3285-20 [ASM-SENSOR]
CLAMP ON AC/DC HiTESTER; 钳形AC / DC测试仪型号: | 3285-20 |
厂家: | ASM GMBH |
描述: | CLAMP ON AC/DC HiTESTER |
文件: | 总2页 (文件大小:1881K) |
中文: | 中文翻译 | 下载: | 下载PDF数据表文档文件 |
(1.30")
200
33
[
Arms, ø mm
max.]
3284
(
2.17")
[
Arms, ø
mm
55
max.]
2000
3285/-20
CLAMP ON AC/DC HiTESTER
Field measuring instruments
2000 A
analysis
(3285 /3285-20)
3284
3285-20
Analysis for
DC to distorted
waves
CAT III 600 V
http://www.hioki.co.jp/
�ꢀꢁꢂꢀꢃꢄꢅꢆꢇꢈꢉꢊꢃꢅꢋꢌꢍꢋꢎꢌꢏꢐꢃꢉꢌꢏꢃꢇꢍꢅꢑꢒꢄꢓꢔꢐꢃꢌꢉꢋꢎꢍꢅꢉꢆꢌꢉꢓꢈꢕꢃꢄꢅꢉꢔꢎꢑꢌꢍꢈꢓꢎꢅꢉꢔꢃꢈꢉꢑ
ꢃꢅꢓꢖꢌꢍꢃꢎꢉꢗꢅꢍꢆꢈꢓꢎꢅꢉꢃꢈꢍꢌꢃꢈꢋꢈꢎꢕꢈꢘꢕꢌꢃꢅꢉꢃꢅꢒꢍꢃꢏꢌꢘꢔꢎꢓꢌꢙ
3285
Waveform and harmonic analysis
Waveform
harmonic
Analysis applications
Model 3284 and 3285 only
9199
BNC TO BANANA
Supporting a wealth of functions
ADAPTER
8807-51
MEMORY HiCORDER
Inrush current crest value
Inrush current
9094
OUTPUT CORD
+
The peak hold function displays
Crest
value
peak
the crest value of the inrush
current occurring when electrical
equipment is started.
Using the external output functions of the 3284 or 3285 in combination
with a HIOKI MEMORY HiCORDER enables recording of current
and frequency fluctuations and recording and harmonic analysis of
instantaneous waveforms. The monitor output (MON) spans a wide
frequency range from DC to 20 kHz (15 kHz in the case of 3285), which
allows the units to be used for electric current analysis in various fields
ranging from batteries to inverters.
0
-
Example of analysis using the MEMORY HiCORDER
RMS value of
current waveform
(inverter output )
▼
full-wave rectified waveforms
full-wave rectified waveforms
+
The AC+DC mode enables
measurement of the RMS value
of full- or half-wave rectified
waveforms used in electrical
machinery.
▲
0
current fluctuations
Harmonic analysis
-
[
] in the Specification :3285,3285-20
Accuracy at 23 ˚C ± 5 ˚C (73 ˚F ± 9 ˚F), 80% rh or less.
■ 3284 / 3285 Specifications
Measurement accuracy applies to input of at least 5% of the current range, and at least 10% of the voltage/frequency range.
Function Mode Display Switch 1
*
Range
Accuracy at DC or 45 to 66 Hz
±1.3%rdg.±3dgt.
Notes
10 to 45Hz or 66 to 2 [ 1 ] kHz
at DC or 45 to 66 Hz
PEAK accuracy
DC Average / PEAK 20.00 / 200.0
[200.0 / 2000]
±1.3%rdg.±7dgt.*6
±1.3%rdg.±7dgt.*6
±1.3%rdg.±7dgt.*6
±1.0%rdg.±7dgt.
±1.0%rdg.±7dgt.*7
±1.0%rdg.±7dgt.*7
Max. allowable current
200Arms cont. 300A peak max.
[2000Arms cont. 2840A peak max.]
Max. allowable voltage
600Vrms cont. 1000V peak max.
Current
(A)
AC
RMS / PEAK
±1.3%rdg.±3dgt.*2
±2.0%rdg.±5dgt.*3 *4
Resolution is
AC+DC RMS / PEAK
±1.3%rdg.
±
13dgt.*2
±2.0%rdg.±
7dgt.*3*4
0.1 [1] A during PEAK
DC Average / PEAK 30.00 / 300.0 / 600
±1.0%rdg.±3dgt.
±1.0%rdg.±3dgt.
±1.0%rdg.±7dgt.*5
±0.3%rdg.±1dgt.
Voltage
(V)
Resolution is decreased by
AC
RMS / PEAK
RMS / PEAK
±1.5%rdg.±5dgt.
