MBC350-1T24L [BEL]
350 W with Forced Air Cooling & 200 W with Convection Cooling;型号: | MBC350-1T24L |
厂家: | BEL FUSE INC. |
描述: | 350 W with Forced Air Cooling & 200 W with Convection Cooling |
文件: | 总6页 (文件大小:552K) |
中文: | 中文翻译 | 下载: | 下载PDF数据表文档文件 |
The MBC350 Series of open frame medical power supplies feature a
wide universal AC input range of 90 – 264 VAC, offering 350 W of
output power in a compact 3 x 5 x 1 inch footprint, with a variety of
isolated single output voltages.
The MBC series is designed and approved to the latest Medical
standards (EN/IEC 60601-1), providing 2 x MOPP isolation for Class I
& Class II applications.
These power supplies are ideal for medical, telecom, datacom,
industrial equipment and other applications.
3 x 5 x 1 Inch Form Factor
•
•
•
•
•
•
•
•
•
•
350 W with Forced Air Cooling & 200 W with Convection Cooling
Approved to EN/IEC 60601-1
Efficiencies up to 94%
-40 to 70°C Operating Temperature
12 V / 0.5 A Fan Output, Thermal Shut-Down Feature
2.56 Million Hours, Telcordia -SR332-Issue 3 MTBF
Standby Power < 0.5 W
Medical (BF) Safety Approvals
RoHS Compliant
Diagnostic
Drug Pump
Dialysis
Home Health Care
Monitoring
•
•
•
•
•
•
Portable Equipment
2
MBC350 Series
MAX. LOAD
(CONVECTION)
MAX. LOAD
POWER
MODEL NUMBER 1
DESCRIPTION
VOLTAGE
12 V
(375 LFM)
MBC350-1T12L
MBC350-1012L
Screw Terminal
Molex Connector
25 A
300 W
225 W
15 A
18.75 A
MBC350-1T15L
MBC350-1015L
Screw Terminal
Molex Connector
21.67 A
18. A
325 W
270 W
15 V
12 A
MBC350-1T24L
MBC350-1024L
Screw Terminal
Molex Connector
24 V
8.33 A
6.67 A
4.17 A
3.45 A
14.60 A
11.67 A
7.30 A
6.04 A
350 W
350 W
350 W
350 W
MBC350-1T30L
MBC350-1030L
Screw Terminal
Molex Connector
30 V
MBC350-1T48L
MBC350-1048L
Screw Terminal
Molex Connector
48 V
MBC350-1T58L
MBC350-1058L
Screw Terminal
Molex Connector
58 V
COVER-350-XBC 2
Metal cover kit (accessory)
1
Class II version available. Add suffix “-2” at the end of the Model Number
When used in Cover Kit, de-rate output power to 70 % under all operating conditions.
2
Specifications are for nominal input voltage, 25°C unless otherwise stated.
PARAMETER
DESCRIPTION / CONDITION
SPECIFICATION
90-264 VAC / 390 VDC
47 - 63 Hz
Universal
Input Voltage
(Derate from 100% at 100 VAC to 90% at 90 VAC)
Input Frequency
Input Current
115 VAC:
230 VAC:
3.6 A max.
1.8 A max.
No Load Power
Typical
>0.5 W
115 VAC:
230 VAC:
264 VAC:
Typical
25 A
45 A
75 A
300 uA
Inrush Current
Leakage Current
Power Factor
Touch Current
<100 uA
Full Load
>0.95
PFC:
PWM:
70 - 130 KHz
50 - 80 KHz
Switching Frequency
tech.support@psbel.com
3
MBC350 Series
PARAMETER
DESCRIPTION / CONDITION
SPECIFICATION
Output Voltage
12 to 58 V
With 375 LFM:
Convection:
350 W
200 W
Output Power 3
Output Adjustment
Hold-up Time
+/-3%
Full Load:
Convection Load:
48 V, 58 V:
> 8 ms typical
> 14 ms typical
94%
Efficiency
24 V, 30 V:
12 V, 15 V:
93%
92%
Line Regulation 5
Load Regulation 5
Minimum Load
+/-0.5%
+/-1%
0.0 A
50-100% step load change, at 0.1A/µS slew rate,
50% duty cycle, 50 Hz = 5% ,
Transient Response
recovery time < 5 ms
Ripple 4, 5
For all outputs
Typical
1.0 % max
55 ms
Rise Time
Set Point Tolerance 5
Over Current Protection
Over Voltage Protection
Short Circuit Protection
+/-1%
Hiccup mode / Auto Recovery
Hiccup mode / Auto Recovery
Hiccup mode / Auto Recovery
>110%
110 to 140%
With 375 LFM forced air cooling at 100 to 264 VAC:
With natural convection cooling at 100 to 264 VAC:
350 W
200 W
Cooling
3
Combined output power of main output, fan supply shall not exceed max. power rating.
4
μ
μ
Ripple is peak to peak with 20 MHz bandwidth and 10 F (Electrolytic capacitor) in parallel with a 0.1 F capacitor at rated line
voltage and load ranges.
