RPS-60 [MEANWELL]
60W Single Output Medical Type; 60W单输出医用型型号: | RPS-60 |
厂家: | MEAN WELL ENTERPRISES CO., LTD. |
描述: | 60W Single Output Medical Type |
文件: | 总2页 (文件大小:287K) |
中文: | 中文翻译 | 下载: | 下载PDF数据表文档文件 |
60W Single Output Medical Type
RPS-60 series
Features :
4"x2" miniature size
Universal AC input/Full range
Low leakage current<200uA
Protections: Short circuit / Overload / Over voltage
Cooling by free air convection
UL60601-1/IEC60601-1/EN60601-1 medical safety approved
UL60950-1/IEC60950-1/EN60950-1 ITE safety approved
No load power consumption<0.75W
Fixed switch frequency at 100KHz
3 years warranty
EN60601-1
IEC60601-1
SPECIFICATION
MODEL
RPS-60-3.3
RPS-60-5
RPS-60-12
RPS-60-15
RPS-60-24
RPS-60-48
DC VOLTAGE
3.3V
5V
12V
15V
24V
48V
RATED CURRENT
CURRENT RANGE
RATED POWER
10A
10A
5A
4A
2.5A
1.25A
0 ~ 11A
33W
0 ~ 11A
50W
0 ~ 5.5A
60W
0 ~ 4.4A
60W
0 ~ 2.75A
60W
0 ~ 1.375A
60W
PEAK LOAD(10sec.) Note.4 36.3W
RIPPLE & NOISE (max.) Note.2 80mVp-p
55W
66W
66W
66W
66W
80mVp-p
4.75 ~ 5.5V
2.0%
120mVp-p
11.4 ~ 13.2V
2.0%
150mVp-p
13.5 ~ 16.5V
2.0%
240mVp-p
22.8 ~ 27.6V
1.0%
300mVp-p
45.6 ~ 52.8V
1.0%
OUTPUT
VOLTAGE ADJ. RANGE
VOLTAGE TOLERANCE Note.3 2.0%
3.1 ~ 3.6V
LINE REGULATION
LOAD REGULATION
SETUP, RISE TIME
0.5%
0.5%
0.5%
0.5%
0.5%
0.5%
1.0%
1.0%
1.0%
1.0%
1.0%
1.0%
500ms, 30ms/230VAC
500ms, 30ms/115VAC at full load
HOLD UP TIME (Typ.)
VOLTAGE RANGE
50ms/230VAC
90 ~ 264VAC
47 ~ 63Hz
74%
13ms/115VAC at full load
127 ~ 370VDC
FREQUENCY RANGE
EFFICIENCY (Typ.)
AC CURRENT (Typ.)
INRUSH CURRENT (Typ.)
LEAKAGE CURRENT
79%
83%
84%
85%
86%
INPUT
1.8A/115VAC
1 A/230VAC
COLD START 60A/230VAC
30A/115VAC
For earth <200uA / 264VAC, For patient <100uA/264VAC
115 ~ 150% rated output power
OVER LOAD
Protection type : Hiccup mode, recovers automatically after fault condition is removed
3.8 ~ 5V 5.75 ~ 6.75V 13.8 ~ 16.2V 17.25 ~ 20.25V
PROTECTION
ENVIRONMENT
28.4 ~ 32.4V
55.2 ~ 64.8V
OVER VOLTAGE
Protection type : Shut down o/p voltage, re-power on to recover
-20 ~ +70 (Refer to output load derating curve)
20 ~ 90% RH non-condensing
WORKING TEMP.
WORKING HUMIDITY
STORAGE TEMP., HUMIDITY
TEMP. COEFFICIENT
VIBRATION
-40 ~ +85 , 10 ~ 95% RH
0.03%/ (0 ~ 45
)
10 ~ 500Hz, 2G 10min./1cycle, period for 60min. each along X, Y, Z axes
UL60950-1, TUV EN60950-1, UL60601-1, TUV EN60601-1, IEC60601-1 approved
I/P-O/P:4KVAC I/P-FG:1.5KVAC O/P-FG:1.5KVAC
I/P-O/P, I/P-FG, O/P-FG:100M Ohms/500VDC
SAFETY STANDARDS
WITHSTAND VOLTAGE
ISOLATION RESISTANCE
EMI CONDUCTION & RADIATION
HARMONIC CURRENT
EMS IMMUNITY
SAFETY &
EMC
(Note 5)
Compliance to EN55011 (CISPR11), EN55022 (CISPR22) Class B
Compliance to EN61000-3-2,-3
Compliance to EN61000-4-2,3,4,5,6,8,11; ENV50204, EN55024, EN60601-1-2, EN61204-3, medical level, criteria A
MTBF
353.6Khrs min. MIL-HDBK-217F (25
101.6*50.8*29mm (L*W*H)
)
OTHERS
NOTE
DIMENSION
0.15Kg; 96pcs/15.4Kg/0.89CUFT
PACKING
1. All parameters NOT specially mentioned are measured at 230VAC input, rated load and 25 of ambient temperature.
2. Ripple & noise are measured at 20MHz of bandwidth by using a 12" twisted pair-wire terminated with a 0.1uf & 47uf parallel capacitor.
3. Tolerance : includes set up tolerance, line regulation and load regulation.
4. 33% Duty cycle maximum within every 30 seconds. Average output power should not exceed the rated power.
5. The power supply is considered a component which will be installed into a final equipment. The final equipment must be re-confirmed that it still meets
EMC directives.
6. Length of set up time is measured at cold first start. Turning ON/OFF the power supply may lead to increase of the set up time.
File Name:RPS-60-SPEC 2008-03-25
60W Single Output Medical Type
RPS-60 series
Mechanical Specification
Unit:mm
101.6
95.25
3.175
SVR1
3
2
CN1
1
1
2
3
4
CN2
4- 3.3
AC Input Connector (CN1) : JST B3P-VH or equivalent
DC Output Connector (CN2) : JST B4P-VH or equivalent
Pin No. Assignment
Mating Housing
Terminal
Pin No. Assignment
Mating Housing
Terminal
1
2
3
AC/N
No Pin
AC/L
1,2
3,4
+V
-V
JST VHR
or equivalent
JST SVH-21T-P1.1
or equivalent
JST VHR
or equivalent
JST SVH-21T-P1.1
or equivalent
Block Diagram
fosc : 100KHz
RECTIFIERS
RECTIFIERS
POWER
+V
-V
EMI
I/P
&
&
SWITCHING
FILTER
FILTER
FILTER
DETECTION
CIRCUIT
FG
CONTROL
O.L.P.
O.V.P.
Output Derating
Output Derating VS Input Voltage
Ta=25
100
90
100
80
80
60
70
60
40
50
40
20
(HORIZONTAL)
-20
0
10
20
30
40
50
60 65 70
90 95 100 115 120 140 160 180 200 220 240 264
AMBIENT TEMPERATURE (
)
INPUT VOLTAGE (VAC) 60Hz
File Name:RPS-60-SPEC 2008-03-25