ML105SM-05RI4 [MPD]

Medical Approved Compact SMT, 1W DC/DC Con vert ers;
ML105SM-05RI4
型号: ML105SM-05RI4
厂家: MICROPOWER DIRECT, LLC    MICROPOWER DIRECT, LLC
描述:

Medical Approved Compact SMT, 1W DC/DC Con vert ers

文件: 总2页 (文件大小:253K)
中文:  中文翻译
下载:  下载PDF数据表文档文件
ML100MRI4  
Medical Approved  
Compact SMT, 1W  
DC/DC Converters  
Electrical Specifications  
Specifications typical @ +25°C, nominal input voltage & rated output current, unless otherwise noted. Specifications subject to change without notice.  
Key Features:  
Input  
Parameter  
EN 60601 3RD Ed. Approved  
1W Output Power  
Conditions  
Min.  
Typ.  
5.0  
Max.  
5.5  
Units  
5 VDC Input  
4.5  
4 kVrms Isolation  
Input Voltage Range  
12 VDC Input 10.8  
24 VDC Input 21.6  
Internal Capacitor  
12.0  
24.0  
13.2  
26.4  
VDC  
Reinforced Insulation  
2 x MOPP per EN 60601-1  
3RD Edition & ANSI/AAMI  
ES 60601-1  
Input Filter  
Output  
Parameter  
Output Voltage Accuracy  
Output Voltage Balance  
Line Regulation  
Conditions  
Min.  
Typ.  
1.0  
0.1  
1.2  
Max.  
3.0  
1.0  
1.5  
10  
Units  
%
%
%
%
Dual Outputs, Balanced Loads  
For VIN Change of 1%  
IOUT = 10% to 100%  
See Note 2  
2 µA Max Leakage Current  
Compact SMT Case  
Single & Dual Outputs  
Load Regulation  
Ripple & Noise (20 MHz)  
Temperature Coefficient  
Output Short Circuit  
General  
100  
0.02  
mV P - P  
%/°C  
0.01  
Continuos (Autorecovery)  
Parameter  
Conditions  
Min.  
60 Seconds 4,000  
250 Vrms  
Typ.  
Max.  
2.0  
Units  
Vrms  
Isolation Voltage  
Reinforced Insulation Working Voltage  
Leakage Current  
Isolation Resistance  
Isolation Capacitance  
Switching Frequency  
EMI Characteristics, See Note 3  
Parameter  
240 VAC, 60 Hz  
500 VDC  
µA  
G  
pF  
10  
100 kHz, 1V  
20  
80  
50  
100  
kHz  
Standard  
Criteria  
Level  
EMC  
EMS  
Complies With EN 55011 4TH Edition  
Complies With EN 60601-1-2  
Environmental  
Parameter  
Conditions  
Min.  
-25  
Typ.  
Max.  
+95  
+105  
+125  
Units  
°C  
°C  
Ambient  
Case  
Operating Temperature Range  
Storage Temperature Range  
Cooling  
Humidity  
-50  
°C  
Free Air Convection  
RH, Non-condensing  
95  
%
Physical  
Case Size  
See Mechanical Drawing (Page 2)  
Case Material  
Weight  
Non-Conductive Black Plastic (UL94-V0)  
0.14 Oz (4.1g)  
Reliability Specifications  
Parameter  
MTBF  
MicroPower Direct  
292 Page Street  
Suite D  
Stoughton, MA 02072  
USA  
Conditions  
Min.  
2.0  
Typ.  
Max.  
Units  
MHours  
MIL HDBK 217F, 25°C, Gnd Benign  
IEC/EN 60601-1, EN 60601-1 3RD Edition, 2xMOPP  
Safety Standards  
ANSI/AAMI ES 60601-1, 2xMOPP Recognition, (UL Certificate)  
ANSI/AAMI ES 60601-1, CAN/CSA-C22.2 No.60601-1  
IPC/JEDEC J-STD-020D.1 Level 2  
T: (781) 344-8226  
F: (781) 344-8481  
E: sales@micropowerdirect.com  
W: www.micropowerdirect.com  
Moisture Sensitivity Level, See Note 4  
Absolute Maximum Ratings  
Parameter  
Conditions  
Min.  
Typ.  
Max.  
9.0  
18.0  
30.0  
245  
Units  
VDC  
°C  
5 VDC Input  
Input Voltage Surge (1 Sec)  
12 VDC Input  
24 VDC Input  
See Note 5  
Peak Reflow Temperature  
Caution: Exceeding Absolute Maximum Ratings may damage the module. These are not continuous operating ratings.  
www.micropowerdirect.com  
Model Selection Guide  
www.micropowerdirect.com  
Input  
Output  
Capacitive  
Fuse Rating  
Slow-Blow  
(mA)  
Model  
Number  
Efficiency  
(%, Typ)  
Voltage (VDC)  
Current (mA)  
Full-Load No-Load  
Load  
(µF, Max)  
Voltage Current  
Current  
(VDC)  
(mA, Max) (mA, Min)  
Nominal  
Range  
ML105SM-05RI4  
ML105SM-12RI4  
ML105SM-15RI4  
ML105DM-12RI4  
ML105DM-15RI4  
ML112SM-05RI4  
ML112SM-12RI4  
ML112SM-15RI4  
ML112DM-12RI4  
ML112DM-15RI4  
ML124SM-05RI4  
ML124SM-12RI4  
ML124SM-15RI4  
ML124DM-12RI4  
ML124DM-15RI4  
Notes:  
5
5
5
5
4.50 - 5.50  
4.50 - 5.50  
4.50 - 5.50  
4.50 - 5.50  
4.50 - 5.50  
10.8 - 13.2  
10.8 - 13.2  
10.8 - 13.2  
10.8 - 13.2  
10.8 - 13.2  
21.6 - 26.4  
21.6 - 26.4  
21.6 - 26.4  
21.6 - 26.4  
21.6 - 26.4  
263  
252  
246  
252  
236  
110  
106  
106  
106  
103  
55  
50  
50  
50  
50  
50  
35  
35  
35  
35  
35  
20  
20  
20  
20  
20  
5.0  
12.0  
15.0  
12.0  
15.0  
5.0  
12.0  
15.0  
12.0  
15.0  
5.0  
200.0  
84.0  
68.0  
42.0  
33.0  
200.0  
84.0  
68.0  
42.0  
33.0  
200.0  
84.0  
68.0  
42.0  
33.0  
4.0  
1.68  
1.36  
0.84  
0.66  
4.0  
1.68  
1.36  
0.84  
0.66  
4.0  
220  
220  
220  
100  
100  
220  
220  
220  
100  
100  
220  
220  
220  
100  
100  
76  
80  
83  
80  
84  
76  
79  
80  
79  
80  
76  
80  
80  
80  
80  
600  
600  
600  
600  
600  
250  
250  
250  
250  
250  
125  
125  
125  
125  
125  
5
12  
12  
12  
12  
12  
24  
24  
24  
24  
24  
53  
53  
53  
52  
12.0  
15.0  
12.0  
15.0  
1.68  
1.36  
0.84  
0.66  
Derating Curve  
0. The specified maximum capacitive load is for each output.  
2. When measuring output ripple, it is recommended that an external ±.47 µF ceramic  
capacitor be placed from the +VOUT pin to the -VOUT pin for single output units and  
from each output to common for dual output units. For noise sensitive applications,  
the use of 3.3 µF capacitors will reduce the output ripple.  
3
Contact the factory for required filter components.  
4. Any units that are not packaged in a vacuum sealed container should be stored in a  
controlled environment. Contact the factory for more information.  
5. The recommended reflow settings are a peak temperature of 245 °C for a maxi-  
mum period (TPK) of 0±S and a time above liquidous (TL) of <6± seconds at 207 °C.  
For more information, please contact the factory.  
6. Operation at no-load will not damage these units. However, they may not meet  
all specifications.  
7. Dual output units may be connected to provide a 24 VDC or 3± VDC output. To do  
this, connect the load across the positive (+VOUT) and negative (-VOUT) outputs and  
float the output common.  
8. The converter should be connected to a low ac-impedance source. An input source  
with a highly inductive impedance may affect the stability of the converter. In appli-  
cations where the converter output loading is high and input power is supplied over  
long lines, it may be necessary to use a capacitor on the input to insure start-up.  
In this case, it is recommended that a low ESR (ESR <0.±at 0±± kHz) capacitor  
be mounted close to the converter. For 5V input units a 2.2 µF is recommended,  
for 02V input units, a 0.± µF; and for 24V units a ±.47 µF.  
9. It is recommended that a fuse be used on the input of a power supply for protection.  
See the table above for the correct rating.  
Mechanical Dimensions  
Pin Connections  
Pin  
Single  
1
-VIN  
NC  
6
Solder Layout  
7
NC  
8
+VOUT  
-VOUT  
+VIN  
9
14  
Pin  
1
Dual  
-VIN  
6
Common  
-VOUT  
7
8
+VOUT  
Common  
+VIN  
9
14  
NC = No Connection  
Mechanical Notes:  
All dimensions are typical in inches (mm)  
Tolerance x.xx = ±±.±0 (±±.25)  
MicroPower Direct  
We Power Your Success - For Less!  
292 Page Street Ste D Stoughton, MA 02072  
TEL: (781) 344-8226  
FAX: (781) 344-8481  
E-Mail: sales@micropowerdirect.com  

