MBC550 [BEL]

12 V / 0.5 A Fan Output, Thermal Shut-Down Feature;
MBC550
型号: MBC550
厂家: BEL FUSE INC.    BEL FUSE INC.
描述:

12 V / 0.5 A Fan Output, Thermal Shut-Down Feature

文件: 总8页 (文件大小:1301K)
中文:  中文翻译
下载:  下载PDF数据表文档文件
The MBC550 Series of open frame medical power supplies feature a  
wide universal AC input range of 90 264 VAC, offering up to 550 W  
of output power with forced air cooling, or 250 W with conduction  
cooling in a compact footprint, with a variety of isolated single output  
voltages.  
The MBC series is designed and approved to the latest Medical  
standards (EN/IEC 60601-1) for Class I applications.  
The MBC550 range is ideally suited to a wide variety of medical  
applications, and where airflow is limited or not available.  
5 x 3 x 1.5 Inch Form Factor  
Up to 550 Watts with Forced Air Cooling  
Efficiencies up to 92%  
-40 to 70°C Operating Temperature  
12 V / 0.5 A Fan Output, Thermal Shut-Down Feature  
Approved to EN/IEC 60601-1  
Medical (BF) Safety Approvals  
MTBF >3 million hours as per Telcordia SR-332, Issue 3  
RoHS Compliant  
CE Marked  
Diagnostic  
Drug Pump  
Dialysis  
Home Health Care  
Monitoring  
Portable Equipment  
2
MBC550 Series  
MBC550-1T12  
MBC550-1T15  
MBC550-1T24  
MBC550-1T30  
MBC550-1T48  
MBC550-1T58  
12 V  
15 V  
24 V  
30 V  
48 V  
58 V  
9.17 A  
7.33 A  
6.25 A  
5.00 A  
3.13 A  
2.59 A  
16.67 A  
13.33 A  
10.42 A  
8.33 A  
5.21 A  
4.31 A  
41.67 A  
33.33 A  
22.92 A  
18.33 A  
11.46 A  
9.48 A  
500 W  
500 W  
550 W  
550 W  
550 W  
550 W  
1
Refer to "Recommended Conduction Plate & Clearance" details on page 8.  
Specifications are for nominal input voltage, 25°C unless otherwise stated.  
Universal  
90-264 VAC / 390 VDC  
47-63 Hz  
Input Voltage  
Input Frequency  
Input Current  
(Derate from 100% at 115 VAC to 78% at 90 VAC)  
115 VAC:  
230 VAC:  
115 VAC:  
230 VAC:  
115 VAC:  
230 VAC:  
264 VAC:  
115 VAC:  
230 VAC:  
Touch current  
6 A max.  
3 A max.  
< 0.5 W  
< 0.7 W  
25 A  
45 A  
75 A  
< 200 µA  
< 400 µA  
<100 µA  
No Load Power  
Inrush Current  
Leakage Current  
Power Factor  
@ Full Load  
> 0.95  
PFC  
Resonant  
70 to 130 KHz  
68 to 80 KHz  
Switching Frequency  
Forced air cooled:  
Conduction cooled:  
Convection cooled:  
48 V:  
up to 550 W  
up to 250 W  
up to 150 W  
92%  
Output Power 2  
24 V:  
12 V, 15 V:  
91%  
90%  
Efficiency typical @ 230 VAC full load)  
(
Full Load  
Convection Load  
Conduction Load  
> 16 ms  
> 55 ms  
> 30 ms  
Hold-up Time (typical)  
Line Regulation 4  
Load Regulation 4  
Minimum Load  
+/-0.5%  
+/-1%  
0.0 A  
50-100% step load change, at 0.1A/µs slew rate,  
50% duty cycle, 50 Hz = 5%  
12 V & 15 V  
Transient Response  
Ripple 3, 4  
recovery time < 5 ms  
2.0 % max  
1.0 % max  
24 V, 30 V, 48 V & 58 V  
Output Voltage Adjustment  
Rise Time  
+/-3%  
Typical  
55 ms  
tech.support@psbel.com  
3
MBC550 Series  
Set Point Tolerance 4  
+/-1%  
Over Current Protection  
Over Voltage Protection  
Short Circuit Protection  
Hiccup mode / Auto recovery  
Hiccup mode / Auto recovery  