5
Fan supply output voltage tolerance including set point accuracy, line and load regulation is +/-10 % and ripple and noise is less
than 10 %.
Conducted Emissions
Radiated Emissions
EN 55011-B, CISPR22-B, FCC PART15-B
Pass
EN 55011 A;
Pass
Level B
with external core (King core K5B RC 25x12x15-M in input cable)
Input Current Harmonics
Voltage Fluctuation and Flicker
ESD Immunity
EN 61000-3-2
EN 61000-3-3
EN 61000-4-2
EN 61000-4-3
EN 61000-4-4
EN 61000-4-5
EN 61000-4-6
EN 61000-4-8
EN 61000-4-11
Class D
Pass
Level 4, Criterion A
Level 3, Criterion A
Level 3, Criterion A
Level 4, Criterion A
Level 3, Criterion A
Level 4, Criterion A
Criterion B
Radiated Field Immunity
Electrical Fast Transient Immunity
Surge Immunity
Conducted Immunity
Magnetic Field Immunity
Voltage Dips, Interruptions
North America
+1 408 785 5200
Asia-Pacific
+86 755 298 85888
Europe, Middle East
+353 61 225 977
© 2020 Bel Power Solutions & Protection
BCD.00561_AG
4
MBC350 Series
Input to Output: (For medical applications)
Input to GND: (Not Applicable For Class II Option)
Output to GND: for type BF
4000 VAC
1500 VAC
1500 VAC
500 VAC
Isolation Voltage
for type B (N/A For Class II Option)
Safety Standard(s)
Agency Approvals
CE mark
EN 60601-1, IEC 60601-1 (ed.3), ANSI / AAMI ES 60601 - 1, CSA C22.2 No. 60601-1
Nemko, UL, C-UL
Complies with LVD Directive
PARAMETER
DESCRIPTION / CONDITION
SPECIFICATION
-40 to +70°C
-40 to 85° C
Operating Temperature 6
Storage Temperature
Altitude
-40 to 0°C startup guaranteed, with spec deviation 7
Operating:
Non-operating:
16,000 ft.
40,000 ft.
Humidity
Reliability
Non-Condensing
5% to 95%
MTBF according to Telcordia - SR332-Issue 3
2.56 million hours
CONNECTOR
PIN
DESCRIPTION / CONDITION
MANUFACTURER / PN
Pin 1
Pin 2
Pin 3
AC Line
Not Fitted
AC Neutral
Molex: 26-60-4030
Mating: 09-50-3031; Pins: 08-50-0106
AC Input Connector
J1
6-32 inches Screw Pan HD
Mating: 16 AWG wire crimped to Ring
Tongue Terminal AMP: 8-31886-1
Molex: 26-60-4080
Mating: 09-50-3081; Pins: 08-50-0106
AMP :640456-2
Mating: 640440-2
Molex: 19705-4301
Mating: 19003-0001
Screw Terminal
(Option 1)
Pin 1
Pin 2
V1 +VE
V1 - VE
DC Output Connector
J2
Molex Connector
(Option 2)
Pin 1,2,3,4
Pin 5,6,7,8
Pin 1
V1 +VE
V1 - VE
FAN +VE
FAN -VE
Aux (Fan) Output
Earth
J3
J4
Pin 2
Weight
300 g
Dimensions
76.2 x 127.0 x 25.4 mm (3 x 5 x 1 inch)
tech.support@psbel.com
5
MBC350 Series
Figure 1. Mechanical Drawing - Screw Terminal (Option 1)
Figure 2. Mechanical Drawing - Molex Header (Option 2)
NOTES: In case the PCB is mounted in a metal enclosure, using metal hardware ensure the following:
1
2
3
Stand off, used to mount PCB has OD of 5.4 mm max.
Screws, used to fix PCB on stand off, have head dia of 6.0 mm max.
Washer, if used, to have dia of 6.5 mm max.
North America
+1 408 785 5200
Asia-Pacific
+86 755 298 85888
Europe, Middle East
+353 61 225 977
© 2020 Bel Power Solutions & Protection
BCD.00561_AG
6
MBC350 Series
Convection load: 180 W up to 50 °C
De-rate above 50 °C @ 2.5% per °C
Forced air cooled load: 300 W up to 50°C
De-rate above 50 °C @ 2.5% per °C
Convection load: 180 W up to 50 °C
De-rate above 50 °C @ 2.5% per °C
Forced air cooled load: 325 W up to 50°C
De-rate above 50 °C @ 2.5% per °C
Convection load: 200 W up to 50 °C
De-rate above 50 °C @ 2.5% per °C
Forced air cooled load: 350 W up to 50°C
De-rate above 50 °C @ 2.5% per °C
NUCLEAR AND MEDICAL APPLICATIONS - Products are not designed or intended for use as critical components in life support systems,
equipment used in hazardous environments, or nuclear control systems.
TECHNICAL REVISIONS - The appearance of products, including safety agency certifications pictured on labels, may change depending on the
date manufactured. Specifications are subject to change without notice.
tech.support@psbel.com
相关型号:
©2020 ICPDF网 联系我们和版权申明