相关型号:

ML105SM-12RI4

Medical Approved Compact SMT, 1W DC/DC Con vert ers
MPD

ML105SM-15RI4

Medical Approved Compact SMT, 1W DC/DC Con vert ers
MPD

ML11

Medical Instrument Lens-End Lamps
GILWAY

ML11-K1

Medical Instrument Lens-End Lamps
GILWAY

ML111PCSX

Audio/RCA Connector, 2 Pole(s), Solder Terminal, Jack, ROHS COMPLIANT
SWITCH

ML111PCX

Audio/RCA Connector, 2 Pole(s), Solder Terminal, Jack, ROHS COMPLIANT
SWITCH

ML112AX

Audio/RCA Connector, 2 Pole(s), Solder Terminal, Jack, ROHS COMPLIANT
SWITCH

ML112B

Audio/RCA Connector, Jack,
SWITCH

ML112BPCSX

Audio/RCA Connector, 3 Pole(s), Solder Terminal, Jack, ROHS COMPLIANT
SWITCH

ML112BPCX

Audio/RCA Connector, 3 Pole(s), Solder Terminal, Jack, ROHS COMPLIANT
SWITCH

ML112D-12RW

Ultra-Compact, 1W Wide 2:1 Input, SMT DC/DC Con vert ers
MPD

ML112D-15RW

Ultra-Compact, 1W Wide 2:1 Input, SMT DC/DC Con vert ers
MPD