User's Guide  
EasyView™  
Digital Hygro-Thermometer  
and Datalogger  
Model EA20  
Introduction  
Congratulations on your purchase of the Extech EasyView™ EA20 Digital Hygro-  
Thermometer. This device measures Humidity, Temperature, Dew Point temperature  
and Wet Bulb temperature. Advanced features include Manual Datalogging of up to 99  
sets, MAX/MIN recording, Relative measurement mode, and elapsed timer. Careful use  
of this meter will provide years of reliable service.  
Specifications  
Function  
Humidity  
Range and Resolution Accuracy  
10.0 to 95.0% RH  
±3% RH (at 25oC, 30 to 95% RH )  
±5% RH (at 25oC, 10 to 30% RH )  
Temperature -4 to 140oF (-20 to 60oC) ±1.8oF (±1oC)  
Display  
Multifunction LCD with bargraph and status indicators  
Sensor Type  
Humidity: Precision capacitance sensor  
Temperature: Thermistor  
Response Time  
Humidity: 45% to 95% < 1 min; 95% to 45% < 3 min.  
Temperature: Approx. 2 seconds for a 10oC change  
2 times per second  
Sampling Rate  
Datalogging Memory  
Operating Conditions  
Storage Conditions  
Power Supply  
99 sets  
32 to 140oF (0 to 60oC); < 95% RH non-condensing  
14 to 140oF (-10 to 60oC); <70% RH non-condensing  
6 x 1.5V ‘AAA’ batteries with programmable auto power off (1  
minute to 59 minutes in 7 selectable times)  
Battery Life  
Approx. 200 hours  
Dimensions / Weight  
5.9 x 2.8 x 1.4” (150 x 72 x 35mm); 8.3 oz. (235g)  
Meter Description  
1. Humidity/Temperature Probe  
2. Multifunction LCD with bargraph  
and status indicators  
1
3. Power ON/OFF button  
4. MX/MN button  
5. DATA HOLD button  
6. Humidity / Temperature select  
button  
2
7. TIME button (for MX/MN elapsed  
timer)  
4
3
5
7
8
8. Temperature units (C/F) select  
button  
6
9
9. DEW POINT temperature select  
button  
10. RELATIVE mode button  
11. WET BULB temperature select  
button  
12. MEM key manual datalogging  
control key  
13. READ key manual memory data reading  
control key  
14. ꢀꢁkeys to select memory  
location  
Note: The battery compartment is located  
on the rear of the instrument  
2
EA20 V1.6 02/06  
Operation  
NOTE: Do not immerse the probe in liquid; it is intended for use in air only.  
Turn Power ON or OFF  
Press the  
turned on.  
button to turn power on or off. The meter will perform a short self-test when  
Taking measurements  
1. Hold the probe in the area to be tested.  
2. Allow adequate time for readings to stabilize.  
3. Read the humidity and temperature on the LCD. The %RH/TEMP button allows the  
user to choose whether the temperature or the humidity is displayed on the larger  
digits. Press the %RH/TEMP button to toggle the display configuration.  
Selecting temperature units of measure (C/F)  
Press the oC/oF button momentarily to toggle the temperature units.  
Maximum (MX) and Minimum (MN) recording  
NOTE: The Automatic Power OFF feature will be disabled while the MX/MN mode is  
selected.  
The MX/MN mode allows the user to view only the highest (MX) or lowest (MN) readings.  
Press the MX/MN button momentarily to start the elapsed timer and begin recording. The  
display will now only show the highest reading (the MX icon will appear). Press the MX/MN  
button again to view the lowest reading (MN will display). The TIME button allows the user  
to view the elapsed time for a particular Max or Min reading in Minutes:Seconds  
(59mins:59secs max) or Days:Hours (99days:24hours max). To check the highest or  
lowest Dew Point or Wet Bulb temperatures, use the DEW and the WET buttons. The  
following data displays are available in the MX/MN mode:  
a. Maximum and Minimum humidity with its associated temperature reading  
b. Maximum and Minimum temperature with its associated humidity reading  
c. Maximum and Minimum Dew point temperature with its associated standard  
temperature or humidity  
d. Maximum and Minimum Wet Bulb temperature with associated standard  
temperature or humidity  
To exit the MX/MN mode, press and hold the MX/MN button until the MX and MN icons  
disappear.  
Data Hold  
Press the H button momentarily to freeze the displayed reading. The 'H' icon will appear on  
the upper left-hand side of the display. Press the H button again to return to normal  
operation (the 'H' hold icon will disappear).  
DEW POINT temperature display  
Dew Point temperature is calculated by the meter based on humidity/temperature  
measurements. To view the Dew Point temperature, press the DEW button momentarily.  
The ‘DEW’ display icon will appear on the right side of the LCD. The Dew Point  
temperature will display on the bottom (larger) digits. The upper (smaller) digits will display  
the regular temperature in °C or °F or the humidity in RH% (use the %RH/TEMP button to  
toggle between the two).  
WET BULB temperature display  
Wet Bulb temperature is calculated by the meter based on humidity/temperature  
measurements. To view the Wet Bulb temperature, press the WET button momentarily.  
The ‘WET’ display icon will appear on the right side of the LCD. The Wet Bulb temperature  
is displayed on the bottom (larger) digits. The upper (smaller) digits will display the regular  
temperature in °C or °F or the humidity in RH% (use the %RH/TEMP button to toggle  
between the two).  
3
EA20 V1.6 02/06  
Manual Datalogging  
Clearing The Memory  
1. Press the  
button to power off the meter.  
2. Press and hold down the MEM key, then press the  
key to power on the meter. The  
display will show “CLr” indicating that the memory has been erased.  
Entering A Reading  
1. Press the MEM key to enter a reading. The display will momentarily show an “M” and the  
memory location number from 1 to 99.  
2. When the memory is full, the display will show “FULL”.  
Data Recall  
1. Press the “READ” key to enter the READ mode. The LCD will show an “R” and the  
memory address number.  
2. Press the ꢀꢁkeys to scroll through the memory locations.  
3. Press “DEW” or “WET” to view the stored Dew Point or Wet Bulb temperature.  
4. Press the “READ” key to exit this mode.  
Relative mode  
Measurements can be displayed as a difference between actual humidity/temperature and  
stored reference values. To store a humidity/temperature reading group as a reference,  
press the REL button when the desired reference readings are on the LCD (the triangle  
icon will appear steady in the upper left-hand corner of the LCD). All subsequent displayed  
readings will be “relative” to the stored reference value. For example, if the reference  
values are 50.0% & 100°F and the actual readings are 53.0% & 105°F, the meter will  
display 3.0% & 5°F. To exit the Relative mode, press and hold the REL button until the  
triangle icon disappears.  
Automatic Power OFF  
The meter automatically shuts off after a programmed period of time. To program the time  
period press and hold the TIME button while turning the meter on. Release the TIME  
button after the meter finishes its self-test. Press the TIME button to scroll through the eight  
time selections (shown on the smaller digits at the center of the LCD). Select from the  
following list: 0 (disabled), 1, 3, 5, 10, 15, 30, or 59 minutes. Press the H button to store the  
desired time.  
Maintenance  
Cleaning and storage  
1. The meter should be cleaned with a damp cloth and mild detergent when necessary.  
Do not use solvents or abrasives.  
2. Store the meter in an area with moderate temperature and humidity (refer to the  
operating and storage range in the specifications chart earlier in this manual).  
Battery Replacement  
When the battery power falls low, the ‘BT’ symbol will appear on the LCD. Replace the six  
(6) 1.5 ‘AAA’ batteries by removing the rear (center) battery compartment screw and  
accessing the battery compartment. Observe polarity when placing the batteries in the  
compartment. Ensure that the compartment cover is securely fastened when finished.  
4
EA20 V1.6 02/06  
Calibration and Repair Services  
Extech offers complete repair and calibration services for all of the products we sell. For  
periodic calibration, NIST certification or repair of this Extech product, call customer service  
for details on services available. Extech recommends that calibration be performed on an  
annual basis to ensure calibration integrity.  
Warranty  
EXTECH INSTRUMENTS CORPORATION warrants the basic instrument to be free of  
defects in parts and workmanship for one year from date of shipment (a six month limited  
warranty applies on sensors and cables). If it should become necessary to return the  
instrument for service during or beyond the warranty period, contact the Customer Service  
Department at (781) 890-7440 EXTENSION 210 for authorization or visit www.extech.com  
for more information. A Return Authorization (RA) number must be issued before any  
product is returned to Extech. The sender is responsible for shipping charges, freight,  
insurance and proper packaging to prevent damage in transit. This warranty does not  
apply to defects resulting from action of the user such as misuse, improper wiring,  
operation outside of specification, improper maintenance or repair, or unauthorized  
modification. Extech specifically disclaims any implied warranties or merchantability or  
fitness for a specific purpose and will not be liable for any direct, indirect, incidental or  
consequential damages. Extech's total liability is limited to repair or replacement of the  
product. The warranty set forth above is inclusive and no other warranty, whether written or  
oral, is expressed or implied.  
Calibration and Repair Services  
Extech offers complete repair and calibration services for all of the products we sell. or  
periodic calibration, NIST certification or repair of any Extech product, call customer  
service for details on services available. Extech recommends that calibration be performed  
on an annual basis to ensure calibration integrity.  
Support Hotline (781) 890-7440  
Tech support: Ext. 200; Email: support@extech.com  
Repair/Returns: Ext. 210; Email: repair@extech.com  
Website: www.extech.com  
Copyright © 2004 Extech Instruments Corporation.  
All rights reserved including the right of reproduction in whole or in part in any form.  
5
EA20 V1.6 02/06  

相关型号:

EA200

字符叠加IC
ETC

EA2007

Consumer IC
ETC

EA2015

13.5 Watts Dual Output Switching Adapter
BOTHHAND

EA2015A

13.5 Watts Dual Output Switching Adapter
BOTHHAND

EA2015B

13.5 Watts Dual Output Switching Adapter
BOTHHAND

EA2025AA08-16.000M

Quartz Crystal Resonator 2.0mm x 2.5mm x 0.60mm 4 Pad Ceramic Surface Mount (SMD) 16.000MHz ±50ppm/100ppm over 0&deg;C to +70&deg;C
ECLIPTEK

EA2025AA08-19.660MTR

QUARTZ CRYSTAL RESONATOR
ECLIPTEK

EA2025AA12-29.939M

Quartz Crystal Resonator 2.0mm x 2.5mm x 0.60mm 4 Pad Ceramic Surface Mount (SMD) 29.939MHz ±50ppm/100ppm over 0&deg;C to +70&deg;C
ECLIPTEK

EA2025AA14-30.720MTR

QUARTZ CRYSTAL RESONATOR
ECLIPTEK

EA2025AA14-34.400M

QUARTZ CRYSTAL RESONATOR
ECLIPTEK

EA2025AA16-40.000MTR

QUARTZ CRYSTAL RESONATOR
ECLIPTEK

EA2025AA18-29.4912MTR

QUARTZ CRYSTAL RESONATOR
ECLIPTEK