43AL1-S/UL [MSYSTEM]
Digital Panel Meters 43 Series;型号: | 43AL1-S/UL |
厂家: | M-SYSTEM |
描述: | Digital Panel Meters 43 Series |
文件: | 总6页 (文件大小:188K) |
中文: | 中文翻译 | 下载: | 下载PDF数据表文档文件 |
MODEL: 43AL1
/SET: Preset according to the Ordering Information Sheet
(No. ESU-9425)
Digital Panel Meters 43 Series
LOOP POWERED DIGITAL PANEL METER
(process meter)
GENERAL SPECIFICATIONS
Construction: Panel flush mounting
Connection
Functions & Features
• Loop powered -1999 to 9999 compact digital meter
• Scaling function
Terminal block "S": Screwless spring terminal
Applicable wire size 1.0 to 1.3 mm
Terminal block "D": Separable screwless spring terminal
Applicable wire size 1.0 to 1.3 mm , stripped length 8 mm
2
, stripped length 8 mm
• Screwless spring terminal
2
48 (1.89)
Housing material: Flame-resistant resin (gray)
Isolation: Input to FE(Functional Earth)
A/D conversion: Σ – Δ
24
(.94)
Sampling rate: 10 times/sec. (100 msec.)
Averaging: None or moving average
Setting: (Front button)
58 (2.28)*
mm (inch)
*42 (1.65) for Terminal Block code "S"
• Scaled range
• Moving average
MODEL: 43AL1–[1][2]
DISPLAY
Display: 4 digits of 10.2 mm (.4”) height, 7-segment, red
ORDERING INFORMATION
• Code number: 43AL1–[1][2]
LED
Display range: -1999 to 9999
Specify a code from below for each of [1] and [2].
(e.g. 43AL1-D/UL/Q)
Scaling range for measurement range: -1999 to 9999
counts
• Specify the specification for option code /Q
(e.g. /C01/SET)
Decimal point position: 10-1, 10-2, 10-3 or none
Zero indication: Higher-digit zeros are suppressed.
Over-range indication: '-1999' or '9999' blinking for display
values out of the display range. 'S.ERR' blinks surpassing
the permissible range.
INPUT
4 – 20 mA DC
Engineering unit indication: Sticker label attached
DC, AC, mV, V, kV, μA, mA, A, kA, mW, W,
kW, var, kvar, Mvar, VA, Hz, Ω, kΩ, MΩ,
cm, mm, m, m/sec, mm/min, cm/min, m/min,
[1] TERMINAL BLOCK
S: Screwless spring terminal
D: Separable screwless spring terminal
m/h, m/s2, inch, ℓ, ℓ/s, ℓ/min, ℓ/h, m3, m3/sec,
m3
/min, m
3
/h, Nm /h, N·m, N/m , g, kg, kg/h,
3
2
N, kN, Pa, kPa, MPa, t, t/h, ℃, °F, %RH, J,
kJ, MJ, rpm, sec, min, pH, %, ppm, etc.
[2] OPTIONS (multiple selections)
Standards & Approvals
blank: CE marking
/UL: UL approval, CE marking
Other Options
INPUT SPECIFICATIONS
■ DC Current
blank: none
Current range: 3.75 to 22 mA DC
/Q: Option other than the above (specify the specification)
Voltage drop: Approx. 3.5 V with 4 mA; approx. 3.7 V with
20 mA (Equivalent input impedance: Approx. 185 Ω)
(Use of the unit causes voltage drop. For 2-wire transmitter,
be sure that the voltage by which 2-wire transmitter can
operate is ensured including the voltage drop by other
devices and wiring resistance.)
SPECIFICATIONS OF OPTION: Q (multiple selections)
COATING (For the detail, refer to M-System's web site.)
Moving parts and indicators are not coated.
/C01: Silicone coating
/C02: Polyurethane coating
/C03: Rubber coating (UL not available)
EX-FACTORY SETTING
43AL1 SPECIFICATIONS
ES-9425 Rev.9 Page 1/6
https://www.m-system.co.jp/
MODEL: 43AL1
INSTALLATION
Operating temperature: -10 to +55°C (14 to 131°F)
Operating humidity: 30 to 90 %RH (non-condensing)
Mounting: Panel flush mounting
Weight: 30 g (1.1 oz)
PERFORMANCE
Accuracy: ±0.1 % rdg ±1 digit× scaling-multiple
(When the scaling-multiple is less than 1, rounded up to 1.)
Temp. coefficient: ±0.3 digits× scaling-multiple/℃
(When the scaling-multiple is less than 1, rounded up to 1.)
