MBC150-1T15G [BEL]
4 x 2 x 1.3 Inch Form Factor;型号: | MBC150-1T15G |
厂家: | BEL FUSE INC. |
描述: | 4 x 2 x 1.3 Inch Form Factor |
文件: | 总6页 (文件大小:730K) |
中文: | 中文翻译 | 下载: | 下载PDF数据表文档文件 |
The MBC150 Series of open-frame medical power supplies, withits
wide universal input range 90-264 VAC, is available at 150 W of
output power with forced-air cooling and a variety of single output
voltages.
The MBC series is designed and approved to the latest Medical
standards (EN/IEC 60601-1), providing 2 x MOPP (Means of Patient
Protection) isolationfor Class I & Class II applications.
These power supplies are ideal for medical, telecom, datacom,
industrial equipment and other applications.
•
•
•
•
•
•
•
•
4 x 2 x 1.3 Inch Form Factor
150 W with Forced-Air Cooling
Approved to EN/IEC 60601-1
12 V @ 0.5 A fan voltage auxiliary output
High Efficiency > 86%
Low conducted and radiated noise
Light weight
IEC Protection Class Options:
−
−
Class I: Earthing Tab J4 (no suffix)
Class II: No Earthing Tab (-2 suffix)
•
•
Medical Safety Agency Approvals
RoHS Compliant
•
•
•
Diagnostic
Drug Pump
Monitoring
•
Dialysis
•
•
Home Health Care
Portable Equipment
2
MBC150 Series
2
MAX LOAD
CONVECTION
OUTPUT
VOLTAGE
MODEL 1
CONNECTOR
MIN LOAD
0.0 A
POWER
300 LFM
MBC150-1005G
MBC150-1T05G
MBC150-1012G
MBC150-1T12G
MBC150-1015G
MBC150-1T15G
MBC150-1024G
MBC150-1T24G
MBC150-1048G
MBC150-1T48G
JST (Header)
Screw Terminal
JST (Header)
16.0 A
20.0 A
80 W
5 VDC
12 VDC
15 VDC
24 VDC
48 VDC
16.0 A
8.33 A
6.67 A
4.17 A
2.08 A
100 W
0.0 A
12.5 A
10.0 A
6.25 A
3.13 A
150 W
150 W
150 W
150 W
Screw Terminal
JST (Header)
0.0 A
Screw Terminal
JST (Header)
0.0 A
Screw Terminal
JST (Header)
0.0 A
Screw Terminal
COVER-150-XBC3 Metal Cover Kit (accessory)
1
Class I products havean Earthing tab. For Class II version (without Earthing tab) add suffix -2 (e.g.: MBC150-1012G-2).
Combined output power fromV1 and VFAN should not exceed the total output power rating.
When used in Cover Kit, de-rate output power to 70 % under alloperating conditions.
2
3
Specifications are for nominal input voltage, 25°C unless otherwise stated.
PARAMETER
DESCRIPTION / CONDITION
SPECIFICATION
90 – 264 VAC / 120 – 390 VDC
47 – 63 Hz
Input Voltage
Universal
Input Frequency
120 VAC:
230 VAC:
1.70 A max.
0.85 A max.
Input Current
No Load Power
Inrush Current
1.2 W
120 VAC:
230 VAC:
120 VAC:
230 VAC:
35 A max.
65 A max.
< 140 µA
< 250 µA
Leakage Current
No Load Power
1.2 W
PFC converter (variable)
Resonant converter (variable)
35 – 250 kHz, 90 kHz typical
35 – 250 kHz, 90 kHz typical
Switching Frequency
tech.support@psbel.com
3
MBC150 Series
PARAMETER
Output Voltage
Output Power4
Fan Output5
DESCRIPTION / CONDITION
SPECIFICATION
5 V to 48 V
150 W
Refer to Model Selection table
Peak Power: 170 W for 2 s
@ 5 A
12 V
120 VAC:
230 VAC:
120 VAC:
230 VAC:
120 VAC:
230 VAC:
84% typical
86% typical
6 ms
10 ms
0.99
Efficiency
Hold Up Time
Power Factor
0.95
Line Regulation
Load Regulation
± 0.5%
± 2.0%
< 10%,
Transient Response
50 to 100% load change, 50 Hz, 50% duty cycle, 0.1A / µs
All outputs
recovery time < 5 ms
Ripple6
1.0 % max
Rise Time
< 100 ms
Set Point Accuracy
Output Voltage Adjustment
Over Current Protection
Over Voltage Protection
Short Circuit Protection
Main output
V1
± 1%
± 3 %
110% typical above rating
110 to 150%
V1
Short term, Automatic recovery
Convection
80 W (5 V model), 100 W (other models)
100 W (5 V model), 150 W (other models)
Cooling
300 LFM
4
Derate output power linearly to 80% from90 VAC to 80 VAC input.
