HICM65 [HITRON]
4x 2 inch Compact size;![HICM65](http://pdffile.icpdf.com/pdf2/p00330/img/icpdf/HICM65_2031173_icpdf.jpg)
型号: | HICM65 |
厂家: | ![]() |
描述: | 4x 2 inch Compact size |
文件: | 总2页 (文件大小:404K) |
中文: | 中文翻译 | 下载: | 下载PDF数据表文档文件 |
![](http://public.icpdf.com/style/img/ads.jpg)
H HiTRON
Universal input AC-DC Medical application open frame internal switching
Convection-cooled power supplies 65 Watts triple output HICM65 series
Features
ꢀ
ꢀ
ꢀ
ꢀ
ꢀ
ꢀ
ꢀ
4x 2 inch Compact size
Convection cooling
Up to 83% high efficiency
-20°C to +70° C wide operating temperature
PCB & Box format optional
Medical application
CE marking compliance
Specification
Input
General
Input Voltage
Input Frequency
Input Current
90-264VAC
Efficiency
Typical 83% (depending on model)
65KHz
47-63Hz
Switching Frequency
Dielectric Withstand
Circuit Topology
Transient Response
Typical 1.1A at 115VAC
Typical 0.55A at 230VAC
Typical 16A rms at 230VAC
V-H connector or equivalent
IEC60601-1
Fix-flyback Circuit
Output voltage returns in less than
0.01mS following a 25% load
change
Inrush Current
Input Connector
Earth Leakage Current Less than 0.24mA
Touch Current
No-load Power
Less than 0.1mA
Less than 1.5W
Power Density
Construction
6.5W / Cubic Inch
PCB & Box optional
Output
Environmental
Operating Temperature
(Refer to derating chart)
Storage Temperature
Cooling
Output Connector
Line Regulation
Load Regulation
Total Regulation
Noise & Ripple
Adjustability
V-H connector or equivalent
Typical 0.5-2%
-20°C to +70°C derate from 100%
load at 50°C to 60% at 70°C
-25°C to +85°C
V1/2 Typical ±2%, V3 ±3%
V1 Typical±4%, V2/3 ±8%
Typical 1% peak to peak
Not available
Convection-Cooled
5000m (?)
Operating Altitude
Operating Humidity
Storage Humidity
10-95% RH, non-condensing
5-95% RH
Hold-up Time
Typical 16mS at 115VAC
Typical 76mS at 230VAC
Safety/EMC
Emissions (conducted)
Harmonic Current
Safety Standard
CISPR EN55011 & FCC Class B
IEC61000-3-2
Protection
Over Voltage
Built-in V1 (Auto-recovery)
Over Load
Typical 130-150% of rating load
IEC60601-1 Class I
Notes:
(1) All measurements are at nominal input, full load, and +25°C unless otherwise specified.
(2) Load regulation is measured at 115VAC or 230VAC in percentage to indicate the change in output voltage as the load varied from half load to full load (±%).
(3) The power supply is considered a component installed into a final equipment. The final equipment must be re-confirmed that it still meets EMC directives.
(4) Due to requests in market and advances in technology, specifications subject to change without notice.
Edited on Feb. 15, 2019
-1-
Output voltage & current rating chart
Single Output
V1 ★
V2
V3
Model No.
(refer to note 1)
Min
Typ. Volt. Max. Min. Typ. Volt. Max. Min
Typ. Volt. Max.
HICM65-T050II-C1P
HICM65-T050KK-C1P
HICM65-T050MI-C1P
1.2A 6.0A +5V 8.0A 0.50A 2.5A +12V 3.0A 0.1A 0.5A -12V 0.5A
1.2A 6.0A +5V 8.0A 0.40A 2.0A +15V 2.5A 0.1A 0.5A -15V 0.5A
1.2A 6.0A +5V 8.0A 0.25A 1.0A +24V 1.5A 0.1A 0.5A -12V 0.5A
Symbol: ★ “OVP” built-in
Notes: (1) Please add suffix to model number to define IEC protection classes: add “-C1” for Class I version (with AC-GND).
Please add suffix to model number to define type: add “-B” for enclosure (metal box) version, and “-P” for PCB version.
For example: HICM65-T050II-C1P is for Class I and PCB version; HICM65-T050II-C1B is for Class I and enclosure (metal box) version.
(2) Derate output power by 20% for enclosure (Metal Box) version.
(3) Please consult the factory if the required output voltage is not listed.
Weight:
Mechanical Dimensions (Note: All dimensions are in mm[inch])
Derating Chart
Note: Derate output power by 20% for enclosure (Metal Box) version.
Pin assignment
Assignment
Pin No.
CNT1-L
CNT1-N
GND
AC-Line
AC-Neutral
AC-Ground
V1
CNT2-5,6
CNT2-2
CNT2-1
CNT2-3,4
V2
V3
DC COM
Edited on Feb. 15, 2019
-2-
相关型号:
©2020 ICPDF网 联系我们和版权申明