AC AUTOMATIC INSULATION/  
WITHSTANDING HiTESTER  
3174  
Safety Standards  
Measuring Instruments  
Featuring contact check functionality  
The effects of test lead wire breaks, erroneous test results caused by faulty contact, and  
fluctuations in test voltage caused by variations in the instrument’s supply voltage on  
withstanding voltage and insulation resistance testing are well known.  
The 3174 (3174-01) AC AUTOMATIC INSULATION/WITHSTANDING HiTESTER is a  
low-cost solution featuring contact check functionality as well as a stabilized power supply  
to prevent the reduced test reliability that can result from these issues.  
To streamline production line test processes, the HiTESTER also features configuration of  
test parameters via RS-232C (GP-IB) and reading of parameters from the EXT I/O interface.  
2
Improved Test Reliability  
NEW  
Contact check function improves test reliability  
If a test lead wire were  
to break during testing with  
a tester that does not offer  
contact check functionality,  
defective test pieces would be  
judged to be non-defective.  
The 3174’s contact check function lets you detect test lead  
wire breaks and faulty contact during testing by sensing  
measurement issues in real time.  
Withstanding voltage tester  
Withstanding voltage tester  
The 3174 improves test  
Voltage  
output  
reliability by monitoring  
voltage to detect lose  
leads during testing  
and faulty wiring during  
measurement.  
Faulty contact  
*Break  
in lead  
Voltage output  
Voltage monitor  
Voltage monitor  
Voltmeter  
Voltmeter  
Contact check  
Contact check  
*Since contact check is performed  
while testing is in progress, use of  
contact check functionality does  
not increase cycle time.  
Test current  
Test current  
Test piece  
Test piece  
Faulty contact  
*Disconnected  
If a test lead were to come lose  
during testing with a measurement  
device that does not offer contact check  
functionality, defective test pieces  
would be judged to be non-defective.  
lead  
Fluorescent display tube  
The display uses a bright,  
easy-to-read fluorescent tube.  
DANGER lamp  
Flashes a warning during testing and  
whenever high voltage is present at the  
terminals.  
The DANGER lamp turns off when  
the voltage at the output terminals is  
no greater than AC 30 V or DC 60 V.  
9613  
9614  
SINGLE-HAND  
REMOTE CONTROL  
(option)  
TWO-HAND  
REMOTE CONTROL  
(option)  
Test mode selection  
External switch  
Select from three test modes:  
Enables start/stop control by means of  
the 9613 SINGLE-HAND REMOTE  
CONTROL or 9614 TWO-HAND  
REMOTE CONTROL.  
1.Manual test mode: W (withstand voltage testing) / I (insulation resistance testing)  
2.Automatic test mode: WI / IW  
(The 9613 and 9614 are optional.)  
NEW  
Judgment output at forced stop  
Ramp timer function  
The ramp-up initial value, ramp-up and ramp-down time  
parameters can be set independently.  
The ability to obtain a judgment even after a forced stop increases testing freedom.  
True effective value display  
NEW  
Eliminate the effects of supply voltage fluctuations  
Continued analytical testing after FAIL judgments  
Delay timer function  
Test pieces can now be analyzed by means of detailed monitoring of the  
test current accompanying FAIL judgments.  
3
Safe, Automated Operation  
Convenient  
Safe  
Interlock function  
Continuous full-auto withstanding voltage and insulation resistance testing  
Signal input from an external device such as an automatic  
tester can be used to disable output and prevent testing,  
ensuring safety during automatic testing and other uses.  
The 3174 lets you set independent test conditions including  
test voltage for withstanding voltage and insulation  
resistance testing, and you can perform these tests  
continuously. Press the WI key to automatically perform  
withstanding voltage followed by insulation resistance tests,  