±1.5%rdg.±7dgt.*5
1 digit except in the 600 V
range during PEAK
AC+DC
±1.0%rdg.±1dgt. (at 1000Hz)
Frequency (Hz)
Resistance(Ω)*8
Continuity Check*8
10.00 / 100.0/1000
1000/10.00k
1000
Effective in the V and A functions
Open terminalVoltage 3.1VDCorless
±1.5%rdg.±5dgt.
Threshold level 30 ±5Ω or less
Ω
*1 The peak hold function works automatically during PEAK measurement. *2 Add ±1.0% rdg. in the 1800 to 2000 A interval for the 3285. *3 Add ±1.0% rdg. in the 1k to 2kHz interval
for the 3284. *4 No assured accuracy in the 1800 to 2000 A interval for the 3285. *5 Add 6 dgt. in the 30 V range. *6 Add ±4.7% rdg. in the 2300 to 2840 A interval for the 3285. *7 Add ±
1.0%rdg.±13dgt when measuring 700 V to 1000V with Model 3285-20. *8 Model 3285-20 only.
Measurable conductor diameter : ø33 (1.30") mm, [ø55 (2.17") mm] max.
Effect of conductor position : within ±0.5[0.7] % in any direction from the center of sensor
Max. usablecircuitvoltage
Functions
Display update rate
:
:
:
600Vrms (insulated conductor)
Effect of external magnetic field
:
:
400A/m corresponds to 0.5 [2] A max.
2.5orless,1.7orlessinthe600Vrange,3284is1.5orlessinthe200Arange[1.42orlessinthe2000Arange]
recording (MAX. MIN.AVE) ,data hold, auto power-off, auto zero
Crest factor
digital indicationFASTapprox. 4times/second, NORMALapprox. 2 times/second,
SLOW1time/ 3seconds; bargraphindicationapprox. 4times/second
1 sec or less
Auto/Manual (Except resistance.)
40 Hz to 2 [1] kHz accuracy
A/Hz only: recording output DC1V/f.s. ±1.3% rdg.±10mV, Response: 250ms
or less (at 0➞90%)/ MONAonly : Waveform output 1Vrms/f.s. ±1.3% rdg.±5mV,
Frequency band: DC to 20 [15] kHz ±3dB
Safety standard : EN61010-1 CAT III600V, EN61010-2-032, EN60529 IP40
EMC
Ambient temperature of use
Storage temperature range
Power supply
:
:
:
:
EN61326, EN61000-3-2, EN61000-3-3
0 to 40˚C (32˚F to 104˚F), 80%rh or less (no condensation)
-10 to 50˚C (14˚F to 122˚F) or less (no condensation)
Display response time
Range switching
Frequency characteristic
External output
:
:
:
:
9445ACADAPTER(ExceptModel3285-20.)or6F22manganesebattery×1
(continuousoperation:Model3284/3285-25hours,Model3285-20-20hours)
(Except model 3285-20)
Dimensions and mass : Approx. 62W×230[260]H×39D mm, 460 [540]g
(Approx 2.45" × 9.06[10.3]" × 1.54" , 16.3 [19.1] oz.)
W
H
D
Options
3284
3285
CLAMP ON AC/DC HiTESTER
CLAMP ON AC/DC HiTESTER
9445-02 AC ADAPTER (UL)
9445-03 AC ADAPTER (EU)
An AC Adapter can only be used with Models 3284 and 3285.
3285-20 CLAMP ON AC/DC HiTESTER
(All include 9207-10 TEST LEADS, 9399 CARRYING CASE(3284), 9345 CARRYING CASE(3285/3285-20), HAND STRAP)
* 9094 OUTPUT CORD(Jack to banana plug)
*No CE marking
DISTRIBUTED BY
Shanghai Representative Office :
1704 Shanghai Times Square Office
93 Huaihai Zhong Road
Shanghai, 200021, P.R.China
TEL +86-21-6391-0090, 0092
FAX +86-21-6391-0360
HEAD OFFICE :
81 Koizumi, Ueda, Nagano, 386-1192, Japan
TEL +81-268-28-0562 / FAX +81-268-28-0568
E-mail: os-com@hioki.co.jp
HIOKI USA CORPORATION :
6 Corporate Drive, Cranbury, NJ 08512 USA
TEL +1-609-409-9109 / FAX +1-609-409-9108
E-mail: hioki@hiokiusa.com
E-mail: info@hioki.cn
All information correct as of Oct. 11, 2005. All specifications are subject to change without notice.
3284E6-5XM-05K Printed in Japan