  
Flat Top windows)  
Replay Mode  
Look back at the history of waveform acquisitions (maximum 2,048 acquisitions)  
Display System  
Display Size  
Display Resolution  
7.5” TFT-LCD Touch-Screen  
640 x 480 VGA  
7
SPECIFICATIONS & ORDERING INFORMATION  
Specifications (cont’d)  
WaveJet Touch 334  
WaveJet Touch 354  
Input/Output Interfaces  
Aux Out  
USB  
Trigger output or pass/fail output  
USB host port for flash drives, USB device port for connecting to PC and direct printer connection  
10/100Base-T Ethernet interface (RJ-45 connector)  
Supports IEE-488.2  
LAN  
GPIB  
Physical  
Dimensions (HWD)  
Weight  
7.5"H x 13"W x 4.9"D (190 mm x 330 mm x 124 mm)  
3.7 kg (8.16 lbs)  
Environment  
Temperature  
Humidity  
Operating: 0 °C to 40 °C; Non-Operating: -20 °C to 60 °C  
Maximum 80% relative humidity (non-condensing) up to ≤ 30 °C, Upper limit derates to 55% relative humidity  
(non-condensing) at 40 °C  
Operating: 3,000 m (9,843 ft) max at 25C; Non-Operating: Up to 12,192 meters (40,000 ft)  
Altitude  
Power Requirements  
Voltage  
Power Consumption (Max)  
100 - 240 V (± 10%) at 50 / 60 Hz (± 5%), 100 - 120 V (± 10%) at 400 Hz (± 5%)  
50 W  
Regulatory  
CE Certification  
Low Voltage Directive 2006/95/EC; EN 61010-1:2010, EN 61010-2-030:2010  
EMC Directive 2004/108/EC; EN 61326-1:2013, EN61326-2-1:2013; RoHS2 Directive 2011/65/EU  
UL 61010-1, UL 61010-2-030:2010, 3rd Edition; CAN/CSA C22.2 No. 61010-1-12  
UL and cUL Listing  
Ordering Information  
Product Description  
Product Code  
Product Description  
General Accessories  
Soft Carrying Case  
Rack Mount Accessory  
16 Channel, 1 GS/s, 100 MHz USB Logic Analyzer  
Product Code  
WaveJet Touch Oscilloscopes  
350 MHz, 1 GS/s, 4 Ch, 2.5 Mpts/Ch with  
7.5” Touch screen Display;  
2 GS/s, 5 Mpts Interleaved  
500 MHz, 1 GS/s, 4 Ch, 2.5 Mpts/Ch with  
WJT-SOFTCASE  
WJT-RACK  
LogicStudio 16  
WaveJet 334T  
WaveJet 354T  
7.5” Touch screen Display;  
2 GS/s, 5 Mpts Interleaved  
Probes  
10:1 500 MHz 10  
Passive Probe  
PP006C  
AP031  
HVP120  
PPE1.2KV  
700 V, 15 MHz High-Voltage Differential Probe  
100:1 400 MHz 50 MΩ 1 kV High-voltage Probe  
10:1/100:1 200/300 MHz, 50 MHigh-voltage Probe  
600 V/1,2 kV Max. Volt. DC  
100:1 400 MHz 50 M2 kV High-voltage Probe  
100:1 400 MHz 50 M4 kV High-voltage Probe  
1000:1 400 MHz 50 M5 kV High-voltage Probe  
1000:1 400 MHz 50 M6 kV High-voltage Probe  
Included with Standard Configurations  
Protective Front Cover  
One Passive Probe per Channel  
Getting Started Guide  
Multi-language User Interface and Context Sensitive Help  
(English, Chinese, French, German, Italian, Japanese, and Russian)  
GPIB, LAN, USB (one host and one device) Ports  
Power Cable for the Destination Country  
Calibration and Performance Certificate  
3-year Warranty  
PPE2KV  
PPE4KV  
PPE5KV  
PPE6KV  
Customer Service  
Teledyne LeCroy oscilloscopes and probes are designed, built, and tested to ensure high reliability. In the unlikely event you experience  
difficulties, our digital oscilloscopes are fully warranted for three years and our probes are warranted for one year. This warranty includes:  
No charge for return shipping  
Long-term 7-year support  
Upgrade to latest software at no charge  
1-800-5-LeCroy  
teledynelecroy.com  
Local sales offices are located throughout the world.  
Visit our website to find the most convenient location.  
wavejet-touch-ds-30sep15  
© 2015 Teledyne LeCroy, Inc. All rights reserved. Specifications, prices, availability, and delivery subject to change without notice.  
Product or brand names are trademarks or requested trademarks of their respective holders.  

相关型号:

WAVESURFER 10

DGTL O-SCOPE 1GHZ 10GS/S 4CH
TELEDYNE

WAVESURFER 10M

1 GHZ, 10 GS/S, 4 CH, 16 MPTS/CH
TELEDYNE

WAVESURFER 510

1 GHZ OSCILLOSCOPE, 10 GS/S, 4 C
TELEDYNE

WAVESURFER 510 DK PACK

OSCOPE 1GHZ W/FREE SOFTWARE PACK
TELEDYNE

WB-841

Butt - Window Splice for 14-16 AWG Wire
MOLEX

WB-841T

Wire Terminal,
MOLEX

WB0130KRM

100mA Surface Mount Schottky Barrier Rectifiers - 30V
WILLAS

WB0130QRL

100mA Surface Mount Schottky Barrier Rectifiers - 30V
WILLAS

WB05

Analog Buffer/Voltage Follower
ETC

WB0610-301Y

THRU-HOLE INDUCTORS BDC & WB Series: Beads & Chokes
RFE

WB0610-301Y/120MHZ

Ferrite Bead, 1 Function(s),
KOA

WB0610-451Y

THRU-HOLE INDUCTORS BDC & WB Series: Beads & Chokes
RFE