Hiccup mode / Auto recovery  
> 110%  
110 to 140%  
Convection Cooled (refer to Derating Curve)  
Conduction Cooled (refer to Derating Curve)  
400 LFM Forced Air Cooled (refer to Derating Curve)  
Up to 150 W max  
Up to 250 W max  
Up to 550 W max  
Cooling  
2
Combined output power of main output, fan supply shall not exceed max. Power rating.  
3
μ
μ
Ripple is peak to peak with 20 MHz bandwidth and 10 F (Electrolytic capacitor) in parallel with a 0.1 F capacitor at rated line  
voltage and load ranges.  
4
Fan supply output voltage tolerance including set point accuracy, line and load regulation is +/-10 % and Ripple and noise is less  
than 10 %.  
Operating Temperature 5  
-40 to +70°C  
-40 to 0°C startup guaranteed, with spec deviation 6  
Non-condensing  
Storage Temperature  
Relative Humidity  
-40 to +85°C  
5% to 95%  
Operating:  
Non-operating:  
16,000 ft.  
40,000 ft.  
Altitude  
5
Thermal shutdown feature: The power supply goes in hiccup mode when the temperature of PCB exceeds 110 °C (+/-10 °C).  
Output ripple can be more than 10% of the output voltage.  
6
Conducted Emissions  
Radiated Emissions  
EN 55011-B, CISPR22-B, FCC PART15-B  
EN 55011 A;  
with external core (King core K5B RC 25x12x15-M in input cable)  
Pass  
Pass  
Level B  
Input Current Harmonics  
Voltage Fluctuation and Flicker  
ESD Immunity  
EN 61000-3-2  
EN 61000-3-3  
EN 61000-4-2  
EN 61000-4-3  
EN 61000-4-4  
EN 61000-4-5  
EN 61000-4-6  
EN 61000-4-8  
EN 61000-4-11  
Class D  
Pass  
Level 4, Criterion A  
Level 3, Criterion A  
Level 3, Criterion A  
Level 3, Criterion A  
Level 3, Criterion A  
Level 4, Criterion A  
Criterion B  
Radiated Field Immunity  
Electrical Fast Transient Immunity  
Surge Immunity  
Conducted Immunity  
Magnetic Field Immunity  
Voltage Dips, Interruptions  
North America  
+1 408 785 5200  
Asia-Pacific  
+86 755 298 85888  
Europe, Middle East  
+353 61 225 977  
© 2020 Bel Power Solutions & Protection  
BCD.00818_AC1  
4
MBC550 Series  
Input to Output: (for medical applications)  
Input to GND:  
Output to GND: for type BF  
for type B  
4000 VAC  
1500 VAC  
1500 VAC  
500 VAC  
Isolation Voltage  
UL /CSA : ANSI/AAMI ES60601-1 (2005 + C1:09 + A2:10) AMD1:2012; CORR1:2009; AMD2:2010,  
(CAN/CSA-C22.2 No. 60601-1 (2008)CAN/CSA C22.2 No. 60601-1:14  
IEC : IEC 60601-1: 2005 + CORR. 1:2006 + CORR. 2:2007 + AM1:2012(or IEC 60601-1: 2012 reprint)  
EN : EN 60601-1:2006;A1  
Safety Standard(s)  
Agency Approvals  
CE mark  
Nemko, UL, C-UL  
Complies with LVD Directive  
Pin 1  
Pin 2  
Pin 3  
AC Line  
Not Fitted  
AC Neutral  
JST : B3P-VH-B(LF)(SN) or equivalent  
Mating: VHR-3M or equivalent Pins : SVH-41T-P1.1 or equivalent  
AC Input Connector  
J1  
J2  
6-32 inches Screw Pan HD  
Mating: Designed to accept Ring Tongue Terminal AMP: 8-31886-1,  
wherein one 16 AWG (max) wire can be crimped.  
Note: One Ring Tongue Terminal with 16 AWG is recommended for  
current up to 11A only.  
DC Output Connector  
(Screw Terminal)  
Pin 1  
Pin 2  
V1 +VE  
V1 -VE  
Use multiple tongue terminals with wire for more current.  
AMP: 640456-2  