or press the IW key to automatically perform insulation  
resistance followed by withstanding voltage tests.  
GP-IB (option)  
Enables remote control.  
RS-232C  
Enables remote control.  
Contact check  
LOW: A break in the lead is determined to have  
occurred when the check current is not detected.  
HIGH: A break in the lead is determined  
to have occurred when the contact check  
measurement voltage falls outside the range  
defined by the upper and lower thresholds.  
(Connectors are locked in place as shown  
on the left to prevent the leads from  
being disconnected.)  
Power cord inlet  
*
Requires an additional pair of  
9615 HIGH-VOLTAGE TEST LEADS  
.
Rear panel voltage output terminals  
The rear panel terminals are connected  
at all times to the front terminals. When  
embedding the 3174 HiTESTER inside  
an automatic tester, connections can be  
made via the rear panel terminals.  
Status out relay terminals  
Maximum input voltage: AC 150 V, DC 30 V  
Maximum contact current: 0.5 A  
An OR operation applied to various enabled conditions  
determines whether relay terminal output is turned on. Devices  
such as the 9616 WARNING LAMP can be connected.  
9616 WARNING LAMP  
(option)  
(AC100V,0.1A)  
Buzzer volume knobs  
The 3174 provides separate  
volume adjustments for the PASS  
and FAIL judgment buzzers.  
Output voltage generation  
Testing in progress  
PASS state  
1. H.V.ON  
2. TEST  
3. PASS  
(Includes warning lamp unit only.  
FAIL state  
4. FAIL  
Refer to the included plans and drill  
installation holes in a base such as a  
metal plate as necessary.)  
Interlocked state  
Ready state  
5. INT.LOCK  
6. READY  
7. EXT.CONT.  
8. POWER.ON  
Under external control  
When the 3174 is on  
External I/O connector (Signal lines have photocoupler isolation.)  
Pin  
1
2
3
4
I/O  
Signa  
READY  
L-FAIL  
U-FAIL  
PASS  
Function  
Interface specifications  
Low in ready state  
OUT  
OUT  
OUT  
OUT  
E X T I / O : Output signals  
Open collector output (with photocoupler isolation)  
Low in FAIL state (lower bound)  
Low in FAIL state (upper bound)  
Low in PASS state  
All active low with a maximum load voltage of DC 30 V  
Maximum output current: DC 100 mA per signal  
Output saturation voltage: DC 1.5 V or less  
Low in test sate  
5
6
7
8
9
10  
11  
12  
13  
14  
15-18  
22-25  
27  
28-29  
33-36  
OUT  
OUT  
IN  
IN  
IN  
TEST  
H.V.ON  
EXT-E  
START  
STOP  
INT.LOCK  
W-MODE  
I-MODE  
W-FAIL  
I-FAIL  
ISO.COM  
MEM-0 to 3  
MEM-E  
MODE-0,1  
ISO.DCV  
Low when voltage is present at output terminals  
When low, external I/O input signals are enabled  
When low, same function as START key  
When low, same function as STOP key  
Interlock on open  
Input signals  
All active low input (with photocoupler isolation)  
Maximum applied voltage: 30 V  
High level voltage: from DC 15 V to 30 V, or open  
Low level voltage: DC 5 V or less (-6 mA typ.)  
IN  
Low during withstanding voltage testing  
Low during insulation resistance testing  
Low in FAIL state during withstand voltage testing  
Low in FAIL state during insulation resistance testing  
Ground inputs for external devices  
Saved test selected pins  
OUT  
OUT  
OUT  
OUT  
IN  
IN  
IN  
OUT  
OUT  
E X T S W : Input signals (contact input)  
START, STOP, SW.EN (external switch terminal enable)  
Output signals  
LED activation signal (maximum load current: 40 mA)  
When low, enables memory selected pins  
Current test mode  
Internal DC 15 V power supply (100 mA)  
AC AUTOMATIC INSULATION/WITHSTANDING HiTESTER 3174  
4
Withstanding Voltage Testing  
Timers  
Setting range  
Operation  
0.3 to 999 s  
:
:
Test Voltage  
When set to on: Display counts down from the set time after start.  
When set to off: Display indicates time elapsed since start.  
AC 0.2 to 5 kV (50/60 Hz), single-range output  
Output voltage  
:
:
PWM switching method (0 V start; voltage can be  
changed while generating output)  
Voltage output method  
0.1 s (0.3 to 99.9 s) ±50 ms  
1 s (100 to 999 s) ±0.5 s  
Setting resolution/accuracy  
:
Digital (setting resolution: 0.01 kV)  
Voltage specification method  
Output voltage accuracy  
Maximum rated output  
Transformer capacity  
Voltmeter  
:
:
:
:
:
Ramp Timers (Withstand voltage testing)  
Setting range  
±1.5% of setting ±20 V  
AC 100 VA (5 kV/20 mA) continuous rating  
100VA  
0.1 to 99.9 s: Ramp-up and ramp-down can be set independently.  
:
:
Ramp-up: The output voltage increases linearly from the  
initial voltage to the test voltage over the ramp-up time.  
Ramp-down: The output voltage decreases from the set  
voltage to 0 V over the ramp-down time after the test time  
elapses, and the display counts down from the set time.  
*The actual ramp-up waveform varies with the load due to the analog response delay.  
Operation  
True effective value display  
Digital meter: AC 0 kV to 5.00 kV  
Accuracy: ±1.5% rdg (1,000 V or less, ±15 V)  
Sine wave  
0.1s  
Setting resolution  
:
Output waveform  
:
:
15% or less (converges to setting within 1 s during  
change from maximum rated load to no load)  
Voltage change rate  
Delay Timers (Insulation resistance testing)  
Setting range  
0.1 to 99.9 s  
:
:
5% or less (tester impedance during measurement with 5  
kV output under 40 MΩ load)  
Distortion factor  
Frequency  
:
:
0.1s  
Setting resolution  
Decision Function  
50/60Hz (±0.2%)  
Window comparison method with upper and lower bound settings (digital specification)  
Decision method  
Decision results  
:
:
Current Detection  
Current measurement range  
Measurement range  
Indicated value  
UPPER-FAIL: The measured current (measured insulation  
value) exceeded the specified upper threshold.  
0.01 mA to 20 mA (2 ranges)  
10mA/20mA  
:
:
:
:
PASS: The measured current (measured insulation value) fell within  
the range defined by the specified upper and lower thresholds.  
LOWER-FAIL: The measured current (measured  
insulation value) was less than the specified lower bound.  
UPPER LOWER-FAIL: A testing error occurred, for  
example due to a failure to generate the set voltage.  
True effective value display  
0.00 mA to 10.00 mA, 0.01 mA (10 mA range)  
0.0 mA to 20.0 mA, 0.1 mA (20 mA range)  
±2% rdg ±0.05 mA (10 mA range)  
Measurement ranges and  
resolutions  
Measurement accuracy  
:
±2% rdg ±0.5 mA (20 mA range)  
Display, buzzer, and EXT I/O signal output is generated according to each decision result.  
Decision processing  
Setting range  
:
:
AC withstanding voltage: 0.1 to 20.0 mA (upper threshold), 0.1 to 19.9 mA (lower threshold)  
DC insulation: 0.2 to 2,000 MΩ (500 V) or 1.0 to 2,000 MΩ (1,000 V) for both upper and lower thresholds  
Insulation Resistance Measurement  
AC withstanding voltage: 0.1 mA  
DC insulation: 0.01 MΩ (0.2 to 2.00 MΩ), 0.1 MΩ (2.10 to  
20.0 MΩ), 1 MΩ (21.0 to 200 MΩ), 10 MΩ (210 to 2,000 MΩ)  
Setting resolution  
:
Test Voltage  
DC 500 V/1,000 V (positive polarity)  
1 to 1.2 times rated voltage  
Rated voltage  
:
:
:
:
:
:
Unloaded voltage  
Rated measurement current  
Short-circuit current  
Measurement range  
Contact Check  
1 to 1.2 mA  
Detection method: Average value detection/effective value conversion  
Accuracy: Setting ±50 V *Inaccuracy may increase when the waveform is distorted.  
Voltmeter accuracy  
:
4 to 5 mA (500 V), 2 to 3 mA (1,000 V)  
0.2 to 2,000 MΩ (500 V), 0.5 to 2,000 MΩ (1,000 V)  
Enables the contact check function (does not increase cycle time).  
LOW: A break in the lead is determined to have occurred when the check current is not detected.  
HIGH: Upper and lower thresholds for the check detection voltage can be set. A  
break in the lead is determined to have occurred when the contact check measurement  
voltage falls outside the range defined by the upper and lower thresholds.  
