26PCDFN6G [HONEYWELL]
Gage Unamplified Compensated Flow-Through; 计无放大补偿流量,通过型号: | 26PCDFN6G |
厂家: | Honeywell |
描述: | Gage Unamplified Compensated Flow-Through |
文件: | 总3页 (文件大小:238K) |
中文: | 中文翻译 | 下载: | 下载PDF数据表文档文件 |
Interactive Catalog Replaces Catalog Pages
Honeywell Sensing and Control has replaced
the PDF product catalog with the new
Interactive Catalog. The Interactive Catalog is
a power search tool that makes it easier to find
product information. It includes more
installation, application, and technical
Click this icon to try the new
information than ever before.
Interactive Catalog.
Sensing and Control
Honeywell Inc.
11 West Spring Street
Freeport, Illinois 61032
Pressure Sensors
26PC Series
Gage Unamplified Compensated Flow-Through
FEATURES
ɀ Measure positive and negative gage ɀ Medical grade ISO10993-1(USP Class
pressures
6) port material
ɀ Flow-through port design fits in-line ɀ Silicon sensor chip
with application
ɀ Popular port sizes:
- 8 mm (.315 in.) OD (1/4 in. ID tubing or ɀ Minimal deadspace — efficient cleans-
standard connectors)
- 0.144 in. OD (1/8 in. ID tubing)
ɀ 24 inch wire harness with splash proof
connector
ing and disinfecting
26PC SERIES PERFORMANCE CHARACTERISTICS at 10.0 ±0.01 VDC
Excitation, 25°C
Min.
---
Typ.
10
Max.
16
Units
VDC
Excitation
Repeatability & Hysteresis
Response Time
---
±0.20
---
---
%Span
msec
ohms
ohms
%Span
grams
---
1.0
Input Resistance
5.5 K
1.5 K
---
7.5 K
2.5 K
±0.5
2
11.5 K
3.0 K
---
Output Resistance
Stability over One Year
Weight
---
---
Total error calculation, see page 105.
ENVIRONMENTAL SPECIFICATIONS
Operating Temperature
Storage Temperature
Compensated Temperature
Shock
−40° to 85°C (−40° to +185°F)
−55° to +100°C (−67° to +212°F)
0° to +50°C (32° to +122°F)
Qualification tested to 150 g
Vibration
MIL-STD-202. Method 213 (150g halfsine,
11 msec)
Media (P1 & P2)
Limited only to those media which will not attack
polyetherimide, silicon, fluorosilicone, silicone,
EPDM, and neoprene seals.
26PC SERIES ORDER GUIDE
Pressure
Range
(psi)
Over-
Sensitivity pressure
Catalog
Listing
Linearity
(% span)
Null Shift
(mV)
Null Offset
(mV)
Span Shift
(% span)
Span
(mV)
mV/psi
Typ.
16.7
10.0
6.67
3.33
1.0
psi
Max.
20
Typ.
Max
0.5
0.5
0.5
Typ.
Max
Min.
Typ. Max. Typ. Max. Min.
Typ. Max.
26PCA TYPE
26PCB TYPE
26PCC TYPE
26PCD TYPE
26PCF TYPE
26PCJ TYPE
26PCK TYPE
1
0.25
0.4
±0.5 ±1.0 −1.5
±0.5 ±1.0 −1.5
±0.5 ±1.0 −1.5
0
0
0
0
0
0
0
+1.5 ±1.0 ±2.0 14.7
16.7
50
18.7
53
5
+1.5 ±1.0 ±1.5
+1.5 ±0.75 ±1.5
+1.5 ±0.75 ±1.5
+2.0 ±0.5 ±1.5
47
97
97
95
20
15
0.25
0.1
100
100
100
39.5
39.5
103
103
105
41.5
41.5
45
30
0.2 ±0.75 ±1.5 −1.5
60
100
38*
38*
0.1
0.2
0.5
0.5
±1.0 ±2.0 −2.0
±0.7 ±1.5 −1.5
±0.7 ±1.5 −1.5
200
60
0.1
+1.5 ±1.0 ±1.5 37.5
+1.5 ±1.0 ±1.5 37.5
2.63
2.63
0.1
60
*Accuracy specifications calculated at 15 psi.
Honeywell ɀ Sensing and Control ɀ1-800-537-6945 USA ɀ ⍣1-815-235-6847 International ɀ1-800-737-3360 Canada
23
Pressure Sensors
26PC Series
Gage Unamplified Compensated Flow-Through
SENSOR SELECTION GUIDE
2
6
PC
Pressure
Transducer
A
F*
Type of
Seal
N
Port
Type
5
G
Product
Family
Circuit
Type
Pressure
Range
Termination
Style
Pressure
Measurement
2 20PC
Family
6 Compen-
sated,
Calibrated
A 1 psi
E EPDM
F Fluoro-
silicone
N Neoprene
S Silicone
G Small
2 4-pin DIP
5 Wire
G Gage
B 5 psi
N Large
C 15 psi
D 30 psi
F 100 psi
J 38 psi
(.350 dia.)
P Large
(.315 dia.)
harness
6 4-pin SIP
K 38 psi
(passivated)
Example: 26PCBFG5G
Compensated, calibrated 5 psi sensor, fluorosilicone seal, small flow-through ports, wire harness, and gage pressure measurements.
*Other media seal materials may be available.
Note: Not all combinations are established. Contact 800 number before final design.
See Accessory Guide, page 27.
24
Honeywell ɀ Sensing and Control ɀ1-800-537-6945 USA ɀ ⍣1-815-235-6847 International ɀ1-800-737-3360 Canada