Scaling-multiple = | (Display Scaling Value B - Display
Scaling Value A) ÷ (default Display Scaling Value B - default
Display Scaling Value A) |
Insulation resistance: ≥ 100 MΩ with 500 V DC
Dielectric strength: 500 V AC @1 minute (input to FE)
STANDARDS & APPROVALS
EU conformity:
EMC Directive
EMI EN 61000-6-4
EMS EN 61000-6-2
RoHS Directive
Approval:
UL/C-UL general safety requirements
(UL 61010-1, CAN/CSA-C22.2 No.61010-1-12)
EXTERNAL VIEW
(1) Front Cover
(2) 4-digit Display
(3) DI Button
(4) SC Button
(6) UP Button
(5) SF Button
■ COMPONENT IDENTIFICATION
No.
(1)
(2)
COMPONENT
Front Cover
FUNCTIONS
Removed at conguration.
4-digit Display
DI Button
4-digit LED display. Range: -1999 to 9999 (not including decimal point)
Used to move on to the display setting modes; or to shift through setting items in each setting
mode.
(3)
(4)
Used to move on to the scaling setting modes; or to shift through setting items in each setting
mode.
SC Button
Used to move on to the setting standby status and shift through display digits in each setting
item.
(5)
(6)
SF Button
UP Button
Used to select setting value.
43AL1 SPECIFICATIONS
ES-9425 Rev.9 Page 2/6
https://www.m-system.co.jp/
MODEL: 43AL1
PARAMETER LIST
■ SCALING SETTING MODE
DEFAULT
VALUE
PARAMETER
DISPLAY
...
FUNCTION
Display value for 4 mA input
To distinguish from B, the first decimal point is blinking.
Display Scaling Value A
Display Scaling Value B
Decimal Point Position
...
Display value for 20 mA input
Decimal point position
10-1 through 10-3 or none
• Normal Scaling
The display value increases when
the input signal increases.
• Inverted Scaling
The display value decreases when
the input signal increases.
Display
B
Display
A
Display Scaling
A
Display Scaling
B
Input
Input
0%
0%
100%
100%
The decimal point position can be set to any digit. Set it according to the 100% value.
■ DISPLAY SETTING MODE
DEFAULT
VALUE
PARAMETER
Moving Average
DISPLAY
FUNCTION
No moving averaging
Moving average with 2 samples
Moving average with 4 samples
Moving average with 8 samples
Non-initialization
Initialization
Initialize settings (change to factory settings) *1
Version number, indication only
Version Indication
–
–
*1. While “
” is shown, pressing DI button or SC button initializes settings.
If “Initialization” is done once, all current parameters will be deleted and overwritten with factory default values. Notice that after this, Ex-factory settings
with “/SET” option will be irrecoverable.
43AL1 SPECIFICATIONS
ES-9425 Rev.9 Page 3/6
https://www.m-system.co.jp/
MODEL: 43AL1
EXTERNAL DIMENSIONS & TERMINAL ASSIGNMENTS unit: mm (inch)
■ TOP VIEW
• Separable terminal
• One block terminal
45 (1.77)
■ FRONT VIEW
■ SIDE VIEW
50.5 (1.99)
48 (1.89)
5.5
35 (1.38)
17.5 (.69) for Separable terminal
1.5 (.06) for One block terminal
(.22)
■ REAR VIEW
• Separable terminal
• One block terminal
Separable screwless
spring terminal
Screwless spring terminal
1
2
3
1
2
3
43AL1 SPECIFICATIONS
ES-9425 Rev.9 Page 4/6
https://www.m-system.co.jp/
MODEL: 43AL1
MOUNTING REQUIREMENTS unit: mm
■ PANEL CUTOUT unit: mm
45 +– 00.6
min. 65
Panel thickness: 0.8 to 3.5 mm
MOUNTING
Just insert the meter body (snap-in method)
SCHEMATIC CIRCUITRY & CONNECTION DIAGRAM
Note: In order to improve EMC performance, bond the FE terminal to ground.
Caution: FE terminal is NOT a protective conductor terminal.
+
–
1
3
2
INPUT
–
+
TWO-WIRE
TRANSMITTER
A/D
Converter
FE
Digital
Computation
Power
Source
Display /
Setting
■ 4-WIRE CONNECTION EXAMPLE
1
3
+
–
INPUT
Controllor, etc.
+
I/P Positioner, etc.
+
4 – 20mA DC
–
–
43AL1 SPECIFICATIONS
ES-9425 Rev.9 Page 5/6
https://www.m-system.co.jp/
MODEL: 43AL1
Specifications are subject to change without notice.
43AL1 SPECIFICATIONS
ES-9425 Rev.9 Page 6/6
https://www.m-system.co.jp/
相关型号: