MB115S36K [SLPOWER]
AC-DC Regulated Power Supply Module,;型号: | MB115S36K |
厂家: | SL Power Electronics |
描述: | AC-DC Regulated Power Supply Module, |
文件: | 总8页 (文件大小:581K) |
中文: | 中文翻译 | 下载: | 下载PDF数据表文档文件 |
MB115S
115W Single Output
LED Medical Series
Features
▪
▪
▪
▪
▪
▪
▪
▪
▪
▪
▪
Small size of 2” x 4” x 1.3”
For 1U Applications
75W convection cooled
115 Watts with 200 LFM
Universal Input 90-264Vac
Approved to UL/CSA/IEC/IEC60601-1, 3rd Edition
Level V Efficiency Compliant
-40˚C start up
-20˚C to 70˚C Operating Temperature Range
3 Year Warranty
Optional LED indicator for power-on
Description
The MB115S LED with Medical Approval Series provide a reliable power source with high power density in 2” x
4” x 1.3” package. Fully compliant to the applicable safety and Global Lighting EMC standards, these models
will allow easy integration into many Lighting fixtures and other medical applications.
Model Selection (only Highlighted ones require at this time)
Model
Output Current
Output Current
Forced air(200 LFM)
(Total Power)
9.00A
Ripple &
Total
OVP
Number
MB115S12K
Volts
12 V
Convection Cooled
6.25 A
Noise*
0.5%RMS,
1.5% pk-pk
0.5%RMS,
1% pk-pk
0.5%RMS,
1% pk-pk
0.5%RMS,
1% pk-pk
0.5%RMS,
1% pk-pk
Regulation
±2%
Threshold
14.0 ± 1.1V
(108 Watts)
4.58A
(110 Watts)
3.19A
MB115S24K
MB115S36K
MB115S48K
MB115S56K
24 V
36V
3.13A
2.08
±2%
±2%
±2%
±2%
28.0 ± 2.5V
42V± 4.0V
55.0 ± 4.0V
63.0 ± 4.0V
48 V
56 V
1.56A
1.34A
2.40A
(115 Watts)
2.05A
(115 Watts)
Notes:
* At -20C, the noise and ripple is 2% of the output.
General & Input Specifications
PARAMETER
SPECIFICATION
NOTES
AC Input Voltage:
90-264Vac, single phase
AC Input Frequency:
AC Input Current:
Inrush Current:
47-63Hz
115Vac: 2A, 230Vac: 1A
65A maximum @ 25C
Earth Leakage Current
<350uA@264Vac, 60 Hz input, NC
(Input–Earth):
Input Fuse:
F1:4A, 250VAC
Fuse provided on all models
www.slpower.com
MB115S 25-AUG-16
1
MB115S
115W Single Output
LED Medical Series
Efficiency
Typical
Measured @ 25˚C
MB115S12K
MB115S24K
MB115S36K
MB115S48K
MB115S56K
89% @230V ac, full load
89% @230V ac, full load
89% @230V ac, full load
90% @230V ac, full load
90% @230V ac, full load
86.5%@115V ac, full load
87%@115V ac, full load
87%@115V ac, full load
88%@115V ac, full load
88%@115V ac, full load
-40˚C start up (full load)
For 12V output, the maximum load is 75%
Operating Temperature
Storage Temperature
-20˚C to 70˚C
-40˚C to 85˚C
<2 Seconds @115Vac(<3S for 12V
output)
<5 Seconds @115Vac for -20˚C ambient
Turn-on Time:
Hold-up Time:
12mS minimum from loss of ac input at
115 Vac
DC Output Specifications
PARAMETER
SPECIFICATION
NOTES
Max of 75 Watts for Convection cooled
Maximum 108 Watts for 12V output
Output Power:
Cooling:
Max of 115 Watts for fan cooled
(48 & 56V Models)
-20 to 50˚C ambient
Convection
Forced Air of 200 LFM
Total regulation is the maximum deviation from
nominal voltage for all loading conditions
Total Regulation:
±2% for all models
120% - 180% of rated output current
value, Hiccup Mode
Overload Protection:
For 12V output, it is 110 to 180%
Short Circuit
Protection:
Short across the output terminals will not
cause damage to the unit. Hiccup Mode
OVP firing reduces output voltage to
<50% of nominal in <50mS. See chart for
trip range
Overvoltage
Protection:
Overtemperature
Protection:
Automatic Power Shutdown
No minimum load is required
Thermistor temperature is 130˚C
Minimum Load:
20 MHz Bandwidth, differential mode.
Measured with noise probe directly across
output terminals, and load terminated with
0.1μF ceramic and 10μF low ESR capacitors
0.5% RMS, 1% pk-pk for all models.
Ripple and Noise:
500s typ. response time for return to
within 0.5% of final value for a 50% load
change, i/t< 0.2A/s. Max. voltage
deviation is 3.5%.
Transient Response:
Measured @ 25˚C
5% overshoot at turn-on, 5% overshoot at
turn-off, under all conditions.
Overshoot:
6% for 12V output
www.slpower.com
MB115S 25-AUG-16
2
MB115S
115W Single Output
LED Medical Series
Safety Standard Compliance
Agency
CONDITIONS
UL
EN/CSA/UL/IEC 60601-1 3rd Edition & EN60950
CSA
Same as above
Demko
Same as above
Yes
CB Report
Isolation Type:
Double/Reinforced between Input and Output
Isolation Specifications
PARAMETER
CONDITIONS
Rating
NOTES
Insulation Safety
Rating:
Basic Insulation
Input to Ground
Input to Output
Input to Ground
Input to Output
Double/Reinforced
2,000Vac
Electric Strength Test
Voltage:
4,000Vac
.
