FJDL-A46-P [MSYSTEM]
Space-saving Plug-in Signal Conditioners F-UNIT;型号: | FJDL-A46-P |
厂家: | M-SYSTEM |
描述: | Space-saving Plug-in Signal Conditioners F-UNIT |
文件: | 总3页 (文件大小:134K) |
中文: | 中文翻译 | 下载: | 下载PDF数据表文档文件 |
MODEL: FJDL
Space-saving Plug-in Signal Conditioners F-UNIT
LINEARIZATION CODE
1: Segment data
2: Square root extraction
3: Thermocouple
4: RTD
DEFAULT
Linear
–––
CURRENT LOOP SUPPLY
(linearizing; field-programmable)
Functions & Features
K
0 – 1000°C
Pt 100 0 – 100°C
• Powering a 4 – 20 mA DC current loop
• Microprocessor based
• Shortcircuit protection
INPUT
Current
• Applicable to smart transmitters
• Field-programmable linearization data
• Loop testing via hand-held programmer PU-2x
• High-density mounting
A: 4 – 20 mA DC (Input resistance 250 Ω)
[1] LINEARIZATION
0: None
Typical Applications
1: Segment data
2: Square root extraction
3: Thermocouple
4: RTD
• Various 2-wire transmitters
• Providing isolation and linearization for a 2-wire
temperature transmitter
• Linearizing weir flowmeter output to provide a linear-to-
volume signal
• Linearizer application
[2] OUTPUT
Current
(4 – 20 mA input)
• Square root extraction for differential pressure transmitter
A: 4 – 20 mA DC (Load resistance 600 Ω max.)
Voltage
26 (1.02)
6: 1 – 5 V DC (Load resistance 500 Ω min.)
[3] POWER INPUT
100
(3.94)
AC Power
K: 85 – 132 V AC
(Operational voltage range 85 – 132 V, 47 – 66 Hz)
L: 170 – 264 V AC
137
(5.39)
mm (inch)
(Operational voltage range 170 – 264 V, 47 – 66 Hz)
DC Power
R: 24 V DC
MODEL: FJDL–A[1][2]–[3][4]
(Operational voltage range 24 V ±10 %, ripple 10 %p-p max.)
P: 110 V DC
(Operational voltage range 85 – 150 V, ripple 10 %p-p max.)
ORDERING INFORMATION
• Code number: FJDL-A[1][2]-[3][4]
Specify a code from below for each of [1] through [4].
(e.g. FJDL-A1A-R/Q)
[4] OPTIONS
blank: none
Default setting (following table) will be used if not otherwise
specified.
/Q: With options (specify the specification)
No linearization data will be programmed if you don't
specify type of linearization and required data.
• Linearization data
SPECIFICATIONS OF OPTION: Q (multiple selections)
COATING (For the detail, refer to M-System's web site.)
/C01: Silicone coating
Code 1 segment data: Use Ordering Information Sheet (No.
ESU-1669) to specify linearization data.
Code 3 T/C, Code 4 RTD: Specify input sensor type and
temperature range.
/C02: Polyurethane coating
/C03: Rubber coating
TERMINAL SCREW MATERIAL
/S01: Stainless steel
• Specify the specification for option code /Q
(e.g. /C01/S01)
FJDL SPECIFICATIONS
ES-1474 Rev.12 Page 1/3
http://www.m-system.co.jp/
MODEL: FJDL
RELATED PRODUCTS
• Thermocouple linearizable range
• Programming Unit (model: PU-2x)
USABLE RANGE
°C
T/C
• PC configurator software (model: JXCON)
Downloadable at M-System’s web site.
A dedicated cable is required to connect the module to the
PC. Please refer to the internet software download site or
the users manual for the PC configurator for applicable
cable types.
°F
(PR)
K (CA)
E (CRC)
0 to 1760
-270 to +1370
-270 to +1000
32 to 3200
-454 to +2498
-454 to +1832
J
(IC)
T (CC)
B (RH)
-210 to +1200
-270 to +400
0 to 1820
-346 to +2192
-454 to +752
32 to 3308
R
S
-50 to +1760
-50 to +1760
-58 to +3200
-58 to +3200
GENERAL SPECIFICATIONS
Construction: Plug-in
Note: For the temperatures that range below 0°C, the
transmitter may partially not satisfy the described accuracy.
Consult factory.