Decision results  
:
0.5 to 999 MΩ (500 V), 1 MΩ to 999 MΩ (1,000 V): ±4% rdg  
1,000 MΩ to 2,000 MΩ: ±8% rdg  
Guaranteed accuracy  
ranges/accuracies  
0.01 MΩ (0.20 MΩ to 19.9 MΩ)  
0.1 MΩ (20.0 MΩ to 199.9 MΩ), 1 MΩ (200 MΩ to 2,000 MΩ)  
Measurement resolution  
:
:
Withstanding voltage testing: 0.20 kV to 5.0 kV (0.01 kV resolution; applies to both  
upper and lower thresholds)Insulation resistance measurement: Upper threshold of  
600 V and lower threshold of 500 V (during 500 V measurement); upper threshold of  
1,200 V and lower threshold of 1,000 V (during 1,000 V measurement) (both fixed)  
Voltage setting range  
:
2 MΩ, 20 MΩ, 200 MΩ, 2,000 MΩ (500 V)  
4 MΩ, 40 MΩ, 400 MΩ, 2,000 MΩ (1,000 V)  
Measured resistance ranges  
General Specifications  
Model  
Fluorescent display tube (digital display)  
Display  
:
:
:
:
:
:
AC AUTOMATIC INSULATION/WITHSTANDING HiTESTER 3174  
AC AUTOMATIC INSULATION/WITHSTANDING HiTESTER 3174-01 (GP-IB Model)  
Output voltage, detected current, insulation resistance  
2 times per second, minimum  
Monitor functions  
Monitor period  
0°C to 40°C, 80% RH or less (non-condensing)  
-10°C to 50°C, 90% RH or less (non-condensing)  
Operating temperature range  
Storage temperature range  
Options  
23 ±5°C, 80% RH or less (non-condensing)  
(With warm-up period of at least 10 min)  
1 year  
Temperature and humidity  
range for guaranteed accuracy  
ELECTRIC SAFETY TESTING SOFTWARE  
SINGLE-HAND REMOTE CONTROL  
TWO-HAND REMOTE CONTROL  
HIGH-VOLTAGE TEST LEADS  
WARNING LAMP (AC100V,0.1A)  
RS-232C CABLE (9-pin to 9-pin/cross/1.8 m)  
9267 (available soon)  
Guaranteed accuracy term  
Operating environment  
Supply voltage  
:
:
:
:
:
:
:
:
:
:
9613  
9614  
9615  
9616  
9637  
Indoors at elevations of up to 2,000 m at a pollution level of 2  
AC 100 to 240 V Designed to tolerate voltage fluctuations of ±10% of the rated supply voltage.  
50Hz/60Hz  
Power supply frequency  
Withstanding voltage  
Maximum rated power  
Dimensions  
Power supply to chassis: 1.39 kV at 10 mA for 15 s  
200VA  
RS-232C CABLE (9-pin to 25-pin/cross/1.8 m) 9638  
GP-IB CONNECTING CABLE (2 m)  
Approx. 320 (W) × 155 (H) × 395 (D) mm (excluding protruding parts)  
Approx. 15 kg  
9151-02  
Weight  
EMC: EN61326 Class A, EN61000-3-2, EN61000-3-3, Safety: EN61010-1  
Applicable standards  
Included accessories  
HIGH-VOLTAGE TEST LEADS 9615 (1 each high-voltage  
and return), power cord, disconnection prevention plate  
HIOKI (Shanghai) Sales & Trading Co., Ltd. :  
DISTRIBUTED BY  
1904 Shanghai Times Square Office, 93 Huai Hai Zhong Road  
Shanghai, P.R.China POSTCODE: 200021  
TEL +86-21-6391-0090/0092 FAX +86-21-6391-0360  
E-mail: info-sh@hioki.cn  
Beijing Office :  
HEAD OFFICE :  
A-2602 Freetown, 58 Dong San Huan Nan Road  
Beijing, P.R.China POSTCODE: 100022  
TEL +86-10-5867-4080/4081 FAX +86-10-5867-4090  
E-mail: info-bj@hioki.cn  
81 Koizumi, Ueda, Nagano, 386-1192, Japan  
TEL +81-268-28-0562 / FAX +81-268-28-0568  
E-mail: os-com@hioki.co.jp  
Guangzhou Office :  
HIOKI USA CORPORATION :  
Room 303, Profit Plaza, No.76, West Huangpu Road  
Guangzhou, P.R.China POSTCODE: 510623  
TEL +86-20-38392673/2676 FAX +86-20-38392679  
E-mail: info-gz@hioki.cn  
6 Corporate Drive, Cranbury, NJ 08512 USA  
TEL +1-609-409-9109 / FAX +1-609-409-9108  
E-mail: hioki@hiokiusa.com  
All information correct as of Jun. 4, 2008. All specifications are subject to change without notice.  
3174E2-86B  

相关型号:

3174-21.4-10

21.4 MHz Bandpass Filter
KR

3174-214-10

21.4 MHz Bandpass Filter
KR

31742

TRANSFORMER, HDSL, 1178 Kbps, 16 mAdc
SCHOTT

31743

Data Line Filter, 2 Function(s), 12A, ROHS COMPLIANT PACKAGE-4
VICOR

3174A-21.4-0.4

21.4 MHz Bandpass Filter
KR

3174A-214-04

21.4 MHz Bandpass Filter
KR

3175

HALL-EFFECT LATCHES
ALLEGRO

317607-001

Express Chipset
INTEL

317610-001

Express Chipset
INTEL
CDE
CDE
CDE