Output to Ground
500Vac
Environmental Specifications
PARAMETER
SPECIFICATION
NOTES
Operating
Temperature:
-20°C to +70°C
-40°C Startup guaranteed
60% derating at 70°C
Convection/ Airflow
Temperature Derating:
Cooling:
75 Watts convection
Storage Temperature:
-40°C to +85°C
Operating: -500 to 3,000 meter
Non-operating: -500 to 40,000 ft.
5% to 95%, non-condensing
Altitude:
Relative Humidity:
Non-Operating: Half-sine, 40 gpk, 10mS, 3 axes, 6
shocks total
Shock:
Random vibration per MIL-STD-810E, Method
514.4, Cat. 1, Figure 514.4-1, 1 hr in each of three
axes
Vibration:
Reliability Specifications
PARAMETER
SPECIFICATION
NOTES
Calculation is done based on
Telcordia. Reports for each
model is available
MTBF:
574K hours, 25°C ambient, full load
Warranty:
3 Years
Limited
HALT Data:
Per SL Power Halt procedure
Report is available
www.slpower.com
MB115S 25-AUG-16
3
MB115S
115W Single Output
LED Medical Series
EMI/EMC Compliance
PARAMETER
SPECIFICATION
NOTES
Conducted Emissions:
Radiated Emissions:
EN55011/22 Class B; FCC Part 15
Also meets EN55015 Class B
EN55011/22 Class A; FCC Part 15
EN61000-3-2, Class A,B,C &D
Harmonic Current
Emissions
Meets class C from 5 to 115 Watts.
This is based on limits set @ 115 Watt
Voltage Fluctuations &
Flicker
EN61000-3-3
Performance criteria are defined as
following:
Static Discharge
Immunity:
EN61000-4-2, Level 4: 6kV contact, 8kV air,
Criteria A
RF Field Susceptibility
Fast Transients/Bursts
EN61000-4-3, Level 3 (3V/m), Criteria A
A – Normal performance
during and after the test
B – Temporary degradation,
self-recoverable
C – Temporary degradation,
operator intervention required
to recover the operation
EN61000-4-4, Level 3 (PS: 2kV-40A, other
lines 1kV-20A), Criteria A
EN61000-4-5, Installation Class 3 (1kV diff.
mode, 2kV common mode), Criteria A
Surge susceptibility
Conducted RF
susceptibility
Power Frequency
Magnetic Field Test
EN61000-4-6, Level 3 (3Vrms), Criteria A
EN61000-4-8, Level 3 (3A/m), Criteria A
EN61000-4-11,
95% dip/0.5 cycle (Criteria A),
60%/5cycles (Criteria B), 30%/25 cycles
(Criteria A).
Voltage Sags & Surges
loading is 70% of 100 watts with 100 Vac
input.
Notes:
1. Specifications subject to change without notice.
2. Specifications are for convection rating at factory settings with 115Vac input and 25°C ambient unless
otherwise stated.
Mechanical Drawing
www.slpower.com
MB115S 25-AUG-16
4
MB115S
115W Single Output
LED Medical Series
Connector Information
Input Connector
J2
DC Output
Connector
J3
Ground
(FG)J1
PIN 1) +Vout
PIN 2) +Vout
PIN 3) +Vout
PIN 4) -Vout
PIN 5) -Vout
PIN 6) -Vout
Mating Connector:
AMP 640250-6
Terminals :
PIN 1) AC NEUTRAL
PIN 2) EMPTY
PIN 3) AC LINE
19-30258-0187 (Keystone 1285)
(Zierick 895)(.187*0.020)
Mating Connector:
Tyco/AMP 640250-3
Terminals :
Mating Connector
Molex 190020005
3-640252-1
3-640252-1
1. All dimensions in inches (mm) undefined tolerance is ±.02” (0.5mm).
2. Mounting holes should be connected together for EMI purpose
3. FG is safety ground connection
4. This power supply requires mounting on metal standoffs 0.20” (5mm) Min. in height
Characteristic Curves
www.slpower.com
MB115S 25-AUG-16
5
MB115S
115W Single Output
LED Medical Series
Output Power vs. Temperature
Efficiency vs. Loading
Ripple & Noise
www.slpower.com
MB115S 25-AUG-16
6
MB115S
115W Single Output
LED Medical Series
To verify that the output ripple and noise does not exceed the level specified in the product specification, measured using a scope probe socket with
0.1uF ceramic and a 10uF electrolytic capacitor connected in parallel across it, 20MHz BW.
Output Overload Characteristic
Overvoltage Protection
Turn On Time
www.slpower.com
MB115S 25-AUG-16
7
MB115S
115W Single Output
LED Medical Series
Hold Up Time
CH1:
CH3:
Vout
Vin:
Iout:
115
2.40
23.2
Vac
Vin
16
Amps
mS
Min_Limit:
Meas:
Data Sheet © 2016 SL Power Electronics Corp. The information and specifications contained herein are believed to be correct at the time of publication. However, SL Power accepts no responsibility for
consequences arising from reproduction errors or inaccuracies. Specifications are subject to change without notice.
www.slpower.com
MB115S 25-AUG-16
8
相关型号:
MB11A10-6P
MIL Series Connector, 6 Contact(s), Aluminum Alloy, Male, Crimp Terminal, Receptacle
AMPHENOL
©2020 ICPDF网 联系我们和版权申明