Fan output voltage tolerance is +/-20 %. Peak current for fan output is 1 A.
Ripple is 2 % up to 20 % load and < 1 % above20 % load. Ripple is peak to peak with 20 MHz bandwidth and 10 µF (Electrolytic
capacitor) in parallel with a 0.1 µF capacitor at rated linevoltage and load ranges.
5
6
Conducted Emissions
Radiated Emissions
EN 55011-B, CISPR22-B, FCC PART15-B
EN 55011 B
Pass
Pass
Input Current Harmonics
Voltage Fluctuation and Flicker
ESD Immunity
EN 61000-3-2
Class D
EN 61000-3-3
Pass
EN 61000-4-2
Level 4, Criterion A
Level 3, Criterion A
Level 3, Criterion A
Level 3, Criterion A
Level 3, Criterion A
Level 4, Criterion A
Criterion A & B
Radiated Field Immunity
Electrical Fast Transient Immunity
Surge Immunity
EN 61000-4-3
EN 61000-4-4
EN 61000-4-5
Conducted Immunity
EN 61000-4-6
Magnetic Field Immunity
Voltage Dips, Interruptions
EN 61000-4-8
EN 61000-4-11
North America
Asia-Pacific
+86 755 298 85888
Europe, Middle East
+353 61 225 977
+1 408 785 5200
© 2021 Bel Power Solutions & Protection
BCD.00288_AF
4
MBC150 Series
PARAMETER
DESCRIPTION / CONDITION
SPECIFICATION
Isolation Voltage
Input to Output:
Min. 5900 VDC
EN 60601-1, IEC 60601-1 (ed.3),
ANSI / AAMI ES 60601 - 1, CSA C22.2 No. 60601-1
Safety Standards
Agency Approvals
CE mark
Nemko, UL, C-UL
Complies with LVD Directive
PARAMETER
DESCRIPTION / CONDITION
SPECIFICATION
-20 to 70°C
-40 to 70° C
95%
Operating Temperature 7
Storage Temperature
Humidity
-20 to 0°C start-up is guaranteed
Non Condensing
Operating:
Non-Operating:
10,000 ft.
40,000 ft.
Altitude
Reliability
MTBF according to Telcordia -SR332-Issue 3
2.4 million hours
7
Refer to derating curves
Convection load: 80 W up to 50 °C
De-rate above 50 °C @ 3.125% per °C
Forced air cooled load: 100W up to 50°C
De-rate above 50 °C @ 2.5% per °C
Convection load: 100 W up to 50 °C
De-rate above 50 °C @ 2.5% per °C
Forced air cooled load : 150 W up to 50°C
De-rate above 50 °C @ 1.67% per °C
Figure 1. Derating Curves
tech.support@psbel.com
5
MBC150 Series
CONNECTOR
PIN
DESCRIPTION / CONDITION
MANUFACTURER / PN
Pin 1
Pin 2
AC Line
AC Neutral
Molex: 26-60-4030 or equivalent
Mating: 09-50-3031; Pins: 08-50-0106
Option 1:
AC Input Connector
J1
Tyco: 1776112-4 or equivalent
Mating: 13 AWG wire
Option 2:
Pin 1, 2
Pin 3, 4
V1
RTN
DC Output Connector
J2
JST: B4P-VH-B (LF) (SN) or B4P-VH (LF) (SN) or equivalent
Mating: VHR-4M; Pins: SVH-41T-P1.1
VFAN (12 V / 0.5 A) Tyco: 640456-2 or equivalent
Pin 1
Pin 2
Fan (Aux) Connector
J3
J4
RTN
Mating: 640440-2
Molex: 19705-4301 or equivalent
Mating: 190030001
Earth 8 (Spade Connector)
8
Class I product only
PARAMETER
Weight
DESCRIPTION / CONDITION
150 g (0.33 lbs.)
Dimensions
101.6 x 50.8 x 33.6 mm (4.0 x 2.0 x 1.3 inch)
Figure 2. Mechanical Drawing
NOTES: In case the PCB is mounted in a metal enclosure, using metal hardware ensure the following:
1
2
3
Stand off, used to mount PCB has OD of 5.4 mm max.
Screws, used to fix PCB on stand off, have head dia of 6.0 mm max.
Washer, if used, to have dia of 6.5 mm max.
North America
Asia-Pacific
+86 755 298 85888
Europe, Middle East
+353 61 225 977
+1 408 785 5200
© 2021 Bel Power Solutions & Protection
BCD.00288_AF
6
MBC150 Series
Figure 3. Mechanical Drawing with Cover Kit
NUCLEAR AND MEDICAL APPLICATIONS - Products are not designed or intended for useas critical components in life support systems,
equipment used in hazardous environments, or nuclear control systems.
TECHNICAL REVISIONS - The appearance of products, including safety agency certifications pictured on labels, may change depending on the
date manufactured. Specifications are subject to change without notice.
tech.support@psbel.com