Mating: 640440-2  
Molex: 19705-4301  
Mating: 19003-0001  
Pin 1  
Pin 2  
FAN +VE  
FAN -VE  
Aux (Fan) Output  
Earth  
J3  
J4  
Weight  
approx. 500 g  
127 x 76.2 x 38.1 mm (5 x 3 x 1.5 inches)  
Dimensions  
Mechanical Drawing  
tech.support@psbel.com  
5
MBC550 Series  
De-rate linearly from 100% at  
110 VAC to 78% at 90 VAC  
Convection load: 110 W up to 50 °C  
De-rate above 50 °C @ 2.5% per °C  
Conduction load: 200 W up to 50 °C  
De-rate above 50 °C @ 2.5% per °C  
Forced air cooled load:  
500 W up to 50°C  
De-rate above 50 °C @ 2.5% per °C  
Convection load: 150 W up to 50 °C  
De-rate above 50 °C @ 2.5% per °C  
Conduction load: 250 W up to 50 °C  
De-rate above 50 °C @ 2.5% per °C  
Forced air cooled load:  
550 W up to 50°C  
De-rate above 50 °C @ 2.5% per °C  
North America  
Asia-Pacific  
Europe, Middle East  
+1 408 785 5200  
+86 755 298 85888  
+353 61 225 977  
© 2020 Bel Power Solutions & Protection  
BCD.00818_AC1  
6
MBC550 Series  
For reliable and safe operation, please make sure the maximum component temperatures given in table below is not exceeded.  
TOP PCB  
BOTTOM PCB  
MAXIMUM TEMP.  
ALLOWED (°C)  
INDENT NO  
DESCRIPTION  
1
2
3
4
Common mode chokes  
Input Bulk Capacitors  
Differential choke  
Boost Choke  
95  
90  
110  
110  
125 (for 12 V & 15 V)  
110 (for 24 V, 30 V,48 V,58 V)  
5
Output Transformer  
6
7
8
9
Output Capacitor  
Bridge Rectifier  
90  
120  
105  
110  
Aluminum Clad PCB  
Output Rectifiers  
tech.support@psbel.com  
7
MBC550 Series  
FIXING OPTION - 2  
FIXING OPTION - 1  
Figure 2. Mounting Options  
North America  
+1 408 785 5200  
Asia-Pacific  
+86 755 298 85888  
Europe, Middle East  
+353 61 225 977  
© 2020 Bel Power Solutions & Protection  
BCD.00818_AC1  
8
MBC550 Series  
Conduction power rating mentioned in the table is with additional aluminum base plate of 3 mm thickness with 177.8 mm (7 in) length  
& 101.6 mm (4 in) width.  
Clearance of minimum 15 mm above the component height is recommended for better thermal management.  
NUCLEAR AND MEDICAL APPLICATIONS - Products are not designed or intended for use as critical components in life support systems,  
equipment used in hazardous environments, or nuclear control systems.  
TECHNICAL REVISIONS - The appearance of products, including safety agency certifications pictured on labels, may change depending on the  
date manufactured. Specifications are subject to change without notice.  
tech.support@psbel.com  

相关型号:

MBC550-1T12

5 x 3 x 1.5 Inch Form Factor
BEL

MBC550-1T15

5 x 3 x 1.5 Inch Form Factor
BEL

MBC550-1T24

5 x 3 x 1.5 Inch Form Factor
BEL

MBC550-1T30

5 x 3 x 1.5 Inch Form Factor
BEL

MBC550-1T48

5 x 3 x 1.5 Inch Form Factor
BEL

MBC550-1T58

5 x 3 x 1.5 Inch Form Factor
BEL

MBC60

50 - 65 W Single to Triple Outputs
BEL

MBC60-1005G

50 - 65 W Single to Triple Outputs
BEL

MBC60-1012G

50 - 65 W Single to Triple Outputs
BEL

MBC60-1012G-2

AC-DC Regulated Power Supply Module, 60W,
BEL

MBC60-1015G

AC-DC Regulated Power Supply Module, 60W,
BEL

MBC60-1024G

AC-DC Regulated Power Supply Module, 60W,
BEL