Connection: M3.5 screw terminals (torque 0.8 N·m)
Screw terminal: Nickel-plated steel (standard) or stainless
steel
• RTD linearizable range
Housing material: Flame-resistant resin (black)
Isolation: Input to output to power
USABLE RANGE
RTD
°C
°F
Overrange output: Approx. -10 to +120 % at 1 – 5 V
Linearization: 16 points max. represented as percentage of
full-scale
JPt 100 (JIS ’89)
Pt 100 (JIS ’89)
Pt 100 (JIS ’97, IEC)
Pt 50Ω (JIS ’81)
Ni 508.4 Ω
-200 to +500
-200 to +650
-200 to +650
-200 to +500
-50 to +200
-328 to +932
-328 to +1202
-328 to +1202
-328 to +932
-58 to +392
Adjustments: Programming Unit (model: PU-2x);
linearization data, zero and span, simulating output, etc.
(Refer to the users manual of JXCON for the adjustments
configurable with JXCON.)
Note:Pt100(JIS’89)isdeviatedfromPt100(JIS’97) only
within the described accuracy.
SUPPLY OUTPUT
INSTALLATION
Output voltage: 24 – 28 V DC with no load
Current rating: ≤ 22 mA DC
• Shortcircuit Protection
Power input
•AC: Approx. 4.5 VA
•DC:
Current limited: 30 mA max.
Protected time duration: No limit
24 V approx. 95 mA
110 V approx. 20 mA
Operating temperature: -5 to +55°C (23 to 131°F)
Operating humidity: 30 to 90 %RH (non-condensing)
Mounting: Surface or DIN rail; Standard Rack Mounting
Frame BX-16H available
INPUT SPECIFICATIONS
■ DC Current: Input resistor incorporated
Weight: 220 g (0.49 lb)
LINEARIZATION
•No linearization: The output is proportional to the input.
•Segment data: 16 points (15 segments) max. within the
range of -15.00 to +115.00 % input or output represented
as percentage of fullscale
PERFORMANCE in percentage of span
Accuracy: ±0.1 % with segment gain ≤ 1 [±0.1 % × gain]
with segment gain > 1
Temp. coefficient: ±0.015 %/°C (±0.008 %/°F)
Response time: ≤ 0.5 sec. (0 – 90 %)
•Square root extraction
Low-end cutout: 5 % (output); curve characteristics as in the
figure below
Line voltage effect: ±0.1 % over voltage range
Insulation resistance: ≥ 100 MΩ with 500 V DC
Dielectric strength
■ Square root extraction
Power input code R:
1000 V AC @ 1 minute (input to output)
2000 V AC @ 1 minute (input or output or power to ground)
500 V AC @ 1 minute (I/O to power)
6%
5%
Power input code K, L, P:
1000 V AC @ 1 minute (input to output)
2000 V AC @ 1 minute (input or output or power to ground)
0
INPUT
FJDL SPECIFICATIONS
ES-1474 Rev.12 Page 2/3
http://www.m-system.co.jp/
MODEL: FJDL
1500 V AC @ 1 minute (I/O to power)
EXTERNAL DIMENSIONS & TERMINAL ASSIGNMENTS unit: mm (inch)
CLAMP
(top & bottom)
7.8 (.31)
1
2
15
(.59)
DIN RAIL
35mm wide
3
4
2–4.5 (.18)
HOLE
15 (.59) deep
5
6
8–M3.5
SCREW
7
8
26 (1.02)
107 (4.21)
137 (5.39)
26 (1.02)
[3.3 (.13)]
•When mounting, no extra space is needed between units.
SCHEMATIC CIRCUITRY & CONNECTION DIAGRAM
2-WIRE
TRANSMITTER
Current
Limiter
24V DC
Supply
+
+
–
1
2
3
4
Isolation
Digital
Computation
Output
Driver
–
5
6
7
8
+
OUTPUT
POWER
+
–
MONITOR
V
(1 – 5V DC)
–
250Ω
U(+)
V(–)
MODULAR JACK
PU-2x
SERIAL TRANSMISSION LINE
Base Socket
■ When Used as Linearizer
No Connection
1
2
3
+
250Ω
INPUT
–
Specifications are subject to change without notice.
FJDL SPECIFICATIONS
ES-1474 Rev.12 Page 3/3
http://www.m-system.co.jp/
相关型号: