MBC550 [BEL]
12 V / 0.5 A Fan Output, Thermal Shut-Down Feature;![MBC550](http://pdffile.icpdf.com/pdf2/p00356/img/icpdf/MBC550_2185235_icpdf.jpg)
型号: | MBC550 |
厂家: | ![]() |
描述: | 12 V / 0.5 A Fan Output, Thermal Shut-Down Feature |
文件: | 总8页 (文件大小:1301K) |
中文: | 中文翻译 | 下载: | 下载PDF数据表文档文件 |
![](http://public.icpdf.com/style/img/ads.jpg)
The MBC550 Series of open frame medical power supplies feature a
wide universal AC input range of 90 – 264 VAC, offering up to 550 W
of output power with forced air cooling, or 250 W with conduction
cooling in a compact footprint, with a variety of isolated single output
voltages.
The MBC series is designed and approved to the latest Medical
standards (EN/IEC 60601-1) for Class I applications.
The MBC550 range is ideally suited to a wide variety of medical
applications, and where airflow is limited or not available.
5 x 3 x 1.5 Inch Form Factor
•
•
•
•
•
•
•
•
•
•
Up to 550 Watts with Forced Air Cooling
Efficiencies up to 92%
-40 to 70°C Operating Temperature
12 V / 0.5 A Fan Output, Thermal Shut-Down Feature
Approved to EN/IEC 60601-1
Medical (BF) Safety Approvals
MTBF >3 million hours as per Telcordia SR-332, Issue 3
RoHS Compliant
CE Marked
Diagnostic
Drug Pump
Dialysis
Home Health Care
Monitoring
•
•
•
•
•
•
Portable Equipment
2
MBC550 Series
MBC550-1T12
MBC550-1T15
MBC550-1T24
MBC550-1T30
MBC550-1T48
MBC550-1T58
12 V
15 V
24 V
30 V
48 V
58 V
9.17 A
7.33 A
6.25 A
5.00 A
3.13 A
2.59 A
16.67 A
13.33 A
10.42 A
8.33 A
5.21 A
4.31 A
41.67 A
33.33 A
22.92 A
18.33 A
11.46 A
9.48 A
500 W
500 W
550 W
550 W
550 W
550 W
1
Refer to "Recommended Conduction Plate & Clearance" details on page 8.
Specifications are for nominal input voltage, 25°C unless otherwise stated.
Universal
90-264 VAC / 390 VDC
47-63 Hz
Input Voltage
Input Frequency
Input Current
(Derate from 100% at 115 VAC to 78% at 90 VAC)
115 VAC:
230 VAC:
115 VAC:
230 VAC:
115 VAC:
230 VAC:
264 VAC:
115 VAC:
230 VAC:
Touch current
6 A max.
3 A max.
< 0.5 W
< 0.7 W
25 A
45 A
75 A
< 200 µA
< 400 µA
<100 µA
No Load Power
Inrush Current
Leakage Current
Power Factor
@ Full Load
> 0.95
PFC
Resonant
70 to 130 KHz
68 to 80 KHz
Switching Frequency
Forced air cooled:
Conduction cooled:
Convection cooled:
48 V:
up to 550 W
up to 250 W
up to 150 W
92%
Output Power 2
24 V:
12 V, 15 V:
91%
90%
Efficiency typical @ 230 VAC full load)
(
Full Load
Convection Load
Conduction Load
> 16 ms
> 55 ms
> 30 ms
Hold-up Time (typical)
Line Regulation 4
Load Regulation 4
Minimum Load
+/-0.5%
+/-1%
0.0 A
50-100% step load change, at 0.1A/µs slew rate,
50% duty cycle, 50 Hz = 5%
12 V & 15 V
Transient Response
Ripple 3, 4
recovery time < 5 ms
2.0 % max
1.0 % max
24 V, 30 V, 48 V & 58 V
Output Voltage Adjustment
Rise Time
+/-3%
Typical
55 ms
tech.support@psbel.com
3
MBC550 Series
Set Point Tolerance 4
+/-1%
Over Current Protection
Over Voltage Protection
Short Circuit Protection
Hiccup mode / Auto recovery
Hiccup mode / Auto recovery
Hiccup mode / Auto recovery
> 110%
110 to 140%
Convection Cooled (refer to Derating Curve)
Conduction Cooled (refer to Derating Curve)
400 LFM Forced Air Cooled (refer to Derating Curve)
Up to 150 W max
Up to 250 W max
Up to 550 W max
Cooling
2
Combined output power of main output, fan supply shall not exceed max. Power rating.
3
μ
μ
Ripple is peak to peak with 20 MHz bandwidth and 10 F (Electrolytic capacitor) in parallel with a 0.1 F capacitor at rated line
voltage and load ranges.
4
Fan supply output voltage tolerance including set point accuracy, line and load regulation is +/-10 % and Ripple and noise is less
than 10 %.
Operating Temperature 5
-40 to +70°C
-40 to 0°C startup guaranteed, with spec deviation 6
Non-condensing
Storage Temperature
Relative Humidity
-40 to +85°C
5% to 95%
Operating:
Non-operating:
16,000 ft.
40,000 ft.
Altitude
5
Thermal shutdown feature: The power supply goes in hiccup mode when the temperature of PCB exceeds 110 °C (+/-10 °C).
Output ripple can be more than 10% of the output voltage.
6
Conducted Emissions
Radiated Emissions
EN 55011-B, CISPR22-B, FCC PART15-B
EN 55011 A;
with external core (King core K5B RC 25x12x15-M in input cable)
Pass
Pass
Level B
Input Current Harmonics
Voltage Fluctuation and Flicker
ESD Immunity
EN 61000-3-2
EN 61000-3-3
EN 61000-4-2
EN 61000-4-3
EN 61000-4-4
EN 61000-4-5
EN 61000-4-6
EN 61000-4-8
EN 61000-4-11
Class D
Pass
Level 4, Criterion A
Level 3, Criterion A
Level 3, Criterion A
Level 3, Criterion A
Level 3, Criterion A
Level 4, Criterion A
Criterion B
Radiated Field Immunity
Electrical Fast Transient Immunity
Surge Immunity
Conducted Immunity
Magnetic Field Immunity
Voltage Dips, Interruptions
North America
+1 408 785 5200
Asia-Pacific
+86 755 298 85888
Europe, Middle East
+353 61 225 977
© 2020 Bel Power Solutions & Protection
BCD.00818_AC1
4
MBC550 Series
Input to Output: (for medical applications)
Input to GND:
Output to GND: for type BF
for type B
4000 VAC
1500 VAC
1500 VAC
500 VAC
Isolation Voltage
UL /CSA : ANSI/AAMI ES60601-1 (2005 + C1:09 + A2:10) AMD1:2012; CORR1:2009; AMD2:2010,
(CAN/CSA-C22.2 No. 60601-1 (2008)CAN/CSA C22.2 No. 60601-1:14
IEC : IEC 60601-1: 2005 + CORR. 1:2006 + CORR. 2:2007 + AM1:2012(or IEC 60601-1: 2012 reprint)
EN : EN 60601-1:2006;A1
Safety Standard(s)
Agency Approvals
CE mark
Nemko, UL, C-UL
Complies with LVD Directive
Pin 1
Pin 2
Pin 3
AC Line
Not Fitted
AC Neutral
JST : B3P-VH-B(LF)(SN) or equivalent
Mating: VHR-3M or equivalent Pins : SVH-41T-P1.1 or equivalent
AC Input Connector
J1
J2
6-32 inches Screw Pan HD
Mating: Designed to accept Ring Tongue Terminal AMP: 8-31886-1,
wherein one 16 AWG (max) wire can be crimped.
Note: One Ring Tongue Terminal with 16 AWG is recommended for
current up to 11A only.
DC Output Connector
(Screw Terminal)
Pin 1
Pin 2
V1 +VE
V1 -VE
Use multiple tongue terminals with wire for more current.
AMP: 640456-2
Mating: 640440-2
Molex: 19705-4301
Mating: 19003-0001
Pin 1
Pin 2
FAN +VE
FAN -VE
Aux (Fan) Output
Earth
J3
J4
Weight
approx. 500 g
127 x 76.2 x 38.1 mm (5 x 3 x 1.5 inches)
Dimensions
Mechanical Drawing
tech.support@psbel.com
5
MBC550 Series
De-rate linearly from 100% at
110 VAC to 78% at 90 VAC
Convection load: 110 W up to 50 °C
De-rate above 50 °C @ 2.5% per °C
Conduction load: 200 W up to 50 °C
De-rate above 50 °C @ 2.5% per °C
Forced air cooled load:
500 W up to 50°C
De-rate above 50 °C @ 2.5% per °C
Convection load: 150 W up to 50 °C
De-rate above 50 °C @ 2.5% per °C
Conduction load: 250 W up to 50 °C
De-rate above 50 °C @ 2.5% per °C
Forced air cooled load:
550 W up to 50°C
De-rate above 50 °C @ 2.5% per °C
North America
Asia-Pacific
Europe, Middle East
+1 408 785 5200
+86 755 298 85888
+353 61 225 977
© 2020 Bel Power Solutions & Protection
BCD.00818_AC1
6
MBC550 Series
For reliable and safe operation, please make sure the maximum component temperatures given in table below is not exceeded.
TOP PCB
BOTTOM PCB
MAXIMUM TEMP.
ALLOWED (°C)
INDENT NO
DESCRIPTION
1
2
3
4
Common mode chokes
Input Bulk Capacitors
Differential choke
Boost Choke
95
90
110
110
125 (for 12 V & 15 V)
110 (for 24 V, 30 V,48 V,58 V)
5
Output Transformer
6
7
8
9
Output Capacitor
Bridge Rectifier
90
120
105
110
Aluminum Clad PCB
Output Rectifiers
tech.support@psbel.com
7
MBC550 Series
FIXING OPTION - 2
FIXING OPTION - 1
Figure 2. Mounting Options
North America
+1 408 785 5200
Asia-Pacific
+86 755 298 85888
Europe, Middle East
+353 61 225 977
© 2020 Bel Power Solutions & Protection
BCD.00818_AC1
8
MBC550 Series
Conduction power rating mentioned in the table is with additional aluminum base plate of 3 mm thickness with 177.8 mm (7 in) length
& 101.6 mm (4 in) width.
Clearance of minimum 15 mm above the component height is recommended for better thermal management.
NUCLEAR AND MEDICAL APPLICATIONS - Products are not designed or intended for use as critical components in life support systems,
equipment used in hazardous environments, or nuclear control systems.
TECHNICAL REVISIONS - The appearance of products, including safety agency certifications pictured on labels, may change depending on the
date manufactured. Specifications are subject to change without notice.
tech.support@psbel.com