User's Guide  
Extech AM600  
600A AC Analog Clamp Meter  
Introduction  
Congratulations on your purchase of the Extech AM600 Analog Clamp Meter. This device  
measures AC Current, AC Voltage, DC Voltage, Resistance and Temperature. This meter  
is shipped fully tested and calibrated and, with proper use, will provide years of reliable  
service.  
Safety  
International Safety Symbols  
This symbol, adjacent to another symbol or terminal, indicates the user must refer  
to the manual for further information.  
This symbol, adjacent to a terminal, indicates that, under normal use, hazardous  
voltages may be present  
Double insulation (Protection Class II)  
SAFETY NOTES  
Do not exceed the maximum allowable input range of any function.  
Do not apply voltage to meter when resistance function is selected.  
Remove the battery if meter is to be stored for longer than 60 days.  
WARNINGS  
Set function switch to the appropriate position before measuring.  
When measuring volts do not switch to current/resistance modes.  
Do not measure current on a circuit whose voltage exceeds 600V.  
When changing ranges always disconnect the test leads from the circuit under test.  
CAUTIONS  
Improper use of this meter can cause damage, shock, injury or death. Read and  
understand this user manual before operating the meter.  
Always remove the test leads before replacing the battery or fuses.  
Inspect the condition of the test leads and the meter itself for any damage before  
operating the meter. Repair or replace any damage before use.  
Use great care when making measurements if the voltages are greater than 25 VAC  
rms or 35 VDC. These voltages are considered a shock hazard.  
Always discharge capacitors and remove power from the device under test before  
performing Resistance tests.  
Voltage checks on electrical outlets can be difficult and misleading because of the  
uncertainty of connection to the recessed electrical contacts. Other means should be  
used to ensure that the terminals are not "live".  
If the equipment is used in a manner not specified by the manufacturer, the protection  
provided by the equipment may be impaired.  
Function  
A AC  
Maximum Input  
600 A  
V AC  
750 V  
V DC  
75 V  
2
AM600 V2.0 6/07  
PER IEC1010 OVERVOLTAGE INSTALLATION CATEGORY  
OVERVOLTAGE CATEGORY I  
Equipment of OVERVOLTAGE CATEGORY I is equipment for connection to circuits in  
which measures are taken to limit the transient overvoltages to an appropriate low level.  
Note – Examples include protected electronic circuits.  
OVERVOLTAGE CATEGORY II  
Equipment of OVERVOLTAGE CATEGORY II is energy-consuming equipment to be  
supplied from the fixed installation.  
Note – Examples include household, office, and laboratory appliances.  
OVERVOLTAGE CATEGORY III  
Equipment of OVERVOLTAGE CATEGORY III is equipment in fixed installations.  
Note – Examples include switches in the fixed installation and some equipment for industrial  
use with permanent connection to the fixed installation.  
OVERVOLTAGE CATEGORY IV  
Equipment of OVERVOLTAGE CATEGORY IV is for use at the origin of the installation.  
Note – Examples include electricity meters and primary over-current protection equipment  
Description  
Meter Description  
1. Current clamp  
2. Clamp opening trigger  
3. 0 Ohms adjustment  
4. Meter zero adjustment  
5. Scale  
6. “COM” Terminal  
7. Range switch  
8. Pointer Lock switch  
9. V, and Temp Terminal  
3
AM600 V2.0 6/07  
Operation  
NOTE: Read and understand all Warning and Caution statements in this operation manual prior to using  
this meter.  
AC Current Measurements  
WARNING: Remove the test leads from the meter before making current clamp  
measurements.  
1.  
2.  
Set the Range switch to the 600A range..  
Set the Pointer Lock  
switch is set to the right-  
most position so that the  
pointer may freely move.  
3.  
4.  
If necessary, turn the  
Zero Adjust screw to  
zero the meter display.  
Press the trigger to open  
jaw. Fully enclose only  
one conductor. For  
optimum results, center  
the conductor in the jaw.  
5.  
6.  
The analog display  
pointer will indicate the current reading.  
When possible, set the Range switch to a lower range to obtain a reading in the  
upper half of the scale.  
AC/DC Voltage Measurements  
1.  
Insert the black test lead into the negative COM terminal and  
the red test lead into the positive V terminal.  
2.  
3.  
Set the Range switch to the highest AC or DC Voltage range.  
Set the Pointer Lock switch is set to the right-most position so  
that the pointer may freely move.  
4.  
If necessary, turn the Zero Adjust screw to zero the meter  
display.  
5.  
6.  
7.  
Connect the test leads in parallel to the circuit under test.  
The analog display pointer will indicate the voltage reading.  
When possible, set the Range switch to a lower range to  
obtain a reading in the upper half of the scale.  
4
AM600 V2.0 6/07  
Resistance Measurements  
1.  
2.  
3.  
4.  
5.  
6.  
7.  
Insert the black test lead into the negative COM terminal and  
the red test lead into the OHMS positive terminal.  
Set the Range switch to the X1or X100position as  
needed.  
Set the Pointer Lock switch is set to the right-most position so  
that the pointer may freely move.  
With the test leads open, adjust the meter zero so that the  
pointer is over the symbol.  
Touch the test lead tips together and adjust the 0 Ohms adjust  
until the pointer is over the 0 ohms position.  
Touch the test probe tips across the circuit or component under  
test.  
The analog display pointer will indicate the resistance reading.  
Temperature Measurements  
Zero procedure (fig. 1)  
1.  
Insert the temperature probe black plug into the COM terminal and the Probe tip  
into the V terminal.  
2.  
Adjust the pointer for a 0 scale reading  
Measurements (fig. 2)  
1.  
Insert the temperature probe black plug into the COM terminal and the red plug  
into the V terminal.  
2.  
3.  
Place the temperature probe into the area to be tested.  
Read the temperature on the °C or °F temperature scale  
fig. 1  
fig. 2  
5
AM600 V2.0 6/07  
Pointer Lock Function  
The Pointer Lock switch freezes the pointer’s position and is used as a ‘data hold’ feature.  
This feature is useful in dimly lit or hard to reach locations.  
Specifications  
Function  
Ranges  
Accuracy  
AC Current  
(50/60Hz)  
6, 15, 60, 150 A  
600 A  
± 3% of Full Scale  
± 5% of Full Scale  
AC Voltage  
(50/60Hz)  
150, 300 and 750V  
± 3% of Full Scale  
± 3% of Full Scale  
DC Voltage  
Resistance  
75 V DC  
2000 (50 mid-scale) (Rx1)  
200 k(5 kmid-scale) (Rx100)  
-4 to 302°F (-20°C to 150°C)  
± 3% of Full Scale Length  
± 3% of arc  
Temperature  
Clamp jaw opening  
Display  
1.3" (34 mm) max.  
Analog with Zero Adjustment  
ACV: 2 k/V, DCV: 4.5 kV  
50 / 60 Hz  
Input Impedance  
AC bandwidth  
Temperature Probe  
Operating conditions  
Operating Altitude  
Battery  
Thermistor  
Temperature 32 - 104°F (0 - 40°C); Relative Humidity <80%  
7000 ft. (2000 m) maximum.  
One (1) 1.5V ‘AA’ Battery  
Fuse  
500 mA, 250V fast  
Dimensions & Weight  
Safety  
8.4 x 3.4 x 1.7" (214 x 85 x 43 mm); 12.4 oz. (350 g)  
For indoor use and in accordance with the requirements for  
double insulation to IEC1010-1 (2001): EN61010-1 (2001)  
Overvoltage Category II 600V, Pollution Degree 2.  
Standards  
C
C
6
AM600 V2.0 6/07  
Maintenance  
WARNING: To avoid electrical shock, disconnect the meter from any circuit and remove the test leads from  
the input terminals before opening the case. Do not operate the meter with an open case.  
Cleaning and Storage  
Periodically wipe the case with a damp cloth and mild detergent; do not use abrasives or  
solvents. If the meter is not to be used for 60 days or more, remove the battery and store it  
separately.  
Battery Replacement  
1. Remove the Phillips head screws from the rear battery compartment cover.  
2. Replace the 1.5V ‘AA’ battery and replace the cover.  
Fuse Replacement  
1. Remove the two (2) Phillips head screws from the rear cover.  
2. Carefully lift off the rear cover.  
3. Replace the 0.5A/250V fuse. Make sure the fuse is centered in the fuse holder.  
4. Replace the rear cover.  
Calibration and Repair Services  
Extech offers repair and calibration services for the products we sell. Extech also  
provides NIST certification for most products. Call the Customer Care Department for  
information on calibration services available for this product. Extech recommends that  
annual calibrations be performed to verify meter performance and accuracy.  
7
AM600 V2.0 6/07  
Warranty  
EXTECH INSTRUMENTS CORPORATION warrants this instrument to be free of defects  
in parts and workmanship for one year from date of shipment (a six month limited warranty  
applies to sensors and cables). If it should become necessary to return the instrument for  
service during or beyond the warranty period, contact the Customer Service Department at  
(781) 890-7440 ext. 210 for authorization or visit our website www.extech.com for contact  
information. A Return Authorization (RA) number must be issued before any product is  
returned to Extech. The sender is responsible for shipping charges, freight, insurance and  
proper packaging to prevent damage in transit. This warranty does not apply to defects  
resulting from action of the user such as misuse, improper wiring, operation outside of  
specification, improper maintenance or repair, or unauthorized modification. Extech  
specifically disclaims any implied warranties or merchantability or fitness for a specific  
purpose and will not be liable for any direct, indirect, incidental or consequential damages.  
Extech's total liability is limited to repair or replacement of the product. The warranty set  
forth above is inclusive and no other warranty, whether written or oral, is expressed or  
implied.  
ꢀꢁ  
Support line (781) 890-7440  
Technical Support: Extension 200; E-mail: support@extech.com  
Repair & Returns: Extension 210; E-mail: repair@extech.com  
Product specifications subject to change without notice  
For the latest version of this User Guide, Software updates, and other up-to-the-  
minute product information, visit our website: www.extech.com  
Extech Instruments Corporation, 285 Bear Hill Road, Waltham, MA 02451  
Copyright © 2007 Extech Instruments Corporation  
All rights reserved including the right of reproduction in whole or in part in any form.  
8
AM600 V2.0 6/07  

相关型号:

AM60007-1

CHANGE PART KIT FOR 71522 CONN
MOLEX

AM60007-3

MANVAL RIBBON CA. TERMR
MOLEX

AM60008-1

MANVAL RIBBON CA. TERMR
MOLEX

AM60026

MANVAL RIBBON CA. TERMR
MOLEX

AM60026-B

MANVAL RIBBON CA. TERMR
MOLEX

AM60026A100

CHANGE PART KIT FOR 71522 CONN
MOLEX

AM60026B100

MANVAL RIBBON CA. TERMR
MOLEX

AM60026C100

MANVUAL RIBBON CA. TERMR
MOLEX

AM60026F100

MANVAL RIBBON CA. TERMR
MOLEX

AM60026G100

MANVAL RIBBON CA. TERMR
MOLEX

AM60026Z

IDT - Manual Press
MOLEX

AM60037-5

90075 PHONE PLUG ARBOR PRESS
MOLEX